SECTRA WORKSTATION, IDS7 VERSION 11.3, IDS5 VERSION 11.1 P2
K081469 · Sectra Imtec AB · LLZ · Jun 12, 2008 · Radiology
Device Facts
Record ID
K081469
Device Name
SECTRA WORKSTATION, IDS7 VERSION 11.3, IDS5 VERSION 11.1 P2
Applicant
Sectra Imtec AB
Product Code
LLZ · Radiology
Decision Date
Jun 12, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
The Sectra Workstation is intended for the manipulation and displaying of medical images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards. Device options make possible mammography reading, telecommunications; fast demonstration; prosthesis CAD; 3-D and angiography, etc.; and teleconferencing. Lossy compressed mammographic images are not intended for diagnostic review. Mammographic images should only be viewed with a monitor approved by FDA for viewing mammographic images. For primary diagnosis, post process DICOM "for presentation" images must be used. Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.
Device Story
Sectra Workstation is a software-based PACS client running on Windows PCs; processes and displays digital medical images from various modalities. Operates as a family of workstations (IDS5/IDS7) for diagnostic, quality assurance, clinical viewing, teleradiology, and document scanning tasks. Features include 3D volume rendering, mammography screening tools, and DICOM printing. Used by radiologists, clinicians, and technicians in clinical settings. Input data received via DICOM/standard interfaces from PACS servers. Output displayed on high-resolution monitors for clinical interpretation. Facilitates diagnostic decision-making, image review, and reporting. Benefits include streamlined radiology workflow, remote access via teleradiology, and specialized tools for mammography and angiography.
Clinical Evidence
Bench testing only. Device complies with ISO 9001:2000 and ACR/NEMA DICOM 3.0 standards. No clinical data provided.
Technological Characteristics
Software-based PACS workstation; runs on Windows 2000/XP/Vista. Supports up to four monitors. Connectivity via DICOM 3.0. Architecture includes .NET-based (IDS7) and legacy (IDS5) versions. Features include 3D rendering, mammography workflow tools, and teleradiology. No patient contact; non-life-sustaining.
Indications for Use
Indicated for trained professionals (physicians, radiologists, nurses, technicians, assistants) for manipulation and display of medical images, including mammograms, from various modalities for diagnostic review and clinical workflow support.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
K051315 — MODIFICATION TO: SECTRA IDS5 WORKSTATION · Sectra Imtec AB · Jun 15, 2005
K063093 — SECTRA WORKSTATION, VERSION 11.1 · Sectra Imtec AB · Nov 7, 2006
K050196 — SECTRA IDS5 WORKSTATION · Sectra Imtec AB · Feb 24, 2005
K040376 — SECTRA IDS5 RADIOLOGY WORKSTATION; SECTRA MPR PACKAGE; SECTRA 3D PACKAGE · Sectra-Imtec AB · May 4, 2004
K033712 — MODIFICATION TO SECTRA IDS5 RADIOLOGY WORKSTATION - VERSION 10.1 · Sectra-Imtec AB · Dec 23, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
K081469
## JUN 12 2008
# 510(k) Summary of Safety & Effectiveness
(as required by 21 CFR 807.92c)
#### Date Prepared
21 May 2008
#### Submitter's Information
Sectra Imtec AB Teknikringen 20 SE-583 30 Linköping Sweden Phone: +1 46 13 23 52 00 Fax: +1 46 13 21 21 85
### Trade Name, Common Name, Classification
| Trade Name: | Sectra Workstation |
|----------------------|---------------------------------------------------|
| Software version: | IDS7 11.3<br>IDS5 11.1 P2 |
| Common Name: | Picture Archiving and Communications System |
| Classification Name: | Image Processing System (LLZ) (21 CFR § 892.2050) |
#### Predicate Device
| Applicant: | Sectra Imtec AB |
|----------------|--------------------|
| 510(k) Number: | K063093 |
| Device: | Sectra Workstation |
#### Device Description
The Sectra Workstation is mainly a software product. It is used for visualization and processing of digital medical images. The Sectra Workstation is used as a client together with a Sectra provided server (Sectra PACS Core, Class I Exempt). The system runs on PCs under the Windows operating systems. Most notably two or more monitors are used.
The Sectra Workstation is a family of devices, including several workstations or types of workstations (see Table 1).
{1}------------------------------------------------
| Workstation Type | Description |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IDS5/dx.net | A diagnostics workstation. It contains tools for assisting the radiologist in making a<br>diagnosis. |
| IDS5/mx.net | A workstation specifically designed to handle the mammography screening<br>workflow. It has all functionality as an IDS5/dx.net but with additional features to<br>enhance the mammography screening workflow. |
| IDS5/qa.net | The quality assurance workstation in the Sectra Workstation family. Mainly used by<br>the technologists to prepare the images for the reviewing radiologist. |
| IDS5/mqa.net | The quality assurance workstation to be used by a mammography technician to<br>prepare images for the reviewing radiologist. |
| IDS5/cl.net | Used by the clinicians within the hospital to view the radiology images and to read<br>the radiology report. |
| IDS5/web | A slim version of the Sectra Workstation that can be used by remote clinics to view<br>images and radiology reports. |
| IDS5/doc | A limited workstation that is used to scan radiology requests. |
| IDS5/send | A teleradiology workstation that is used to send examinations to teleradiology<br>destinations. |
| IDS7/dx | A diagnostics workstation, based on the .NET architecture. It contains tools for<br>assisting the radiologist in making a diagnosis. |
| IDS7/cl | A clinician's workstation, based on the .NET architecture. Used by the clinicians<br>within the hospital to view the radiology images and to read the radiology report. |
#### Table 1 Sectra Workstation types.
. .
To handle specific user needs it is also possible to activate Sectra developed add-ons for the Sectra Workstation. An add-on is thus an additional feature set that can be used on the workstation. Some features are add-ons on Sectra Workstation IDS5 but are included as ordinary features for IDS7 (see Table 2 for examples).
| Add-on | Sectra Workstation IDS5 | Sectra Workstation IDS7 |
|--------------------------------------------------------|-------------------------|---------------------------------|
| Volume rendering (3D) | X | X |
| Clinical Application Interface slots (2<br>additional) | X | |
| Dual monitor support | X | Included in IDS7 by<br>default. |
| Multiframe support | X | Included in IDS7 by<br>default. |
| Image cache | X | Included in IDS7 by<br>default. |
| DICOM print support | X | |
| Image scanner support | X | |
| Basic document scanner support | X | |
| Demonstration scheduling | X | |
| Nuclear medicine image controls | X | |
| Patient CD | X | |
| Teleradiology | X | |
| WISE/doc | X | |
Examples of add-ons for Sectra Workstation. Table 2
{2}------------------------------------------------
#### Indications for Use
The Sectra Workstation is intended for the manipulation and displaying of medical images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards.
Device options make possible mammography reading, telecommunications; fast demonstration; prosthesis CAD; 3-D and angiography, etc.; and teleconferencing.
Lossy compressed mammographic images are not intended for diagnostic review. Mammographic images should only be viewed with a monitor approved by FDA for viewing mammographic images. For primary diagnosis, post process DICOM "for presentation" images must be used.
Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.
#### Technological Characteristics
The Sectra Workstation runs under the Windows 2000 Professional . Windows XP Professional, and Windows Vista Business2 operating systems for PCs (as a minimum and depending upon system configuration)3. The requirements on hardware are quite ordinary for a system used for displaying images. Most notably up to four monitors can be used.
#### Performance Data
The subject device is developed according to ISQ 9001:2000 and complies with ACR/NEMA Digital Imaging Communications in Medicine version 3.0.
#### Conclusion
Similar to the predicate device, the Sectra Workstation does not contact the patient, nor does it control any life sustaining devices. Images and information being reviewed, processed, relayed, and or transmitted are interpreted by a physician or trained medical personnel, providing ample opportunity for competent human intervention. The device and the predicate device share the same certification or conformance to performance standards and both function as Image Processing System (LLZ). Device failures, which might result in partial or failed transmissions, images, or data, may be recovered from storage or re-transmission after correcting the problem(s). Passwords are required for operation and to protect against unauthorized use.
Based on the information supplied in this Special 510(k), we conclude that the subject device is safe, effective, and substantially equivalent to the predicate device.
Sectra Workstation IDS7 is not supported on Microsoft Windows 2000 Professional.
<sup>2</sup> Windows Vista is supported for IDS7 only.
<sup>3</sup> Windows NT is supported for IDS5/web.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN 12 2008
Sectra Imtec AB % Mr. Carl Alletto Official Correspondent OTech, Inc. 1600 Manchester Way CORINTH TX 76210
Re: K081469
Trade/Device Name: Sectra Workstation by Sectra Imtec AB Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 21, 2008 Received: May 27, 2008
#### Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours,
Nancy Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number: K081469
Device Name: Sectra Workstation by Sectra Imtec AB
Indications For Use:
The Sectra Workstation is intended for the manipulation and displaying of medical images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards.
Device options make possible mammography reading, telecommunications; fast demonstration; prosthesis CAD; 3-D and angiography, etc ; and teleconferencing.
Lossy compressed mammographic images are not intended for diagnostic review. Mammographic images should only be viewed with a monitor approved by FDA for viewing mammographic images. For primary diagnosis, post process DICOM "for presentation" images must be used.
Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.
Prescription Use
AND/OR
(Per 21 CFR 801.109)
(21 CFR 807 Subpart C)
Over -The-Counter Use ----------
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hubert Reiner
on Sign-Off) (Division of Reproductive, Abdominal an Radiological Devices 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.