K102148 · Moksha Digital Software Pvt Limited · LLZ · Aug 17, 2010 · Radiology
Device Facts
Record ID
K102148
Device Name
CURIEPACS (DICOM PACS SOFTWARE)
Applicant
Moksha Digital Software Pvt Limited
Product Code
LLZ · Radiology
Decision Date
Aug 17, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
CuriePACS (Dicom PACS Software) Software is a software device that receives digital images and data from various sources (Computed Radiography, Magnetic Resonance Imaging, Ultrasound, Endoscopy, Computed Tomography, Digital X-Ray Mammography, Nuclear Medicine Imaging, Secondary Capture, Radio Fluoroscopy, X-Ray or other Angiograms such as CT Angio). Images and data can he captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor that offers at least 5 Megapixel resolution and meets other technical specifications reviewed and accepted by FDA.
Device Story
CuriePACS is a scalable PACS solution for DICOM imaging; consists of MD Athena (server for storage/connectivity) and MD Vision (viewer for retrieval/workflow). Inputs include DICOM-compliant images from CR, CT, DX, MG, MR, NM, RF, SC, US, XA, and ES modalities. Operates on network-connected PCs; used by medical professionals for electronic viewing, storage, and communication. Provides image processing tools: window/level, zoom, pan, magnifying glass, probe, spatial locator, shutter, and various ROI/measurement tools. Output displayed on FDA-approved medical-grade monitors. Facilitates clinical decision-making by enabling remote access and standardized review of diagnostic images across distributed networks.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence consists of software validation, risk analysis, and demonstration of DICOM 3 compliance.
Technological Characteristics
Software-based PACS; runs on Windows XP/Vista; requires .NET Framework 3.5 SP1, Intel Core 2 Duo, 2GB RAM, 320GB HDD, and Gigabit Ethernet. Supports DICOM, JPEG, and JPEG 2000. Connectivity via network-connected server. Requires FDA-approved self-calibrating medical-grade monitor with DVI input for diagnostic interpretation.
Indications for Use
Indicated for medical professionals to receive, store, communicate, process, and display digital medical images from various modalities (CR, MRI, US, Endoscopy, CT, Mammography, Nuclear Medicine, Secondary Capture, R/F, X-Ray/Angio). Contraindicated for primary interpretation of lossy compressed mammographic images and digitized film screen images. Mammographic interpretation requires FDA-cleared 5MP monitors.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
## 510(K) Summary K102148
Moksha Digital Software PVT Limited #1611, Janardhana Towers, 7th Cross, 19th Main, Sector 1, HSR Layout, Bangalore 560034, India Phone : +91 (80) 4110 1208 Fax : +91 (80) 4110 1824 E-mail : sales@mokshadigital.com Contact: Deepak Sharma, Managing Director Date prepared: August 12, 2010
AUG 1 7 2010
- 1 . Trade Name: CuriePACS Dicom PACS Software Common Name: PACS Software Classification Name: System, image processing, radiological
- 2. Regulation Description Picture archiving and communications system, product code LLZ, Regulation: 892.2050 Class of device: Class II.
- 3. The legally marketed device to which we are claiming equivalence Voyager PACS System, Voyager Imaging, K062062.
- 4. Description of device: CuriePACS is a comprehensive solution for Dicom imaging needs. This scalable PACS solution provides electronic viewing, storage and communication in a secure environment. The device consists of: 1. Dicom Server (MD Athena) - For storage and connectivity with all the modalities and 2. Image Viewer (MD Vision) - To connect to the server and retrieve images based on search criteria and other workflow requirements. Image modalities supported: CR. CT. DX. MG. MR. NM. RF. SC. US. XA (x-ray), and ES.
- న. Indications for use: CuriePACS (Dicom PACS Software) Software is a software device that receives digital images and data from various sources (Computed Radiography, Magnetic Resonance Imaging, Ultrasound, Endoscopy, Computed Tomography, Digital X-Ray Mammography, Nuclear Medicine Imaging, Secondary Capture, Radio Fluoroscopy, X-Ray or other Angiograms such as CT Angio). Images and data can he captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor that offers at least 5 Megapixel resolution and meets other technical specifications reviewed and accepted by FDA.
- Technological characteristics: The software is installed on a network connected PACS server 6. with the following minimum requirements:
- Microsoft® Windows® XP Service Pack 2 or above .
- Microsoft® Windows® Vista Service Pack 1 or above .
- . Microsoft .Net Framework 3.5 SP1
- Microsoft Internet Explorer 7 or above .
- Intel Core 2 Duo Processor .
- 320 GB HDD .
{1}------------------------------------------------
- 2 GB RAM (Min) •
- Dedicated or Onboard Graphics card with minimum 1440 x 900 resolution and DVI output. •
- . Gigabit Ethernet Adaptor
- . Display: The display used for diagnosis purposes should be FDA approved self calibrating medical grade monitor which is driven by a digital input (DVI output from graphics card).
| Characteristic | Predicate Device<br>Voyager PACS System, Voyager<br>Imaging, K062062 | CuriePACS (Dicom PACS Software)<br>(This submission) |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | Voyager PACS System is a software based device that receives digital images and data from<br>various sources (i.e. CT scanners, MR scanners,<br>ultrasound systems, R/F Units,<br>computed & direct radiographic devices.<br>secondary capture dev ices, scanners, imaging<br>gateways, etc.). Images and data can be<br>captured, stored, communicated, processed and<br>displayed within the system and or across<br>computer networks at distributed locations.<br>Lossy compressed mammographic images and<br>digitized film screen images must not be<br>reviewed for primary image interpretation.<br>Mammographic images may only be interpreted<br>using an FDA approved monitor that offers at<br>least 5 Mpixel resolution and meets other<br>technical specifications reviewed and accepted<br>by FDA. | CuriePACS (Dicom PACS Software) Software<br>is a software device that receives digital<br>images and data from various sources<br>(Computed Radiography, Magnetic Resonance<br>Imaging, Ultrasound, Endoscopy, Computed<br>Tomography, Digital X-Ray Mammography,<br>Nuclear Medicine Imaging, Secondary<br>Capture, Radio Fluoroscopy, X-Ray or other<br>Angiograms such as CT Angio). Images and<br>data can be captured, stored, communicated,<br>processed and displayed within the system and<br>or across computer networks at distributed<br>locations. Lossy compressed mammographic<br>images and digitized film screen images must<br>not be reviewed for primary image<br>interpretation. Mammographic images may<br>only be interpreted using an FDA cleared<br>monitor that offers at least 5 Megapixel<br>resolution and meets other technical<br>specifications reviewed and accepted by FDA. |
| Use, key feature<br>Connection | Medical professionals, WEB based<br>WEB and Ethernet | SAME |
| Multimodality<br>Connectivity,<br>DICOM<br>compatibility | Voyager PACS can receive with ease<br>any DICOM compliant image,<br>irrespective of the source | Connect all DICOM Modalities CR,<br>CT, DX, MG, MR, NM, RF, SC,<br>US, XA (x-ray), and ES. |
| Image Sources | Dicom, JPEG and JPEG 2000 compliant | SAME |
| Target hardware | PC compatible | SAME |
| Image Processing<br>Tools | • Brightness/ Contrast<br>• Mouse button operation for quick<br>image manipulation • Selectable<br>multiple image view• Magnify, zoom<br>and pan• Flip (left/ right, top/bottom)<br>and rotate• Image inversion<br>• Measurement and angles | All processing tools Stack,<br>Window/Level, Zoom, Pan,<br>Magnifying Glass, Probe, Spatial<br>Locator, Draw Shutter, Ruler,<br>Elliptical ROI, Rectangular ROI,<br>Polygonal ROI, Text Area, Text<br>Callout, Protractor. |
| Security | SSL Encryption | SAME |
#### 7. Comparison Table
Performance: The results of nonclinical tests submitted showing Dicom 3 compliance, 8. along with bench testing (software validation and risk analysis) shows that this new device has equivalent indications and performs in an equivalent fashion to the named predicate. Therefore this device poses no new issues of safety or effectiveness, and is substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, representing the department's mission to protect the health of all Americans.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Moksha Digital Software PVT Limited % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25" Street NW BUFFALO MN 55313
AUG 1 7 2010
Re: K102148
Trade/Device Name: CuriePACS (Dicom PACS Software) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and Communications System Regulatory Class: II Product Code: LLZ Dated: July 29, 2010 Received: July 30, 2010
### Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
Enclosure
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{4}------------------------------------------------
# Indications for Use
K102148
510(k) Number (if known): K102148.
Device Name: CuriePACS (Dicom PACS Software)
Indications For Use:
CuriePACS (Dicom PACS Software) Software is a software device that receives digital images and data from various sources (Computed Radiography, Magnetic Resonance Imaging, Ultrasound, Endoscopy, Computed Tomography, Digital X-Ray Mammography, Nuclear Medicine Imaging, Secondary Capture, Radio Fluoroscopy, X-Ray or other Angiograms such as CT Angio). Images and data can he captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor that offers at least 5 Megapixel resolution and meets other technical specifications reviewed and accepted by FDA.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Page 1 of 1
510K K102148
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.