DEMASQ IMAGING SPFTWARE DEVICE

K090481 · Demasq Limited · LLZ · Mar 19, 2009 · Radiology

Device Facts

Record IDK090481
Device NameDEMASQ IMAGING SPFTWARE DEVICE
ApplicantDemasq Limited
Product CodeLLZ · Radiology
Decision DateMar 19, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device, 3rd-Party Reviewed

Indications for Use

DEMASQ Imaging Software Device™ is a software device that receives medical images and data from various imaging sources (including but not limited to CT, MR, US, RF units, computed and direct radiographic devices, and secondary capture devices, (scanners, imaging gateways or imaging sources). Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations. Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography. Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Typical users of this system are trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants.

Device Story

DEMASQ Imaging Software Device™ (DISD) is a PACS DICOM viewing software for IBM-compatible PCs (Windows OS). It receives medical images from various modalities (CT, MR, US, RF, radiography); stores, processes, and routes data via network. Users (radiologists, clinicians) utilize a GUI to manipulate images (color, crop, zoom, rotate, measure). Output is displayed on monitors for diagnostic interpretation. Supports DICOM 3.0 and TCP/IP. Facilitates remote diagnosis via telecommunications networks. Benefits include replacement of traditional film handling, improved workflow, and access to diagnostic data for clinical decision-making.

Clinical Evidence

No clinical data provided. Bench testing only.

Technological Characteristics

Software-only PACS; runs on Windows (98 SE, NT 4.0, XP, Vista). Connectivity via Ethernet (TCP/IP). Supports DICOM 3.0 standard. Features include image manipulation (contrast, magnification, measurement), remote transmission, and online help. Does not contact patient; no life-sustaining function.

Indications for Use

Indicated for trained professionals (physicians, radiologists, nurses, technicians) to receive, store, communicate, process, and display medical images (CT, MR, US, RF, radiographic) for diagnostic review. Mammography requires pre-processed DICOM for presentation images and 5 Mpixel FDA-approved monitors; lossy compressed or digitized film-screen mammography images are contraindicated for primary diagnosis.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ 510(k) Summary of Safety and Effectiveness This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990. Date Prepared: November 20, 2008 Submitter's Information: 21 CFR 807.92(a)(1) Mr. Stuart Gall, CEO DEMASQ Limited 8th Floor Eastgate House 35-43 Newport Road Cardiff, Wales CF240AB United Kingdom +44 114 275555 Tel: +44 292 0491275 Fax: Email: stuartgall(@demasq.com Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2) | Trade Name: | DEMASQ Imaging Software Device™ | |------------------------|-----------------------------------------| | Common Name: | Picture Archiving Communications System | | Classification Name: | system, image processing, radiological | | Product code: | LLZ | | Device Classification: | 892.2050 | Predicate Device: 21 CFR 807. 92(a)(3) | 510(k) Number | K052545 | K071894 | |----------------------------|-------------------------------------------|-------------------------------------------| | Device Classification Name | system, image processing,<br>radiological | system, image processing,<br>radiological | | Regulation Number | 892.2050 | 892.2050 | | Device Name | INFINITT G3 PACS | XEBRA DICOM IMAGE<br>BROWSER | | Applicant | INFINITT CO., LTD. | HX TECHNOLOGIES, INC. | | Product Code | LLZ | LLZ | | Decision Date | 11/08/2005 | 08/16/2007 | Device Description: 21 CFR 807 92(a)(4) DEMASQ Imaging Software Device™ (DISD) is a user network and or a web-based PACS DICOM viewing software running on an IBM Compatible PC with a Windows 98 SE, Windows NT 4.0 (with SP 5 or later), XP or Vista operating system. DISD is adapted for, storing, processing, routing and report generating. DISD supports the DICOM standard. To assist interpretation a user-friendly interface will allow the enduser to color, crop, zoom, rotate, measure and save the image. Images can be saved to a desktop or network as a standard file format. DISD allows users to take full advantage of the radiographic images from digital X-ray and various modalities in order to obtain invaluable mission critical diagnostic data and images. l {2}------------------------------------------------ ### 510(k) Summary of Safety and Effectiveness #### Indications for Use: 21 CFR 807 92(a)(5) DEMASQ Imaging Software Device™ is a software device that receives medical images and data from various imaging sources (including but not limited to CT, MR, US, RF units, computed and direct radiographic devices, and secondary capture devices, (scanners, imaging gateways or imaging sources). Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations. Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography. Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Typical users of this system are trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants. #### Technological Characteristics: 21 CFR 807 92(a)(6) DEMASQ Imaging Software Device™ is a software device that does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed. DEMASQ Imaging Software Device ™ and its predicates have essentially the same technological characteristics. The predicate devices and the new device are software PACS for use in medical image acquisition, management, manipulation, and distribution. All systems have been developed to replace traditional film handling in radiology. The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use. The predicate device and the new device include the following similar features and functionality: - Advanced reviewing features as contrast adjustments, magnification and more. . - Evaluation of general parameters such as pixel values and measurements of . distances and angles in the images. - Possibility of remote diagnoses through transmission via public or private . telecommunications network (including the Internet) of an examination. - Image data acquisition based on Ethernet communication standard with TCP/IP file . transfer protocol and DICOM 3.0 radiology image file standards - A user interface using Microsoft Windows operating systems. . - Internal handling of images in DICOM 3.0 format. . - On-line help system t #### Conclusion: 21 CFR 807 92(b)(1) The 510(k) Pre-Market Notification for DEMASQ Imaging Software Device™ contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device. The subject device has been designed and will be delivered in accordance with the voluntary standards. {3}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three lines forming its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service # MAR 1 9 2009 ### Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEMASQ Limited % Mr. Jeff D. Rongero Senior Project Engineer Underwriters Laboratories, Inc. 12 Laboratory Drive Research Triangle Park, NC 27709 Re: K090481 Trade/Device Name: DEMASQ Imaging Software Device™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 20, 2009 Received: February 24, 2009 ### Dear Mr. Rongero: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 | |----------------|----------------------------------|----------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 | | 21 CFR 892.xxx | (Radiology) | (240) 276-0120 | | Other | | (240) 276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division-of-Surveillance-Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppor/index.html. Sincerely yours, Laura M. Morris anine M. Morris Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ ### Indications for Use 510(k) Number (if known): K09048 Device Name: DEMASQ Imaging Software Device™ Indications for Use: DEMASQ Imaging Software Device™ is a software device that receives medical images and data from various imaqing sources (including but not limited to CT. MR. US, RF units, computed and direct radiographic devices, and secondary capture devices, (scanners, imaging gateways or imaging sources). Images and data can be stored. communicated, processed and displayed within the system or across computer networks at distributed locations. Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography. Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Typical users of this system are trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED). Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Division Sign-Off Office of In Vitro Øiagnostic Device Evaluation and Safety 510(k) Hedy Lamarr (Division Sign-Off) Division of Reproductive. Abdominal Radiological Devices 510(k) Number Page 1 of 1 Updated June 4, 2007
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...