K013098 · Medical Metrics, Inc. · LLZ · Dec 7, 2001 · Radiology
Device Facts
Record ID
K013098
Device Name
KIMAX REMOTE VIEWING STATION
Applicant
Medical Metrics, Inc.
Product Code
LLZ · Radiology
Decision Date
Dec 7, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
The KIMAX ™ Remote Viewing Station software and hardware (RVS) are intended to allow physicians to view static or dynamic digital images from radiographic and other medical imaging equipment. The RVS includes a high-performance computer, data storage devices, a high-resolution monitor and software to: 1) accept AVI images or import other image formats; 2) allow physician to quickly locate and display images; 3) adjust contrast and brightness, apply filters to enhance the images, zoom in on selected features, pan to select features of interest, and annotate images; 4) print out selected images; and 5) create reports that include text and annotated images. The RVS is intended to be used in the clinical workspace of a practicing physician.
Device Story
KIMAX Remote Viewing Station (RVS) is a workstation for medical image management; inputs include static or dynamic digital images from radiographic and other imaging equipment (e.g., AVI files); system comprises high-performance computer, data storage, and high-resolution monitor; software enables image retrieval, display, contrast/brightness adjustment, filtering, zooming, panning, annotation, printing, and report generation; used in clinical workspace by physicians; output allows clinicians to visualize and document diagnostic findings; facilitates clinical decision-making through enhanced image review and reporting.
Technological Characteristics
System includes high-performance computer, data storage, and high-resolution monitor; software-based image processing for display and manipulation; supports AVI and other image formats; intended for clinical workspace use.
Indications for Use
Indicated for use by physicians to view, manipulate, and report on static or dynamic digital medical images from radiographic and other imaging equipment in a clinical workspace.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Submission Summary (Full Text)
{0}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
## Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 0 7 2001
Mr. George T. Knackstedt Director Medical Metrics, Inc. 10540 Rockley Road, Suite 200 HOUSTON TX 77099
Re: K013098 K013098
Trade/Device Name: KIMAX™ Remote Viewing Station
Trade/Device Name: KIMAX™ 200 2050 TraderDevice Namber: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II
Product Code: 90 LLZ Dated: August 27, 2001 Received: September 17, 2001
Dear Mr. Knackstedt:
We have reviewed your Section 510(k) premarket notification of intent to market the device
ty the indication of the said to since is substantially equivalent (for the indica We have reviewed your Section 5 IQ(k) prematice in univation in muirvalent (for the indications
referenced above and have detectived the devices is substantially equivale referenced above and have determined the devices marketed in interstate
for use stated in the enclosure) to legally marketed prodicale devices marketed in interstate for use stated in the enclosure) to legally market predical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food, Drug, commerce prior to May 28, 1976, the enactment with the provisions of the Federal Food, Drug,
devices that have been reclassified in accordance with the provisions of the Rede devices that have been reclassified in accordance will the proval application (PMA).
and Cosmetic Act (Act) that do ice approval of a premalces approval application (PMA). and Cosmetic Act (Act) that do not require approval or a provisions of the Act. The
You may, therefore, market the device, subject to the general controls provisions of You may, therefore, market the device, subject to may begistration, listing of
general controls provisions of the Act include requirements for annual registrations and general controls provisions of the Act include requirements of the Groupen in Maranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (see above) into extrem confections affecting your device can be
it may be subject to additional controls. Existing major regulations affective it may be subject to additional controls. Existing major regulations and and of your of the may
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In add found in the Code of Federal Regulations, Trilo 21, I accords
publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean
ts of the Act Please be advised that FDA's issuance of a substanted doler requires with other requirements of the Act
that FDA has made a determination that your device complies with of th that FDA has made a determinations administered by other Federal agencies. You must
or any Federal statutes and regulations anningiting by other Federal on and listi or any Federal statutes and regulations and including, but not listing the mand listing
comply with all the Act's requirements, including, but not listing requirements as comply with all the Act's requirements, finciding, our not minted to reguirements as se
(21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirement (21 CFR Part 807); labeling (21 CFR Part 801); good manataotande for mobile, the electronic
forth in the quality systems (QS) regulation (21 CFR 1000-1050. forth in the quality systems (QS) regulation (21 CFR Part 25) -542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (sections 531-542 of the Act); 21 CFR 1
{1}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket This letter will allow you to begin marketing your device of your device to a legally marketed
notification. The FDA finding of substantial equivaler and thus nemits your d notification. The FDA finding of substantial equivalence or your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
r in the regulation If you desire specific advice for your device on our laboring agains.
contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on the promotion and and one one of essent of the mating.
Office of Compliance at (301) 594-4639. Also, please noneral information on Office of Compliance at (301) 394-4057. Also, production) Other general information of Small Manufacture by reference to premarket notification (21 Cr F r a F orision of Small Manufacturers,
your responsibilities under the Act may be obtained from the Division of Small Manufactu your responsibilities under the Act may or obtails nonemal (800) 638-2041 or (301) 443-6597
International and Consumer Assistance at 11 1 14 modemons btml International and Consumer Tissistance at the form of the main.html.
Sincerely yours,
Vernie C. Snodgin
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 2013098 510(k) Application
Medical Metrics, Inc. Medical Metrics, te Viewing Station
### Statement of Intended Use 1.0
The KIMAX ™ Remote Viewing Station software and hardware (RVS) are intended
ed to the Remote with and mamic digital images from radiographic and The KIMAX ™ Remote Viewstatic or dynamic digital images from radiographic and
to allow physicians to view static or dynamic digital images from radiographic and to allow physicians to view station of dynamic digh-performance computer,
other medical imaging equipment. The RVS includes a high-performance computer, other medical imaging equipment. The R V > includes a magn-pons intended to be used
data storage devices, a high-resolution monitor and software to: l) data storage devices, a high-resolution monitor and Socware to: 1)
in the clinical workspace of a practicing physician. The RVS includes software to: 1)
Pressure of a the tha in the clinical workspace of a practicing physician: 2) allow physician to quickly
accept AVI images or import other image formats; 2) allow physician to quickly accept AVI images or import other image formals, 27 and prightsess, apply filters to enhance
locate and display images; 3) adjust contrast and brightness, apply filters to en locate and display images; 3) adjust contrast and offectures of interest, and the images, zoom in on selected teatures, pan to select reading of more of the rest
annotate images; 4) print out selected images; and 5) create reports that include text and annotated images.
Tämit le. A. yper
Prescription Use K0/3098
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.