K102000 · Keosys Sas · LLZ · Oct 12, 2010 · Radiology
Device Facts
Record ID
K102000
Device Name
IMAGYS
Applicant
Keosys Sas
Product Code
LLZ · Radiology
Decision Date
Oct 12, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Imagys (KSNRMAVWRxx, KSNERDxx) is a medical solution for the upload, the secure transfer, the management, the viewing, the reviewing, the storage and archiving of imaging data, and the related documentations, acquired from a variety of imaging devices. Imagys can also support the management of the imaging parts of a clinical trial workflow including the following steps : quality control, data processing and audit trails. Imagys runs on any standard PC and servers compliant with the Imagys specifications. Typical end users are trained medical professionals and clinical trial actors. The imaging data can be CT (Computed Tomography), MR (Magnetic Resonance), CR (Computed Radiography), NM (Nuclear Medecine), PET (Positon Emission Tomography), SC (Secondary Capture), US (Ultrasound), XA (Angiographic), video and more DICOM, or others imaging standards, data accepted by the Imagys solution. It is the user responsibility to check that the ambient luminosity conditions, the images compression ratio and the interpretation monitor specifications are consistent with a clinical diagnostic use of the data. This device is not indicated for mammography use. Lossy compressed mammography-images and digitized film screen images must not be used for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
Device Story
Imagys is a software-based Picture Archiving and Communication System (PACS) designed for clinical trial workflows. It ingests medical imaging data (DICOM and other standards) from various modalities including CT, MR, CR, NM, PET, US, and XA. The system facilitates secure transfer, quality control, data processing, audit trails, storage, and archiving. It operates on standard PCs and servers. Medical professionals and clinical trial personnel use the software to view and review images for clinical trial management. The device provides tools for data management and visualization, assisting clinicians in assessing imaging data within the context of clinical trials. It does not perform automated diagnostic interpretation but supports the workflow of imaging data management. The system is not intended for mammography interpretation.
Clinical Evidence
Bench testing only. The device underwent performance and functional testing as part of the software development process, documented in a Software Test Plan.
Technological Characteristics
Software-based PACS running on standard PCs and servers. Supports DICOM and other imaging standards. Connectivity via network for data transfer. No specific hardware materials or energy sources defined beyond standard computing infrastructure. Software-based image processing and management.
Indications for Use
Indicated for trained medical professionals and clinical trial actors for the management, viewing, and archiving of medical imaging data (CT, MR, CR, NM, PET, SC, US, XA, video) in clinical trial workflows. Not indicated for mammography use; lossy compressed mammography images and digitized film screen images are contraindicated for primary interpretation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo for Keosys Medical Imaging. The logo consists of a black circle with a white "K" inside, followed by the word "KEOSYS" in black, block letters. Below "KEOSYS" is the phrase "MEDICAL IMAGING" in smaller, black letters.
K102000
OCT 12 2010
# 510(k) Summary of Safety and Effectiveness
This summary of safety and effectiveness is provided as part of the Premarket Notification for Imagys in conformance with 21 CFR 807.92.
| Date Prepared: | July 2010. |
|-----------------------------|------------------------------------------------------------|
| Submitter: | Keosys S.A.S. |
| | 1, impasse Augustin Fresnel, Z.A. du Moulin Neuf, B.P. 227 |
| | 44815 Saint-Herblain cedex, FRANCE |
| Contact name: | Mr. Anthony MOTTIER |
| Contact Email: | anthony.mottier@keosys.com |
| Contact Telephone: | 00 33 (0)2 40 92 26 13 |
| Contact Fax: | 00 33 (0)2 40 92 26 26 |
| Device Trade Name: | Imagys |
| Device Common Name: | Picture Archiving and Communication System |
| Device Classification Name: | Picture Archiving and Communication System |
| Classification Panel: | 21 CFR §892.2050 |
| Product code: | LLZ |
| Device Classification: | Class II |
| Predicate Devices: | |
| Name | Manufacturer | 510(k) number |
|--------------------------------|-----------------------|---------------|
| INTEGRADWeb MPR/MIP | Dynamic Imaging, Inc. | K042313 |
| eFilm Worksattion with Modules | eFilm Medical Inc. | K020995 |
| Visio+ CD Viewer | Keosys S.A.S. | K070606 |
#### Device Description:
Imagys is a computing and software solution designed by experts in medical imaging and telecommunications for the collection, quality control, processing, analysis and visualization of imaging data. Imagys can help to execute olinical trials with secure transfer, management, viewing, printing, storage and archiving of data.
#### Indications for Use:
Imagys (KSNRMAVWRxx, KSNERDxx) is a medical solution for the upload, the secure transfer, the management, the viewing, the reviewing, the storage and archiving of imaging data, and the related documentations, acquired from a variety of imaging devices. Imagys can also support the management of the imaging parts of a clinical trial workflow including the following steps : quality control, data processing and audit trails. Imagys runs on any standard PC and servers compliant with the Imagys specifications. Typical end users are trained medical professionals and clinical trial actors.
The imaging data can be CT (Computed Tomography), MR (Magnetic Resonance), CR (Computed Radiography), NM (Nuclear Medecine), PET (Positon Emission Tomography), SC (Secondary Capture), US (Ultrasound), XA (Angiographic), video and more DICOM, or others imaging standards, data accepted by the Imagys solution.
It is the user responsibility to check that the ambient luminosity conditions, the images compression ratio and the interpretation monitor specifications are consistent with a clinical diagnostic use of the data.
This device is not indicated for mammography use. Lossy compressed mammography-images and digitized film screen images must not be used for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
#### Testing:
Imagys is tested according to the specifications that are documented in a Software Test Plan. Performance and functional testing are an integral part of Keosys's software development process.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized image of an eagle or bird with its wings spread, formed by three overlapping profiles of human faces.
MAR 16 2011
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Mr. Anthony Mottier Quality System Manager Keosys S.A.S. I, ImpasseAugustin Fresnel, Z.A. du Moulin Neuf, B.P. 227 Saint-Herblsin, 44815 FRANCE
Re: K102000
Trade/Device Name: Imagys Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 12, 2010 Received: July 15, 2010
Dear Mr. Mottier:
This letter corrects our substantially equivalent letter of October 12, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{2}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely Yours,
Mary S Patil
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{3}------------------------------------------------
## Indications for Use
### 510(k) Number (if known):
Device Name: Imagys
Indications For Use:
lmagys (KSNRMAVWRxx, KSNERDxx) is a medical solution for the upload, the secure transfer, the O management, the viewing, the printing, the printing, the storage and archiving of imaging data, and the O related documentations, acquired from a variety of imaging devices. Imaging dual, and the U of the inaging parts of a clinical trial workflow including steps : quality control, data processing C and audit trails. Imagys runs on any standard PC and servers compliant with the Imagys specifications. Typical ( end users are trained medical professionals and clinical trial actors.D 0
The imaging data can be CT (Computed Tomography), MR (Magnetic Resonance), CR (Computed O Radiography), NM (Nuclear Medecine), PET (Positon Emission Roomano), SC (Secondary Capture), US D (Ultrasound), XA (Anglographic), video and more DICOM, or others imaging standards, data accepted by the D Imagys solution.D
D
It is the user responsibility to check that the ambient luminosity conditions, the images compression ratio and O the interpretation monitor specifications are consistent with a dinical to the data. This device is I not indicated for mammography use. Lossy compressed mammography-images and digitized films area. O images must not be used for primary image interpretations. Mammography images may only handed on a using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical O specifications approved by the FDA.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, @fficeref-Beviews Concertion (CDFL) ○エレ
(Division Sign-Off)
510K
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Page 1 of 1
3
ij
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.