← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K082240

# SNTRYPACS (K082240)

_Qstar Technologies, Inc. · LLZ · Nov 20, 2008 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K082240

## Device Facts

- **Applicant:** Qstar Technologies, Inc.
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Nov 20, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

SntryPACS™ is a device that receives digital images and data from various sources (i.e. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways, etc.). Images and data can be captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. SntryPACS™ can be integrated with an institution's existing Hospital Information System (HIS) or Radiology Information System (RIS) providing access to reports for fully-integrated electronic patient records. Typical users of the system are trained medical professionals (e.g. physicians, radiologists) Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

## Device Story

SntryPACS is a DICOM-compliant Picture Archiving and Communication System (PACS) server. It receives digital images from modalities (CT, MRI, PET, Ultrasound) via DICOM C-Store; stores data in a central database; and facilitates retrieval by client workstations via DICOM Query/Retrieve. The system runs as Linux daemons using a DBMS and file indexers. It integrates with hospital HIS/RIS for electronic patient records. Used in clinical environments by radiologists and physicians to view and manage medical images. The system provides scalable storage and web-based administration. It does not contact patients or control life-sustaining equipment; clinicians perform all image interpretation. Benefits include centralized, efficient access to diagnostic images across distributed networks.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

DICOM-compliant server; Linux-based OS; utilizes DBMS and file indexers for data management. Supports DICOM C-Store and Query/Retrieve protocols. Scalable storage architecture with support for removable media. Web-based administration interface. No patient-contacting components. Software-based image processing and archival system.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Voyager PACS System ([K062062](/device/K062062.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Kofzzyo

## 510(k) Summary of Safety and Effectiveness

NOV 20 2008

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

Date Prepared: August 2, 2008

Submitter's Information: 21 CFR 807.92(a)(1) Mr. Jim Wheeler Director of Marketing and Business Development QStar Technologies Inc 2175 West HWY 98 Mary Esther, FI 32569 Fax: (850) 243-4234 Email jwheeler@gstar.com

Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)

| Trade Name:            | SntryPACS™                              |
|------------------------|-----------------------------------------|
| Common Name:           | Picture Archiving Communications System |
| Classification Name:   | system, image processing, radiological  |
| Product code:          | LLZ                                     |
| Device Classification: | 892.2050                                |

Predicate Device: 21 CFR 807. 92(a)(3)

| Edge PACS™ is substantially equivalent to: |                                        |
|--------------------------------------------|----------------------------------------|
| Manufacturer:                              | Voyager Imaging                        |
| Device Name:                               | Voyager PACS System                    |
| 510(k) Number:                             | K062062                                |
| Product Code:                              | LLZ                                    |
| Device Classification Name:                | SYSTEM, IMAGE PROCESSING, RADIOLOGICAL |
| Regulation Number:                         | Class II - 892.2050                    |

#### Device Description: 21 CFR 807 92(a)(4)

SntryPACS Server is a fully integrated DICOM Storage and DICOM Application Server, compatible with any DICOM modality and most DICOM viewing workstations.

SntryPACS server runs under the QStar storage server environment and provides PACS services such as storage provider and query retrieve provider. Both of these services runs as Linux daemons and utilizes DBMS system and file indexers for managing the medical images and reports. A typical SntryPACS network consists of a central SntryPACS server which stores a database containing the images, and of multiple clients that can retrieve and display these images on medical imaging software. The images are stored in DICOM format. The modalities (MRI, CT, PET, Ultrasounds, etc.) send the images to the PACS Server by using a DICOM "push" (DICOM C-Store). The server and the client workstations communicate by using the DICOM protocol (DICOM C-Store or Query & Retrieve). The clients display the images by using medical imaging software such as a DICOM viewer or a radiology review workstation client software system.

#### Key features

- Central storage server with complete DICOM protocol support to handle all different ਼ types of DICOM images and reports.
- Web-based user interface for system administration. ਂ

{1}------------------------------------------------

## 510(k) Summary of Safety and Effectiveness

- o Fast transfer and access time (limited by the network bandwidth) for store and retrieval.
- o Supports patient root and study root query retrieval.
- Support concurrent storage and retrieval of multiple modalities and clients. o
- Provides scalable storage architecture for images/report data and includes support ೧ for removable media solutions.

Indications for Use: 21 CFR 807 92(a)(5)

SntryPACS™ is a device that receives digital images and data from various sources (i.e. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways, etc.). Images and data can be captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. SntryPACS™ can be integrated with an institution's existing Hospital Information System (HIS) or Radiology Information System (RIS) providing access to reports for fully-integrated electronic patient records. Typical users of the system are trained medical professionals (e.g. physicians, radiologists) Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

Technological Characteristics: 21 CFR 807 92(a)(6)

SntryPACS™ is a device that does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.

Conclusion: 21 CFR 807 92(b)(1)

The 510(k) Pre-Market Notification for SntryPACS™ contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device. The subject device has been and will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey.

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing right, with flowing lines beneath them.

### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## NOV 2 0 2008

QStar Technologies, Inc. c/o Mr. Herman Oosterwijk Consultant OTech, Inc. 1600 Manchester Wav CORINTH TX 76210

Re: K082240

Trade/Device Name: SntryPACS Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 11, 2008 Received: November 17, 2008

Dear Mr. Oosterwijk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

## Page 2

Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial cquivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jorgu M. Whang

Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number:

Device Name:

Indications for Use:

SntryPACS is a device that receives digital images and data from various sources (i.e. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways, etc.), Images and data can be captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. SntryPACS™ can be integrated with an institution's existing Hospital Information System (HIS) or Radiology Information System (RIS) providing access to reports for fully-integrated electronic patient records. Typical users of the system are trained medical .professionals (e.g. physicians, radiologists)

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

Prescription Use × Over-The-Counter Use AND (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Arnimhth

(Division Sign-Off) ( Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K082240](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K082240)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
