K083740 · Philips Medical Systems Nederland B.V. · LLZ · Feb 13, 2009 · Radiology
Device Facts
Record ID
K083740
Device Name
MAMMODIAGNOST VU
Applicant
Philips Medical Systems Nederland B.V.
Product Code
LLZ · Radiology
Decision Date
Feb 13, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
Philips MammoDiagnost VU workstation software package is intended for viewing, manipulation, reporting and communication of digital mammography images (DICOM "For Presentation" images) as well as other modality images. It interfaces to mammography acquisition stations and image storage, printing and CAD devices using DICOM or similar interface standards. Philips MammoDiagnost VU is used with high-resolution monitors suitable for screening and diagnostic mammography, for use in the USA these monitors must be FDA cleared. It is used by trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians and assistants.
Device Story
Software workstation for digital mammography; inputs DICOM 'For Presentation' images from acquisition stations; provides tools for viewing, manipulation, reporting, and communication. Features include DICOM support, generic and mammography hanging protocols, CAD overlays, double-blind read workflow, automatic breast tissue alignment, and background suppression at inversion. Operates on standard IT hardware using Microsoft OS; interfaces with image storage, printing, and CAD devices via TCP/IP, DICOM, and HL7. Used by radiologists and clinicians in clinical settings to facilitate diagnostic and screening mammography interpretation. Output enables clinical decision-making through enhanced visualization and workflow management.
Clinical Evidence
No clinical data. Bench testing only; device validated under ISO 13485 and 21 CFR Part 820 compliant Quality System. Compliance with NEMA XR 22-2006 and NEMA PS 3.2 DICOM standards confirmed.
Technological Characteristics
Software-based PACS workstation; runs on standard off-the-shelf IT hardware/Microsoft OS. Connectivity via TCP/IP, DICOM, and HL7. Features include automatic breast tissue alignment, CAD overlays, and background suppression. Complies with NEMA XR 22-2006 and NEMA PS 3.2. Requires FDA-cleared 5 Mpixel resolution monitors for mammography.
Indications for Use
Indicated for viewing, manipulation, reporting, and communication of digital mammography (DICOM "For Presentation") and other modality images by trained professionals including physicians, radiologists, nurses, and technicians. In the USA, requires FDA-cleared monitors with at least 5 Mpixel resolution for mammographic interpretation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
이는 데 그 이번 말에서 그 나 그 부르는 대전 남 4 1. 1978 19. 3 31 ' 1 - 3 : 4 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : 1 : .. . . .
## Philips Healthcare
גמרים בין או יוצר מאי 10 אינטער שלו במשפחת Construction Construction and Comparis (Spanding 3 2001 1000
# 510(k) Summary
・・・・・・
Philips MammoDiagnost VU------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
විකාශය සිංහ මහතා විශ්වාදය විසින් පවත්වය පැවැත්වයි. ආශ්රිත
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. Sec. Saland . Sale and Controllar
| General Information | Bath : 1 : 1 : 143 : 13 : 26 : 26 : 26 : 26 : 26 : 26 : 26 : 20 : 26 : 26 : 20 : 20 : 26 : 20 : 20 : 20 : 20 : 20 : 20 : 20 : 20 : 20 : 20 : 20 : 20 : 20 : 20 : 20 : 20 : 20<br>2017/08/2018 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Name: | Philips Medical Systems North America Company<br>. |
| Address: | 22100 Bothell Everett Highway<br>Bothell Washington 98021-8431<br>USA 39 000 000 000 000 000 000 000 0000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000<br>1 M C = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = |
| Contact Person | Lynn. T. Harmer |
| Telephone Number: | itaa da persona personalise de la primer de la r<br>4254478-7312666 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 12 - 1 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12<br>的新的公司在位于一次的一个成功的主要的一个问题。<br>1978 - 1970 - 1991 - 1991 - 1991 - 19 |
| Prepared (date): | 2008 November 14 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1<br>1993 11 11 165 11 |
| Manufacturing Site: | Philips Medical Systems Nederland B.V. J . J<br>Veenpluis 4-6 2000 2 2 3 4 2 4 4 4 4 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1<br>5684 PC Best (1) 300 Best Bearing Beach (1946) 1<br>The Netherlands of the many of the comment of the free |
| Device Name: | Philips MammoDiagnost VU - 11 - 11 - 11 - 11 - 11 - 11 - 11 -<br>1000 133 133 133 13 11 11 11 11 13 |
| Classification Name: | Picture Archiving and Communication System<br>WE CLES IS OF IS IN DOS COME OF |
| Regulation number: | 11 VOIDU. 1191051. |
| Classification: | mon<br>r barberting Rei<br>: 1975年: 1979年 1998年 1998 |
| ProCode: | 1:11 - 1. 1. 1. 1. 1.<br>사람들 전 행<br>: 上一篇: 家 |
| Common/Usual Name: | PACS Workstation 1 1 1 1 1 1 2 0 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -<br>objection in other than the more and captive and portugal por |
| Predicate Devices: | ೆ ಇವರ ಇದೆ.<br>Stentor Inc. (now Philips Healthcare Inc.), Arento<br>iSite PACS v 4.x 1 1 1 1 1 1 1 1 1 1 1 1 1 1 |
| | General Electric Medical Systems of Childer Systems and<br>Seno-Advantage of Santa Childer of Children Charless Pick,<br>Hologio Inc. Spino Led oct 100 and the formatio<br>iləyə dəmə, mətbə y<br><br>SecurView DX |
| | း မို့အိပ္စားရွိပုံ<br>ﺪ ﺍﻟﻤﻨﺎﺳﺒﺔ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﺘﻲ ﺗﻌﻠﻴﻘﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﺘﻲ ﺗﻌﺘﺒﺮ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﺘﻲ ﺗﻌﺘﺒﺮ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ |
| | ਪ੍ਰਾਫ਼ ਕੀਕੀ ਨਾਮਾਕੀ ਮੰਗਰ<br>能 |
{1}------------------------------------------------
.
#### Information Supporting Substantial Equivalence Determination II
#### System Description:
Philips MammoDiagnost VU workstation software package is intended for viewing, manipulation, reporting and communication of digital mammography images (DICOM "For Presentation" images) as well as other modality images. It interfaces to mammography acquisition stations and image storage, printing and CAD devices using DICOM or similar interface standards. It uses and builds further on the Stentor iSite PACS v4.x software (Stentor is now a Philips owned company).
is now a Philips owned company).
is now a Philips owned company).
Philips MammoDiamore VII includes and Sun Marc
Philips MammoDiagnost VU workstation software package runs on standard information technology hardware and software. The proprietary software in the Philips MammoDisonost VU workstation software package uses a standard Microsoft Operating System and user interface. Communication and data exchange are done using standard TCP/IP, DICOM and HL7 protocols.
#### Intended Use:
Philips MammoDiagnost VU workstation software package is intended for viewing, manipulation, reporting and communication of digital mammography images (DICOM "For Presentation" images) as well as other modality images. It interfaces to mammography acquisition stations and image storage, printing and CAD devices using DICOM or similar interface standards.
Philips MammoDiagnost VU is used with high-resolution monitors suitable for screening and diagnostic mammography, for use in the USA these monitors must be FDA cleared. It is used by trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians and assistants.
#### General Safety and Effectiveness:
The MammoDiagnost VU software is specified, validated and tested under a registered ISO 13485 and 21 CFR Part 820 compliant Quality System.
The MammoDiagnost VU complies with the NEMA XR 22-2006, ("Quality Control Manual" Template for Manufactures of Displays and Workstations Labeled for Final Interpretation in Full-field Digital Mammography) and to the NEMA PS 3.2 DICOM set
The device labelling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device. It is the user's responsibility to ensure that display quality, environmental lighting and other possible distractions are consistent with the clinical environment. lunakes (1917) .
The hardware components specified are all "off the shelf" computer components.
#### Conclusion:
The Philips MammoDiagnost VU does not introduce new indications for use, nor does the use of the device result in any new potential hazard. Philips Medical Systems considers the Philips MammoDiagnost VU to be substantially equivalent to the Stentor (Philips) iSite PACS v 4.x (K063267), the GE, Seno Advantage (K033400) and the Hologic, Secur View DX (K062107)
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird with three curved lines representing its wings.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Philips Medical Systems Nederland BV % Ms. Melissa J. DeGuia Senior Project Engineer Underwriters Laboratories, Inc. 2600 NW Lake Road CAMAS WA 98607
Re: K083740
Trade/Device Name: Philips MammoDiagnost VU Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 2, 2009 Received: February 3, 2009
#### Dear Ms. DeGuia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
FEB 1 3 2009
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean
that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.
Sincerely yours,
Janine M. Morris
anine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
### ications for
510(K) Number (if known): K083740
Device Name:
Philips MammoDiagnost VU
1 2019 30 33
्
#### Indications for Use:
Philips MammoDiagnost VU is a software workstation for viewing, manipulati reporting and communication of digital mammography images (DICOM "For Presentation" images) and other modality images. It interfaces to image storage, printing and CAD devices using DICOM or similar interface standards
The functional components include
- · DICOM support
- Generic hanging protocols
- · Mammography hanging protoce
- CAD overlays
- Double blind read workflow
- Automatic breast tissue alignment
- Background suppression at inversion
Philips MammoDiagnost VU software nuns on standard information technology hardware and software. The proprietary software in the MammoDiagnost VU product uses a standard Microsoft Operating System and user interface. Communication and data exchange are done using standard TCPMP, DICQMand HL7 protocols
14-401-02-2015 11 13:15:53:33 In the USA, mammographic images may only be interpreted using FDA cleared monitors that offer at least 5 Mpixel resolution and meet other technical specifications reviewed and accepted by FDA.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) िस्तारिए 14 (Division (Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number