ENSPHERE TEMPO JETSTREAM WORKSPACE CARDIOLOGY MODULE

K040142 · Adac Laboratories · LLZ · Feb 6, 2004 · Radiology

Device Facts

Record IDK040142
Device NameENSPHERE TEMPO JETSTREAM WORKSPACE CARDIOLOGY MODULE
ApplicantAdac Laboratories
Product CodeLLZ · Radiology
Decision DateFeb 6, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

ENsphere™ Tempo is a nuclear medicine image processing and display workstation that provides cardiology specific software applications used to process, analyze, and display medical images. When interpreted by a nuclear physician, the results obtained by using the ENsphere Tempo applications may be used as a tool in determining a patient's diagnosis. ENsphere Tempo should only be operated by qualified healthcare professionals trained in the use of nuclear medicine equipment. This workstation is intended to be operated in a hospital, clinic, or doctor's office. Patient data may be transferred via DICOM over Local Area Networks (LAN), or Wide Area Networks WAN. The primary users of this product include nuclear medicine technologists who process the data and physicians who display, review, and interpret the processed image data.

Device Story

ENsphere™ Tempo is a Windows-based nuclear medicine workstation for cardiology; utilizes Hewlett Packard XW4100 hardware. Inputs consist of medical image data; system processes, analyzes, and displays these images for interpretation. Operated by nuclear medicine technologists and physicians in hospitals, clinics, or offices. Supports DICOM data transfer via LAN/WAN. Provides tools for image review, post-processing, hardcopy production, and report generation. Output serves as a diagnostic tool for physicians to interpret nuclear cardiology procedures; aids clinical decision-making by facilitating efficient workflow and data management.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological comparison and system performance.

Technological Characteristics

Windows-based workstation (HP XW4100 or equivalent). Supports DICOM connectivity for LAN/WAN data transfer. Software-based image processing and display for nuclear medicine cardiology. No specific materials or sterilization methods applicable as this is a software/workstation system.

Indications for Use

Indicated for use by qualified healthcare professionals in hospitals, clinics, or doctor's offices to process, analyze, and display nuclear medicine cardiology images to assist physicians in determining patient diagnosis.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K040142 CONFIDENTIAL 510(k) Premarket Notification Section B. Administrative Information ## ADMINISTRATIVE INFORMATION ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS I. | A. | Submitted By: | ADAC Laboratories A Philips Medical<br>Systems Company<br>540 Alder Drive<br>Milpitas, California 95035 | | |----|----------------------|---------------------------------------------------------------------------------------------------------|---------------| | | | Tel: (408) 468-3051<br>Fax: (408) 468-3050 | | | | Contact Person: | Coleen Coleman<br>At address above | | | B. | Device Trade Name: | ENsphere™ Tempo Cardiology Workstation | | | | Common Name: | Picture Archive and Communication<br>Systems (PACS) | | | | Classification Name: | Image Processing System | | | C. | Predicate Device(s): | | | | | Manufacturer | Product Name | 510(k)<br>No. | | ADAC Laboratories | Pegasys Ultra™ | K993940 | |-------------------|--------------------------------------|---------| | ADAC Laboratories | ENsphere™ (Physician Desktop Review) | K021669 | #### D. Device Description: ENsphere™ Tempo is a Windows®-based Nuclear Mcdicine workstation specific for the Cardiology market segment. The computer system will consist of a Hewlett Packard XW4100 workstation or equivalent. The comprehensive tools and features provided with this product, will allow the technologist and/or physician to perform image review, processing of source data, post processing, hardcopy production, interpretation, report gencration and contains the utilities necessary to support the workflow and data management between those activities. The system will support connectivity aspects necessary to import and export data as required to accomplish daily work scenarios. - E. Intended Use: ENsphere™ Tempo is a nuclear medicine image processing and display workstation that provides cardiology specific software applications used to {2}------------------------------------------------ process, analyze, and display medical images. When interpreted by a nuclear physician, the results obtained by using the ENsphere Tempo applications may be used as a tool in determining a patient's diagnosis. ENsphere Tempo should only be operated by qualified healthcare professionals trained in the use of nuclear medicine equipment. This workstation is intended to be operated in a hospital, clinic, or doctor's office. Patient data may be transferred via DICOM over Local Area Networks (LAN), or Wide Area Networks WAN. The primary users of this product include nuclear medicine technologists who process the data and physicians who display, review, and interpret the processed image data. - Technological Comparison: F. The ENsphere™ (Physician Desktop Review K021669), Pegasys Ultra™ (K993946), and ENsphere™ Tempo have similar indications for use and overall function and perform in a similar manner with respect to, display, review and processing applications, data storage, and system utilities. #### II. CONCLUSION ENsphere™ Tempo is substantially equivalent to the following predicate devices, Pegasys Ultra™ (K993946) and ENsphere™ (Physician Desktop Review K021669) based on similar intended use, technological comparison, and system performance. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized depiction of an eagle with three lines representing its wings. The emblem is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB - 6 2004 ADAC Laboratories % Ms. Ellen Fickewirth Reviewer Underwriters Laboratoies, Inc. Santa Clara Division 1655 Scott Boulevard SANTA CLARA CA 95050-4169 Re: K040142 Trade/Device Name: ENsphere™ Tempo Cardiology Workstation Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: January 21, 2004 Received: January 22, 2004 Dear Ms. Fickewirth: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892,2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmaldsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Page 2 Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE STATEMENT 510 (k) NUMBER (if KNOWN): _ K 040142 DEVICE NAME: ENsphere™ Tempo Processing Workstation SPONSOR NAME: ADAC Laboratories ### INDICATIONS FOR USE: ENsphere™ Tempo is a nuclear medicine image processing and display workstation that provides cardiology specific software applications used to process, analyze, and display medical images. The applications included with ENsphere Tempo may be used as a tool, by the Physician in interpreting Nuclear Cardiology procedures. ENsphere Tempo should only be operated by qualified healthcare professionals trained in the use of nuclear medicine equipment. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter-Use (Optional Format 1-2-96) 000013 Nancy C. Hoddon **Division Sign-Off** Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K040142
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...