Presentation quality mammography images sent via the DICOM standard can be displayed using features that are native to Emageon UV Inc.'s software. Other standard tools include intuitive tools for real-time pan, zoom, window/level and scroll; drag and drop series thumbnails for intuitive navigation; comprehensive set of grayscale and pseudo-color lookup tables; fully configurable magnifying glass; customizable window/level and zoom presets stored by modality and user; rotate images in 90-degree increments; flip image horizontally or vertically; customize display of on-screen user and patient demographics; sharpen, edge and blur filters; key image creation for communicating with other physicians, and multiple display support. Images can be displayed in any configuration or format the user specifies and associates with that specific user's profile. All display protocols and user configurable settings follow the user throughout the enterprise. Lossy compressed or digitized screen film mammographic images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using a FDA approved monitor that offers at least 5 Mpixel resolutions and other technical specifications reviewed and accepted by FDA. Intended users of the image distribution system include radiologists, referring physicians, tertiary care physicians, medical technologists, and information technology professionals.
Device Story
Ultravisual is a client-server software package for Picture Archiving and Communication Systems (PACS). It receives digital mammography images via DICOM standard. The software provides visualization tools including real-time pan, zoom, window/level, scroll, rotation, flipping, and image filters (sharpen, edge, blur). It supports multi-monitor configurations and user-specific display profiles that roam across the enterprise. Used by radiologists, referring physicians, and technologists in clinical settings to view and interpret medical images. The device facilitates clinical decision-making by providing configurable, high-resolution image display and communication tools (key image creation). It benefits patients by enabling efficient, enterprise-wide access to diagnostic-quality mammography images for timely review.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and validation procedures consistent with the predicate device (K012097).
Technological Characteristics
Integrated client-server software architecture. Operates on off-the-shelf computer hardware. Supports DICOM standard for image communication. Features include grayscale/pseudo-color lookup tables, image filters (sharpen, edge, blur), and multi-display support. Designed for use with FDA-approved 5 Mpixel resolution monitors for mammography interpretation.
Indications for Use
Indicated for the display of presentation quality mammography images for interpretation by radiologists and other clinicians. Contraindicated for primary image interpretation of lossy compressed or digitized screen film mammographic images. Requires use of FDA-approved monitors with at least 5 Mpixel resolution.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K042008
SEP - 3 2004
# 510(k) SUMMARY
In accordance with the provisions of the Safe Medical Device Act of 1990, Emageon UV.. Inc. is providing a summary of safety and effectiveness information regarding the Ultravisual™ software.
- l . I Company Identification
Emageon UV, Inc. 131 W. Wilson Street Suite 700 Madison WI 53703 Establishment Registration Number: 2135350 Contact: Inger Couture Tel: 608 256 7775 Fax: 608 256 7779 Email: Inger.Hanson@emageon.com
- 1.2 Official Correspondent
Inger Couture Director of Regulatory Affairs and Quality Assurance Emageon UV, Inc. 131 W. Wilson Street Suite 700 Madison WI 53703 Establishment Registration Number: 2135350 Tel: 608 256 7775 Fax: 608 256 7779 Email: Inger.Hanson@emageon.com
- 1.3 Date of Submission
June 29th, 2004
- Device Name 1.4
Classification Name: Image Processing System, 21 CFR \$892.2050, ProCode LLZ
{1}------------------------------------------------
Common/Usual Name: Picture Archiving and Communication System
Proprietary Name: Ultravisual™ (formerly Vortex TM)
Cleared Device(s): Vortex™ , 510(k): K012097
#### Device Description and Intended Use 1.5
Emageon UV, Inc.'s Ultravisual™ software is integrated client-server software package comprised of features that were previously cleared in Vortex TM 510(k): K012097. The main difference is that the software will now allow display of presentation quality digital mammography images, sent via the DICOM standard in order to make viewing of these images more convenient for the user.
- 1.6 Software Development
Emageon UV, Inc. certifies that the Ultravisual™ software is designed. developed, tested and validated in accordance with the same procedures as documented in the initial Vortex™ 510(k). K012097.
{2}------------------------------------------------
#### 1.8 Safety and Effectiveness
## General Safety and Effectiveness Concerns:
The device labeling contains instructions for use and indications for use.
The hardware components specified (and/or optionally supplied) are all "off the shelf" computer components.
## Validation and Effectiveness:
As documented in the initial Vortex™ 510(k): K012097.
### Substantial Equivalence:
The intended use of Emageon UV, Inc.'s Ultravisual™ software is substantially equivalent in the opinion of Emageon UV, Inc. to the feature set described in the original Vortex 130 software, 510(k) K012097 and do not pose any new issues of safety and effectiveness
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 3 2004
Emageon UV, Inc. % Ms. Laura Danielson Responsible Third Party TÜV Product Service 1775 Old Highway 8 NEW BRIGHTON MN 55112-1891 Re: K042008
Trade/Device Name: UltraVisual™ Image Processing, Communication and Visualization Workstation Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system
Regulatory Class: II Product Code: 90 LLZ Dated: August 19, 2004 Received: August 20, 2004
Dear Ms. Danielson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I result et notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Brydon
Nancy C. Brogdor Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Device Name: Emageon UV, Inc., Ultravisual ™ Image Processing, Communication and Visualization Software Supporting Workstation
Indications for use:
Presentation quality mammography images sent via the DICOM standard can be displayed using features that are native to Emageon UV Inc.'s software. Other standard tools include intuitive tools for real-time pan, zoom, window/level and scroll; drag and drop series thumbnails for intuitive navigation; comprehensive set of grayscale and pseudo-color lookup tables; fully configurable magnifying glass; customizable window/level and zoom presets stored by modality and user; rotate images in 90-degree increments; flip image horizontally or vertically; customize display of on-screen user and patient demographics; sharpen, edge and blur filters; key image creation for communicating with other physicians, and multiple display support. Images can be displayed in any configuration or format the user specifies and associates with that specific user's profile. All display protocols and user configurable settings follow the user throughout the enterprise.
Lossy compressed or digitized screen film mammographic images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using a FDA approved monitor that offers at least 5 Mpixel resolutions and other technical specifications reviewed and accepted by FDA.
Intended users of the image distribution system include radiologists, referring physicians, tertiary care physicians, medical technologists, and information technology professionals.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
510(k) Number
| Prescription Use | <div> <span style="text-decoration: overline;">✓</span> </div> |
|-----------------------------|----------------------------------------------------------------|
| AND/OR Over-The-Counter Use | <div> <span style="text-decoration: overline;"></span> </div> |
David A. Repass
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices 10/1/2008
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.