← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K070618

# DICOMPACS, MODEL 5 (K070618)

_Oehm Und Rehbein GmbH · LLZ · Apr 25, 2007 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K070618

## Device Facts

- **Applicant:** Oehm Und Rehbein GmbH
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Apr 25, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

dicomPACS® is intended to be used for the acquisition, storage, communication and viewing of medical images. dicomPACS® receives images from modalities including but not limited to digital X-ray, CT, MRI, ultrasound, scanners and video sources. It archives and displays these images for the use of medical specialists who are qualified to operate radiological equipment and to record and diagnose medical images. It provides the user with a range of tools to assist them in viewing the images, such as zoom, filters and measurements, and with facilities to exchange images with other specialists. dicomPACS® can be integrated with an existing RIS or KIS system for an integrated electronic patient record.

## Device Story

dicomPACS is a modular Picture Archiving and Communications System (PACS) software. It acquires medical images from modalities (X-ray, CT, MRI, ultrasound, scanners, video) via DICOM or direct import. The system archives images in a database and provides a viewer for display on computer monitors. It includes tools for window leveling, rotation, zoom, measurements, and annotations. Used by radiologists and orthopaedists in clinical settings to assist in diagnosis and image exchange. It can integrate with RIS or KIS for electronic patient records. The software does not control life-sustaining devices; qualified medical personnel perform all interpretations. Benefits include centralized image management, improved workflow, and diagnostic support through image processing tools.

## Clinical Evidence

Bench testing only. The device was tested according to specifications documented in the notification and conforms to the DICOM standard. No clinical data provided.

## Technological Characteristics

Autonomous software system; no hardware. Runs on standard hardware platforms using MS Windows XP or MS Windows 2003 server. Conforms to DICOM standard. Modular architecture includes viewer, control center, database management, and acquisition modules. No life-sustaining control functions.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- HIPAX Medical Imaging Software ([K052411](/device/K052411.md))
- eFilm Workstation with Modules ([K020995](/device/K020995.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/12 description: The image contains the logo for dicomPACS. The logo includes the text "dicomPACS" in a stylized font, with the "PACS" portion in a lighter shade. Below the text, there are the words "Digital Image Management" in a smaller, lighter font. To the left of the text, there are two abstract graphic elements, one of which appears to be an image of a hand or bones.

by. www.oehm-rehbein.com

# 510(k) Summary

K070618

Summary of functions of the device and its major components provided as part of the Premarket Notification for dicomPACS®.

| Date:                  | 28 February 2007                                   | APR 25 2007 |
|------------------------|----------------------------------------------------|-------------|
| Company Name:          | Oehm und Rehbein                                   |             |
| Address:               | Waldemarstr. 20 g/h<br>18057 Rostock<br>Germany    |             |
| Contact Person:        | Franziska Günther                                  |             |
| E-Mail:                | franziska.guenther@oehm-rehbein.de                 |             |
| Tel:                   | +49-381-20361291                                   |             |
| Fax:                   | +49-381-2036111                                    |             |
| Device Trade Name:     | dicomPACS®                                         |             |
| Device Common Name:    | Picture Archiving and Communications System (PACS) |             |
| Product Code:          | LLZ                                                |             |
| Regulation No.:        | 892.2050                                           |             |
| Device Classification: | Class II                                           |             |

### Description

Predicate Devices:

dicomPACS® is a Picture Archiving and Communications System (PACS), as are the two Predicate Devices HIPAX and eFilm. All of them have been developed to acquire, store, communicate, display and process medical images, for example X-ray, CR, CT, MRI, and ultrasound images etc. They offer features (e.g. window levelling, zoom, measurements, annotations etc.) routinely used by medical professionals, such as radiologists and orthopaedists, and required of all PACS solutions.

HIPAX Medical Imaging Software (K052411) eFilm Workstation with Modules (K020995)

dicomPACS® has a modular system architecture. It consists of the basic application dicomPACS-Viewer for image viewing and processing, the dicomControlCenter as application for image storage and communication and a number of other modules for database management, image acquisition, printing etc.

dicomPACS® conforms to the DICOM (Digital Imaging and Communications in Medicine) standard.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains a logo with the text "dicomPACS" in a stylized font. Below the logo, there are smaller words that say "Digital Image Management". To the left of the text, there is a graphic of a hand with rays emanating from it, set against a black background. At the bottom of the image, the text "by www.oehm-rehbein.com" is displayed.

#### Intended Use

dicomPACS® is intended to be used for the acquisition, storage, communication and viewing of medical images.

dicomPACS® receives images from modalities including but not limited to digital X-ray, CT, MRI, ultrasound, scanners and video sources. It archives and displays these images for the use of medical specialists who are qualified to operate radiological equipment and to record and diagnose medical images. It provides the user with a range of tools to assist them in viewing the images, such as zoom, filters and measurements, and with facilities to exchange images with other specialists. dicomPACS® can be integrated with an existing RIS or KIS system for an integrated electronic patient record.

All these tools are also offered by both HIPAX and eFilm.

#### Technological Characteristics

dicomPACS® is an autonomous software and involves no hardware. This largely applies to the predicate devices HIPAX and eFilm. There is no difference to eFilm.

dicomPACS® runs on any hardware platform meeting the minimum system requirements.

dicomPACS® can be used with the server operating systems MS Windows XP and the MS Windows 2003 server.

dicomPACS® does not control any life-sustaining devices. Specialists with adequate expert knowledge for competent human intervention interpret displayed or printed images and information.

#### Testing

dicomPACS® has been tested according to the specifications documented in this notification (Section 14 g). It conforms to the DICOM standard as laid out in the included DICOM Conformance Statement.

#### Conclusion

dicomPACS® is a medical device. dicomPACS® provides functionality comparable to that of its Predicate Devices and is intended for the same user and patient groups as its Predicate Devices.

This premarket notification contains sufficient information to establish substantial equivalence to the Predicate Devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus symbol. The image is in black and white and appears to be a scan or photocopy.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Ms. Franziska Günther Assistant to the Management Oehm und Rehbein GmbH Waldemarstr. 20 g/h 18057 Rostock GERMANY

APR 2 5 2007

Re: K070618

Trade/Device Name: dicomPACS® Regulation Number: 21 CER 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 28, 2007 Received: March 5, 2007

#### Dear Ms. Günther:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image shows a circular logo with the letters FDA in the center. The logo also includes the words "Centennial" and the years "1906-2006". The logo appears to be a commemorative emblem for the centennial anniversary of the Food and Drug Administration. The image also contains some text below the logo, but it is not clear enough to read.

{3}------------------------------------------------

Page 2 --

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy Cbrogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): K070618

dicomPACS® Device Name:

Indications For Use:

dicomPACS® is a software system for the administration, archiving, improvement and compression of medical image data for diagnosis. The images are either acquired from imaging modalities via DICOM or imported directly. All images are archived in a database as DICOM compliant files. The data is displayed on a computer monitor for diagnosis. dicomPACS® also provides services for administering the data.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 MP resolution and meets other technical specifications reviewed and accepted by FDA.

Functions to be carried out using dicomPACS® are, for example, but not limited to, adjustment of window leveling, rotation, zoom, and measurements.

dicomPACS® is meant to be used by qualified medical personnel only. All users must be qualified to create and diagnose radiological image data.

Prescription Use: × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use: (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) .

David A. Symm

(Division Sian-Off) Division of Reproductive. Abdominal, and Radiological Devices 510(k) Number

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K070618](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K070618)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
