K041162 · Medis Medical Imaging Systems BV · LLZ · Jul 27, 2004 · Radiology
Device Facts
Record ID
K041162
Device Name
ORTHO-CMS
Applicant
Medis Medical Imaging Systems BV
Product Code
LLZ · Radiology
Decision Date
Jul 27, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Ortho-CMS is an orthopaedic analysis software tool. It has been developed to optimize preoperative planning through digital prosthesis templating and to enable preoperative and postoperative measurements in digital or digitized X-Ray images. Ortho-CMS software is meant solely for use by trained medical personnel. The intended purposes of Ortho-CMS are: Displaying of X-Ray images Supporting planning of joint replacement operations Supporting clinical diagnoses on implant loosening Enabling preoperative and postoperative measurements for clinical and research purposes
Device Story
Ortho-CMS is standalone orthopaedic analysis software; inputs digital/digitized X-ray images (DICOM CR/DX) from PACS. Software facilitates digital prosthesis templating; performs measurements including radiolucent lines, bone growth/resorption, prosthesis position/motion, and polyethylene wear. Used by radiologists and orthopaedic specialists in clinics or core labs for pre-surgical planning and post-surgical evaluation. Output displayed to clinician; operator interprets results to accept/reject findings. Benefits include increased objectivity, speed, and reduced inter/intra-observer variation compared to manual measurements.
Clinical Evidence
No clinical data. Safety and effectiveness supported by software risk management, hazard analysis, and verification/validation testing. Evaluations by hospitals and literature review provided to support performance.
Technological Characteristics
Standalone software for digital image viewing and measurement. Operates on digital/digitized X-ray images (DICOM). Software-based analysis tool. Level of concern: minor.
Indications for Use
Indicated for use by trained medical personnel for orthopaedic analysis, including preoperative planning via digital prosthesis templating and preoperative/postoperative measurements in digital or digitized X-ray images for patients undergoing joint replacement or requiring assessment of orthopaedic disorders, joint conditions, bone fractures, or implant loosening.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Ortho-CMS 510(k) Pre-market Notification
# JUL 2 7 2004
## 12. 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
K041162
Submission in accordance with the requirements of 21 CFR Part 807.87(h)
- : Medis medical imaqinq systems bv 1. Submitter : Schuttersveld 9, 2316 XG Leiden, The Netherlands Address : +31 71 522 3244 Telephone : +31 71 521 5617 Fax : J.I. Hollander, Quality Coordinator Contact Person : April 26, 2004 Prepared : Medis Ortho-CMS 2. Device Name : Ortho-CMS Common Name : System Image Processing, Radiological Device Class. Name : 21 CFR 892.2050 (90 LLZ; Class II) Requlation Number : Meridian Technique Ltd: 510(k) K032401 3. Predicate Device(s) eFilm Medical Inc.: 510(k) K020995
#### 4. Description of the device:
In orthopaedics, radiographs are used to diagnose and analyse various kinds of orthopaedic disorders, generative joint conditions, and bone fractures, and to evaluate orthopaedic treatments such as osteotomies and total joint arthroplasty. These arthroplasties need to be evaluated in order to assess the quality of the procedure. Measurements on radiographs of endoprostheses include the assessment of radiolucent lines around the prosthesis, bone growth or bone resorption, position of the prosthesis, motion of the prosthesis relative to the surrounding bone, and the determination of wear of the polyethylene components.
Since the measurements on radiographs are commonly performed manually, considerable intra-observer and inter-observer variation exists. Automation of the measurements might increase the objectivity and speed of the analysis, and decrease the variation of the results. In radiology, digital roentgen imaging techniques are increasingly being used over plain film radiographs. The digital roentgen images (DICOM CR or DX) are easily accessible from a medical picture archive (PACS) through a network connection.
Ortho-CMS supports the radiologist by facilitating the diagnosis of orthopaedic digital images and allows the orthopaedic specialist to perform a pre-surgical planning and a post-surgical evaluation on these images. Further, Ortho-CMS can be deployed as a measurement tool for core-labs that focus on quality assessment of orthopaedic implants or it can be used by bone centres that need to make measurements in donor bone images for joints replacement purposes.
#### 5. Intended use:
Ortho-CMS is an orthopaedic analysis software tool. It has been developed to optimize preoperative planning through digital prosthesis templating and to enable preoperative and postoperative measurements in digital or digitized X-Ray images. Ortho-CMS software is meant solely for use by trained medical personnel.
- · The intended purposes of Ortho-CMS are:
- · Displaying of X-Ray images
- · Supporting planning of joint replacement operations
- · Supporting clinical diagnoses on implant loosening
- · Enabling preoperative and postoperative measurements for clinical and research purposes
{1}------------------------------------------------
#### 6. Substantial equivalence Information:
Ortho-CMS is substantially equivalent to the Predicate Devices of Meridian Technique Ltd. 6132401 "Orthoview™" and of eFilm Medical Inc., K020995 "eFilm"14 Workstation™, using the same technological technique for the same intended use.
### Conclusion respecting safety and effectiveness:
It is the opinion of Medis medical imaging systems bv that Ortho-CMS is safe and potential hazards are controlled by a risk management plan for the software development process (See Appendix C), including hazard analysis (See Appendix D), verification and validation tests (See Appendix E). Evaluations by hospitals and literature (See Appendix F) support this statement. The software package Ortho-CMS itself will not have any adverse effects on health. This tool supports the radiologist by facilitating the diagnosis of orthopaedic digital images and allows the orthopaedic specialist to perform a pre-surgical planning and a post-surgical evaluation on orthopaedic images. The analyses results will be interpreted by the operator, who can choose to accept or reject the tools results.
It is the opinion of Medis medical imaging systems bv that the level of concern for the stand alone software to view images is 'minor' and that the use of Ortho-CMS software does not change the intended use of X-ray equipment in practice, nor does the use of software result in any new potential hazards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. J.I. Hollander Quality Coordinator Medis Medical Imaging Systems bv Schuttersveld 9, 2316 XG Leiden P.O. Box 384, 2300 AJ Leiden THE NETHERLANDS
Re: K041162 Trade/Device Name: Ortho-CMS Regulatory Number: 21 CFR 892.2050 Regulatory Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: April 26, 2004
Received: May 3, 2004
Dear Mr. Hollander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibit.ons against misbranding and adulteration.
JUL 27 2004
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter with and your your of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brydon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{4}------------------------------------------------
Page 4
Page
510(k) Number (if known):
#### Device Name
1 2017 11:00:00 10 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
Indications For Use: 。 保安了 Ortho-CMS is an orthopaedic analysis software tool. It has been developed to がなくなっていますが optimize preoperative planning through digital prosthesis templating and to enable preoperative and postoperative measurements in digitized X-Ray images. Ortho-CMS software is meant solely for use by trained medical personnel.
- · The intended purposes of Ortho-CMS are:
- - · Displaying of X-Ray images
- Supporting planning of joint replacement operations
- Supporting clinical diagnoses on implant loosening
- · Enabling preoperative and postoperative measurements for clinical and research purposes
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Optional Format 3-10-98)
Nancy C. Broadlon
(Division Sign (Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.