K053155 · Cybermed, Inc. · LLZ · Nov 23, 2005 · Radiology
Device Facts
Record ID
K053155
Device Name
VIMPLANT
Applicant
Cybermed, Inc.
Product Code
LLZ · Radiology
Decision Date
Nov 23, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
The Vimplant is intended for use as a software interface for the transfer of imaging information from a CT scanner and also as a pre-operative software for simulation and evaluation dental implant placement and surgical treatment options.
Device Story
Vimplant is dental implant simulation software for desktop PCs; used by dentists/implantologists for pre-operative planning. Inputs: DICOM 3.0 CT scan data. Processing: 2D image reformation (panoramic/cross-sectional); 3D surface rendering from axial images; nerve identification/creation; bone density evaluation. Outputs: 3D models; implant placement simulations; collision warnings (nerve/implant or implant/implant); sinus bone graft volume estimation; measurement reports (length, angle, density). Clinical utility: assists in selecting implant position, angle, diameter, and length; reduces surgical risk; aids in surgical treatment decision-making. Benefits: improved surgical accuracy and planning efficiency.
Clinical Evidence
No clinical data. Bench testing only. Reliability and validity of 3D image construction verified via comparison to V-works™ (K013878).
Technological Characteristics
Software-based dental implant simulation; DICOM 3.0 compliant; surface rendering for 3D construction; Hounsfield unit-based thresholding; collision detection algorithms; runs on standard desktop PCs; outputs DCM, BMP, JPG formats.
Indications for Use
Indicated for use by dentists and implantologists as a pre-operative software interface for CT imaging data to simulate and evaluate dental implant placement and surgical treatment options.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K053/55
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: __ Aug 29, 2005_
1. Company and Correspondent making the submission:
Name - CyberMed, Inc.
Address - #504 SJ Technoville, Gasan-dong 60-19, Geumcheon-gu, Seoul, 153-710, Korea
Telephone -- +82-2-3397-3970
Fax -- +82-2-3397-3971
Contact - Mr. Song Nak Choi/ Manager
Internet - http://www.cybermed.co.kr
# 2. Device :
| Trade/proprietary name | : Vimplant |
|------------------------|-----------------------------------------------|
| Common Name | : Dental implant simulation software |
| Classification Name | : Picture archiving and communications system |
#### 3. Predicate Devices :
| Manufacturer | : MATERIALISE N.V. |
|---------------|----------------------------------------|
| Device | : SimPlant System |
| 510(k) Number | : K033849 (Decision Date - 05/25/2004) |
| Manufacturer: | CyberMed, Inc. |
|----------------|--------------------------------------|
| Device: | V-works™ |
| 510(k) Number: | K013878 (Decision Date - 12/07/2001) |
# 4. Classifications Names & Citations :
21CFR 820.2050, LLZ, Picture archiving and communications system, Class2
# 5. Description :
1) General Description
CyberMed, Inc.
0000026
{1}------------------------------------------------
VImplant™ is a dental implant simulation software for dentists and implantologists. By use of VImplant™, dental practicians can plan and practice their surgery in advance so they can reduce some risks which can happen during their real surgery.
Vimplant™ provides very useful and needful functions. It contains functions of implant simulation, manipulation with 2D and 3D mode, nerve identification, and evaluation of bone density.
By using Vimplant TM, dental practicians can quickly and easily simulate implant surgery on their desktop PCs. This enables them to conduct implant surgery more effectively by isolating the exact implant position site and angle and to assist in deciding the proper implant diameter, length, etc.
#### 2) Main Function
- a. Import DICOM 3.0 data
Users can import DICOM 3.0 data taken from CT to VImplant's database and manage them very easily on the patient list window of VImplant.
## b. 2D image reformation
Vimplant™ has the capability of reformatting panoramic images and cross-sectional images. So dental practicians can get the accurate anatomical knowledge of the patient on each view and know the exact relation of implant with nerve and bone before the implant surgery.
# c. 3D image construction
VImplant™ has the capability of constructing 3D model from original axial images. The method of 3D image construction is Surface rendering and the reliability and validity were already verified in Vworks™. The 3D image can be edited by CT number(Hounsfield value) threshold.
# d. Nerve Creation and Display
Vimplant™ has nerve creation function to reduce some risks in placing an implant. Users can draw a nerve line on axial images, then arrange it on Panoramic view. And then users can see the intersection shape of nerve on each view. Also the nerve line is displayed in 3D model.
- e. Implant Simulation
CyberMed, Inc.
{2}------------------------------------------------
VImplant™ provides users very useful functions for Implant simulation. Users can create and manipulate implants as they want. In the concrete, users can place an implant very easily by the mouse operation and also move and rotate it as they want. And also users can change some properties of the implant such as diameter, color and shape of the fixture and abutment part for their specific implant and it to Implant database. In addition, Vimplant™ shows the implant in 3D model. So users can get the accurate position of implant in 3D model as well as 2D images.
#### f. Collision detection
VImplant™ has Collision detection function to prevent some collisions between nerve and implant or implant and implant as type of warning message.
# g. Sinus bone graft
VImplant™ has Sinus bone graft function which estimate the volume of the material needed for bone grafting surgery.
# h. Measurement and information
VImplant™ provides users needful measurement functions for implant planning such as bone density profile, length, angle. Specially users can measure the density distribution around a placed implant by use of Bone density function. And also users can see the distance between implant and implant or implant and nerve by use of 3D distance calculation function.
# i. Report and Image Library
VImplant™ has the capabilliy of capturing all images and provides simple report form. Users can manage their images very easily in Image Library and also can send them to some storage such as PACS server.
# 3) Information of the image format
Vimplant™ can load only DCM files and save results as DCM, BMP and JPG files.
- DCM : DICOM (Digital Image Communication in Medicine) is a Standard Protocol to exchange and transfer the data acquired by Medical Image devices such as a CT, MR, 3D US, etc. It is designated as a Standard Protocol by ACR-NEMA (American College of Radiology-National Electrical Manufacturers Association) and now adopted by most
CyberMed, Inc.
11111: 111: 18
{3}------------------------------------------------
Medical Imaging Devices. VimplantTM 2.0 is adaptable technically for all data of DICOM 3.0.
Reference : Digital Imaging and Communications in Medicine (DICOM) ACR-NEMA Standards Publication PS 3.1~PS 3.16 2003.
- BMP : The standard bit-mapped graphics format used in the Windows environment. By convention, graphics files in the BMP format end with a BMP extension. BMP files store graphics in a format called device-independent bitmap (DIB).
- JPG/JPEG : Short for "Joint Photographic Experts Group", the original name of the committee that wrote the standard. JPG is one of the formats supported on the Web. JPG is a lossy compression technique that is designed to compress color and grayscale continuous-tone images, but VImplant™ does not compress original graphics any more, that is to say lossless compression.
JPG images support 16 million colors and are best suited for photographs and complex graphics.
- 6. Indication for use :
The Vimplant is intended for use as a software interface for the transfer of imaging information from a CT scanner and also as a pre-operative software for simulation and evaluation dental implant placement and surgical treatment options.
- 7. Comparison with predicate device :
CyberMed, Inc., believes that the Vimplant Dental implant simulation software is substantially equivalent to SimPlant System of MATERIALISE N.V..
- 8. Conclusions :
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification CyberMed, Inc. concludes that Vimplant is safe and effective and substantially equivalent to predicate devices as described herein.
- 9. CyberMed, Inc. will update and include in this summary any other information deemed seasonably necessary by the FDA.
END
CyberMed, Inc.
Curit: 11 / 2 8 - 1
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a stylized caduceus symbol. The caduceus is a symbol of medicine and health, and it is depicted as a staff with two snakes coiled around it and wings at the top. The seal is black and white.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# NOV 2 3 2005
#
1
CyberMed, Inc. % Mark M. Mouser Senior Project Engineer/Program Reviewer Underwriters Laboratories, Inc. 2600 NW Lake Road CAMAS WA 98607-9526
Re.: K053155 Trade/Device Name: Vimplant™ Dental implant simulation software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 19, 2005 Received: November 14, 2005
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis icier will anow you're ough maing of substantial equivalence of your device to a legally premarked notheadon: "The PDF mining sification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific actrice for your be of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Asso, process note to regulation entrass," cormation on your responsibilities under the Act from the 807.77). Tod may ooually ourers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number(if known): ¥053155
Device Name: Vimplant Dental implant simulation software
Indications for Use:
The Vimplant is intended for use as a software interface for the transfer of imaging information from a CT scanner and also as a pre-operative software for simulation and evaluation dental implant placement and surgical treatment options.
Prescription Use (Part 21 CFR 801 Subpart D)
్లు
AND/OR
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
David R. Grayson
Page 1 of _1
(Division Sign-Off) Division of Reproductive, Abdomis and Radiological Devic 510(k) Number
00000036
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.