21.3-inch (54cm) Monochrome LCD Monitor with sub-pixel driving techniques enabling 6M subpixels to be driven independently, MS21i2 (MDL2127A) is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is not meant to be used in digital mammography.
Device Story
21.3-inch monochrome LCD monitor; displays medical images for diagnostic review by trained practitioners. Input via DVI; utilizes sub-pixel driving techniques to independently drive 6 million subpixels. Used in clinical settings; supports landscape (1200x4800) and portrait (4800x1200) orientations. Provides high-resolution visualization for diagnostic decision-making; excludes digital mammography applications.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
21.3-inch monochrome LCD; 1200x4800/4800x1200 resolution; DVI connectivity; sub-pixel driving technology for 6M subpixel control.
Indications for Use
Indicated for trained medical practitioners to display and view medical images for diagnostic purposes. Not indicated for digital mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
K081056 — 21.3 (54CM) 3M COLOR LCD MONITOR CCL354I2 (CDL2126A) · Totoku Electric Co., Ltd. · Jun 4, 2008
K090947 — 21.3 (54 CM) MONOCHROME DIGITAL MAMMOGRAPHY LCD MONITOR, MODEL MS3112 (MDL2128A) · Totoku Electric Co., Ltd. · Aug 16, 2010
K063199 — 21.3-INCH (54CM) COLOR LCD MONITOR CDL2120A (CCL252I2) · Totoku Electric Co., Ltd. · Dec 4, 2006
K063198 — 21.3-INCH (54CM) MONOCHROME LCD MONITOR MDL2121A (ME253I2) · Totoku Electric Co., Ltd. · Nov 30, 2006
K081055 — 21.3 (54CM) 2M COLOR LCD MONITOR CCL254I2 (CDL2125A) · Totoku Electric Co., Ltd. · Jun 4, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
K08:i945
#
SEP 1 2 2008
# 510(k) SUMMARY
TOTOKU ELECTRIC CO., LTD. Submitter Information: 300 Oya, Ueda Nagano 386-0192 Japan
> Contact Person: Mikio Hasegawa, General Manager Email: hasegawam@totoku.co.jp Tel: +81.268.34.5469 Fax: +81.268.34.5548
Date Prepared: June 30, 2008
21.3-inch (54 cm) Monochrome LCD Monitor MS21i2 (MDL2127A) Device Name:
MS21i2 (MDL2127A) Common Name:
Classification Name: Class II (Part 892 Radiology Devices Sec. 892.2050 Picture Archiving and Communication System)
Predicate Device: 21.3-inch (54cm) Monochrome LCD Monitor MDL2121A (ME253i2) (K063198) 21.3-inch (53cm) Monochrome LCD Monitor MS51i2 (MDL2123A) (K081242)
Device Description: MS21i2 (MDL2127A) is a 21.3-inch (54 cm) Monochrome LCD monitor whose display resolution is 1200 x 4800 (landscape), 4800 x 1200 (portrait) supporting DVI (digital visual interface).
21.3-inch (54 cm) Monochrome LCD Monitor with sub-pixel driving Indended Use: techniques enabling 6M subpixels to be driven independently, MS21i2 (MDL2127A) is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is not meant to be used for digital mammography.
MS21i2 (MDL2127A) shares the same characteristics with our Substantial Equivalence: predicate devices, ME253i2 (MDL2121A) (K063198) and MS51i2 (MDL2123A) (K081242) except for bezel, enclosure and main board (driver board).
MDL2127A Premarket Notification
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 2 2008
Mr. Mikio Hascgawa General Manager TOTOKU Electric Co., Ltd. Intelligent Devices and Solutions Division 300 Oya, Ueda, Nagano 386-0192 JAPAN
Re: K081945
Trade/Device Name: 21.3-inch (54 cm) Monochrome LCD MS21i2 (MDL2127A) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 30, 2008 Received: July 8, 2008
Dear Mr. Hasegawa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and . adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
## Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Jorque M. Whang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number: Not Known
Device Name: 21.3-inch (54 cm) Monochrome LCD Monitor MS21i2 (MDL2127A)
Indications for Use:
21.3-inch (54cm) Monochrome LCD Monitor with sub-pixel driving techniques enabling 6M subpixels to be driven independently, MS21i2 (MDL2127A) is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is not meant to be used in digital mammography.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Halina Pluma
(Division Sidn-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.