K082628 · Circle Cardiovascular Imaging, Inc. · LLZ · Nov 20, 2008 · Radiology
Device Facts
Record ID
K082628
Device Name
CMR42 CARDIAC MRI IMAGING SYSTEM, 2.1.23
Applicant
Circle Cardiovascular Imaging, Inc.
Product Code
LLZ · Radiology
Decision Date
Nov 20, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
cmr42 is intended to be used for viewing, post-processing and quantitative evaluation of cardiovascular magnetic resonance (MR) images in a Digital Imaging and Communications in Medicine (DICOM) Standard format. It enables: Importing Cardiac MR Images in DICOM format Supporting clinical diagnostics by qualitative analysis of the cardiac MR images using display functionality such as panning, windowing, zooming, navigation through series/slices and phases. Supporting clinical diagnostics by quantitative measurement of the heart and adjacent vessels in cardiac MR images, specifically distance, area, volume and mass Supporting clinical diagnostics by using area and volume measurements for measuring LV function and derived parameters cardiac output and cardiac index in long axis and short axis cardiac MR images. Flow quantifications based on velocity encodes images It shall be used by qualified medical professionals, experienced in examining and evaluating cardiovascular MR images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. cmr42 is a software application that can be used as a stand-alone product or in a networked environment. The target population for the cmr42 is not restricted, however the image acquisition by a cardiac magnetic resonance scanner may limit the use of the device for certain sectors of the general public. cmr42 shall not be used to view or analyze images of any part of the body except the cardiac magnetic resonance images acquired from a cardiovascular magnetic resonance scanner.
Device Story
cmr42 is a software application for qualitative and quantitative analysis of cardiovascular MR images in DICOM format. Input consists of cardiac MR images acquired from a scanner. The software provides tools for image display (panning, windowing, zooming, navigation) and quantitative measurements (distance, area, volume, mass, LV function, cardiac output, cardiac index, and flow quantification from velocity-encoded images). Used in hospital or private practice environments by qualified medical professionals. Output is visualized on a workstation, assisting clinicians in diagnostic decision-making. Benefits include objective, reproducible analysis of cardiac function and anatomy.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness were demonstrated through non-clinical testing according to a Master Software Test Plan.
Technological Characteristics
Software-only device; runs on MacOS and Microsoft Windows. Supports DICOM networking. Provides task-specific modules for qualitative/quantitative analysis of cardiac MR images. No hardware components.
Indications for Use
Indicated for use by qualified medical professionals to view, post-process, and quantitatively evaluate cardiovascular MR images. Target population is unrestricted, though limited by patient suitability for cardiac MR scanning. Contraindicated for non-cardiac MR images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
MRI-MAGNETIC RESONANCE ANALYTICAL SOFTWARE SYSTEM (MASS) (K994283)
{1}------------------------------------------------
-
# 510(k) Summary (REVISED) - K082628 cmr42 Cardiac MR Software Application
| Submitter's Name | Circle Cardiovascular Imaging Inc. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | Suite 130, 3553 31 Street NW, Calgary, AB, Canada T2L 2K7 |
| Establishment<br>Registration Number | Not available |
| Date of Summary | November 05, 2008 |
| Telephone Number | 1 403 775 1857 |
| Fax Number | 1 403 270 2384 |
| Email | shirantha@circlecvi.com |
| Contact Person | Shirantha Samarappuli |
| Name of the Device | cmr42 |
| Common or Usual Name | Image Processing System |
| Classification Name | Classification Name: Picture Archiving and Communications<br>System<br>Device Class: II<br>Product Code: LLZ<br>Regulation Number: 21 CFR 892.2050 |
| Indications for Use | cmr42 is intended to be used for viewing, post-processing and<br>quantitative evaluation of cardiovascular magnetic resonance<br>(MR) images in a Digital Imaging and Communications in<br>Medicine (DICOM) Standard format.<br>It enables:<br>Importing Cardiac MR Images in DICOM format Supporting clinical diagnostics by qualitative analysis of the cardiac MR images using display functionality such as panning, windowing, zooming, navigation through series/slices and phases. Supporting clinical diagnostics by quantitative measurement of the heart and adjacent vessels in cardiac MR images, specifically distance, area, volume and mass Supporting clinical diagnostics by using area and volume measurements for measuring LV function and |
| Document No. | N/A |
| Rev | 00 |
| File name: | cmr42 Traditional 510k Submission - 510k Summary |
| Sheet | Page 1 of 7 |
| This document contains information, which is the property of CIRCLE Cardiovascular Imaging, Inc. This<br>document may not, in whole or in part, be duplicated, disclosed, or used for design or manufacturing purposes<br>without the prior written permission of CIRCLE Cardiovascular Imaging, Inc. | |
{2}------------------------------------------------
cmr42 Traditional 510(k) K082628
derived parameters cardiac output and cardiac index in long axis and short axis cardiac MR images.
• Flow quantifications based on velocity encodes images
It shall be used by qualified medical professionals, experienced in examining and evaluating cardiovascular MR images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. cmr<sup>42</sup> is a software application that can be used as a stand-alone product or in a networked environment.
The target population for the cmr<sup>42</sup> is not restricted, however the image acquisition by a cardiac magnetic resonance scanner may limit the use of the device for certain sectors of the general public.
cmr<sup>42</sup> shall not be used to view or analyze images of any part of the body except the cardiac magnetic resonance images acquired from a cardiovascular magnetic resonance scanner.
| Identification of the Legally Marketed Device (Predicate Device) | MRI-MAGNETIC RESONANCE ANALYTICAL SOFTWARE SYSTEM (MASS) |
|------------------------------------------------------------------|----------------------------------------------------------|
| Classification Name: | System, Nuclear Magnetic Resonance Imaging |
| Device Class: | II |
| Product Code: | LNH |
| Regulation Number: | 21 CFR 892.1000 |
| 510k #: | K994283 |
| | MRI-Flow Analytical Software (FLOW) |
|----------------------|--------------------------------------------|
| Classification Name: | System, Nuclear Magnetic Resonance Imaging |
| Device Class: | II |
| Product Code: | LNH |
| Regulation Number: | 21 CFR 892.1000 |
| 510K #: | K994282 |
#### Device Description
cmr42 is a dedicated software application for cvaluating cardiovascular images in a DICOM Standard format. The software
| Document No. | Rev | File name: | Sheet |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|-----------------------------------|-------------|
| N/A | 00 | cmr42 Traditional 510k Submission | Page 2 of 7 |
| This document contains information, which is the property of CIRCLE Cardiovascular Imaging, Inc. This<br>document may not, in whole or in part, be duplicated, disclosed, or used for design or manufacturing purposes<br>without the prior written permission of CIRCLE Cardiovascular Imaging, Inc. | | | |
{3}------------------------------------------------
CIRCLE Cardiovascular Imaging Inc.
can be used as a stand-alone product that can be integrated into a hospital or private practice environment. cmr42 has a graphical user interface which allows users to qualitatively and quantitatively analyze cardiac images for volume/mass, and flow quantification. It provides a comprehensive set of tools for the analysis of Cardiovascular Magnetic Resonance (CMR) images.
| Document No. | Rev | File name: | Sheet |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|-----------------------------------|-------------|
| N/A | 00 | cmr42 Traditional 510k Submission | Page 3 of 7 |
| This document contains information, which is the property of CIRCLE Cardiovascular Imaging, Inc. This<br>document may not, in whole or in part, be duplicated, disclosed, or used for design or manufacturing purposes<br>without the prior written permission of CIRCLE Cardiovascular Imaging, Inc. | | | |
{4}------------------------------------------------
# 510(k) SUMMARY, continued
## Indications for Use Comparison
| DEVICE | INDICATIONS FOR USE |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| cmr42 CARDIAC MR<br>SOFTWARE<br>K082628 | cmr42 is intended to be used for viewing, post-processing and<br>quantitative evaluation of cardiovascular magnetic resonance<br>(MR) images in a Digital Imaging and Communications in<br>Medicine (DICOM) Standard format.<br>It enables; |
| | • Importing Cardiac MR Images in DICOM format |
| | • Supporting clinical diagnostics by qualitative analysis<br>of the cardiac MR images using display functionality<br>such as panning, windowing, zooming, navigation<br>through series/slices and phases. |
| | • Supporting clinical diagnostics by quantitative<br>measurement of the heart and adjacent vessels in<br>cardiac MR images, specifically distance, area,<br>volume and mass |
| | • Supporting clinical diagnostics by using area and<br>volume measurements for measuring LV function and<br>derived parameters cardiac output and cardiac index in<br>long axis and short axis cardiac MR images. |
| | • Flow quantifications based on velocity encodes<br>images |
| | It shall be used by qualified medical professionals,<br>experienced in examining and evaluating cardiovascular MR<br>images, for the purpose of obtaining diagnostic information as<br>part of a comprehensive diagnostic decision-making process.<br>cmr42 is a software application that can be used as a stand-<br>alone product or in a networked environment. |
| | The target population for the cmr42 is not restricted, however<br>the image acquisition by a cardiac magnetic resonance<br>scanner may limit the use of the device for certain sectors of<br>the general public. |
| | cmr42 shall not be used to view or analyze images of any part<br>of the body except the cardiac magnetic resonance images<br>acquired from a cardiovascular magnetic resonance scanner. |
| Document No. | Rev | File name: | Sheet |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|-----------------------------------|-------------|
| N/A | 00 | cmr42 Traditional 510k Submission | Page 4 of 7 |
| This document contains information, which is the property of CIRCLE Cardiovascular Imaging, Inc. This<br>document may not, in whole or in part, be duplicated, disclosed, or used for design or manufacturing purposes<br>without the prior written permission of CIRCLE Cardiovascular Imaging, Inc. | | | |
{5}------------------------------------------------
:
| DEVICE | INDICATIONS FOR USE |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MRI-MAGNETIC<br>RESONANCE<br>ANALYTICAL<br>SOFTWARE<br>SYSTEM<br>(MASS)<br>K994283 | MASS, including its option, has been developed for the<br>objective and reproducible analysis of multi-slice, multi-phase<br>left and right ventricular function from cardiac MR data sets.<br>The software enables the display of images for use by trained<br>medical personnel.<br>Intended purposes are:<br>1. Supporting clinical diagnoses about the status of the<br>global and regional function and anatomy of the cardiac<br>chambers;<br>2. Supporting the subsequent clinical decision making<br>processes;<br>3. Supporting the use in clinical research trials, directed at<br>studying changes in function and anatomy of the heart<br>chambers as a result of interventions; |
| MRI-FLOW<br>ANALYTICAL<br>SOFTWARE<br>K994282 | Flow has been developed for the objective and reproducible<br>analysis of velocity-encoded cine MR imaging studies of<br>arterial vessels and heart valves. Intended purposes are:<br>1. Supporting clinical diagnoses about the status of the<br>function of the cardiac chambers;<br>2. Supporting clinical diagnoses about the status of the flow<br>velocity and volume flow through cardiac and peripheral<br>vessels, both under basal and increased flow conditions;<br>3. Supporting subsequent clinical decision making purposes;<br>4. Supporting the use in clinical research trials, directed at<br>studying changes in function of the heart chambers and in<br>the flow through cardiac and peripheral vessels as a result<br>of interventions. |
| Document No. | N/A |
|--------------|-----------------------------------------------|
| Rev | 00 |
| File name: | cmr <sup>42</sup> Traditional 510k Submission |
| Sheet | Page 5 of 7 |
This document contains information, which is the property of CIRCLE Cardiovascular Imaging, Inc. This document may not, in whole or in part, be duplicated, disclosed, or used for design or manufacturing purposes without the prior written permission of CIRCLE Cardiovascular Imaging, Inc.
{6}------------------------------------------------
# 510(k) SUMMARY, continued
Device Comparison Table
| | Submission<br>cmr42<br>Cardiac MR<br>Software<br>K082628 | Predicate<br>MRI-MAGNETIC<br>RESONANCE<br>ANALYTICAL<br>SOFTWARE<br>SYSTEM (MASS)<br>K994283 | Predicate<br>MRI-Flow Analytical<br>Software<br>K994282 |
|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Images from all MRI scanner<br>vendors supported | X | X | X |
| Workstation operating system | MacOS,<br>Microsoft<br>Windows | Microsoft Windows,<br>Unix, Linux | Microsoft Windows,<br>Unix, Linux |
| Import and display magnetic<br>resonance images | X | X | X |
| DICOM compliant networking | X | X | X |
| Images can be displayed by<br>study and series | X | | |
| Store images | X | X | X |
| Quantitative assessment of<br>cardiac function | X | X | X |
| Task specific modules with<br>corresponding tool sets | X | | |
| Analysis of velocity-encoded<br>images | X | | X |
| Dynamic display of ventricular<br>contractions | X | X | |
| Reports containing<br>visualization of images and<br>quantitative parameters | X | X | X |
| Analyzes long and short-axis<br>views of the heart for<br>quantitative assessment of<br>cardiac function | X | X | |
| Document No. | Rev | File name: | Sheet | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|-----------------------------------|-------------|--|
| N/A | 00 | cmr42 Traditional 510k Submission | Page 6 of 7 | |
| This document contains information, which is the property of CIRCLE Cardiovascular Imaging, Inc. This<br>document may not, in whole or in part, be duplicated, disclosed, or used for design or manufacturing purposes<br>without the prior written permission of CIRCLE Cardiovascular Imaging, Inc. | | | | |
{7}------------------------------------------------
CIRCLE Cardiovascular Imaging Inc.
#### 510(k) SUMMARY, continued
Description and Testing: cmr42 have been tested according to the specifications that are Conclusion of Testing documented in a Master Software Test Plan. Testing is an integral part of Circle Cardiovascular Imaging Inc software development process as described in the company's product development process.
#### Conclusion:
The successful non-clinical testing demonstrates the safety and effectiveness of the cm42 when used for the defined indications for use and demonstrates that the device for which the 510(k) is submitted performs as well as or better than the legally marketed predicate device.
| Document No. | Rev | File name: | Sheet |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|-----------------------------------|-------------|
| N/A | 00 | cmr42 Traditional 510k Submission | Page 7 of 7 |
| This document contains information, which is the property of CIRCLE Cardiovascular Imaging, Inc. This<br>document may not, in whole or in part, be duplicated, disclosed, or used for design or manufacturing purposes<br>without the prior written permission of CIRCLE Cardiovascular Imaging, Inc. | | | |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a symbol. The symbol is a stylized representation of three overlapping human figures, suggesting collaboration and support. The seal is in black and white.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Shirantha Samarappuli Director-Regulatory Affairs & Quality Assurance Circle Cardiovascular Imaging, Inc. Suite 130, 31 Street NW Calgary, Alberta, T2L 2K7 CANADA
NOV 2 0 2008
Re: K082628
Trade/Device Name: cmr42 Cardiac Magnetic Resonance Imaging Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 11, 2008 Received: November 12, 2008
Dear Mr. Samarappuli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
#### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part.807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
loque An Whang
Joyce M. Whang, Ph.D. Acting Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
## INDICATIONS FOR USE
## 510(k) Number (if known): K082628
# Device: cmr42 Cardiac Magnetic Resonance Imaging Software
#### Indications for Use:
cmr is intended to be used for vicwing, post-processing and quantitative evaluation of cardiovascular magnetic resonance (MR) images in a Digital Imaging and Communications in Medicine (DICOM) Standard format. It enables;
- . Importing Cardiac MR Images in DICOM format
- Supporting clinical diagnostics by qualitative analysis of the cardiac MR images . using display functionality such as panning, windowing, zooming, navigation through series/slices and phases.
- Supporting clinical diagnostics by quantitative measurement of the heart and . adjacent vessels in cardiac MR images, specifically distance, area, volume and mass
- Supporting clinical diagnostics by using area and volume measurements for . measuring LV function and derived parameters cardiac output and cardiac of long axis and short axis cardiac MR images.
- . Flow quantifications based on velocity encodes images
It shall be used by qualified medical professionals, experienced in examining and evaluating cardiovascular MR images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. cmr3 is a software application that can be used as a stand-alone product or in a networked environment.
The target population for the cmr44 is not restricted, however the image acquisition by a cardiac magnetic resonance scanner may limit the use of the device for certain sectors of the general public.
cmr2 shall not be used to view or analyze images of any part of the body except the cardiac magnetic resonance images acquired from a cardiovascular magnetic resonance scanner.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X |
|-------------------------------------------------|--------|
| | AND/OR |
| Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF
NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|--------------------------------------------------------------|--|
| Division of Reproductive, Abdominal and Radiological Devices | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.