K041990 · Fujifilm Medical System U.S.A., Inc. · LLZ · Aug 6, 2004 · Radiology
Device Facts
Record ID
K041990
Device Name
FUJI MEDICAL CR CONSOLE, MODEL FLASH IIP
Applicant
Fujifilm Medical System U.S.A., Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 6, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
The Fuji CR Console (Flash IIP) is a CR modality workstation intended to associate FCR images (except mammography images) with patient and exam information, apply image processing to facilitate diagnosis, display the image, and output the resulting image and exam data for further display, distribution, or archiving.
Device Story
Fuji CR Console (Flash IIP) is a medical image processing workstation used by radiographers in clinical settings. It interfaces with Fuji Image Readers via network to receive CR images; allows entry of patient ID, exam info, and IP barcode data. The device performs image processing and optimization for quality assurance before transferring images to external systems for archiving or display. It runs on off-the-shelf PCs (Windows 2000/XP) with either high-resolution (Plus) or standard-resolution (Lite) monitors. Healthcare providers use the console to verify image quality and associate clinical data, facilitating diagnostic workflows and patient management.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness supported by hazard analysis and compliance with UL 60950, IEC/CISPR 22, and IEC/CISPR 24 standards.
Technological Characteristics
Off-the-shelf PC (Pentium 4, 512MB RAM, 40GB HDD); Windows 2000/XP OS. Displays: 19" 2MP LCD (Plus) or 15" 1MP LCD (Lite). Connectivity: LAN, DICOM 3.0, Fuji DMS network. Input: Barcode scanner, keyboard, mouse. Software: Enhanced image processing and viewing functions. Standards: UL 60950, IEC/CISPR 22, IEC/CISPR 24, DICOM.
Indications for Use
Indicated for use as a CR modality workstation to associate FCR images (excluding mammography) with patient/exam data, apply image processing, display images, and output data for clinical diagnosis, distribution, or archiving. Prescription use only.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Fuji CR DMS CRT Image Console HI-C654
Submission Summary (Full Text)
{1}------------------------------------------------
K04 1990
AUG = 6 2004
Image /page/0/Picture/2 description: The image contains the Fujifilm logo. The logo consists of a stylized square shape on the left, followed by the word "FUJIFILM" in bold, sans-serif font. Below the wordmark, there is a line of text that reads "I&I - Imaging & Information" in a smaller font size.
FUJIFILM MEDICAL SYSTEMS
419 WEST AVENUE STAMFORD, CT 06902 Telephone: 203/324-2000 Fax: 203/353-0926
## 510(K) Summary
In accordance with the requirements of the Safe Medical Device Act, FUJIFILM Medical Systems, USA, Inc. herewith submits a 510(K) summary of safety and effectiveness for the following device.
| SUBMITTER NAME / ADDRESS: | FUJIFILM Medical Systems, USA, Inc.<br>419 West Avenue<br>Stamford, CT 06902 |
|---------------------------|------------------------------------------------------------------------------|
| CONTACT PERSON / TEL NO: | Frank Gianelli<br>Regulatory Coordinator |
| DATE SUMMARY PREPARED: | June 11, 2004 |
| ESTABLISHMENT NO.: | 2443168 |
| TRADE/PROPRIETARY NAME: | Fuji Medical CR Console, Flash IIP |
| COMMON/USUAL NAME: | Medical Image Processing Workstation |
| CLASSIFICATION NAME: | Picture archiving and communications system |
| CLASS/PANEL: | Class II, 90-LLZ, 21CFR 892.2050 |
| PREDICATE DEVICE(S): | Fuji CR DMS CRT Image Console HI-C654 (for networks) |
### DEVICE DESCRIPTION:
The CR Console (Flash IIP) is used by a radiographer to view CR images for final quality assurance (QA) checking and image processing/optimization prior to transfer of those images to external (&A) oncelling and intage problem. Furthermore, when connected to Fuji Image Readers via a network, the CR Console is used to enter patient ID information, exposure information and register Inage Plate (IP) barcode numbers. The CR Console application software runs on an "off-the-shell" personnel computer under the Windows 2000 or Windows XP operating system with a choice of a high resolution monitor (Plus configuration) or standard resolution monitor (Lite configuration) for image display.
### INTENDED USE:
The Fuji CR Console (Flash IIP) is a CR modality workstation intended to associate FCR images (except mammography images) with patient and exam information, apply image processing to (xxcept manimography image, and output the resulting image and exam data for further display, distribution, or archiving.
{2}------------------------------------------------
.
Image /page/1/Picture/1 description: The image shows the Fujifilm logo. The logo consists of a stylized graphic to the left of the word "FUJIFILM" in bold, sans-serif font. Below the wordmark is the text "I&I - Imaging & Information" in a smaller, lighter font.
FUJIFILM MEDICAL SYSTEMS USA,
419 WEST AVENUE STAMFORD, CT 06902 Telephone: 203/324-2000 Fax: 203/353-0926
## TECHNOLOGICAL CHARACTERISTICS:
Fuji CR Console (Flash IIP) is considered comparable and substantially equivalent to the Fuji CR DMS CRT Image Console HI-C654 manufactured by Fuji.
| Feature | CR Console Plus<br>(Flash Plus IIP) | CR Console Lite<br>(Flash Lite IIP) | Fuji HI-C654 |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Minimum Basic<br>Computer<br>Configuration | Computer "Off the Shelf"<br>Desktop or Tower<br>CPU: Pentium 4<br>Bus: ISA/PCI<br>RAM: 512MB<br>Hard Drive: 40 GByte<br>Floppy Drive: 3¼"<br>CD-ROM<br>Keyboard<br>Mouse<br>Barcode scanner | Computer "Off the Shelf"<br>Desktop or Tower<br>CPU: Pentium 4<br>Bus: ISA/PCI<br>RAM: 512MB<br>Hard Drive: 40 GByte<br>Floppy Drive: 3½"<br>CD-ROM<br>Keyboard<br>Mouse<br>Barcode scanner | Computer (custom design):<br>Desktop<br>CPU: 68020<br>Bus: Fuji-Bus<br>RAM: 5MB<br>Hard Drive: 4 Gbyte<br>Floppy Drive: 3½"<br>Keyboard<br>Mouse |
| Operating System<br>Software | Microsoft Windows 2000<br>or Windows XP | Microsoft Windows 2000<br>or Windows XP | proprietary |
| Ethernet Capability &<br>Type | Yes: LAN | Yes: LAN | Yes: LAN |
| Image Transfer | via DICOM 3.0 &<br>via Fuji DMS Network | via DICOM 3.0 &<br>via Fuji DMS Network | via Fuji DMS network |
| Image Display | 19" Monochrome 2MP LCD<br>with Touch screen | 15" color 1MP LCD<br>with Touch screen | 20" 1MP Monochrome CRT |
| Fuji Image processing<br>functions | Yes, enhanced | Yes, enhanced | Yes |
| Image viewing &<br>orientation functions | Yes, enhanced | Yes | Yes |
| Connects to Fuji CR<br>Image Readers | Yes | Yes | Yes |
| Connects to Fuji Image<br>Recorders (Printers) | Yes | Yes | Yes |
### SAFETY INFORMATION:
Fuji CR Console (Flash IIP) introduces no new safety and efficacy issues other than those already identified with the predicate device. The results of a hazard analysis, combined with the appropriate lientifice measures taken indicate that the device is of minor level of concern as per the May 29, 1998 issue of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
Fuji CR Console (Flash IIP) comply with the following mandatory and voluntary standards:
- Information Technology Equipment Part 1: General Requirements for Safety UL Standard . 60950
- Information Technology Equipment, Radio Disturbance (Emissions) Characteristics-Limits . and Methods of Measurements, IEC/CISPR 22 (EN55022)
- Information Technology Equipment, Immunity Characteristics-Limits and Methods of . Measurement, IEC/CISPR 24 (EN55024)
- DICOM (Digital Imaging and Communications in Medicine) Developed by the American . College of Radiology and the National Electrical Manufacturers Association.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## AUG - 6 2004
FUJIFILM Medical Systems USA, Inc. % Mr. William J. Sammons Project Engineer Underwriters Laboratories, Inc. Research Triangle Park Division 12 Laboratory Drive P.O. Box 13995 Research Triangle Park, NC 27709-3995 Re: K041990
Trade/Device Name: Fuji Medical CR Console, Model Flash IIP Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II
Product Code: 90 LLZ Dated: July 20, 2004 Received: July 23, 2004
Dear Mr. Sammons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreation 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will and in you to organizatial equivalence of your-device of your-device to a legally promance noticated. " results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, roll and (301) 594-4639. Also, please note the regulation entitled, "Misbranding Other or othernation to the ation" (21CFR Part 807.97) you may obtain. Other general of receive to premainted notifies under the Act may be obtained from the Division of Small intormation on your respondent Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brydon.
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications For Use
510(k) Number (if known): _ Ko4 1990
Device Name: Fuji CR Console (Flash IIP)
Indications For Use:
The Fuji CR Console is a CR modality workstation intended to associate FCR images (except mammography images) with patient and exam information, apply image processing to facilitate diagnosis, display the image, and output the resulting image and exam data for further display, distribution, or archiving.
X Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Bergdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number Kc41990
Page 1_ of _1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.