← Product Code [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ) · K023533

# BARCO MEDIS 5MP2 AURA (K023533)

_Barco N.V. · LLZ · Jan 17, 2003 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K023533

## Device Facts

- **Applicant:** Barco N.V.
- **Product Code:** [LLZ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ.md)
- **Decision Date:** Jan 17, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

The Barco MeDis 5MP2Aura Medical Image Processing System is intended to be used in displaying and viewing digital images for review by trained medical practitioners.

## Device Story

MeDis 5MP2 Aura is a digital image display system; provides high-resolution visualization of medical images. Used by trained medical practitioners in clinical settings for review and analysis of digital diagnostic images. Device functions as a workstation monitor to render images for clinical decision-making.

## Technological Characteristics

Digital image display system; high-resolution visualization components.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Barco NV Display Systems MeDis 5MP2 Quad-Head Medical Diagnostic Display System ([K001748](/device/K001748.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JAN 1 7 2003

## 510(K) SUMMARY

## ん c235 33

| Manufacturer:                  | Barco NV Barcoview<br>Theodoor Sevenslaan 106<br>8500 Kortrijk<br>Belgium                                                                                                      |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By:                  | Ferguson Medical<br>Consultant to Barco NV                                                                                                                                     |
| Contact Information:           | Phone: +32(0) 56 23 32 11<br>FAX: +32(0) 56 23 3 74                                                                                                                            |
| Classification Name:           | System, image processing                                                                                                                                                       |
| Common/Usual Name:             | Image display system, medical image<br>workstation, image monitor/display, and others                                                                                          |
| Proprietary Name:              | Barco MeDis 5MP2 Aura                                                                                                                                                          |
| Classification Number:         | 21 CFR 892.2050/Procode 90LLZ                                                                                                                                                  |
| Substantial Equivalence:       | Barco NV Display Systems MeDis 5MP2 Quad-<br>Head Medical Diagnostic Display System<br>(K001748)                                                                               |
| Device Description:            | The MeDis 5MP2 Aura device is a digital image<br>display system                                                                                                                |
| Intended Use:                  | The Barco MeDis 5MP2Aura Medical Image<br>Processing System is intended to be used in<br>displaying and viewing digital images for<br>review by trained medical practitioners. |
| Technological Characteristics: | The Barco MeDis 5MP2 Aura device consists of<br>components to provide high resolution<br>visualization of digital images.                                                      |

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing its body and wings. The eagle is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the circumference of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## JAN 1 7 2003

Barco NV c/o Mr. Frank Ferguson Official Correspondent Ferguson Medical P.O. Box 12038 LA JOLLA CA 92039

Re: K023533

Trade/Device Name: MeDis 5MP2 Aura Regulation Number: 21 CFR 8892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: September 25, 2002 Received: October 21, 2002

Dear Mr. Ferguson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx. 1xxx                        | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely vours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page 1 of 1

510(k) Number (If known):

Device Name: MeDis 5MP2 Aura

Indications For Use:

The MeDis 5MP2 Aura is intended to be used in displaying and viewing digital images for review and analysis by trained medical practitioners.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ramona H. Hare for NB

(Division Sign-Of)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K023533

Prescription Use _ XX (Per 21 CFR 801.109)

OR

Over-The- Counter Use ________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K023533](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/LLZ/K023533)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com