K072258 · Almen Laboratories, Inc. · LLZ · May 16, 2008 · Radiology
Device Facts
Record ID
K072258
Device Name
BI-RADS COMPANION
Applicant
Almen Laboratories, Inc.
Product Code
LLZ · Radiology
Decision Date
May 16, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
BI-RADS Companion™ ("BRC") is a computer-aided image viewing and report generating tool designed to assist radiologists in the analysis of ultrasound images of breast masses by making the ACR "BI-RADS Classification Form and Lexicon®" ("ABCFL") available in electronic format. The user can view any selected image from a case and segment the lesion from the background by applying a user selected threshold or outlining the lesion manually with a cursor. The user can annotate, tag, measure, and record selected views. Under user control and review, the software electronically assembles reports from inputs selected by the user from the ultrasound ABCFL during the image interpretation process. The assembled report complies with the ABCFL for ultrasound. The output may be viewed and sent to electronic storage media, or standard film or paper printers or other document manipulation programs such as Adobe or Microsoft. When used by a skilled radiologist, this device provides information that may be useful in image interpretation. Patient management decisions should not be based solely on forms generated by BRC. The ultrasound images displayed by BRC must not be used for primary diagnostic interpretation.
Device Story
Computer-aided image viewing and report generation tool; assists radiologists in analyzing ultrasound images of breast masses. Inputs: ultrasound images. Operation: user views images; segments lesions via manual cursor outlining or user-selected thresholding; annotates, tags, and measures findings. Software assembles reports based on ACR BI-RADS Classification Form and Lexicon (ABCFL) inputs selected by the radiologist. Output: electronic reports, printable or exportable to standard document formats. Used in clinical settings by radiologists. Benefits: standardizes reporting according to ACR guidelines; facilitates efficient documentation and record-keeping during image interpretation.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Windows 2000/XP-based software; DICOM File Set Reader (FSR) compatible; HL7 compliant. Operates on PC, notebook, or workstation. Features include manual and threshold-based image segmentation, annotation, measurement, and automated report generation based on ACR ABCFL standards.
Indications for Use
Indicated for use by radiologists to assist in the analysis of ultrasound images of breast masses in female patients. Not for primary diagnostic interpretation; patient management decisions should not be based solely on generated reports.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K072258
MAY 1 6 2008
## 510(k) Summary of Safety and Effectiveness As required by 807.92 for BI-RADS Companion Prepared on August 7, 2007 Revised October 31, 2007, January 29, 2008, April 18, 2008, May 16, 2008
Submitted by: Almen Laboratories, Inc. 1672 Gil Way Vista, CA 92084
> Fax: 760-806-7577 Tel. 760-806-0040
Contact Person: Roger H. Schneider 6319 Massachusetts Avenue Bethesda, MD 20816-1139
> Fax: 301-229-8485 Tel. 301-229-8442
Device Trade Name: BI-RADS Companion
Common Name: Medical Image Processing System
Classification: Class II, 21 CFR 892.2050 PACS
B-CAD System, Version 1.0 (K050846) Predicate Device:
MEDIPATTERN CORPORATION Manufactured by: 2300 Sheppard Avenue West Suite 204. Toronto, Ontario M9M 3 A4
#### Indications for Use
BI-RADS Companion™ ("BRC") is a computer-aided image viewing and report generating tool designed to assist radiologists in the analysis of ultrasound images of breast masses by making the ACR "BI-RADS Classification Form and Lexicon® ("ABCFL") available in electronic format. The user can view any selected image from a case and segment the background by applying a user selected threshold or outlining the lesion manually with a cursor.
The user can annotate, tag, measure, and record selected views. Under user control and review, the software electronically assembles reports from inputs selected by the user from the ultrasound ABCFL during the image interpretation process. The assembled report complies with the ABCFL for ultrasound. The output may be viewed and sent to electronic storage media, or standard film or paper printers or other document manipulation programs such as Adobe or Microsoft.
When used by a skilled radiologist, this device provides information that may be useful in image interpretation. Paticnt management decisions should not be based solely on forms generated by BRC.
The ultrasound images displayed by BRC must not be used for primary diagnostic interpretation.
{1}------------------------------------------------
#### Description of the Device:
BI-RADS Companion™ ("BRC") is a computer-aided image viewing and report generating tool designed to assist radiologists in the analysis of ultrasound images of breast masses by making the ACR "BI-RADS Classification Form and Lexicon® ("ABCFL") available in electronic format. The user can view any selected image from a case and segment the background by applying a user selected threshold or outlining the lesion manually with a cursor.
The user can annotate, tag, measure, and record selected views. Under user control and review, the software electronically assembles reports from inputs selected by the ultrasound ABCFL during the image interpretation process. The assembled report complies with the ABCFL for ultrasound. The output may be viewed and sent to electronic storage media, or standard film or paper printers or other document manipulation programs such as Adobe or Microsoft.
{2}------------------------------------------------
Substantial Equivalence to Predicate Device: BI-RADS Companion is substantially equivalent to: B-CAD System, Version 1.0. (K050846)
The following table is a comparison of the functions of the device.
| | Function | BI-RADS<br>Companion,<br>Version 1.0<br>(K072258) | B-CAD<br>System,<br>Version<br>1.0.<br>(K050846) |
|----|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|--------------------------------------------------|
| 1 | Intended as aide to radiologists' reviewing of images of<br>masses in the human female breast | yes | yes |
| 3 | Windows 2000/XP based DICOM File Set Reader (FSR)<br>compatible platform may be installed on a PC, Notebook or<br>Workstation with HL7 compatibility | yes | yes |
| 4 | User interface designed to follow typical clinical workflow<br>patterns to process, review and analyze digital images | yes | yes |
| 5 | User may annotate, tag, calibrate, measure and<br>automatically record selected views | yes | yes |
| 6 | Automatically generates reports with user annotations<br>during image analysis process. Allows saving custom<br>annotations of images of confirmed cases to be used for<br>future reference and education | yes | yes |
| 7 | Annotations is based on ACR ABCFL Breast Imaging Atlas | yes | yes |
| 8 | Report format supports compliance with ACR ABCFL | yes | yes |
| 9 | Outputs may be viewed and sent to standard film or paper<br>printers or sent electronically to intranet web server or<br>other HL7 compliant media format | yes | yes |
| 10 | Display multiple and full study views | yes | yes |
| 11 | Radiologist's BI-RADS assessment based on visual<br>impressions of the user ABCFL | yes | yes |
The intended use, design, function, and performance characteristics for BRC are substantially equivalent to the predicate device listed above. It is the opinion of Almen Laboratories, Inc., that BRC raises no new issues of safety and effectiveness compared to these predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines representing the wings and a stylized representation of the body and legs.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ALMEN LABORATORIES, Inc. % Mr. Roger Schneider Consultant Medical & Radiation Technology 6319 Massachusetts Avenue BETHESDA MD 20816
MAY 1 6 2008
Re: K072258
Trade/Device Name: BI-RADS Companion Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 20, 2008 Received: March 21, 2008
Dear Mr. Schneider:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K072258
Device Name: BI-RADS Companion
Indications For Use:
BI-RADS Companion™ ("BRC") is a computer-aided image viewing and report generating tool designed to assist radiologists in the analysis of ultrasound images of breast masses by making the ACR "BI-RADS Classification Form and Lexicon®" ("ABCFL") available in electronic format. The user can view any selected image from a case and segment the lesion from the background by applying a user selected threshold or outlining the lesion manually with a cursor.
The user can annotate, tag, measure, and record selected views. Under user control and review, the software electronically assembles reports from inputs selected by the user from the ultrasound ABCFL during the image interpretation process. The assembled report complies with the ABCFL for ultrasound. The output may be viewed and sent to electronic storage media, or standard film or paper printers or other document manipulation programs such as Adobe or Microsoft.
When used by a skilled radiologist, this device provides information that may be useful in image interpretation. Patient management decisions should not be based solely on forms generated by BRC.
The ultrasound images displayed by BRC must not be used for primary diagnostic interpretation.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Nancy C Brogdon
ctive. A Radiological De 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.