← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K070244

# AEGIS (K070244)

_Sentinelle Medical, Inc. · LLZ · Feb 9, 2007 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K070244

## Device Facts

- **Applicant:** Sentinelle Medical, Inc.
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Feb 9, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Intended Use

Aegis is a software application intended for use in analyzing magnetic resonance imaging (MRI) medical images as well as other modality images. Aegis includes software to support the use of interventional breast coils and MR fiducial localization devices to perform MR-guided breast interventional procedures (BRIGHT"). Its primary goal is to identify where and how deep a biopsy or localization needle should be inserted into an imaged breast in order to strike a targeted lesion or region of interest, as chosen by a trained medical professional. Aegis receives images and data from various sources (including CT, MR, US, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways and imaging sources). Aegis can be used to commonicate, process, and display medical images. Users have access to various image processing and measurement tools to assist them in wiewing images. These user-defined post-processing functions include image subtractions, multiplanar reformats, maximum intensity projections, and segmenting of regions based on enhancement characteristics. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mamnographic images may only be interpreted using a FDA approved monitor that offers at least 5 MPixel resolution and meets other technical specifications reviewed and accepted by the FDA. Typical users of Aegis are trained medical professionals, including radiologists, technologists and clinicians. When interpreted by a skilled user, this device provides information that may be useful in screening and diagnosis. Patient management decisions should not be made solely of the results of Aegis analysis.

## Device Story

Aegis is a stand-alone PACS visualization software for breast imaging and interventional procedures. It receives DICOM 3.0 images from various sources (CT, MR, US, scanners). The software provides tools for image subtraction, multiplanar reformatting, maximum intensity projections, and segmentation based on enhancement characteristics. It is operated by trained medical professionals (radiologists, technologists) on standard laptops or computers. The primary function is to calculate biopsy needle trajectory and depth to target lesions identified by the clinician. The output is displayed for the clinician to guide interventional procedures. It does not control life-sustaining devices or contact the patient. It serves as an adjunct to clinical decision-making; patient management decisions should not rely solely on its output.

## Clinical Evidence

No clinical data provided. Safety and effectiveness were established through product development validation, risk analysis, and planned verification and validation testing processes.

## Technological Characteristics

Stand-alone software package for installation on laptops or computers. Functions as a PACS component for digital image processing and measurement. Receives images via medical imaging network. Does not contact patient. No specific hardware requirements other than standard computer specifications. Software-based image processing.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- CADStream™ Version 4.0 ([K043216](/device/K043216.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070244

#### 9 2007 FEB

## 510(k) Summary of Safety and Effectiveness

This summary of safety and effectiveness is provided as part of the Premarket Notification for Aegis, in accordance with SMDA 1990.

| Date Prepared: | January 19th/2007 |
|----------------|-------------------|
|----------------|-------------------|

### Identification of Submitter:

Submitted By:

Sentinelle Medical Inc. 3080 Yonge St., Suite 6020 Toronto, Ontario, Canada M4N 3N1

Contact Name: Contact Email: Contact Telephone: Contact Fax:

Stephen Leung stephen.leung@sentinellemedical.com (416) 482-3831 (416) 482-3807

### Identification of Product:

Device Trade Name: Device Common Name: Regulation Number: Device Classification: Classification Name:

Aegis Picture Archiving Communications System (PACS) CFR 892.2050 Class II Image Processing System

#### Marketed Device:

Predicate Device: Predicate Device Manufacturer: Predicate Device 510(k) Number: Date Received: Decision Date: Decision: Panel Code Device Reviewed by: Panel Code Device Classified by: Product Code: Regulation Number: Device Classification:

CADStream™ Version 4.0 Confirma, Inc. K043216 11/19/2004 11/22/2004 Substantially Equivalent Radiology Radiology LLZ 892.2050 Class II

## Device Description

Aegis is one of the components of a PACS (Picture Archiving and Communications System), Aegis is visualization software designed for breast imaging and intervention procedures. Aegis receives DICOM 3.0 images over a medical imaging network where its primary goal is to identify where and how deep a biopsy needle should be inserted into an imaged breast in order to strike a targeted lesion or region of interest, as chosen by a trained medical professional.

1

{1}------------------------------------------------

## Indications for Use

Aegis is a software application that is intended for use in analyzing magnetic resonance imaging (MRI) medical images as well as other multi-modality images. Its primary goal is to identify where and how deep a biopsy or localization needle should be inserted into an imaged breast in order to strike a targeted lesion or region of interest, as chosen by a trained medical professional. Aeris receives images and data from various sources (including but not linited to CT, MR, US, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). Aegis can be used to communicate, process, and display medical images. Users have access to various image processing and measurement tools to assist them in viewing images. These user-defined post-processing functions include image subtractions, multiplanar reformats, maximum intensity projections, and segmenting of regions based on enhancement characteristics.

The digitized mammographic images and/or ultrasound images displayed by Aegis on the laptop display (or on any display not approved by the FDA for such purposes) must not be used for primary diagnostic interpretation.

Typical users of Aegis are trained medical professionals, including but not limited, to radiologists, technologists and clinicians. When interpreted by a skilled user, this device provides information that may be useful in screening and diagnosis. Patient management decisions shoulding be made solely on the results of Aegis analysis.

## Technological Characteristics

Aegis is a stand-alone software package that is pre-installed onto laptops before being sold to customers. It can also be installed on any computer as long as it meets the ning wom requirements stated in the User Manual.

The system allows digital image processing and measurement capability and can transmit images from remote devices to itself over a medical imaging network.

Aegis does not contact the patient, nor does it control any life-sustaining devices. A physician providing ample opportunity for competent human intervention interprets images and information being displayed.

## Comparison with Predicate Devices

Aegis is substantially equivalent to the following image processing system used by radiologists:

| Confirma CADStream Version 4.0 |               |
|--------------------------------|---------------|
| Manufacturer:                  | Confirma Inc. |
| 510(k) Number:                 | K043216       |

This workstation allows easy selection, review, processing, and media interchange of multimodality images from a variety of diagnostic imaging systems.

## Conclusion

Aegis provides additional features to further integrate radiological and interventional workflow. The potential hazards have been studied and controlled as part of the product development validation houring risk analysis, test and design considerations, and planned verification and validation testing processes. Aegis provides images and functionslity comparable to the breadicate device.

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

# 9 2007

Sentinelle Medical, Inc. c/o Mr. Daniel W. Lehtonen Intertek Testing Services 2307 East Aurora Road Unit B7 TWINSBURG OH 44087

Re: K070244

Trade/Device Name: Aegis Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 23, 2007 Received: January 25, 2007

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image is a circular logo or emblem. The text "1906-2006" is at the top of the logo. The letters "PDA" are in the center of the logo. The word "Centennial" is below the letters. Three stars are below the word "Centennial".

Protecting and Promoting Public Health

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indication(s) For Use

Ko70244 510(k) Number:

Device Name: Aegis

#### Indications for Use:

Aegis is a software application intended for use in analyzing magnetic resonance imaging (MRI) medical images as well as other modality images.

Aegis includes software to support the use of interventional breast coils and MR fiducial localization devices to perform MR-guided breast interventional procedures (BRIGHT"). Its primary goal is to identify where and how deep a biopsy or localization needle should be inserted into an imaged breast in order to strike a targeted lesion or region of interest, as chosen by a trained medical professional.

Aegis receives images and data from various sources (including CT, MR, US, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways and imaging sources). Aegis can be used to commonicate, process, and display medical images. Users have access to various image processing and measurement tools to assist them in wiewing images. These user-defined post-processing functions include image subtractions, multiplanar reformats, maximum intensity projections, and segmenting of regions based on enhancement characteristics.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mamnographic images may only be interpreted using a FDA approved monitor that offers at least 5 MPixel resolution and meets other technical specifications reviewed and accepted by the FDA.

Typical users of Aegis are trained medical professionals, including radiologists, technologists and clinicians. When interpreted by a skilled user, this device provides information that may be useful in screening and diagnosis. Patient management decisions should not be made solely of the results of Aegis analysis.

Prescription Use (Part 21 CFR 801 Subpart D):

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

# Concurrence of CDRH, Office of Device Evaluation (ODE)

David G. Seymour

(Division Sign-Off)
Division of Reproductive, Abdominal
and Radiological Devices
510(k) Number K070244

Page 1 of

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K070244](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K070244)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
