UltraPACS™ system is device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. The device can also assists orthopedic surgeons when doing preoperative planning and post-operative follow-up. Options make possible reading (including mammography), telecommunications; fast demonstration; etc.; and teleconferencing. Lossy compressed or digitized screen film mammographic images must not be reviewed for primary image interpretations. Only FFDM manufacturer processed images in DICOM "For Presentation" format can be displayed for primary interpretation. Mammographic images must only be interpreted using a FDA approved monitor that offers at least 5Mpixel resolutions and other technical specifications reviewed and accepted by the FDA. Typical users of this system are trained professionals, physicians, nurses, and technicians.
Device Story
UltraPACS™ is a Picture Archiving and Communications System (PACS) for medical imaging. It receives digital images from modalities like CT, MR, ultrasound, and radiographic devices; utilizes DICOM gateways to ingest non-DICOM or film-digitized data. System stores, distributes, and networks images across clinical locations. Orthopedic application provides tools for prosthesis template overlay, repositioning, and measurements to assist surgeons in preoperative planning and post-operative follow-up. Operated by physicians, nurses, and technicians in clinical settings. Healthcare providers view images on workstations to support clinical decision-making. Device does not contact patients or control life-sustaining equipment; relies on human interpretation.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is based on technological characteristics and intended use comparisons to predicate devices.
Technological Characteristics
PACS software for digital medical image management. Connectivity via DICOM standards and network protocols. Supports orthopedic measurement and template overlay tools. Does not contact patients. Software hazard level classified as 'Minor'.
Indications for Use
Indicated for use by trained professionals, physicians, nurses, and technicians to receive, store, communicate, process, and display digital medical images from various sources (CT, MR, ultrasound, R/F, CR/DR). Includes orthopedic preoperative planning and post-operative follow-up tools. Supports mammography interpretation provided specific FDA-approved 5Mpixel monitors and DICOM 'For Presentation' formats are used.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
# AUG 1 5 2005
Image /page/0/Picture/2 description: The image shows a logo with the word "ultra" stacked on top of the word "RAD". The word "ultra" is in a smaller, italicized font, while "RAD" is in a larger, bold font. The logo is black and white and has a simple, clean design.
## 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
Date Prepared: June 27, 2005
Submitter's Information: 21 CFR 807.92(a)(1) Tom Goliash, President / CEO UltraRAD Corporation 801 West Bay Drive, Suite 424 Largo, FL 33770 Tel: 800-719-9379 Email: tgoliash@ultraradcorp.com
Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2) Trade name - UltraPACS™ by UltraRAD Corporation Common name - Picture, Archive, and Communications System Classification name - system, image processing, radiological (21 CFR 880.2050, Product Code LLZ)
# Predicate Device: 21 CFR 807. 92(a)(3)
| Device Classification Name | System, Image Processing,<br>Radiological | System, Image Processing,<br>Radiological |
|----------------------------|-------------------------------------------|-------------------------------------------|
| Regulation Number | 892.2050 | 892.2050 |
| 510(k) Number | K042311 | K041500 |
| Device Name | PACSPartner | StarPACS Ortho |
| Applicant | Medical Standard Co. Ltd. | Infinitt Co. Ltd. |
| Product Code | LLZ | LLZ |
| Decision Date | 09/09/2004 | 06/21/04 |
### Device Description: 21 CFR 807 92(a)(4)
UltraPACS™ makes possible the capturing, storage, distribution, and networking of medical images at distributed locations. In cases where DICOM images are not directly available to UltraPACS™, the system can acquire medical images using a DICOM gateway, which generates DICOM-type files. For example, film digitizers obtain images from old film and convert them to meet DICOM standards and stored in an archive. Stored files are transmitted using a network and can be viewed or manipulated from an imaging workstation. The orthopedic application is intended to assist orthopedic surgeons when doing preoperative planning and post-operative follow-up. The orthopedic application of the device is used to overlay prosthesis templates on radiological images, tools for repositioning the templates. and tools for measurements in the images.
### Indications for Use: 21 CFR 807 92(a)(5)
UltraPACS™ system is device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the word "ultra" stacked on top of the word "RAD". The word "ultra" is in a cursive font and is smaller than the word "RAD". The word "RAD" is in a bold, sans-serif font. The logo is black and white.
imaging sources). Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. The device can also assists orthopedic surgeons when doing preoperative planning and post-operative follow-up.
Options make possible reading (including mammography), telecommunications; fast demonstration; etc.; and teleconferencing.
Lossy compressed or digitized screen film mammographic images must not be reviewed for primary image interpretations. Only FFDM manufacturer processed images in DICOM "For Presentation" format can be displayed for primary interpretation. Mammographic images must only be interpreted using a FDA approved monitor that offers at least 5Mpixel resolutions and other technical specifications reviewed and accepted by the FDA.
Typical users of this system are trained professionals, physicians, nurses, and technicians.
# Technological Characteristics: 21 CFR 807 92(a)(6)
UltraPACS™ is a Picture Archiving and Communications device that handles digital medical images. The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.
## Conclusion: 21 CFR 807 92(b)(1)
The 510(k) Pre-Market Notification for UltraPACS™ contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.
UltraPACS™ has been and will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The submission contains the results of a hazard analysis and the "Level of Concern for potential hazards has been classified as "Minor".
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image contains the text 'Public Health Service'. The text is in a bold, sans-serif font. The text is black and the background is white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
UltraRAD Corporation % Mr. Carl Alletto Official Correspondent 1600 Manchester Way CORINTH TX 76210
Re: K051813
Trade/Device Name: UltraPACSTM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system
Regulatory Class: II Product Code: LLZ Dated: June 27, 2005 Received: July 13, 2005
Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling. (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
# AUG 1 5 2005
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
(Indications for Use Form)
510(k) Number: K051813
Device Name: UltraPACS™
Indications for Use:
UltraPACS™ system is device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. The device can also assists orthopedic surgeons when doing preoperative planning and post-operative follow-up.
Options make possible reading (including mammography), telecommunications; fast demonstration; etc ; and teleconferencing.
Lossy compressed or digitized screen film mammographic images must not be reviewed for primary image interpretations. Only FFDM manufacturer processed images in DICOM "For Presentation" format can be displayed for primary interpretation. Mammographic images must only be interpreted using a FDA approved monitor that offers at least 5Mpixel resolutions and other technical specifications reviewed and accepted by the FDA.
Typical users of this system are trained professionals, physicians, nurses, and technicians.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <span style="text-decoration: overline;">✓</span> | AND/OR | Over-The-Counter Use | <span style="text-decoration: overline;"></span> |
|-----------------------------|---------------------------------------------------|--------|------------------------|--------------------------------------------------|
| (Part 21 CFR 801 Subpart D) | | | (21 CFR 807 Subpart C) | |
*David B. Seymore*
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number: K051813
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.