← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K080770

# IMAGENMD WITH IMAGENQ (K080770)

_Cardiovascular Imaging Technologies, LLC · LLZ · Apr 9, 2008 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K080770

## Device Facts

- **Applicant:** Cardiovascular Imaging Technologies, LLC
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Apr 9, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

The ImagenMD™ with ImagenQ™ system is software that allows the user to visualize raw PET and/or PET/CT data, assist in evaluating the quality of PET scans, and perform quantitative measurements of tracer uptake to aid in the interpretation of myocardial perfusion PET images.

## Device Story

ImagenMD™ with ImagenQ™ is a Windows-based software application for clinical use by radiologists, nuclear cardiologists, or nuclear medicine physicians. It processes raw PET and PET/CT image data to assist in myocardial perfusion assessment. The software enables quality assessment of PET scans, automated and manual selection of myocardial boundaries and alignment, and visualization of quantitative perfusion calculations. By providing automated measurements of tracer uptake, the device aids clinicians in interpreting myocardial perfusion images, supporting clinical decision-making regarding cardiac health. The device functions as a diagnostic aid for interpreting nuclear medicine imaging.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on comparison of technological characteristics and intended use to predicate devices.

## Technological Characteristics

Windows-based software application. Performs image processing and quantitative analysis of PET/CT data. Utilizes automated and manual boundary selection and alignment algorithms. Standalone software.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- ECAT LSO PET/CT 16
- EMORY Cardiac Toolbox

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K080770

APR - 9 2008

## 5. 510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with 21 CFR 807.92.

### General Information:

| A. | Submitted By:        | Cardiovascular Imaging Technologies<br>4320 Wornall Road, Suite 55<br>Kansas City, MO 64111<br>Tel: 816-531-2842<br>Fax: 816-531-0643 |
|----|----------------------|---------------------------------------------------------------------------------------------------------------------------------------|
|    | Contact Person:      | James A. Case                                                                                                                         |
|    | Date Prepared:       | March 17, 2008                                                                                                                        |
| B. | Device Trade Name:   | ImagenMD™ with ImagenQ™                                                                                                               |
|    | Classification Name: | System, Emission Computed Tomography<br>21 CFR 892.1200 (KPS)<br>System, Image Processing, Radiological<br>21 CFR 892.2050 (LLZ)      |
| C. | Predicate Devices:   | ECAT LSO PET/CT 16<br>EMORY Cardiac Toolbox                                                                                           |

#### Device Description: D.

ImagenMD™ with ImagenQ™ is a Windows software application which allows physicians and healthcare professionals to inspect, quantitatively and automatically perform calculations on myocardial perfusion PET images. The user can perform quality assessments, automatically and manually select myocardial boundaries and alignment, and visualize the results of quantitative perfusion calculations. The use of this system is limited to qualified, licensed healthcare providers (radiologists, nuclear cardiologists or nuclear medicine physicians) trained in the use of nuclear medicine imaging devices.

- Indications for Use: E.
The ImagenMD™ with ImagenQ™ system is software that allows the user to visualize raw PET and/or PET/CT data, assist in evaluating the quality of PET scans and perform quantitative measurements of tracer uptake to aid in the interpretation of myocardial perfusion PET images.

Page 13 of 95

{1}------------------------------------------------

- Comparison of Technical Characteristics to Predicate Device: F.
The ImagenMD™ with ImagenQ™ system and its predicates, the ECAT LSO PET/CT 16 and the Emory Cardiac Toolbox™ utilize the same type of data sets for analysis and calculation of data.

- H. Summary:
Testing and comparison of technological characteristics and intended uses found that all components of the ImagenMD™ with ImagenQ™ system are equivalent to the predicates.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### APR - 9 2008

Cardiovascular Imaging Technologies % Ms. Melanie Hasek Manager, Regulatory Affairs PRA International 9755 Ridge Drive LENEXA KS 66219

Re: K080770

Trade/Device Name: ImagenMD™ with ImagenQTM Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: LLZ Dated: March 17, 2008 Received: March 25, 2008

Dear Ms. Hasek:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

# 4. Indications for Use:

Indications for Use

Page 1 of 1

2080770 510(k) Number (if known): _

Device Name: ImagenMDTM with ImagenQTM

Indications For Use:

The ImagenMD™ with ImagenQ™ system is software that allows the user to visualize raw PET and/or PET/CT data, assist in evaluating the quality of PET scans, and perform quantitative measurements of tracer uptake to aid in the interpretation of myocardial perfusion PET images.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Atmor M. Whing
(Division/Sign-Off)

Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

Page 12 of 95

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K080770](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K080770)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
