PACS SOFTWARE SOLUTION 2003

K033624 · Canon USA, Inc. · LLZ · Dec 15, 2003 · Radiology

Device Facts

Record IDK033624
Device NamePACS SOFTWARE SOLUTION 2003
ApplicantCanon USA, Inc.
Product CodeLLZ · Radiology
Decision DateDec 15, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

PACS Software Solution 2003 (PSS03) is intended to be used by radiologists and other medical professionals for diagnosis, or for clinical review, of medical images on digital display devices or hard copy media. PSS03 provides an efficient interface. Additionally, PSS03 is intended to be used with imaging equipment (e.g., Magnetic Resonance Imaging (MRI), Computerized Tomography (CT), Digital Radiography (DR), Ultrasound (US), etc.) and patient information systems (e.g., Hospital Information Systems (HIS), Radiology Information Systems (RIS)) approved for medical use.

Device Story

PACS Software Solution 2003 (PSS03) is a software-based Picture Archiving and Communication System (PACS) designed to manage medical imaging data. It ingests images from various modalities (MRI, CT, DR, Ultrasound) via DICOM protocols. The system operates on Unix-based hardware, providing tools for image display, processing, rendering, archiving, and distribution. It supports user-specific presentation preferences and redundant data storage. Healthcare providers use workstations to view images, utilizing filtering tools (smooth/sharp) and JPEG compression. The system integrates with RIS/HIS for patient data management, assigning unique exam keys to ensure patient integrity. PSS03 is intended for use by radiologists and clinicians in clinical settings to facilitate diagnostic decision-making and clinical review, potentially improving workflow efficiency and image accessibility.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Software-based PACS running on Unix operating systems. Supports DICOM, TCP/IP, and Ethernet communication protocols. Features include image server, database server, and workstations. Image processing includes 8-bit and 12-bit JPEG compression and low-pass blurring kernels for image smoothing. Display devices calibrated via SMPTE patterns or graphics card supplier software. Scalable architecture for small to large facilities.

Indications for Use

Indicated for radiologists and medical professionals to acquire, process, display, evaluate, archive, print, and distribute DICOM-compliant medical images from modalities including CR, CT, MR, X-Ray, and Ultrasound for clinical diagnosis or review.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 1633624 Image /page/0/Picture/1 description: The image shows the word "Canon" in a bold, stylized font. The letters are thick and have a slightly rounded appearance. The word is in black and the background is white. CANON U.S.A., INC. 15955 Alton Parkway Irvine, CA 92618-3731 Tel.: (949) 753-4000 www.usa.canon.com DEC 1 5 2003 ### 510(k) Summary | Submitter: | Canon U.S.A., Inc.<br>Canon Medical Systems Division<br>15955 Alton Parkway<br>Irvine, CA 92618 | |-------------------------------------|-------------------------------------------------------------------------------------------------| | Contact person: | Sean M. Curry<br>16787 Bernardo Center Drive, Suite A1<br>San Diego, CA 92128 | | Phone: | (858) 675-8200 | | FAX: | (858) 675-8201 | | Proprietary name: | PACS Software Solution 2003 | | Common name: | Picture Archiving and Communication System (PACS) | | Classification: | 892.2050 | | Product Code: | LLZ | | Classification name: | System, Imaging Processing, Radiological | | Substantial equivalence claimed to: | | Substantial equivalence claimed to: K940664: Paper Film - PACS ### Description: PACS Software Solution 2003 (PSS03)' is proprietary software used in conjunction with computer hardware to provide a Picture Archiving and Communication System. PSS03 has been designed to manage medical imaging data from various imaging devices (e.g. Magnetic Resonance Imaging, Computerized Tomography, Digital Radiography, Ultrasound, etc.). Management of medical imaging data from these devices, enabled by PSS03, includes display tools, the support of industry standard communication protocols (e.g. Transfer Control Protocol/Internet Protocol, Digital Communication in Medicine, Ethernet, etc.) and archiving of data using hierarchical protocols. The system provides redundant storage of the exam data. PSS03 accommodates individual user preferences by automatically displaying medical images in a presentation format unique to the user and to the type of exam selected. All referenced product names mentioned throughout this document, and other marks, are or may be trademarks of their respective owners. {1}------------------------------------------------ PSS03 is highly scalable, providing a small facility or a large enterprise of facilities the same functionality. ## Intended use: PACS Software Solution 2003 (PSS03) is intended to be used by radiologists and other I ACD Bottware Dolunion 2008 y medical images on digital display devices or hard copy motived professionals of diagnosis, or for clinical review, PSS03 provides an efficient princis. For parposes or clag.com, nanages the archiving of medical images for short used linerface. Additionally, I BO% interf. PSS03 in intended to be used with imaging equipment (e.g., Magnetic Resonance Imaging (MRI), Computerized Tomography (CT), Digital Radiography (DR), Ultrasound (US), etc.) and patient information systems (e.g., Hospital Information Systems (HIS), Radiology Information Systems (RIS)) approved for medical use. # Summary of technological characteristics in comparison to predicate device: The PACS Software Solution 2003 and the Paper Film - PACS, are Picture Archiving and Communication Systems (PACS) that operate under Open Windows on a Unix based operating system. They both consist of one or more of the following: image server, database server, and workstations for acquiring, processing, rendering, archiving, printing and distributing images. PACS Software Solution 2003 has been updated to take advantage of newer faster systems such as monitor resolution and tape drives as well as allowing for multiple external storage devices. The viewing applications in both systems provide "smooth" and "sharp" filtering of images. A low-pass blurring kernel is used for smoothing. Both systems also utilize 8 bit and 12 bit compression using JPEG format. The new system ensures patient integrity by assigning a Radiology Information System key to cach exam, in addition to the Medical Record Number and the patient's name (which was used with the predicate device). As with the predecessor, the PACS Software Solution 2003 display devices are calibrated using a standard SMPTE pattern and/or calibration software provided by the graphics card supplier. The PACS Software Solution 2003 system has the same Indications for Use, equivalent design features, methods of use, and equivalent functional characteristics as the predicate device and therefore raises no new safety or effectiveness issues. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular fashion around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 5 2003 Canon U.S.A., Inc. % Mr. Sean Curry Chief Operating Officer Certified Software Solutions, Inc. 16787 Bernardo Center Drive, Suite A1 SAN DIEGO CA 92128 Re: K033624 Trade/Device Name: PACS Software Solution 2003 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: November 11, 2003 Received: November 18, 2003 Dear Mr. Curry: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely vours. Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): Ko3 362 4 Device Name: PACS Software Solution 2003 Indications for Use: PACS Software Solution 2003 is used to acquire, process, display, evaluate, archive, print and distribute DICOM compliant images for clinical purposes. Images may be acquired from imaging modalities such as CR, CT, MR, X-Ray, and other devices. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-the-Counter Use Daniel A. Szyrmer (Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number .
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...