K023476 · Universal Pacs, Inc. · LLZ · Jan 7, 2003 · Radiology
Device Facts
Record ID
K023476
Device Name
UNIPACS WORKSTATION
Applicant
Universal Pacs, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 7, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
UniPACS Workstation software is intended to be used as a fully-functional, PCbased PACS software application, to review, interpret, manipulate, network, print and visualize medical multi-modality image data in DICOM format. When interpreted by a trained physician, reviewed images can be used as an element for diagnosis.
Device Story
UniPACS Workstation is a Windows-based software application functioning as a standalone Picture Archiving and Communications System (PACS). It operates on standard off-the-shelf PC hardware. The system ingests medical image data in DICOM format; provides tools for physician review, interpretation, manipulation, printing, and visualization. Networking capabilities allow exchange of image data with DICOM or FTP-compatible servers via TCP/IP. A web component enables remote review and reporting via web-browser interface. Used by physicians, nurses, and technicians in clinical settings. Output assists physicians in clinical decision-making and diagnosis. Device does not contact patients or control life-sustaining equipment.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Windows-based software application; runs on standard off-the-shelf PC hardware; supports DICOM and FTP networking protocols; TCP/IP connectivity; web-browser interface for remote access.
Indications for Use
Indicated for trained professionals (physicians, nurses, technicians) to review, interpret, manipulate, network, print, and visualize multi-modality medical images in DICOM format for diagnostic support.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Ko2 3476
JAN 0 7 2003
# B. Administrative Information
### 1. 510(k) Summary of Safety and Effectiveness
### Date Prepared
October 8, 2002
#### Submitter's Information
Universal PACS, Inc. 127 Albert Hart Drive Baton Rouge, LA 70803 Phone: (225) 766-9381
## Contact Person
John M. Tyler, CEO, Universal PACS, Inc. Phone: (225) 766-9381 (225) 578-2198 Email: tyler@bit.csc.lsu.edu
#### Device
Trade Name: UniPACS Workstation Picture Archiving and Communications System Common Name: PACS (per 21 CFR 892.2050), LLZ, Class II Classification Name:
### Substantially Equivalent To
eFilm Workstation (k012211, see also k020995)) eFilm Medical Inc. 500 University Ave, Suite 300 Toronto, Ontario Canada M5G 1V7 www.efilm.ca
{2}------------------------------------------------
#### Device Description
The UniPACS Workstation is a Windows-based software application acting as stand alone Picture Archiving and Communications System (PACS). It may be marketed as software only, as well as packaged with standard "off-the-shelf" PC hardware.
The comprehensive set of tools provided in this software allows the physician to review, interpret, manipulate, print and visualize medical image data stored in DICOM format. The networking component of the product allows the physician to exchange image data with any other DICOM-compatible or FTP-compatible server over a standard TCP/IP network. The web component of the product displays workstation images in the web-browser interface, for remote review and reporting.
The device does not contain the patient, nor does it control any life sustaining devices.
#### Indications for Use
UniPACS Workstation software is intended to be used as a fully-functional, PCbased PACS, to review, interpret, manipulate, network, print and visualize medical multi-modality image data in DICOM format. When interpreted by a trained physician, reviewed images can be used as an element for diagnosis.
Typical users of this system are trained professionals, including but not limited to physicians, nurses and technicians.
#### Technological Comparison to Predicate Device
The proposed (UniPACS) and predicate (eFilm) devices are both software programs that can be used for manipulation of DICOM-compliant images on PC hardware. The proposed and predicate software can be operated from a personal computer, and provide a standard set of features with respect to image processing, storage and networking. The detailed feature comparison between the UniPACS Workstation and the eFilm Workstation demonstrates that, except a few minor, unimportant differences, the UniPACS Workstation provides the users with functionality, equivalent to that of the eFilm Workstation predicate device. This is a strong proof that the UniPACS and eFilm PACS Workstations are substantially equivalent.
#### Conclusion
Universal PACS Inc. considers the UniPACS Workstation to be substantially equivalent to the eFilm Workstation (k012211) device. The 510(k) Premarket Notification for the UniPACS Workstation contains adequate information and data to enable FDA-CDRH to determine substantial equivalence to the predicate device.
The UniPACS Workstation PACS software will be manufactured in accordance with the voluntary standards and development techniques listed in the submission. The software does not result in any new potential safety risks and performs as stated in its intended use.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines suggesting movement or connection.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 0 7 2003
Mr. John M. Tyler CEO Universal PACS, Inc. 127 Albert Hart Drive BATON ROUGE LA 70803 Re: K023476
Trade/Device Name: UniPACS Workstation Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communication system Regulatory Class: II
Product Code: 90 LLZ Dated: October 8, 2002 Received: October 16, 2002
Dear Mr. Tyler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
2. FDA Statement of Indications for Use
Kaz 3476 510(k) Number (if known)
Device Name: UniPACS Workstation
#### Indications for Use:
UniPACS Workstation software is intended to be used as a fully-functional, PCbased PACS software application, to review, interpret, manipulate, network, print and visualize medical multi-modality image data in DICOM format. When interpreted by a trained physician, reviewed images can be used as an element for diagnosis.
Typical users of this system are trained professionals, including but not limited to physicians, nurses and technicians.
### (PLEASE DO NO WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
-OR-
Prescription Use . (Per 21 CFR 801.109) Over-The-Counter Use
Daniel A. Searson
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K023476
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.