K093234 · GE Medical Systems SCS · LLZ · Oct 30, 2009 · Radiology
Device Facts
Record ID
K093234
Device Name
INTEGRATED REGISTRATION
Applicant
GE Medical Systems SCS
Product Code
LLZ · Radiology
Decision Date
Oct 30, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
INTEGRATED REGISTRATION permits comparison of three-dimensional (3D) images from Computed Tomography (CT), Magnetic Resonance Imaging (MRI), Emission Tomography (PET or SPECT) and X-Ray Angiography (XA) to help physicians in diagnostic radiology or therapy planning.
Device Story
Integrated Registration is a software tool extension for GE Advantage Workstation 4.5+; functions as a multi-modality image registration and fusion application. Inputs include volumetric datasets from CT, MR, PET, SPECT, and 3D XA. Device performs 2D, 3D, and hybrid 2D/3D fusion; enables loading multiple series/modalities in a single session; provides optimized registration algorithms for modality pairs and anatomies. Output is a registered/fused image set for clinical review, saved as a DICOM Registration Object. Used by radiologists and clinicians in diagnostic or therapy planning workflows. Benefits include improved usability via automated default registration at load, manual/regional registration tools, and integration with GE Radiotherapy simulation software (Advantage Sim MD).
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical testing, including risk analysis, requirements reviews, design reviews, performance verification, and final acceptance validation.
Technological Characteristics
Software-based image registration tool running on GE Advantage Workstation 4.5+. Employs optimized registration algorithms for multi-modality image pairs (CT, MR, PET, SPECT, XA). Supports DICOM standard NEMA PS 3.1-3.18 (2008). Features include 2D/3D/hybrid fusion, automated registration at load, and manual/regional registration tools. Connectivity via DICOM networking.
Indications for Use
Indicated for physicians in diagnostic radiology or therapy planning requiring comparison of 3D volumetric images from CT, MRI, PET, SPECT, or XA modalities. No specific age or gender restrictions.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
093234
## GE Healthcare 510(k) Premarket Notification Submission
## 510(k) Summary
OC1 3 0 2009
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | October 8th, 2009 |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare, (GE Medical Systems SCS)<br>283, rue de la Minière<br>78530 Buc - FRANCE |
| Primary Contact Person: | Sophie Le Loarer<br>Regulatory Affairs Specialist<br>GE Healthcare, (GE Medical Systems SCS)<br>TEL: (33) 1 30 70 43 31 - FRANCE<br>FAX: (33) 1 30 70 41 00 - FRANCE |
| Secondary Contact Person: | Stephen G. Slavens, RAC<br>Global Regulatory Affairs Director<br>GE Healthcare, (GE Medical Systems LLC)<br>TEL: (262) 548-4992<br>FAX: (262) 548-3884 |
| Device: Trade Name: | INTEGRATED REGISTRATION |
| Common/Usual Name: | INTEGRATED REGISTRATION |
| Classification Name: | 21CFR 892.2050 Picture archiving and communications system |
| Product Code: | 90LLZ |
| Predicate Devices: | K010336: Advantage Windows CT/PET Fusion<br>K983256: Advantage Windows (CT/MR) Fusion<br>K022310: Advantage Windows X-Ray/MR Fusion |
| Device Description: | The INTEGRATED REGISTRATION tool runs on Advantage<br>Workstation 4.5 or higher versions. This product is an extension<br>to the Volume Viewer application, dedicated to the registration of<br>multi-modality images, and comparison of volumetric datasets<br>from Computed Tomography (CT), Magnetic Resonance (MR),<br>Positron Emission Tomography (PET) or Single Photon Emission<br>Computed Tomography (SPECT), and 3D X-Ray Angiography<br>(XA). Note that the INTEGRATED REGISTRATION licenses<br>control which algorithms are available and which modalities can<br>be saved. |
| Intended Use: | INTEGRATED REGISTRATION permits comparison of three-<br>dimensional (3D) images from Computed Tomography (CT),<br>Magnetic Resonance Imaging (MRI), Emission Tomography<br>(PET or SPECT) and X-Ray Angiography (XA) to help<br>physicians in diagnostic radiology or therapy planning. |
| Technology: | INTEGRATED REGISTRATION consists of optimized<br>registration algorithms for pairs of modality images,<br>configuration options as licensed by the end user, and protocols<br>for use within each configuration option.<br>INTEGRATED REGISTRATION is an optimized combination<br>of its predicate devices and employs the same fundamental<br>scientific technology as these predicate devices.<br><br>Major improvements to the device include:<br>2D, 3D and hybrid 2D/3D Fusion capabilities The ability to load multiple (more than two) series and<br>multiple modalities (CT, MR, PET, SPECT, XA) in the same<br>registration session Registration results propagation to all the supported<br>modalities Optimized algorithms for modality pairs and anatomies Default registration at load Optimized tools for manual and regional registration, Reset, center and undo/redo tools to improve the usability, Protocols and layouts dedicated to typical clinical use cases, Saving ability for the new DICOM Registration Object Easier access to GE Radiotherapy simulation software:<br>Advantage Sim MD |
| Determination of<br>Substantial Equivalence: | Summary of Non-Clinical Tests:<br>INTEGRATED REGISTRATION complies with DICOM<br>Standard NEMA PS 3.1 - 3.18(2008). The following quality<br>assurance measures were applied to the development of the<br>system:<br>Risk Analysis Requirements Reviews Design Reviews Performance testing (Verification) Safety testing (Verification) Final acceptance testing (Validation)<br>Summary of Clinical Tests:<br>The subject of this premarket submission, INTEGRATED<br>REGISTRATION, did not require clinical studies to support<br>substantial equivalence. |
| Conclusion: | GE Healthcare considers the INTEGRATED REGISTRATION<br>Software application to be as safe, as effective, and performance<br>is substantially equivalent to the predicate devices. |
Image /page/0/Picture/6 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular border. The letters and the border are in black, while the background is white. The logo has a vintage or classic appearance.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. There are three teardrop-shaped elements positioned around the circle, one at the top and two at the bottom, adding a distinctive touch to the design.
510(k) Premarket Notification Submission
·
・
.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized, curved lines.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
General Electric Medical Systems SCS % Mr. Daniel W. Lehtonen Senior Staff Engineer-Medical Devices Intertek Testing Services NA, Inc. 2307 East Aurora Road, Unit B7 TWINSBURG OH 44087
Re: K093234
OCT 3 0 2009
Trade/Device Name: Integrated Registration Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 14, 2009 Received: October 15, 2009
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 -
Enclosure
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined within a circle. The letters and the circle are black, while the background is white.
GE Healthcare 510(k) Premarket Notification Submission
510(k) Number (if known): K093234
Device Name: INTEGRATED REGISTRATION
Indications for Use:
INTEGRATED REGISTRATION provides easy means for comparison of three-dimensional (3D) images from Computed Tomography (CT), Magnetic Resonance Imaging (MRI), Emission Tomography (PET or SPECT) and X-Ray Angiography (XA). To help physicians in diagnostic radiology or therapy planning, INTEGRATED REGISTRATION allows 3D registration between volumetric acquisitions that may come from the same acquisition modality or from different acquisition modalities.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Torya M. Whiz
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number K093234
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.