← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K061995

# XCELERA (K061995)

_Philips Medical Systems North America Co. · LLZ · Sep 6, 2006 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K061995

## Device Facts

- **Applicant:** Philips Medical Systems North America Co.
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Sep 6, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Intended Use

Philips Xcelera software product is an integrated multimodality image and information system designed to perform the necessary functions required for import, export, storage, archival, review, analysis, quantification, reporting and database management of digital medical images. Philips Xcelera software is an integrated multimodality image and information system, designed to perform the necessary functions required for import /export/ storage / archiving / review / analysis/ quantification (for example: area, circumference, volume, velocity, length, percent, time, ejection fraction, pressure gradient, LV volumes) / reporting and database management of digital cardiovascular images, waveforms and data related to cardiology. Xcelera offers support for third party plug-ins in order to enable the use of commercially available tools and for analysis, quantification and reporting Xcelera offers support to launch specified 3th party programs from the user interface (Desktop Integration). It allows multiple users fast access to, and exchange of specific and/or multiple cardiology exams. Philips Xcelera software runs on standard information technology hardware and software. The Xcelera proprietary software product utilizes the standard Microsoft Windows Operating System and user interface. Communication and data exchange are done using standard TCP/IP, DICOM and HL7 protocols. Philips Xcelera will also be made available for use on specified Cardiovascular Monitoring Systems, which use suitable hardware components. The modular design allows configurability to tailor the image import, archive and communications solution to one's particular budgetary and performance needs. The number of modalities and reporting and/or viewing sites can be configured per system.

## Device Story

Philips Xcelera is an integrated multimodality image and information system (PACS) for cardiology. It ingests digital cardiovascular images, waveforms, and data; performs storage, archival, review, analysis, and quantification (e.g., area, volume, ejection fraction, pressure gradient); and generates reports. The system runs on standard IT hardware using the Microsoft Windows OS. It supports third-party plug-ins and desktop integration to launch external analysis tools. Used in clinical environments, it allows multiple users to access and exchange cardiology exams. The system utilizes standard TCP/IP, DICOM, and HL7 protocols for data exchange. It is configurable to support various modalities and viewing sites, aiding clinicians in diagnostic decision-making and workflow management.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics, intended use, and compliance with established standards (DICOM/HL7).

## Technological Characteristics

Integrated multimodality PACS software. Runs on standard IT hardware with Microsoft Windows OS. Connectivity via TCP/IP, DICOM, and HL7 protocols. Modular design supports third-party plug-ins and desktop integration. No specific materials or energy sources defined as it is a software-based system.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Philips Medical Systems, Harmony ([K022788](/device/K022788.md))
- Siemens Medical Systems Inc., LEONARDO syngo Cardiology Workstation ([K042203](/device/K042203.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Koli995

Image /page/0/Picture/1 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are large and black, contrasting with the white background. The word is presented in all capital letters and is horizontally oriented.

## Philips Medical Systems

SEP - 6 2006

# 510(k) Summary

Philips Xcelera

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

I General Information

| Company Name:        | Philips Medical Systems North America Company                                                                 |  |
|----------------------|---------------------------------------------------------------------------------------------------------------|--|
| Address:             | 22100 Bothell Everett Highway<br>Bothell Washington 98021-8431<br>USA                                         |  |
| Contact Person       | Lynn T. Harmer                                                                                                |  |
| Telephone Number:    | 425-478-7312                                                                                                  |  |
| Prepared (date):     | 2006 May 12                                                                                                   |  |
| Manufacturing Site:  | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6<br>5684 PC Best, The Netherlands                      |  |
| Device Name:         | Philips Xcelera                                                                                               |  |
| Classification Name: | Picture Archiving and Communication System                                                                    |  |
| Regulation number    | 892.2050                                                                                                      |  |
| Classification:      | Class: II                                                                                                     |  |
| ProCode:             | 90 LLZ                                                                                                        |  |
| Common/Usual Name:   | Workstation                                                                                                   |  |
| Predicate Devices:   | Philips Medical Systems,<br>Harmony<br>Siemens Medical Systems Inc.,<br>LEONARDO syngo Cardiology Workstation |  |

{1}------------------------------------------------

#### I I Information Supporting Substantial Equivalence Determination

#### System Description:

Philips Xcelera software is an integrated multimodality image and information system, designed to perform the necessary functions required for import /export/ storage / archiving / review / analysis/ quantification / reporting and database management of digital cardiovascular images, waveforms and data related to cardiology,

Xcelera offers support for third party plug-ins in order to enable the use of commercially available tools and for analysis, quantification and reporting Xcelera offers support to launch specified 3th party programs from the user interface (Desktop Integration). It allows multiple users fast access to, and exchange of specific and/or multiple cardiology exams.

Philips Xcelera software runs on standard information technology hardware and software. The Xcelera proprietary software product utilizes the standard Microsoft Windows Operating System and user interface. Communication and data exchange are done using standard TCP/IP, DICOM and HL7 protocols.

Philips Xcelera will also be made available for use on specified Cardiovascular Monitoring Systems, which use suitable hardware components

The modular design allows configurability to tailor the image import, archive and communications solution to one's particular budgetary and performance needs. The number of modalities and reporting and/or viewing sites can be configured per system.

### Intended Use:

Philips Xcelera software product is an integrated multimodality image and information system designed to perform the necessary functions required for import, export, storage, archival, review, analysis, quantification, reporting and database management of digital medical images.

### General Safety and Effectiveness:

The Xcelera complies with ACR/NEMA DICOM digital imaging communication standard.

### Conclusion:

The Philips Xcelera does not introduce new indications for use, nor does the use of the device result in any new potential hazard. Philips Medical Systems considers the Xcelera to be substantially equivalent to the Philips Medical Systems, Harmony (K022788) and the Siemens Medical Systems Inc., LEONARDO syngo Cardiology Workstation (K042203).

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the eagle symbol.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

## SEP - 6 2006

Philips Medical Systems North America c/o Mr. Marc M. Mouser Program Reviewer Underwriters Laboratories, Inc. 2600 N. W. Lake Road CAMAS WA 98607-8542

K061995 Re:

Trade/Device Name: Philips Xcelera Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 4, 2006 Received: August 21, 2006

Dear Mr. Mouser:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Booton's ro(t) predantially equivalent (for the indications for relerenced above and nave decemined anceted predicate devices marketed in interstate commerce use stated in the encrosury manations p Medical Device Amendments, or to devices that prob io May 26, 1770, the chaomistic date of the Federal Food, Drug, and Cosmetic
have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosm have been recrassmed in accordance with and portication (PMA). You may, ACt (Act) that do flor require approval or a peneral controls provisions of the Act. The general therefore, mailier the device, subjoct to the general for annual registration, listing of devices, good controls provisions of also ing, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket II your device is classifica (see above) into entrols. Existing major regulations affecting your Approval), it thay be subject to such adamons. Title 21, Parts 800 to 898. In addition, FDA active can be round in the cements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image shows a circular logo with the text "FDA Centennial" and the years "1906-2006". The logo also features three stars at the bottom. The text is arranged around the circle's perimeter, and the FDA acronym is prominently displayed in the center.

d Promoting Public He

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r toase of acressionation that your device complies with other requirements of the Act that I Drederal statutes and regulations administered by other Federal agencies. You must or any I bathar all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI IT Fart 6077, adoling (21 CFR Part 820); and if applicable, the electronic form in the quant) Bystems (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) rins letter wif anon you to ough finding of substantial equivalence of your device to a legally premaired predicated on: "The Dr. In a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rf you desire spoorned at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)         | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                     | 240-276-0120 |
| Other          |                                 | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" / tion, prease note and regain other general information on your responsibilities under (21 CF R F art 60 : 37). I coof Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.

Sincercly yours,

Nancy C. Brogdon

Nancy C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(K) Number (if known):

Device Name: Philips Xcelera

Philips Xcelera software is an integrated multimodality image and information system, designed to perform the necessary functions required for import /export/ storage / archiving / review / analysis/ quantification (for example: area, circumference, volume, velocity, length, percent, time, ejection fraction, pressure gradient, LV volumes) / reporting and database management of digital cardiovascular images, waveforms and data related to cardiology.

Xcelera offers support for third party plug-ins in order to enable the use of commercially available tools and for analysis, quantification and reporting Xcelera offers support to launch specified 3th party programs from the user interface (Desktop Integration). It allows multiple users fast access to, and exchange of specific and/or multiple cardiology exams.

Philips Xcelera software runs on standard information technology hardware and software. The Xcelera proprietary software product utilizes the standard Microsoft Windows Operating System and user interface. Communication and data exchange are done using standard TCP/IP, DICOM and HL7 protocols.

Philips Xcelera will also be made available for use on specified Cardiovascular Monitoring Systems, which use suitable hardware components.

The modular design allows configurability to tailor the image import, archive and communications solution to one's particular budgetary and performance needs. The number of modalities and reporting and/or viewing sites can be configured per system.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Hepson

(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K061995](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K061995)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
