K031562 · Ramsoft, Inc. · LLZ · Aug 11, 2003 · Radiology
Device Facts
Record ID
K031562
Device Name
RAMSOFT PACS, MODEL 3.0
Applicant
Ramsoft, Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 11, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
RamSoft PACS software is used with general purpose computing hardware to acquire, transmit, store and display images for clinical purposes. RamSoft PACS consists of one or more of image server, web server, database server, and workstations for interpretation, clinical review, quality control, video capturing, film digitizing, document scanning. Images may be acquired from imaging devices such as CR, CT, DX, MR, NM, PT, RF, US, XA, and other devices.
Device Story
RamSoft PACS is a stand-alone software package operating on general-purpose computing hardware. It functions as a Picture Archiving and Communication System (PACS) to acquire, transmit, store, and display medical images. Inputs include digital images from modalities like CR, CT, DX, MR, NM, PT, RF, US, and XA, as well as video captures, film digitizing, and document scanning. The system utilizes a network (TCP/IP) to manage data between image servers, web servers, and workstations. Physicians use the software for clinical review, interpretation, and quality control. Features include image measurement, cine tools, comparison modes, MPR, MIR, and annotation. The system provides clinicians with tools to manipulate and view images, facilitating diagnostic decision-making. It is intended for use in clinical environments where a physician provides human intervention during image interpretation.
Clinical Evidence
Bench testing only. No clinical data provided. The device was tested according to a documented System Test Plan as part of the software development process.
Technological Characteristics
Stand-alone software for general-purpose computing hardware. Supports DICOM 3.0, Bitmap, JPEG, TIFF, and TWAIN inputs. Network protocol: TCP/IP. Compression methods: JPEG, RLE. Features include MPR, MIR, image measurement, cine tools, and annotation. Software level of concern: minor.
Indications for Use
Indicated for clinical image acquisition, transmission, storage, and display for diagnostic interpretation and review. Used by clinicians in radiology and other clinical settings. No specific age or gender restrictions.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Tel: (416) 674-1347 Fax: (416) 674-7147 E-mail: sales@ramsoft.biz
Image /page/0/Picture/2 description: The image shows a logo with the word "RAMSOFT" inside a circle. The circle is surrounded by a design that looks like two curved shapes. The logo is black and white and has a simple design.
uanian mu 16 Four Seasons Place, Unit 215 Toronto, ON M9B 6E5 CANADA Web Site: http://www.ramsoft.biz
## 510(k) Summary
Contact Name: Lucian Popescu, QA Manager Trade Name - RamSoft PACS Common Name - Picture Archiving and Communication System (PACS) Classification Name - LLZ - Image Processing System (892.2050)
Substantially Equivalent devices:
Ultrapro 98 (RamSoft Inc) - K982563
Kodak AutoRad (Kodak) - K955092
GE Advantage (General Electric Medical Systems) - K94120
#### Description:
RamSoft PACS is a stand-alone software package that is used on general purpose computing hardware. As long as minimum hardware requirements are met, the user or system integrator is free to choose his/her own hardware platform.
The software allows digital image processing, measurement, communication, and storage.
RamSoft PACS is tested according to the specifications that are documented in a RamSoft PACS System Test Plan. Testing is an integral part of RamSoft's software development process.
The software does not contact the patient, nor does it control any life-sustaining devices. A physician has ample opportunity for competent human intervention while interpreting images and clinical information.
Intended Use:
RamSoft PACS software is used with general purpose computing hardware to acquire, transmit, store and display images for clinical purposes. RamSoft PACS consists of one or more of image server, web server, and workstations for interpretation, dinical review, quality control, video capturing, film digitizing, document scanning. Images may be acquired from imaging devices such as CR, CT, DX, MR, NM, PT, RF, US, XA, and other devices.
PD0003
RAMSO
{1}------------------------------------------------
K031542
Page 2 of 2
Image /page/1/Picture/1 description: The image shows a close-up of a horseshoe. The horseshoe is made of metal and has a curved shape. The horseshoe appears to be old and worn, with some rust and discoloration. The background is blurred, which helps to focus attention on the horseshoe.
# Comparison to Predicate Devices
| Product Name | GE Advantage<br>Windows (K942120) | Kodak AutoRad<br>(K955092) | RamSoft Ultrapro 98<br>K982563 | RamSoft PACS<br>(this<br>submission) |
|-----------------------------------------------------|-----------------------------------|----------------------------|--------------------------------|--------------------------------------|
| Print to paper | Yes | Yes | Yes | Yes |
| Graphical UI | Yes | Yes | Yes | Yes |
| Windows OS | Yes | Yes | Yes | Yes |
| Uses<br>standard<br>displays | Yes | Yes | Yes | Yes |
| Image input | DICOM 3.0 | DICOM 3.0 | DICOM 3.0 | DICOM 3.0 |
| | | | Bitmap | Bitmap |
| | | | JPEG | JPEG |
| | | | TIFF | TIFF |
| | | | Film | Film |
| | | | Video | Video |
| | | | TWAIN | TWAIN |
| Images stored on<br>remote NT server | Yes | Yes | Yes | Yes |
| Network protocol | TCP/IP | TCP/IP | TCP/IP | TCP/IP |
| Compression | Wavelet | JPEG | JPEG | JPEG |
| | | | RLE | RLE |
| Annotation | Yes | Yes | Yes | Yes |
| Image<br>Measurement | Yes | No | Yes | Yes |
| | | | | |
| Cine tool | Yes | Yes | Yes | Yes |
| Comparison Mode | Yes | Yes | Yes | Yes |
| Review RIS Report | Yes | Yes | Yes | Yes |
| Designed for use<br>inside and outside<br>radiology | Yes | Yes | Yes | Yes |
| | | | | |
| Flip/Rotate images | Yes | Yes | Yes | Yes |
| User Log in | Yes | Yes | Yes | Yes |
| Multiple<br>layout<br>options | Yes | Yes | Yes | Yes |
| Window<br>control<br>and presets | Yes | Yes | Yes | Yes |
| | | | | |
| Patient and Study<br>Browser | Yes | Yes | Yes | Yes |
| | | | | |
| MPR | Yes | No | No | Yes |
| MIR | Yes | No | No | Yes |
## Conclusion
In conclusion, RamSoft PACS is substantially equivalent to RamSoft Ultrapro 98, GE Advantage, and Kodak AutoRad. The determination of substantial equivalence is not based on an assessment of performance tests. It is our conclusion that there is no software component in RamSoft PACS product or hardware component which would be used in conjunction with RamSoft PACS product that we know of whose failure or latent design flaw would be expected to result in death or injury to a patient. Thus, the "Level of Concern" of RamSoft PACS product is "minor."
Image /page/1/Picture/6 description: The image shows the word "RAMSOFT" in large, bold, black letters. Above the word "RAMSOFT" is the text "PD0003" in a smaller font size. There is a horizontal line above the text "PD0003".
May 16, 2003
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo with a stylized bird in the center. The bird is depicted with three curved lines forming its body and wings, giving it a sense of motion. Encircling the bird is text, arranged in a circular fashion around the bird. The text appears to be in all capital letters.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 1 2003
Mr. Lucian Popescu QA Manager RamSoft, Inc. 215-16 Four Seasons Place Toronto, ON M9B 6E5 CANADA
Re: K031562 Trade/Device Name: RamSoft PACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: May 16, 2003 Received: May 21, 2003
Dear Mr. Popescu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
510(k) Number (if known): K031562
Device Name: RamSoft PACS
Indications for Use:
RamSoft PACS software is used with general purpose computing hardware to acquire, transmit, store and display images for clinical purposes. RamSoft PACS consists of one or more of image server, web server, database server, and workstations for interpretation, clinical review, quality control, video capturing, film digitizing, document scanning. Images may be acquired from imaging devices such as CR, CT, DX, MR, NM, PT, RF, US, XA, and other devices.
# (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
X Prescription Use
Over-The-Counter Use -OR-
David G. Bergman
ivision Sign-Off Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.