← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K052411

# HIPAX MEDICAL IMAGING SOFTWARE (K052411)

_Steinhart Medizinsysteme GmbH · LLZ · Sep 21, 2005 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K052411

## Device Facts

- **Applicant:** Steinhart Medizinsysteme GmbH
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Sep 21, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Intended Use

The Hipax Medical Imaging Software is intended to be used for medical image processing and communication. Medical Images (single images, series or sequences) and corresponding patient data can be received from various sources, e.g. from CR, CT, MRI, US, R/F units, secondary capture devices as scanners, video sources, etc. Images can be administrated, displayed, transmitted and stored on the local disk of a workstation as well as on distributed locations across a network, or on optical or digital media, e.g. CD or DVD. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Functions to be carried out using the Hipax Medical Imaging Software are, for example, but not limited to, adjustment of window leveling, defining region of interest, image stacking, MPR, rotation, zoom, measurements. The Hipax Medical Imaging Software can be integrated into a patient administration system.

## Device Story

Hipax Medical Imaging Software is an autonomous PACS workstation application running on Windows 2000/XP. It receives medical images (CT, MRI, CR, US, endoscopy) via DICOM network, video sources, or digitizers. The software provides tools for image management, display, and processing, including window leveling, ROI definition, cine-loop, measurements, and Multiplanar Reconstruction (MPR). It supports DICOM worklists, image transmission, and storage on local disks, networks, or optical media (CD/DVD). Used by radiologists, orthopedists, and clinicians in clinical settings to facilitate image review and clinical decision-making. The software does not contact the patient or control life-sustaining equipment; it serves as a tool for expert human interpretation of medical images.

## Clinical Evidence

Bench testing only. No clinical data provided. Software functionality was verified through internal software test procedures and documented test forms as part of the development process.

## Technological Characteristics

Stand-alone software; runs on Windows 2000/XP; hardware-agnostic (user-provided). Supports DICOM 3.0 connectivity. Features include window leveling, ROI, MPR, zoom, and rotation. Copy protection via hardware dongle. No patient contact; no life-sustaining control.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Radworks Medical Imaging Software ([K962699](/device/K962699.md))
- eFilm Workstation ([K012211](/device/K012211.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Kc52411

### SEP 2 1 2005

Image /page/0/Picture/2 description: The image shows the word "HIPAK" in a stylized font. The word is written in all capital letters and appears to be slightly curved. The letters are bold and have a textured appearance, possibly indicating a rough or grainy surface. The word is surrounded by a dark outline, which gives it a three-dimensional effect.

Steinhart Medizinsysteme GmbH

## 510(k) Summary

Summary of functions of the device and its major components provided as part of the Device Summary of functions of the device and its major componing to the Guidance for the Submission of Description for Thir 70 : Modical Image Management Devices, 2000.

| Date:                  | July 27th 2005                                                             |
|------------------------|----------------------------------------------------------------------------|
| Company Name:          | Steinhart Medizinsysteme GmbH                                              |
| Address:               | Grubstrasse 6-8<br>D-79279 Voerstetten<br>Germany                          |
| Contact Person:        | Martina Steinhart                                                          |
| E-Mail:                | sales@hipax.de                                                             |
| Phone:                 | +49-(0)-7666-90070                                                         |
| Fax:                   | +49-(0)-7666-900711                                                        |
| Device Trade Name:     | HIPAX Medical Imaging Software                                             |
| Device Common Name:    | Picture Archiving Communications System (PACS                              |
| Product Code:          | LLZ                                                                        |
| Regulation No.         | 892.2050                                                                   |
| Device Classification: | Class II                                                                   |
| Predicate Devices:     | Radworks Medical Imaging Software (K962699)<br>eFilm Workstation (K012211) |

#### Device Description

The Hipax Medical Imaging Software is an autonomous software with The Hipax Medical Imaging Soltware is an autonomicas oringer Microsoft Windows 2000/XP
the exception of a dongle for copy protection. It is running under requirements and the exception of a dongle for oopy protocion. It is ranimaly inimum system requirements and supporting the Windows 2000/XP operating system.

The Hipax Medical Imaging Software has an open system architecture consisting of a basic The Hipax Medical miaging Ooltware his an open of the research and the mage acquisition, storage and communication to be added as an option.

The functions of the Hipax Medical Imaging Software correspond to the features described for the The functions of the Hipax Medical image, for example, from CT, MRI, CR, US, Predicate Devices. Displaying any medical Imaging Software offers features endoscopy, gastroscopy, and other medical specialists (e.g. window leveling, ROIs, edge routinely used by Tadiologists and other modifal op Scalless, cine-loop, measurement, writing and marking, Childrioument, 2001Itiplanar Reconstruction is available as an option.

histogram etc.). Multipland Reconstruction is division as an Sparare or the eFilm Workstation, the Like the Predicale Devices, the Rudions Moulan mage acquisition from video sources (Video Hipax Medical Innaging Soltware onero for Intege beep.
Module) digitizers (X-ray Digitizing Module), or CR systems (CR-Connection Module), Within a Module) digitzels (X-ray Digitzing Modic, of UN costs and received using the DICOM
network DICOM worklists can be received and images can be sent and liney programs was nework DICOM worklists can be recorved and inisges who inistations or other Hipax programs via protocol. Image exchange between two remote hipax of using the Hipax DICOM Communication phone lines, broadband, salente, cle. can be carrou out as oftware. The DICOM Email module, which is a modile, which is also part of the Nadworks Modes can be compressed and encrypted. The DICOM
is available to transmit images as Emails. Images can be images an he virithe is available to transmit mages as Emails: images can be written in the written Print Module supports DICOM 3.0 primary. Osing the Patient Station of Children on CD to be handed out to

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the word "HIPAA" in a stylized font. The word is set inside of an oval shape. The letters are bold and slightly slanted to the right. 

# 510(k) Summary

on digital media, e.g. DVD. To burn CDs or DVDs automatically, a CD/DVD robot can be connected.

### Intended Use

The Hipax Medical Imaging Software is intended to be used for medical image processing and communication. Medical Images (single images, series or sequences) and corresponding patient data can be received from various sources, e.g. from CR, CT, MRI, US, R/F units, secondary capture devices as scanners, video sources, etc. Images can be administrated, displayed, transmitted and stored on the local disk of a workstation as well as on distributed locations across a network, or on optical or digital media, e.g. CD or DVD. Lossy compressed mammographic mages and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Functions to be carried out using the Hipax Medical Imaging Software are, for example, but not limited to, adjustment of window leveling, defining region of interest, image stacking, MPR, rotation, zoom, measurements. The Hipax Medical Imaging Software can be integrated into a patient administration system.

Users of the Hipax Medical Imaging Software are typically trained medical professionals, for example, radiologists, orthopedists, clinicians, technologists, and others.

#### Technological Characteristics

The Hipax Medical Imaging Software is a stand-alone software, like the Predicate Devices. It can be used on more than one hardware platform. As long as minimum hardware requirements are met, the user is free to choose his/her own hardware platform.

The Hipax Medical Imaging Software as well as the Predicate Devices allow digital image processing and measurement capability.

The Hipax Medical Imaging Software nor contacts the patient, neither controls any life-sustaining devices. Physicians providing the adequate expert knowledge for competent human intervention interpret images and information being displayed and/or printed.

#### Testing

The Hipax Medical Imaging Software is tested according to the specifications that are documented in an own description (Description of the software test procedures) and the corresponding Softwaretest forms. Testing is an integral part of the software development process of Steinhart Medizinsysteme GmbH (see documents in G 2 and G 4).

#### Conclusion

The Hipax Medical Imaging Software is a medical device. It has the same indications for use and target population as the legally marketed Predicate Devices.

The Hipax Medical Imaging Software has the same technological characteristics as the Predicate Devices.

This premarket notification describes the characteristics of the Hipax Medical Imaging Software in sufficient detail to assure substantial equivalence.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

.
.
.
.

SEP 2 1 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Steinhart Medizinsysteme GmbH % Mr. Stefan Preiss Responsible Third Party Official TÜV America, Inc. TÜV Product Service 1775 Old Highway 8 NEW BRIGHTON MN 55112-1891 · Re: K052411 Trade/Device Name: Hipax Medical Imaging Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 22, 2005 Received: August 29, 2005

Dear Mr. Preiss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced w ove and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to the chercours) to e enactment date of the Medical Device Amendments, or to devices that have been rear 20, 1770, and characteriting of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the do not require upproval or a provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket If your de riot is onestilia (200 as 10) additional controls. Existing major regulations affecting your Apple rally it after of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I tease or as not is a determination that your device complies with other requirements of the Act or any I Drinas made and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k)
Continued and consequence The PDA finding of substantial courvalence of your dev This letter will allow you to begin marketing your device of your device of your device to a legally
premarket notification. The FD finding of substantial equivalence of your premarket notification. The FDA finding of substantial equivalence of your device to angle your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your device on our labeling regulation (1) - 11 Feb 2017)
contact the Office of Compliance at one of the following numbers, based on the reg the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation information on your responsibilities under the Act from the 807.97). You may obtain other general information on your responsible in the more
Division of Small Manufacturers, International and Consumer Assistance at its toll-free num Division of Small Manafastar at its Internet address. 036-2041 of (501) of (501) 11301) or vindex.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Nancy of Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Office of Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

Kosta di li 510(k) Number (if known):

Hipax Medical Imaging Software

Indications for Use:

Device Name:

The Hipax Medical Imaging Software is intended to be used for medical image processing and communication. Medical Images (single images, series or sequences) and corresponding patient data can be received from various sources, e.g. from CR, DR, CT, MRI, US, RF units, secondary capture devices as scanners, video sources, etc. lmages can be administrated, displayed, transmitted and stored on the local disk of a workstation as well as on distributed locations across a network, or on optical or digital media, e.g. CD or DVD.

. . .
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

Functions to be carried out using the Hipax Medical Imaging Software are, for example, but not limited to, adjustment of window leveling, defining region of interest, image stacking, MPR, rotation, zoom, measurements. The Hipax Medical Imaging Software can be integrated into a patient administration system.

Users of the Hipax Medical Imaging Software are typically trained medical professionals, for example, radiologists, orthopedists, clinicians, technologists, and others.

Over-The-Counter Use Prescription Use × AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Syverson

Page 1 of 1

(Division Sign-Off) #
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K052411

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K052411](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K052411)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
