IMAGE-ARENA APPLICATIONS AND RESEARCH-ARENA APPLICATIONS
K071232 · Tomtec Imaging Systems GmbH · LLZ · Jun 20, 2007 · Radiology
Device Facts
Record ID
K071232
Device Name
IMAGE-ARENA APPLICATIONS AND RESEARCH-ARENA APPLICATIONS
Applicant
Tomtec Imaging Systems GmbH
Product Code
LLZ · Radiology
Decision Date
Jun 20, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Image-Arena and Research-Arena Platform Software is intended to serve as a data management platform for clinical application packages. As the Image-Arena and Research-Arena Applications software tool package is modular structured, the clinical applications packages are indicated as software packages for analysis of the left ventricle in heart failure patients, to analyze pathologies related to the Mitral Valve and for analysis of the right ventricle in all patients with a need of right heart function diagnosis.
Device Story
Software platform for retrieval, storage, analysis, and reporting of digital ultrasound studies (2D, 3D, 4D). Inputs include DICOM, HPSONOS, GE, and TomTec file formats, plus analog video. Operates on standard PC hardware (Windows XP/2000). Used by clinicians/researchers in hospitals or physician practices for cardiology, radiology, neurology, gastroenterology, urology, surgery, obstetrics, and gynecology. Performs tomographic reconstructions and surface rendering. Provides offline measurements and report generation. Facilitates clinical decision-making by organizing and analyzing ultrasound data across multiple imaging modes (B-mode, M-mode, Doppler, etc.).
Clinical Evidence
No clinical data provided. The device relies on non-clinical software verification and validation testing performed at the module and system level according to internal protocols.
Technological Characteristics
Software-based image management and analysis platform. Runs on Intel Pentium-based PCs with Windows XP/2000. Supports DICOM and proprietary file formats. Modular architecture for 2D/3D/4D image processing, tomographic reconstruction, and surface rendering. Connectivity via standalone or networked workstations.
Indications for Use
Indicated for analysis of the left ventricle in heart failure patients, analysis of mitral valve pathologies, and analysis of the right ventricle in patients requiring right heart function diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
TomTec Image-Arena Applications and Research-Arena Applications (K040546)
Submission Summary (Full Text)
{1}------------------------------------------------
K071232
Image /page/0/Picture/2 description: The image shows the logo for Tomtec Imaging Systems. The logo is in black and white, with the word "Tomtec" in large, bold letters. Below the word "Tomtec" is the phrase "Imaging Systems" in smaller letters. The logo is simple and modern, and it is likely used to represent the company's brand.
Jan. 29th, 2007
# Special 510(k) Summary
Image-Arena Platform 3,x Research-Arena Platform 2 x Echo-Com 3.x lmage-Com 3.x 4D Cardio-View 2.x 4D LV-Analysis 2.5x 4D RV-Function 1.x 4D MV-Assessment 1.x 4D LV-Function 2.x
#### Owner's Name and Address
TomTec Imaging Systems GmbH Edisonstrasse 6 D-85716 Unterschleissheim
#### Contact Person
Inge Scheidt QM & RA Manager ++49-89-32175-515 Phone ++49-89-32175-750 Fax
## Common, Classification & Proprietary Names
| Common Name: | Various Ultrasound Image Analysis Software &<br>System |
|----------------------|--------------------------------------------------------|
| Classification Name: | Ultrasonic Pulsed Echo Imaging System |
| Proprietary Name(s): | Image-Arena and Research-Arena Applications |
Image-Arena Platform 3.x Research-Arena Platform 2.x Echo-Com 3.x Image-Com 3.x 4D Cardio-View 2.x 4D LV-Analysis 2.5x 4D RV-Function 1.x 4D MV-Assessment 1.x 4D LV-Function 2.x
Image /page/0/Picture/13 description: The image shows a collection of black, angled lines arranged in a somewhat curved pattern. The lines are short and thick, and they are oriented at various angles, creating a sense of disarray. The overall arrangement of the lines gives the impression of a winding path or a scattered collection of objects. The lines are all black, and the background is white.
page 1 of 5
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for TOMTEC Imaging Systems. The logo is in black and white, with the word "TOMTEC" in large, bold letters. Below the word "TOMTEC" is the phrase "IMAGING SYSTEMS" in smaller letters. The logo is simple and modern, and it is likely used to represent the company's brand.
#### Predicate Device
TomTec K040546
Image-Arena Applications Research-Arena Applications
### Device Description
The hardware requirements are based on an Intel Pentium high performance computer system and Microsoft® Windows XP Professional™ or Microsoft® Windows 2000 Professional™ Operating System standards.
The Image-Arena/Research-Arena Applications are a software tool package designed for analysis, documentation and archiving of ultrasound studies in multiple dimensions. The Image-Arena/Research-Arena Applications software tools are modular structured and consist of different software modules, combining the advantages of the previously FDA 510(k) cleared TomTec software product line Image-Arena Applications and Research-Arena Applications. The different modules can be combined on the demand of the users to fulfil the requirements of a clinical researcher or routine oriented physician.
The new Image-Arena/Research-Arena Applications offer features to import different digital 2D, 3D and 4D (dynamic 3D) image formats based on defined file format standards (DICOM-, HPSONOS-, GE-,TomTec- file formats) as well as analogue video acquisition in one system, thus making image analysis independent of the ultrasound-device or other imaging devices used. Offline measurements, documentation in standard report forms, the possibility to implement user-defined report templates and instant access to the stored data through digital archiving make it a flexible tool for image analysis and storage of different imaging modalities data including B-mode, M-mode, Pulsed (PW) Doppler mode, Continuous (CW) wave Doppler mode, Power Amplitude Doppler mode, Color Doppler mode, Doppler Tissue Imaging and 3D/4D imaging modes.
#### Intended Use
The Image-Arena/Research-Arena Applications software tool package is intended to retrieve, store, analyze and report digital ultrasound studies. The Image-Arena Platform and the Research-Arena Platform are based on a SQL - database intended as image management system especially for medical ultrasound studies. The Image-Arena/Research-Arena Applications software can import certain digital 2D or 3D image file formats for 2D/3D and 4D tomographic reconstructions and surface rendering.
The software is suited to stand-alone workstations as well as networked multisystem installations and is therefore an image management system for research and routine use in both physician practices and hospitals. It is intended as a general purpose digital medical image processing tool for cardiology, radiology, neurology, gastroenterology, urology, surgery, obstetrics and gynecology.
Image /page/1/Picture/11 description: The image shows a collection of black, angled lines arranged in a somewhat curved formation. The lines are of similar length and thickness, and they are positioned close together, creating a dense pattern. The overall impression is abstract, with the lines resembling a series of connected, angular shapes.
summary
page 2 of 5 0014
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Tomtec Imaging Systems. The logo is in black and white, with the word "Tomtec" in large, bold letters. Below the word "Tomtec" are the words "Imaging Systems" in smaller letters. The logo is simple and modern, and it is likely used to represent the company's brand.
# Technological Characteristics Comparison
The Image-Arena/Research-Arena Applications software tool package is modular structured and consists of different software modules, combining the advantages of the previously FDA cleared software product:
Image-Arena Applications Research-Arena Applications K040546
Image /page/2/Picture/4 description: The image shows a series of black, right-angle shapes arranged in a curved line. The shapes are oriented in a way that suggests movement or direction, creating a dynamic visual effect. The arrangement of the shapes gives the impression of a winding path or a chain of interconnected elements. The overall composition is simple yet visually engaging due to the repetition and pattern of the shapes.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for TOMTEC Imaging Systems. The logo is black and white and features the word "TOMTEC" in large, bold letters. Below the word "TOMTEC" is the phrase "IMAGING SYSTEMS" in smaller letters. The logo is simple and modern.
Image /page/3/Figure/1 description: The image shows a diagram of the predicate devices and a new device. The predicate devices include Image-Arena Applications and Research-Arena Applications on Image-Arena Platform 3.x and Research-Arena Platform 1.x. The new device includes Image-Arena Applications and Research-Arena Applications on Image-Arena Platform 3.x and Research-Arena Platform 2.x. The diagram also shows the generic interface, database and data management system, and device modification.
0016
page 4 of 5
summary
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Tomtec Imaging Systems. The logo is in black and white, with the word "Tomtec" in large, bold letters. Below the word "Tomtec" are the words "Imaging Systems" in smaller letters. The logo is simple and modern, and it is likely used to represent the company's brand.
#### Discussion according non-clinical performance data testing
Testing was performed according to internal company procedures. Software testing and validation were done at the module and system level according to written test protocols established before testing was conducted. Test results were reviewed by designated technical professionals before software proceeded to release.
#### Discussion according clinical performance data testing
The overall product concept was clinically accepted and the clinical test results support the conclusion that the device is as safe as effective, and performs as well as or better than the predicate device.
#### Test Conclusions of non-clinical and clinical performance data
Test results support the conclusion, that the device is as safe as effective, and performs as well as or better than the predicate device.
Munich, Jan. 2981, 2007
J. Sibert
Image /page/4/Picture/10 description: The image shows a collection of black, angled lines arranged in a somewhat curved pattern. The lines are short and thick, and they appear to be clustered together. The overall impression is abstract, with no clear objects or figures discernible.
page 5 of 5
summary
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is black. The background is white. The text is centered in the image.
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized abstract symbol resembling a bird or a person reaching upwards. The symbol is positioned to the right of the text, which is arranged in a circular pattern around the symbol. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA".
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Ms. Inge Scheidt QM & RA Manager TomTec Imaging Systems GmbH Edisonstrasse 6 Unterschleissheim, Bavaria D-85716 GERMANY
Re: K071232
Trade/Device Name: Image-Arena Platform 3.0; Research-Arena Platform 2.0; Echo-Com 3.x; Image-Com 3.x; 4D Cardio-View 2.x; 4D LV-Analysis 2.5x; 4D RV-Function 1.x; 4D MV-Assessment 1.x; and, 4D LV-Function 2.x Regulation Number: 21 CFR §892.2050
Regulation Name: Picture archiving and communications systems Regulatory Class: II Product Code: LLZ Dated: April 19, 2007 Received: May 7, 2007
Dear Ms. Scheidt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Aot include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/5/Picture/11 description: The image is a circular logo or emblem. The text "1906-2006" is at the top, and the letters "FDA" are in the center. Below the letters, the word "Centennial" is written in a cursive font. Three stars are at the bottom of the logo. The text around the circle says "FDA Centennial: A Century of Protecting the Public Health".
Protesting and Promoting Public Health
{7}------------------------------------------------
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) ( premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
> 21 CFR 876.xxxxx 21 CFR 884.xxxx 21 CFR 892 xxxxx Other
(Gastroenterology/Renal/Urology) (Obstetrics/Gynecology) (Radiology);
240-276-0115 240-276-0115 240-276-0120 240-276-0100
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
| | Indications for Use |
|--------------------------|---------------------|
| 10(k) Number (if known): | K071232 |
Device Name:
5
Image-Arena Platform 3.0 Research-Arena Platform 2.0 Echo-Com 3.x Image-Com 3.x 4D Cardio-View 2.x 4D LV-Analysis 2.5 x 4D RV-Function 1.x 4D MV-Assessment 1.x 4D LV-Function 2.x
Indications for Use:
The Image-Arena and Research-Arena Platform Software is intended to serve as a data management platform for clinical application packages.
As the Image-Arena and Research-Arena Applications software tool package is modular structured, the clinical applications packages are indicated as software packages for analysis of the left ventricle in heart failure patients, to analyze pathologies related to the Mitral Valve and for analysis of the right ventricle in all patients with a need of right heart function diagnosis.
Hulbert
(Division Sian-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number .
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
0012
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.