The 3-D Imaging Workstation is intended to be used by physicians in the clinic or hospital for 2-D and 3-D visualization of ultrasound images of the prostate gland. Additional software features include patient data management, multi-planar reconstruction, segmentation, image measurement and 3-D image registration.
Device Story
3-D Imaging Workstation processes 2-D live video from standard ultrasound machines to reconstruct 3-D prostate images. System includes mechanical tracker for end-fire TRUS probe; PC-based workstation with video digitizing card; image processing software. Physician performs manual ultrasound procedure; workstation tracks probe position/orientation during scanning. Outputs include 3-D rendered surface models, volume estimations, and biopsy site planning. Used in clinics/hospitals by physicians. Enables visualization of prostate abnormalities; assists in biopsy needle guidance by displaying trajectory relative to 3-D model. Benefits include enhanced anatomical assessment and procedural planning for prostate biopsies.
Clinical Evidence
Bench testing only. System performance verified and validated using test phantoms incorporating simulated prostates. No clinical data provided.
Technological Characteristics
PC-based workstation; video digitizing card; mechanical assembly for ultrasound probe tracking. Inputs: analog NTSC video from ultrasound systems. Features: multi-planar reconstruction, segmentation, 3-D registration, volume estimation. Software-based image processing.
Indications for Use
Indicated for use by physicians in clinical or hospital settings for 2-D and 3-D visualization of prostate ultrasound images, including image measurement, segmentation, and registration for prostate assessment and biopsy planning.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
XELERIS 2 Processing and Review Workstation (K051673)
Submission Summary (Full Text)
{0}------------------------------------------------
# 081093
510(k) Summary
| Submitter's Name: | Eigen LLC | | | | | | |
|----------------------------|-----------------------------------------------------------------------------------------------------|--|--|--|--|--|--|
| Submitter's Address: | 13366 Grass Valley Avenue, Grass Valley, CA 95945 | | | | | | |
| Submitter's Telephone: | 530-274-1240 | | | | | | |
| Contact Name: | Mark A. Hoffman | | | | | | |
| Date Summary was Prepared: | March 25, 2007 | | | | | | |
| Trade or Proprietary Name: | 3-D Imaging Workstation | | | | | | |
| Common or Usual Name: | Medical Image Processing Workstation | | | | | | |
| Classification Name: | System, Image Processing, Radiological, LLZ<br>Picture Archiving and Communications, 21CFR 892.2050 | | | | | | |
| Predicate Devices: | Device Name 510(k) Number 3-Dnet Suite K063107 XELERIS 2 Processing and K051673 | | | | | | |
Description of the Device and Summary of the Technological Characteristics The 3-D Imaging Workstation is designed to display the 2-D live video received from commercially available ultrasound machines and use this 2-D video to reconstruct a 3-D ultrasound image. The system has been designed to work with the clinicians' existing ultrasound machine, end fire TRUS probe, commercially available needle guide and needle gun combination. Additional software features include patient data management, multiplanar reconstruction, segmentation, image measurement and 3-D image registration.
Review Workstation
The 3-D Imaging Workstation is comprised of a mechanical assembly that holds the ultrasound probe and tracks probe position while the physician performs a normal ultrasound imaging procedure of the subject prostate. The mechanical tracker is connected to a PC-based workstation containing a video digitizing card and running the image processing software. Control of the ultrasound probe and ultrasound system is done manually by the physician, just as it would be in the absence of the 3-D Imaging Workstation. However, by tracking the position and orientation of the ultrasound probe while capturing the video image, the workstation is able to reconstruct and display a 3-D image and 3-D rendered surface model of the prostate.
The reconstructed 3-D image can be further processed to perform various measurements including volume estimation, and can be examined for abnormalities by the physician. Patient information, notes, and images may be stored for future retrieval.
{1}------------------------------------------------
Locations for biopsies may be selected by the physician, displayed on the 3-D image and 3-D rendered surface model, and stored. Previously stored 3-D models may be recalled and a stored 3-D model may be aligned or registered to the current 3-D model of the prostate.
Finally, the physician may attach a commercially available biopsy needle guide to the TRUS probe and use the probe and biopsy needle to perform tissue biopsy. Whenever the ultrasound machine is turned on by the physician, the live 2-D ultrasound image is displayed on the screen of 3-D Imaging Workstation during the biopsy. As the TRUS probe with attached needle guide is maneuvered by the physician, the position and orientation of the probe is tracked. The 3-D Imaging Workstation is able to add, display and edit plans for biopsy sites as well as an estimate of the probe position and needle trajectory relative to the 3-D image and 3-D rendered surface model of the prostate.
The 3-D Imaging Workstation offers the physician additional 3-D information for assessing prostate abnormalities, planning and implementing biopsy procedures. The additional image processing features are generated with minimal changes to previous TRUS probe based procedures, and the physician always has access to the live 2-D ultrasound image during prostate assessment or biopsy procedure.
### Substantial Equivalence
5
The product's technical features are substantially equivalent to the 3-Dnet Suite (K063107) and the XELERIS 2 Processing and Review Workstation (K051673). The 3-D Imaging Workstation is primarily a software product that runs on a PC-based workstation. Image data is input to the devices and used to generate 3-D views and perform image processing. Like the two predicate devices, the software has image measurement, multi-planar reformatting, segmentation and image registration abilities, image storage and retrieval, as well as patient information management functions.
The 3-D Net Suite and XELERIS 2 are software products that accept multiple image data types including magnetic resonance, computed tomography and ultrasound. The 3-D Imaging Workstation has been designed for ultrasound imaging, using standard analog NTSC video input from commercially available transrectal ultrasound systems. The 3-D Imaging Workstation incorporates a mechanical device to track the position of an end fire ultrasound probe. Ultrasound system operation and transrectal probe movement all remain under manual control of the physician.
{2}------------------------------------------------
## Testing and Performance Data
All product and engineering specifications were verified and validated. Test phantoms incorporating simulated prostates were developed and used to verify system performance through verification, validation (Appendix D) and benchmarking (Appendix F).
#### Conclusion
5
The results of comparing the intended use, function, technological characteristics, mode of operation and specifications of the 3-D Imaging Workstation with those of the two predicate devices demonstrate that the 3-D Imaging Workstation is substantially equivalent to existing products in the market today.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the emblem.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Eigen LLC % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
# MAY - 1 2008
Re: K081093
Trade/Device Name: 3-D Imaging Workstation Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: LLZ Dated: April 16, 2008 Received: April 17, 2008
#### Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/odri/industry/support/index.html.
Sincerely vours.
Nancy C. Brigdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): pending
Device Name: 3-D Imaging Workstation
Indications for Use:
4
Use
The 3-D Imaging Workstation is intended to be used by physicians in the clinic or hospital for 2-D and 3-D visualization of ultrasound images of the prostate gland. Additional પ્રતિ software features include patient data management, multi-planar reconstruction, segmentation, image measurement and 3-D image registration.
Arry Thithy
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number __
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
. :
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.