JETSTREAM WORKSPACE

K061029 · Adac Laboratories · LLZ · May 1, 2006 · Radiology

Device Facts

Record IDK061029
Device NameJETSTREAM WORKSPACE
ApplicantAdac Laboratories
Product CodeLLZ · Radiology
Decision DateMay 1, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

JETStream® Workspace is a nuclear medicine image display and processing workstation that provides software applications used to process, analyze, and display medical images/data. The results obtained may be used as a tool, by a nuclear physician, in determining the diagnosis of patient disease conditions in various organs, tissues, and other anatomical structures. The data processed may be derived from any nuclear medicine gamma camera. The JETStream® Workspace system should only be operated by qualified healthcare professionals trained in the use of nuclear medicine equipment.

Device Story

JETStream® Workspace is a Windows-based nuclear medicine workstation (HP XW4300 or HP Compaq nc6230 hardware). It ingests image data from any nuclear medicine gamma camera. The system provides tools for image review, processing, post-processing, hardcopy production, interpretation, and report generation. It supports data management and connectivity for workflow integration. Used in clinical settings by nuclear physicians and technologists to assist in diagnosing patient disease conditions. The output is displayed for clinical review, aiding the physician in determining diagnoses. Benefits include streamlined workflow, standardized image processing, and enhanced diagnostic interpretation capabilities.

Clinical Evidence

No clinical data provided; substantial equivalence is based on technological comparison and system performance.

Technological Characteristics

Windows-based workstation (HP XW4300 or HP Compaq nc6230). Functions as a PACS-related image processing system. Connectivity for data import/export. Software-based image processing and analysis tools.

Indications for Use

Indicated for nuclear medicine image display and processing for patients undergoing diagnostic evaluation of organs, tissues, and anatomical structures. Operated by qualified healthcare professionals trained in nuclear medicine equipment.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 06/02 ADAC Laboratories JETStream® Workspace CONFIDENTIAL 510(k) Premarket Notification Section B. Administrative Information # ADMINISTRATIVE INFORMATION ### I. 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS - A. Submitted By: ADAC Laboratories A Philips Medical Systems Company 540 Alder Drive Milpitas, California 95035 Tel: (408) 468-3042 Fax: (408) 468-3050 Contact Person: - Lori R. Peterson At address above - B. Device Trade Name: JETStream® Workspace Workstation Common Name: Image Processing System Classification Name: Picture Archive and Communication Systems (PACS) ். Predicate Device(s): | Manufacturer | Product Name | 510(k) No. | |-------------------|--------------------------------|------------| | ADAC Laboratories | Pegasys Ultra™ | K993946 | | ADAC Laboratories | Predicate JETStream® Workspace | K042880 | #### D. Device Description: JETStream® Workspace is a Windows®-based Nuclear Medicine workstation for the Nuclear Medicine market segment. The computer system will consist of a Hewlett Packard XW4300 workstation or HP Compaq nc6230 Notebook or their equivalents. The comprehensive tools and features provided with this product, will allow the technologist and/or physician to perform image review, processing of source data, post processing, hardcopy production, interpretation, report generation and contains the utilities necessary to support the workflow and data management between those activities. The system will support connectivity aspects necessary to import and export data as required to accomplish daily work scenarios. E. Intended Use: > JETStream® Workspace is a nuclear medicine image display and processing workstation that provides software applications used to {1}------------------------------------------------ process, analyze, and display medical images/data. The results obtained may be used as a tool, by a nuclear physician, in determining the diagnosis of patient disease conditions in various organs, tissues, and other anatomical structures. The data processed may be derived from any nuclear medicine gamma camera. The JETStream® Workspace system should only be operated by qualified healthcare professionals trained in the use of nuclear medicine equipment. - F. Technological Comparison: The Pegasys Ultra™ (K993946), JETStream® Workspace (Predicate - K042880) and the JETStream® Workspace (modified) have similar indications for use and overall function and perform in a similar manner with respect to, display, review and processing applications, data storage, and system utilities. ## II. CONCLUSION JETStream® Workspace is substantially equivalent to the following predicate devices, Pegasys Ultra™ (K993946) and JETStream® Workspace (Predicate - K042880) based on similar intended use, technological comparison, and system performance. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the department's name around the perimeter. Inside the circle is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 MAY 1 2006 2006 ADAC Laboratories % Mr. Morten Simon Christensen Staff Engineer & FDA Office Coordinator Underwriters Laboratories, Inc. 455 East Trimble Road SAN JOSE CA 95131 Re: K061029 Trade/Device Name: JETStream® Workstation Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 13, 2006 Received: April 14, 2006 Dear Mr. Christensen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/2/Picture/11 description: The image shows a logo for the FDA Centennial 1906-2006. The logo is circular with the words "FDA Centennial" in the center. The words "Protecting and Promoting Public Health" are written in a fancy font below the logo. The logo is black and white. {3}------------------------------------------------ Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K061029 Device Name: JETStream® Workspace Indications For Use: JETStream® Workspace is a nuclear medicine image display and processing workstation that provides software applications used to process, analyze, and display medical images/data. The results obtained may be used as a tool, by a nuclear physician, in determining the diagnosis of patient disease conditions in various organs, tissues, and other anatomical structures. The data processed may be derived from any nuclear medicine gamma camera. The JETStream® Workspace system should only be operated by qualified healthcare professionals trained in the use of nuclear medicine equipment. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE.) Page 1 of David A. Swanson (Division Sian-Off Division of Reproductive, and Radiological Device 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...