VIZDEXTER VERSION 2.0; DEXTROSCOPE MK8; DEXTROBEAM MK3
Applicant
Volume Interactions Pte, Ltd.
Product Code
LLZ · Radiology
Decision Date
Oct 25, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
Volume Interactions Pte Ltd's Image Processing System is a medical device for the display and 3D visualization of medical image data derived from CT and MRI scans. It is intended to be used by gualified and trained medial professionals, after proper installation. Volume Interactions Pte Ltd's Image Processing System is not intended to be used in direct contact with the patient nor is it intended to be connected to equipment that is used in direct contact with the patient.
Device Story
System processes DICOM 3.0 tomographic data (CT/MRI) to generate 3D reconstructions for surgery planning and visualization. Comprises VizDexter 2.0 software, Dextroscope interactive console, and DextroBeam collaborative projection system. Operates on Windows NT/2000 PCs. Users interact with 3D virtual objects using two-handed 3D trackers (joystick/stylus) while wearing stereoscopic glasses. Dextroscope uses mirror-based virtual image viewing; DextroBeam uses stereoscopic screen projection for group discussion. Clinicians manipulate, segment, and measure structures in 3D space. Output includes 3D renderings, distance/volume measurements, and TIFF projections. Facilitates surgical planning by providing enhanced spatial visualization of anatomy. No direct patient contact; physician interprets output for clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence based on technological comparison, risk management, and verification/validation testing.
Technological Characteristics
Runs on IBM PC compatible hardware (Windows NT/2000). Inputs: DICOM 3.0, SGI, TIFF, RAW. Outputs: TIFF, proprietary 3D formats. Rendering: Direct volume rendering, polygonal surface display, multi-modal fusion. Interface: Stereoscopic glasses, mirror-based display (Dextroscope), stereoscopic projection (DextroBeam), 3D tracker (joystick/stylus). Connectivity: TCP/IP. Storage: SCSI, CD-ROM. Software: Standalone application.
Indications for Use
Indicated for display and 3D visualization of CT and MRI medical image data. Intended for use by qualified, trained medical professionals. Not for direct patient contact or connection to patient-contact equipment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
# 510(K) Summary
The following safety and effectiveness summary has been prepared pursuant to requirement for 510(K) summaries specified in 21CFR 807.92(a)
# Submitter Information
Volume Interactions Pte Ltd 5 Shenton Way, #37-04 UIC Building, Singapore 068808
+ 65 6226962 Phone: + 65 62226215 Facsimile:
### Contact Person
John Baby, Process Development Manager, QA/RA
#### Date Prepared
August 26, 2002
# Device Name
- Trade Name Dextroscope™, DextroBeam™, VizDexter™ 2.0
### Common Name
lmage Processing System
#### Classification Name
Picture Archiving and Communications System (per 21 CFR 892.2050)
#### Classification Number
LLZ
## Devices to which substantial equivalence is being claimed
The Volume Interactions Image Processing System is substantially equivalent to
| 510(k)<br>Number | Device Name | Manufacturer |
|------------------|-----------------------------|-------------------|
| K992654 | Plug n View 3D, Version 1.0 | Voxar Limited |
| K002519 | Vitrea 2, Version 2.1 | Vital Images Inc. |
# Device Description
Volume Interactions Image Processing System reads DICOM 3.0 format medical image data sets (and other formats) and displays 3D image reconstructions of these data sets through various user selectable industry standard rendering methods and algorithms. The clinical users can spatially manipulate, process to highlight structures and volumes of interest, and measure distances and volumes in the 3D image reconstructions. The processed data can be
{1}------------------------------------------------
#### Image Processing System Volume Interactions Pte Ltd
stored either as 3D image data in a proprietary format. or as 2D picture projections of the 3D image data in TIFF image format. The system runs on commercially available IBM PC compatible computers and hardware components with the Microsoft Windows NT and 2000 operating systems.
The system consists of three product modules namely, VizDexter™ 2.0, Dextroscope™ and DextroBeam™. The modules are described as follows:
VizDexter™ 2.0 is software that processes tomographic (e.g.: Computer Tomography, Magnetic Resonance Imaging) data and produces stereoscopic 3D renderings for surgery planning and visualization purposes. The software user selectable industry standard rendering methods and algorithms.
Dextroscope™ is an interactive console and display system that allows the user to interact with two hands with the 3D images generated by the VizDexter™ software. The Dextroscope™ user works seated, with both forearms positioned on armrests. Wearing stereoscopic glasses, the user looks into a mirror and perceives the virtual image within comfortable reach of both hands for precise hand-eye coordinated manipulation. The hardware uses various industry standard components.
DextroBeam™ is an interactive console intended for group collaborative discussions with 3D images using a stereoscopic projection system. The DextroBeam™ system uses the base of the Dextroscope™ as the 3D interaction interface with the virtual objects. The monitor of the Dextroscope™ is replaced by a screen projection system, so instead of looking into the mirror of the Dextroscope™, the user looks at large stereoscopic screen projections while working with the virtual data in reach of his hands. This enables the discussion of 3D data sets with other specialists in stereoscopic 3D. The hardware uses various industry standard components.
# Intended Use
Volume Interactions Pte Ltd's Image Processing System is a medical device for the display and 3D visualization of medical image data derived from CT and MRI scans. It is intended to be used by gualified and trained medial professionals, after proper installation.
Volume Interactions Pte Ltd's Image Processing System is not intended to be used in direct contact with the patient nor is it intended to be connected to equipment that is used in direct contact with the patient
# Comparison of Technological Characteristics
The Volume Interactions Image Processing System is substantially equivalent to Voxar Limited Plug n View 3D, Version 1.0 (K992654) for non measurement features and Vital Images Vitrea 2, Version 2.1 (K002519) for measurements.
| Feature | Volume Interactions<br>Image Processing System<br>(VizDexter ™ 2.0,<br>Dextroscope™,<br>DextroBeam™) | Voxar Plug n View 3D,<br>version 1.0 |
|--------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | | K992654 |
| Intended use | Intended for the display and<br>3D visualization of medical<br>image data derived from CT<br>and MRI scans. It is intended<br>to be used by qualified and<br>trained medical professionals,<br>after proper installation. | Intended for use by<br>radiologists, clinicians and<br>referring physicians to<br>acquire, process, render,<br>review, store, print and<br>distribute DICOM 3.0<br>complaint image studies,<br>utilizing standard PC |
| | | hardware. |
| Type of software<br>Program | Stand alone application<br>program | Stand alone application<br>program |
| Software platform used | Windows NT, Windows 2000<br>professional | Windows 95/98/NT |
| Stereoscopic Interactive<br>console | Dextroscope or DextroBeam | None |
| Communications | TCP/IP | TCP/IP |
| Image Format In | DICOM 3.0, Classic SGI, TIFF,<br>RAW | ACR NEMA 2.0, DICOM 3.0 |
| Image Format out | TIFF image format | DICOM 3.0 |
| Image Display | Color CRT or<br>Stereoscopic Projection<br>System | Color/Grayscale, CRT or<br>Laptop LCD |
| Image Archive | SCSI 10-30 Gbytes, CD-ROM | SCSI 2-20 Gbytes, CD ROM |
| Printing | No | Printing to standards window<br>printers |
| Prescriptive Device | Prescription use only | Prescription use only |
| Segmentation by<br>Threshold selection | Yes | Yes (Window level Presets) |
| Morphology<br>Segmentation Manager | Yes | Yes (Limited) |
| Segmentation by<br>Contour Editor | Yes | Yes (Shape & Sculpt) |
| Volume Rendering,<br>Triplanar Mode and 3D<br>view | | |
| Simultaneous display<br>of coronal + saggital<br>+ axial plane | Yes (Triplanar Mode) | Yes (Multi Planar Reformatting<br>(MPR) view) |
| Translate 3 Planes | Yes | Yes |
| Display lines of<br>intersection between<br>planes | Yes (shown as actual 3D plane<br>intersections) | No |
| Patient orientation<br>Labels | No | Yes |
| Patient Labels | Yes | Yes |
| Adjustable text on<br>labels | Yes | Yes |
| Borders to highlight<br>"current" plane | Yes | Yes |
| Reset<br>pan/zoom/orientation | Yes | Yes |
| Show Volume<br>Rendering and<br>Triplanar view<br>concurrently | Yes | Yes (shows Volume Rendering<br>and 3 quadrant views) |
| Adjustable clipping in<br>Triplanar Mode | Yes | Yes (in 3 quadrant views) |
| Volume Rendered 3D<br>Display mode | Yes | Yes |
| Polygonal Surface<br>object display mode | Yes | No |
| Multi-modality<br>Volume Rendered<br>Display mode | Yes | No |
| MIP Display Mode | No | Yes |
| Basic Tools | | |
| Volume | Yes | Yes |
| Crop | Yes | Yes (Shape tool) |
| Cut | Yes | No (only orthogonal cuts) |
| Pick | Yes | Yes (Selection tool) |
| Zoom | Yes | Yes |
| Pitch | Yes | Yes (resolution varies during<br>image movement) |
| View Box | Yes | Unknown |
| Color and<br>transparency editing | Yes (Volume Console) | Yes (Active Color Presets) |
| Image Processing | | |
| Rotation | Yes | Yes |
| Pan | Yes | Yes |
| Zoom | Yes | Yes |
| Flip according to<br>patient orientation | Yes (Manual Flipping) | Yes |
| Opacity | Yes | Yes |
| Orthogonal clipping | Yes | Yes |
| Sculpting | Yes (Contour Editor) | Yes (Shape & Sculpt tool) |
| Window level<br>protocols | No | Yes |
| Image Layout | | |
| Show Single Screen | Yes | Yes |
| Show four quadrants<br>view (coronal +<br>saggital + axial+ 3D) | No | Yes |
| Multi-modal volume rendering | | |
| 3D registration | Yes | No |
| Multimodal fusion of CT and MR scans | Yes | No |
| Image Storage | | |
| Add captured 3D rendered image to user profiles | Yes (as TIFF files) | Yes |
| Record movie clips | Yes | Yes |
| Hardcopy generation | | |
| Print to DICOM and NON-DICOM printers | No | Yes |
| Image Editing Tools | Erase and Restore | Tools for removal of obscuring anatomy |
| Measurements | | |
| Distance | Yes | Yes |
| Area | No | No |
| Volume | Yes | Yes |
| Feature | Volume Interactions<br>Image Processing System<br>(VizDexterTM 2.0,<br>DextroscopeTM,<br>DextroBeamTM) | Vital Images Vitrea 2,<br>Version 2.1 |
| 510(k) number | | K002519 |
| Intended use | Intended for the display and 3D<br>visualization of medical image<br>data derived from CT and MRI<br>scans. It is intended to be used<br>by qualified and trained medial<br>professionals, after proper<br>installation. | Intended for processing /<br>analyzing 2D/3D images from<br>CT / MR scanners. |
| Type of software program | Stand alone application<br>program | Stand alone application<br>program |
| Software platform<br>required | Windows NT, Windows 2000<br>professional | Windows NT |
| Stereoscopic Interactive<br>console | Dextroscope or DextroBeam | None |
| Medical Image Modalities | | |
| All DICOM 3.0<br>recognized modalities<br>where datasets are<br>comprised of parallel,<br>2D images with know<br>inter-image spacing | Yes | Yes |
| Measurements | | |
| Distance | Yes | Yes |
| Area | No | Yes |
| Volume | Yes | Yes |
| Interactive contour<br>definition of lesions | Yes (Contour Editor) | No |
| Safety: Clinician<br>review/editing of data | Yes (distance and volume<br>measurements are displayed,<br>can be modified and be<br>accepted or rejected by<br>clinician) | Yes (clinician interactive<br>reviews and edits the data) |
| Data Source | CT/MR scanners | CT/MR scanners |
| Stereoscopic view | Yes | No |
| Dextroscope Technology<br>(direct hand access of 3D<br>objects using 3D joystick<br>and 3D stylus) | Yes | No |
{2}------------------------------------------------
{3}------------------------------------------------
.
{4}------------------------------------------------
# Image Processing System
Volume Interactions Pte Ltd
{5}------------------------------------------------
,
.
{6}------------------------------------------------
# Discussion of similarities and differences
The software module VizDexter™ 2.0 of the Volume Interactions Image Processing System utilizes similar technological characteristics as the predicate devices. All provide multiview user interfaces with combinations of 2D and 3D views correlated together for enhanced visualization. All provide measurement tools for analysis of the observed structures (volumes); and allow virtual lighting parameters to emphasize details.
VizDexter™ 2.0 and all predicates, provide 3D views. As with Vitrea 2 and Voxar Plug n View 3D, VizDexter 2.0 utilizes direct volume rendering for all its 3D views, including transparent volume images and visible surface views. For changing the mapping to opacity during translucent views, VizDexter 2.0 is similar to Voxar Plug n View 3D and Vitrea 2.
VizDexter™ 2.0 has functions of multi-modal volume rendering and a registration facility, which are not present in Plug n View 3D and Vitrea 2.
Volume Interactions Image Processing System and all predicates use identical technology to manipulate and display the 2D image views of the data sets (industry standard mouse and keyboard and a computer monitor). VizDexter™ 2.0 differs from all predicates however in the user interface utilized to manipulate the 3D image data, and in the display of the 3D image data. The user interfaces of the Dextroscope and the DextroBeam utilize an industry standard 3D tracker to position and orient the 3D image data with both hands, for ease of manipulation. The display of the 3D data is achieved by wearing stereoscopic glasses to view the 3D images of a computer monitor through a normal mirror (Dextroscope ™) or to view the 3D images displayed by a stereoscopic projector (DextroBeam ™). Voxar Plug n View 3D and Vital Images Vitrea 2.0 do not provide stereoscopic display nor hand controlled 3D trackers and follow the conventional keyboard and mouse interface for 2D and 3D interactions.
# Technological characteristics
The device does not contact the patient, nor does it control any life sustaining devices. A physician providing ample opportunity for competent human intervention, interprets images and information being displayed.
# Conclusion
We conclude that the Volume Interactions Image Processing System (VizDexter™ 2.0, Dextroscope™ and DextroBeam ™) is substantially equivalent to its predicate devices in its ability to render 3D images for use in medical analysis. In comparison of the predicate devices Volume Interactions Image Procession System provides a user interface that utilizes both hands to manipulate the 3D image data, and a stereoscopic display to view the 3D image data, which the predicate devices do not provide.
Regarding safety and effectiveness, Volume Interactions Pte Ltd is an ISO 9001;2000 certified company. All processes are controlled and monitored. Risk management including risk analysis, verification and validation tests. and evaluation by hospitals control potential hazards in both software and hardware. The level of concern for the Volume Interactions Image Processing System is minor.
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 25 2002
Mr. John Baby Process Development Manager Volume Interactions, Pte., Ltd. 5 Shenton Way #37-04 UIC Building 068808 SINGAPORE
Re: K022938
Trade/Device Name: VizDexter Version 2.0; Dextroscope MK8; Dextrobeam MK3 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ
Dated: August 30, 2002 Received: September 4, 2002
Dear Mr. Baby:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
Image /page/7/Picture/12 description: The image shows a partial view of a logo or seal, focusing on the upper left quadrant. A portion of the text "DEPARTMENT OF..." is visible, arranged vertically along the left edge. To the right of the text, there are three stylized, curved lines that appear to be a design element of the logo. The overall impression is that the image captures a fragment of an official emblem or seal.
{8}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Snogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
#### Indications for Use 2.
510(k) Number: Ko 2 2938
Device Name:
lmage Processing System
Indications for Use:
Volume Interactions Pte Ltd's Image Processing System is a medical device for the display and visualization of 3D medical image data derived from CT and MRI scans. It is intended to be used by qualified and trained medial professionals, after proper installation.
Volume Interactions Pte Ltd's Image Processing System is not intended to be used in direct contact with the patient nor is it intended to be connected to equipment that is used in direct contact with the patient.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _____________________________________________________________________________________________________________________________________________________________
(Per 21 CFR 801.109) (Optional Format 1-2-96)
Oerist A. Seymour
(Division Sign-Off) Division of Reproductive, Abdomina anc Padiological Devices 5 ! (k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.