← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K094064

# ILUMA VISION MODEL VERSION 2.2 (K094064)

_3M Imtec · LLZ · Apr 30, 2010 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K094064

## Device Facts

- **Applicant:** 3M Imtec
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Apr 30, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

ILUMAVision is a software application used for the display and 3D visualization of medical image files from scanning devices, such as CT, MRI, PET or 3D Ultrasound. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, review, store, print, and distribute DICOM 3.0 compliant images, utilizing standard PC hardware. Additionally, ILUMA Vision is a preoperative software application used for the simulation and evaluation of dental implants, orthodontic planning and surgical treatments. ILUMAVision is not intended for use with mammography.

## Device Story

ILUMAVision is a PC-based software application for medical image management and 3D visualization. It ingests DICOM 3.0 compliant images from CT, MRI, PET, or 3D ultrasound scanners via TCP/IP network or local storage. The software performs image filtering, 3D reconstruction, and quantitative analysis to support clinical review, measurement, and annotation. Key features include multi-tissue opacity control, volume sculpting, segmentation, and specialized modules for temporal bone isolation and virtual endoscopy. Clinicians use the software to monitor treatment progress, plan dental implants, and perform orthodontic/surgical planning. Output includes rendered 3D visualizations, measurements, and reports, which are used by physicians to inform clinical decision-making and surgical workflows. The device facilitates data export and PACS integration, aiding in preoperative assessment and patient treatment planning.

## Clinical Evidence

Bench testing only. No clinical data provided. Verification and validation of software features were performed to confirm performance of new modules (temporal bone, virtual endoscopy) and PACS integration.

## Technological Characteristics

PC-based software application; runs on Intel-based PCs with Microsoft Windows. Connectivity via TCP/IP network. Supports DICOM 3.0 input/output and JPEG export. Features include 3D volume rendering (radiographic projection, surface rendering, fly-through), segmentation, and measurement tools. No specific hardware requirements beyond standard PC components.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- ILUMAVision, version 2.0 ([K081347](/device/K081347.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K094064

APR 3 0 2010

# Premarket Notification (510(k)) Summary

### 1. Sponsor Information

3M IMTEC

2401 North Commerce

Ardmore, OK 73401

Ginger Cantor, RAC Contact Person: Regulatory Affairs Specialist Phone Number: (651) 733-1317 FAX Number: (651) 737-9665 e-mail: gcantor@mmm.com Date of Summary: December 29, 2009

### 2. Device Name and Classification:

| Common or Usual Name:  | Software application for medical purposes (diagnostic imaging)                 |
|------------------------|--------------------------------------------------------------------------------|
| Proprietary Name:      | ILUMAVision, v. 2.2                                                            |
| Classification Name:   | Picture Archiving and Communications System, Class II device, 21 CFR §892.2050 |
| Performance Standards: | None                                                                           |

## 3. Predicate Device:

ILUMAVision, version 2.0 (K081347)

ILUMAVision 2.2 510(k) 29 Dec 2009

{1}------------------------------------------------

#### 4. Description of Device:

ILUMAVision is an image management software application used for the display and 3D visualization of medical image files obtained from scanning devices, such as computed tomography (CT), magnetic resonance imaging (MRI), positron emission tomography (PET) or three-dimensional (3D) ultrasound.

ILUMAVision uses image filtering, 3D reconstruction and quantitative algorithms to view, measure, and annotate images. ILUMA Vision can be used to make panoramic images and to monitor treatment progress, capture images, bookmark certain items in a treatment, generate and edit reports, and export datasets. The application can also query and import images directly from a Picture Archiving and Communication System (PACS) over a TCP/IP network.

It distributes DICOM 3.0 compliant images, using standard personal computer (PC) hardware. Images can also be saved in JPEG format.

The software is intended to run on a personal computer (PC).

{2}------------------------------------------------

#### 5. Intended Use/Indications for Use:

ILUMAVision is a software application used for the display and 3D visualization of medical image files from scanning devices, such as CT, MRI, PET or 3D Ultrasound.

It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, review, store, print, and distribute DICOM 3.0 compliant images, utilizing standard PC hardware.

Additionally, ILUMAVision is a preoperative software application used for the simulation and evaluation of dental implants, orthodontic planning and surgical treatments.

ILUMAVision is not intended for use with mammography.

#### Substantial Equivalent Determination 6.

Version 2.2 includes additional features of a temporal bone module, a virtual endoscope module and the functionality of exporting directly from the application

The additional software features have been suitably verified and validated; the table below identifies the equivalence between the modified ILUMAVision software (v. 2.2) and the cleared predicate ILUMA Vision software (v. 2.0).

While new features have been added, ILUMA Vision version 2.2 as modified in this pre-market notification submission has the same intended use, same intended users and indications for use as the predicate device ILUMA Vision version 2.0.

{3}------------------------------------------------

| Feature             | ILUMAVision<br>version 2.0<br>(Predicate- K081347)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | ILUMAVision<br>version 2.2<br>(This submission)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Comment                            |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| Indications for Use | A software application used for the display and 3D visualization of medical image files from scanning devices, such as CT, MRI, PET or 3D Ultrasound.<br><br>Intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, review, store, print, and distribute DICOM 3.0 compliant images, utilizing standard PC hardware.<br><br>Additionally, ILUMA Vision is a preoperative software application used for the simulation and evaluation of dental implants, orthodontic planning and surgical treatments.<br><br>ILUMA Vision is not intended for use with mammography. | A software application used for the display and 3D visualization of medical image files from scanning devices, such as CT, MRI, PET or 3D Ultrasound.<br><br>Intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, review, store, print, and distribute DICOM 3.0 compliant images, utilizing standard PC hardware.<br><br>Additionally, ILUMA Vision is a preoperative software application used for the simulation and evaluation of dental implants, orthodontic planning and surgical treatments.<br><br>ILUMA Vision is not intended for use with mammography. | No change from original 510(k)     |
| Computer Platform   | Minimum Requirement:<br>Intel®-based PC running<br>Microsoft® Windows®                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Minimum Requirement:<br>Intel®-based PC running<br>Microsoft® Windows®                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | No change from original 510(k)     |
| Communications      | TCP/IP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | TCP/IP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | No change from original 510(k)     |
| DICOM Compliance    | DICOM 3.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | DICOM 3.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | No change from original 510(k)     |
| JPEG Compliance     | Images may be saved in JPEG format                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Images may be saved in JPEG format                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | No change from original 510(k)     |
| Input Image Format  | DICOM 3.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | DICOM 3.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | No change from original 510(k)     |
| Output Image Format | DICOM 3.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | DICOM 3.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | No change from original 510(k)     |
| Image Archive       | Computer hard drive, CD, DVD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Computer hard drive, CD, DVD or PACS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Added PACS capability to archiving |
| Image Display       | Color/Grayscale CRT or LCD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Color/Grayscale CRT or LCD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | No change from original 510(k)     |

ЗМ ІМТЕС

{4}------------------------------------------------

| Substantial Equivalence Summary Table                   |                                                                                       |                                                                                                                                              |                                   |  |  |
|---------------------------------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|--|--|
| Feature                                                 | ILUMAVision<br>version 2.0<br>(Predicate- K081347))                                   | ILUMAVision<br>version 2.2<br>(This submission)                                                                                              | Comment                           |  |  |
| Printing                                                | Print to standard PC<br>connected printers                                            | Print to standard PC<br>connected printers                                                                                                   | No change from<br>original 510(k) |  |  |
| Import from PACS                                        | Not in original 510k                                                                  | New functionality                                                                                                                            | New feature                       |  |  |
| Export to PACS                                          | Not in original 510k                                                                  | New functionality                                                                                                                            | New feature                       |  |  |
| Volume Rendering                                        | Radiographic Projection,<br>Surface rendering, Fly-<br>through                        | Radiographic Projection,<br>Surface rendering, Fly-<br>through                                                                               | No change from<br>original 510(k) |  |  |
| Image Edit                                              | Multi-tissue opacity<br>control, volume<br>sculpting, segmentation                    | Multi-tissue opacity<br>control, volume<br>sculpting, segmentation                                                                           | No change from<br>original 510(k) |  |  |
| Region of Interest<br>(ROI)/Volume of Interest<br>(VOI) | 2D region and 3D<br>volume of interest<br>selection tools                             | 2D region and 3D<br>volume of interest<br>selection tools                                                                                    | No change from<br>original 510(k) |  |  |
| 2D Measurements                                         | 2D measurement tools<br>including distance and<br>angle                               | 2D measurement tools<br>including distance and<br>angle                                                                                      | No change from<br>original 510(k) |  |  |
| 3D Measurements                                         | 3D measurement tools<br>including distance and<br>angle                               | 3D measurement tools<br>including distance and<br>angle                                                                                      | No change from<br>original 510(k) |  |  |
| Implant Planning                                        | Tools for pre-surgical<br>planning of dental<br>implant placement                     | Tools for pre-surgical<br>planning of dental<br>implant placement                                                                            | No change from<br>original 510(k) |  |  |
| Stent Fabrication                                       | Tools for pre-surgical<br>fabrication of stents to<br>aid dental implant<br>insertion | Not a current feature-<br>Note: Feature  was never<br>incorporated into final<br>release of original cleared<br>ILUMA Vision, version<br>2.0 | No change from<br>original 510(k) |  |  |
| Orthodontic Treatment<br>Planning                       | Tools for planning<br>orthodontic treatment                                           | Tools for planning<br>orthodontic treatment                                                                                                  | No change from<br>original 510k   |  |  |
| Temporal Bone Module                                    | This module is used to<br>isolate and examine the<br>temporal bone                    | New feature- not in<br>original 510(k)                                                                                                       | New feature                       |  |  |
| Endoscope Module                                        | This module allows the<br>user to perform a virtual<br>endoscopy                      | New feature - not in<br>original 510(k)                                                                                                      | New feature                       |  |  |

.

:

:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three parallel lines instead of a single staff.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring. MD 20993-0002

3M IMTEC % Ms. Ginger Cantor, RAC Regulatory Affairs Specialist 3M ESPE Dental Products 3M Center, Building 275-2W-08 ST. PAUL MN 55144

Re: K094064

APR 30 2010

Trade/Device Name: ILUMA Vision (Version 2.2) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 29, 2009 Received: December 31, 2009

Dear Ms. Cantor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of

{6}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

tmill

Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

# Indications for Use

510(k) Number (if known): K094064

Device Name: ILUMA Vision (Version 2.2)

## Indications for Use:

ILUMAVision is a software application used for the display and 3D visualization of medical image files from scanning devices, such as CT, MRI, PET or 3D Ultrasound.

It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, review, store, print, and distribute DICOM 3.0 compliant images, utilizing standard PC hardware.

Additionally, ILUMA Vision is a preoperative software application used for the simulation and evaluation of dental implants, orthodontic planning and surgical treatments.

ILUMAVision is not intended for use with mammography.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Tobe B. Becker
(Division Sign-Off)

(Division Sign 7)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

510K K094064

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K094064](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K094064)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
