K082607 · Vpdiagnostics, Inc. · LLZ · Dec 5, 2008 · Radiology
Device Facts
Record ID
K082607
Device Name
MRI-PLAQUE VIEW, VERSION 1.1.2003
Applicant
Vpdiagnostics, Inc.
Product Code
LLZ · Radiology
Decision Date
Dec 5, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The MRI-PlaqueView software provides a set of post-processing tools to assist trained cardiologists and radiologists in the quantitative analysis of atherosclerotic carotid arteries from 1.5T or 3.0T magnetic resonance imaging (MRI) studies acquired with a combination of one or more contrast weightings such as T1, T2, Proton Density, and Time of Flight. Users of MRI-Plaque View perform semiautomatic delineation of lumen and outer vessel wall boundaries. Users may also perform semi-automatic, user configurable segmentation or manual drawing for delineation of atherosclerotic plaque components within the vessel wall. Users may edit the results on the fly. The software enables length, thickness, area, and volume measurements of the vessel wall as well as quantification of user-indicated areas. MRI-PlaqueView further aids in the visualization of atherosclerotic arteries (for example, through maximum intensity projections) of MRI data and color-coded maps of segmented arteries. When interpreted by a trained physician, the output of this image analysis toolbox can be used to support the decision-making process in clinical practice or to support investigations in clinical research and trials.
Device Story
MRI-PlaqueView is an image analysis software toolbox for carotid artery assessment. Input: 1-6 standard MRI series (T1, T2, Proton Density, TOF) from DICOM-compliant scanners. Operation: User selects region of interest; software employs semi-automatic boundary detection or manual tools to delineate lumen and outer vessel wall; additional semi-automated or manual segmentation labels internal vessel wall structures (calcified/soft plaque). Registration functions map contours across different contrast weightings. Output: Quantitative measurements (volumes, thicknesses, maximal areas, area ratios), maximum intensity projections, and color-coded maps. Usage: Installed on standalone PC, PACS, or imaging system; operated by cardiologists/radiologists in clinical or research settings. Physician reviews output to support clinical decision-making regarding atherosclerotic disease.
Clinical Evidence
No clinical data provided. Bench testing only.
Technological Characteristics
Software-based image analysis toolbox. Operates on Windows XP platform. Compatible with DICOM-compliant 1.5T/3.0T MRI scanners. Features include semi-automatic boundary detection, manual editing, image registration, and rendering (e.g., maximum intensity projections).
Indications for Use
Indicated for trained cardiologists and radiologists to perform quantitative analysis of atherosclerotic carotid arteries using 1.5T or 3.0T MRI studies (T1, T2, Proton Density, or Time of Flight contrast weightings).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K082607
# DEC 0 5 2008
## 510(k) Summary
This 510(k) summary is submitted in accordance with the requirements of 21 CFR Part 807.92, Subpart E.
## 1. Identification of Submitter:
| Submitter: | VPDiagnostics |
|------------|------------------------------------------------|
| Address: | 819 Virginia Street |
| | Suite 2209 |
| | Seattle, WA |
| Phone: | 206-428-6160 |
| Fax: | 866-278-3299 |
| Contact: | Michael Hartmann |
| Title: | Vice President, Business & Product Development |
| Phone: | Corporate: 206-428-6160 |
| | Mobile: 857-928-4551 |
| Fax: | 866-278-3299 |
Summary Date: June 19, 2008
#### 2. Identification of Product:
| Device Name: | MRI-PlaqueView |
|------------------------|----------------------------------------|
| Device Common Name: | System, Image Processing, Radiological |
| Device Classification: | 21 CFR 892.2050, Class II, LLZ |
| Manufacturer: | VPDiagnostics |
#### 3. Predicate Devices
MRI-PlaqueView provides viewing and post-acquisition image analysis of userselected regions of interest on MR images. This software medical device is substantially equivalent to the following devices:
Model: Vitrea Version 4.0 with SUREPlaque Manufacturer: Vital Images 510K Number: K071331
| Model: | QPlaque-MR |
|---------------|------------|
| Manufacturer: | Medis |
| 510K Number: | K073156 |
{1}------------------------------------------------
#### 4. Device Description:
MRI-PlaqueView is an image analysis software toolbox that facilitates visualization and quantitative analysis of MR images of carotid arteries acquired on 1.5T or 3.0T scanners. MRI-PlaqueView provides a set of image review, delineation, registration, editing, visualization, and measurement tools. The MRI-Plaque View user interface is designed to follow typical clinical workflow patterns to review, process, edit, validate, analyze, and visualize digital images and to present, distribute and save the findings.
The software application operates on a Windows XP platform that may be installed on a standalone PC, PACS or other medical imaging system. The software is designed to read MRI images from DICOM-compliant medical devices distributed by various OEM vendors.
The input to the system consists of one to six standard MRI series, which provide information through use of different MRI contrast weightings, such as T1, T2, Proton Density, and TOF weighted images
Once the user has selected a region of interest (ROI) for analysis, the MRI-PlaqueView tools will help to delineate the lumen and outer wall boundaries using either semi-automatic boundary detection or manual drawing/editing tools
The software provide the user an option to further analyze the multiple contrast weighted MRI data using either semi-automated algorithms or manual drawing/editing tools to segment and label the internal structure of the vessel wall, delineating calcified and soft plaque regions.
To facilitate the use of multiple contrast weighted data, the software displays the identified contours simultaneously on original images from all available series. Automated and user-controlled registration functions enable mapping of contours from one contrast weighting to another.
The quantitative measurements this software provides include volumes, thicknesses, maximal areas, and area ratios of the different regions outlined. The software also supports review and visualization of the data using standard rendering techniques, such as maximum intensity projections. The analysis results (both contours and quantitative measurements) can be saved in separate files to facilitate restoring previous works for subsequent sessions, and for reporting results.
{2}------------------------------------------------
#### 5. Indications for Use
The MRI-PlaqueView software provides a set of post-processing tools to assist trained cardiologists and radiologists in the quantitative analysis of atherosclerotic carotid arteries from 1.5T or 3.0T magnetic resonance imaging (MRI) studies acquired with a combination of one or more contrast weightings such as T1, T2, Proton Density, and Time of Flight. Users of MRI-Plaque View perform semiautomatic delineation of lumen and outer vessel wall boundaries. Users may also perform semi-automatic, user configurable segmentation or manual drawing for delineation of atherosclerotic plaque components within the vessel wall. Users may edit the results on the fly. The software enables length, thickness, and area and volume measurements of the vessel wall as well as quantification of user-indicated areas. MRI-Plaque View further aids in the visualization of atherosclerotic arteries (for example, through maximum intensity projections) of MRI data and color-coded maps of segmented arteries. When interpreted by a trained physician, the output of this image analysis toolbox can be used to support the decision-making process in clinical practice or to support investigations in clinical research and trials.
#### 6. Comparison with Predicate Devices
MRI-PlaqueView is substantially equivalent to several software applications that analyze and measure regions of interest in magnetic resonance or CT images. The predicate devices include
K071331- Vital Images Vitrea Version 4.0 with SUREPLAOUE (Appendix A) K073156 -Medis OPlaque-MR) (Appendix B)
Vital Images Vitrea Version 4.0 has been classified under 21 CFR 892.2050 as a Class II medical device
Medis OPlaque-MR has been classified under 21 CFR 892.1000 as a Class II medical device.
Both of these devices provide automated and interactive analysis tools to delineate and make measurements of size (volume, area, thickness) of regions of interest (ROL) in magnetic resonance (Medis OPlaque MR) or CT images (Vital Images Vitrea) and provide visualization and reporting utilities for the results. Both of these devices specifically address ROI measurements in blood vessels and specifically address measurement of ROIs within atherosclerotic plaque for CT (Vital Images Vitrea) and magnetic resonance images (Medis OPlaque-MR) respectively. Both provide substantially equivalent indications for use and have been developed with substantially equivalent mathematical algorithms.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus is depicted with three wavy lines forming a shape resembling a bird or abstract figure. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - (USA)" is arranged in a circular pattern around the caduceus.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### DEC 0 5 2008
Mr. Michael Hartmann VP, Product and Business Development VPDiagnostics, Inc. 819 Virgina St., Suite 2209 SEATTLE WA 98101
Re: K082607
Trade/Device Name: MRI-PlaqueView Regulation Number: 21 CFR 892.2050
Regulation Name: Picture archiving and communications system Regulatory Class: II
Product Code: LLZ
Dated: September 4, 2008 Received: September 8, 2008
#### Dear Mr. Hartmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device , can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part.807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Arputh Thang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indication(s) for Use Statement
510(k) Number:
4
To be assigned LOB 2607
Device Name:
MRI-PlaqueView
Indications for Use:
The MRI-PlaqueView software provides a set of post-processing tools to assist trained cardiologists and radiologists in the quantitative analysis of atherosclerotic carotid arteries from 1.5T or 3.0T magnetic resonance imaging (MRI) studies acquired with a combination of one or more contrast weightings such as T1, T2, Proton Density, and Time of Flight. Users of MRI-Plaque View perform semiautomatic delineation of lumen and outer vessel wall boundaries. Users may also perform semi-automatic, user configurable segmentation or manual drawing for delineation of atherosclerotic plaque components within the vessel wall. Users may edit the results on the fly. The software enables length, thickness, area, and volume measurements of the vessel wall as well as quantification of user-indicated areas. MRI-PlaqueView further aids in the visualization of atherosclerotic arteries (for example, through maximum intensity projections) of MRI data and color-coded maps of segmented arteries. When interpreted by a trained physician, the output of this image analysis toolbox can be used to support the decision-making process in clinical practice or to support investigations in clinical research and trials.
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Haber Resmer
(Division Sign-Off) Division of Reproductive, Abdominal. and Radiological Devices
510(k) Number K08260
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.