NORDICBRAINEX SOFTWARE, VERSION 1.0

K092102 · Nordicimaginglab AS · LLZ · Aug 25, 2009 · Radiology

Device Facts

Record IDK092102
Device NameNORDICBRAINEX SOFTWARE, VERSION 1.0
ApplicantNordicimaginglab AS
Product CodeLLZ · Radiology
Decision DateAug 25, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

nordicBrainEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. nordicBrainEx provides both viewing and analysis capabilities of functional and dynamic imaging datasets acquired with MRI or other relevant modalities, including blood oxygen level dependent (BOLD) fMRI, diffusion weighted MRI (DWI) / fiber tracking and dynamic analysis. BOLD fMRI: BOLD analysis is used to highlight small magnetic susceptibility changes in the human brain in areas with altered blood-flow resulting from neuronal activity. DWI/Fiber Tracking: Diffusion analysis is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data. Fiber tracking utilize the directional dependency of the diffusion to display the white matter structure in the brain. Dynamic Analysis: Dynamic analysis is used for visualization and analysis of dynamic imaging data of the brain, showing properties of changes in contrast over time where such techniques are useful or necessary.

Device Story

Post-processing software for functional MRI data; runs on standard Windows PC workstation. Inputs: DICOM-compliant imaging data (BOLD fMRI, DWI, dynamic imaging). Processing: Analyzes BOLD signals for blood-flow changes; calculates water diffusion properties for fiber tracking; performs dynamic contrast analysis over time. Outputs: Visualizations and reports for clinical review. Used in clinical environments by physicians and technicians to assist in evaluating brain structure and function. Benefits: Enables efficient clinical workflow for complex neuroimaging analysis; aids in visualizing white matter structure and neuronal activity-related blood flow changes.

Clinical Evidence

Bench testing only. Verification and validation activities were performed to ensure the software meets design specifications and user needs.

Technological Characteristics

Software-based image processing application; runs on standard off-the-shelf Windows PC workstation; DICOM-compliant; supports BOLD fMRI, DWI/fiber tracking, and dynamic imaging analysis.

Indications for Use

Indicated for use by trained professionals (physicians, medical technicians) to view, process, and analyze medical images (MRI/other modalities) for BOLD fMRI, DWI/fiber tracking, and dynamic brain imaging analysis.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K092102 Page 1 of 2 ## 510(k) Summary NordicImagingLab AS nordicBrainEx Software AUG 2 5 2009 Submitter: NordicImagingLab AS Møllendalsveien 65C N-5009 Bergen Norway Phone: +47 55 70 70 95 +47 55 70 70 96 Fax: Primary Contact: Chandana Gurung Bhandari (Chandana@nordicimaginglab.com) | Proprietary Name: | nordicBrainEx Software | |----------------------------|--------------------------------------------| | Device Common Name: | PACS | | Device: | System, image processing, radiological | | Classification Name: | Picture archiving and communication system | | Classification Regulation: | 892.2050 | | Class: | II | | Panel: | Radiology | | Product Code: | LLZ | Predicate device name: Nordic Image Control and Evaluation (nordicICE) Software, K090546 ### Device Description The nordicBrainEx·Software is a post-processing application for functional MRI data developed with a view to ease of use and high performance on a standard Windows workstation. The software provides comprehensive functionality for processing and analyzing BOLD fMRI and DTI data, in addition to efficient and easy to use tools for visualization and report generation. The application workflow has been optimized to ensure efficiency and high throughput in a clinical environment. #### Intended Use nordicBrainEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. 510(k) Summary Page 5-2 {1}------------------------------------------------ nordicBrainEx provides both viewing and analysis capabilities of functional and dynamic imaging datasets acquired with MRI or other relevant modalities, including blood oxygen level dependent (BOLD) fMRI, diffusion weighted MRI (DWI) / fiber tracking and dynamic analysis. BOLD fMRI: BOLD analysis is used to highlight small magnetic susceptibility changes in the human brain in areas with altered blood-flow resulting from neuronal activity. DWI/Fiber Tracking: Diffusion analysis is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data. Fiber tracking utilize the directional dependency of the diffusion to display the white matter structure in the brain. Dynamic Analysis: Dynamic analysis is used for visualization and analysis of dynamic imaging data of the brain, showing properties of changes in contrast over time where such techniques are useful or necessary. #### Technological Characteristics and Substantial Equivalence The nordicBrainEx Software is substantially equivalent to the nordicICE Software (K090546) in intended use, indications for use, technological characteristics and operational characteristics. ### Performance Testing Prospectively defined verification and validation activities for the nordicBrainEx Software assure that the nordicBrainEx Software is substantially equivalent to the cleared nordicICE Software and meets design and performance specifications as well as user needs when operated according to the operating instructions. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles a stylized caduceus or a representation of human figures. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Ms. Chandana G. Bhandari Quality Manager NordicImagingLab AS Møllendalsveien 65C Bergen Norway, 5009 AUG 2 5 2009 Re: K092102 Trade/Device Name: nordicBrainEx Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 10, 2009 Received: July 14, 2009 Dear Ms. Bhandari: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical {3}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrth/indr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use KO92102 510(k) Number (if known): Device Name: nordicBrainEx Software Indications for Use: nordicBrainEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities. nordicBrainEx provides both viewing and analysis capabilities of functional and dynamic imaging datasets acquired with MRI or other relevant modalities, including blood oxygen level dependent (BOLD) fMRI, diffusion weighted MRI (DWI) / fiber tracking and dynamic analysis. BOLD fMRI: BOLD analysis is used to highlight small magnetic susceptibility changes in the human brain in areas with altered blood-flow resulting from neuronal activity. DWI/Fiber Tracking: Diffusion analysis is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data. Fiber tracking utilize the directional dependency of the diffusion to display the white matter structure in the brain. Dynamic Analysis: Dynamic analysis is used for visualization and analysis of dynamic imaging data of the brain, showing properties of changes in contrast over time where such techniques are useful or necessary. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices | NordicImagingLab AS – Traditional 510(k) nordicBrainEx Software | | |-----------------------------------------------------------------|------------------------------| | 510(k) Number | K092102 | | | Indications for Use Page 4-2 |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...