SIMPLANT ORTHO ; VISTADENT 3D

K081402 · Materialise Dental NV · LLZ · Jul 18, 2008 · Radiology

Device Facts

Record IDK081402
Device NameSIMPLANT ORTHO ; VISTADENT 3D
ApplicantMaterialise Dental NV
Product CodeLLZ · Radiology
Decision DateJul 18, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Materialise Dental's SimPlant Ortho; Vistadent 3D software is indicated for use as a medical front-end software that can be used by medically trained people for the purpose of visualizing gray value images. It is intended for use as a software interface and image segmentation system for the transfer of imaging information from a medical scanner such as a CT scanner or a Magnetic Resonance scanner. It is also used as a software system for simulating/evaluating orthodontic treatment i.e. dental bite options.

Device Story

SimPlant Ortho/Vistadent 3D is a medical front-end software system for dental orthodontic treatment planning. It accepts CT or MRI scan data as input. The software provides tools for image segmentation, cephalometric analysis, and 3D visualization of anatomical structures. It allows clinicians to simulate orthodontic procedures, including the use of temporary anchorage devices (TADs), and visualize osteotomies, distractions, and the resulting impact on jaw relations and soft tissue profiles. The system is intended for use by medically trained professionals in a clinical setting. Output is used in conjunction with diagnostic tools and expert clinical judgment to assist in treatment planning and evaluation. The device benefits patients by allowing non-invasive assessment of anatomical structures and pre-surgical simulation of orthodontic outcomes.

Clinical Evidence

No clinical data. Software validation performed via documented test plan derived from specifications and end-user testing.

Technological Characteristics

Software-based image processing system. Programmed in C++ language. Operates on Windows OS. Functions include image segmentation, cephalometric analysis, and 3D visualization of CT/MRI data.

Indications for Use

Indicated for use by medically trained personnel to visualize gray value images, perform image segmentation of CT/MRI data, and simulate or evaluate orthodontic treatment options, including dental bite adjustments.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ KoriYoz | SUBMITTER'S NAME AND ADDRESS: | MATERIALISE DENTAL N.V.<br>Technologielaan 15<br>B-3001 Leuven, Belgium | |--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------| | ESTABLISHMENT REGISTRATION NO: | 3006638827 | | CONTACT PERSON: | Carl Van Lierde, Materialise Dental N.V.<br>Quality Manager<br>+32 163 967 14 (tel)<br>+32 163 966 22 (fax)<br>carl.vanlierde@materialise.be | | SUMMARY PREPARATION DATE: | April 21, 2008 | | TRADE NAME: | SimPlant Ortho; Vistadent 3D | | COMMON NAME: | Image processing system and software for simulating<br>/evaluating dental orthodontic treatment i.e. dental bite<br>options | | CLASSIFICATION NAME: | System, Image Processing (C.F.R. Section 892.2050,<br>Product code: LLZ) | ## PREDICATE DEVICE SimPlant System; SimPlant Dr. James (K053592) ## FUNCTION Simplant Ortho provides a method of segmenting CT images. This file allows the individual patient's CT image to be assessed in a three-dimensional way, to see the anatomical structures without patient contact or surgical insult. It includes features for cephalometric analysis of the patient and orthodontic treatment simulation. Additional information about the exact geometry of the tooth surfaces can be visualized together with the CT data and orthodontic procedures with TADs (temporary anchorage devices) can be simulated. Osteotomies and distractions can be visualized to simulate the desired relation of both jaws and the result on the soft tissue profile of the patient can be visualized. The output file is intended to be used in conjunction with diagnostic tools and expert clinical judgment. ## INTENDED USE Materialise Dental's SimPlant Ortho; Vistadent 3D software is indicated for use as a medical front-end software that can be used by medically trained people for the purpose of visualizing gray value images. It is intended for use as a software interface and image segmentation system for the transfer of imaging information from a medical scanner such as a CT scanner or a Magnetic Resonance scanner. It is also used as a software system for simulating/evaluating orthodontic treatment i.e. dental bite options. ## TESTING & VALIDATION The software is thoroughly tested in accordance with a documented test plan. This test plan is derived from the specifications and ensures that all controls and features are functioning properly. The software is validated together with end-users. {2}------------------------------------------------ ## TECHNOLOGICAL CHARACTERISITICS . : and the same of the same of the states of the states and the comments of the country SimPlant Ortho; Vistadent 3D is software for simulating/evaluating orthodontic treatment (i.e. dental bite options) programmed in C++ language and running on the Windows operating system. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. ## Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUL 1 8 2008 Mr. Carl Van Lierde Quality Manager Materialise Dental NV Technologielaan 15, 3001 Leuven BELGIUM Re: K081402 Trade/Device Name: SimPlant Ortho; Vistadent 3D Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 21, 2008 Received: May 19, 2008 ## Dear Mr. Lierde: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act, The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the seconomic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to 10(L) marketed predicate device results in a classification for your device of your device to re device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Houlth's (CDRH's) Office of Comptiment on the of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRF s Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 24/1276-3474. For questions regarding the reporting of device adverse events (Medical Device Reportion (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. Boveon reportain other general information on your responsibilities under the Act from the Division of Mary of Manufacturers, International and Consumer Assistance at its toll-free number (800) 63-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html . Sincerely vours. Nancy C brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K081402 Device Name: SimPlant Ortho; Vistadent 3D Indications for Use: Materialise Dental's SimPlant Ortho; Vistadent 3D software is indicated for use as a medical front-end software that can be used by medically trained people for the purpose of visualizing gray value images. It is intended for use as a software interface and image segmentation system for the transfer of imaging information from a medical scanner such as a CT scanner or a Magnetic Resonance scanner. It is also used as a software system for simulating/evaluating orthodontic treatment i.e. dental bite options. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Argin McWhan (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number _
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...