The AquariusNET Server acquires, stores, transmits, and enables compatible medical images from medical scanning devices computers on a network to display patient reports of various types. Teleradiology, such as EDT, CT of MRI and archiving, image manipulation, 3D and 4D visualization are supported. Calcium scoring from whole body computed tomography derived measurements, for non-invasive detection and quantification of atherosclerotic plaque. Tools for histogram analysis of the density distribution of certain regions of interest are provided. A database management and report generation tool is included.
Device Story
AquariusNET is a DICOM server and teleradiology system; receives/stores images from PACS or scanning modalities; archives data; serves images to remote thin-client PCs/notebooks. Features integrated 2D/3D streaming engine for interactive review, 3D reconstruction, and full-color volume rendering. Includes semi-automated calcium scoring tools for atherosclerotic plaque quantification; allows user override of identified regions; generates reports. Used in clinical settings by healthcare providers to review images and perform measurements. Output facilitates clinical decision-making by providing visualization and quantitative analysis of CT data, aiding in non-invasive assessment of cardiovascular disease.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Utilizes off-the-shelf PC hardware with Pentium III-class processor; Windows 2000 OS. DICOM-compliant server architecture. Supports 2D/3D streaming, volume rendering, and calcium scoring. Software developed per ISO/IEC 12207:1995. Networked connectivity via DICOM protocols.
Indications for Use
Indicated for medical imaging workstations to acquire, store, transmit, and display medical images for diagnostic review. Supports non-invasive detection and quantification of atherosclerotic plaque via calcium scoring from whole-body CT scans. Intended for prescription use.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Imatron Ultra Access Workstation with Cardiac Software Extensions (K972903)
TeraRecon IiVS™ Integrated image Viewing Station (K994329)
Submission Summary (Full Text)
{1}------------------------------------------------
SEP 1 3 2001
## 510(k) Premarket Notification Summary of Safety and Effectiveness Information
AquariusNET Server Trade Name:
Image communication and storage system Common Name:
## Classification
System, Digital Image Communication, Teleradiology System Name:
Establishment Name & Registration Number: TeraRecon, Inc. Name: Number: Pending
### Classification:
\$ 892.2020, System, Image Processing. Class I, proposed exempt, final rule pending.
8 892.2050, Picture Archiving and Communication System. Class I, proposed exempt, final rule pending.
90-LLZ & 90-LMD ProCode(s):
#### Equivalent Device(s):
- Imatron Ultra Access Workstation with Cardiac Software Extensions (K972903). 1.
IiVS™ Integrated image Viewing Station (K994329). 2.
These devices are substantially equivalent in terms of basic design, features and intended use.
### Description of the Device:
AquariusNET is a device consisting of a DICOM server that receives and stores images from a PACS or other image giving modalities. It archives images in a scalable storage medium and delivers them in response to DICOM Query/Retrieve requests from other DICOM devices on the network (not part of AquariusNET). It also serves image requests to its remote "thin clients", which act as the graphical user interface to the AquariusNET server. The server can host multiple concurrent sessions from remote "thin clients". AquariusNET features an integrated 2D/3D streaming engine which allows regular PCs or notebooks to control the server, and to review 2D images and 3D reconstructions interactively over a network. AquariusNET is capable of image review, communications, archiving, database maintenance, reporting and basic 3D capabilities described elsewhere in this document. It is also capable of full-color Volume Rendering and Calcium Scoring.
### Applicant/Sponsor Name / Address:
TeraRecon, Inc. 2955 Campus Drive, Suite 325, San Mateo, CA 94403 650.372.2668
### Contact Person:
Mr. Robert Taylor Executive Vice President TeraRecon Inc., 2955 Campus Drive, Suite 325, San Mateo, CA 94403 650.372.2668
{2}------------------------------------------------
#### Submission Correspondent:
Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 925.356.2640 / 925.356.2854 FAX
### Hardware & Software Information:
The AquariusNET Server utilizes standard "off the shelf" personal computer systems and s The AquariasNET Server utinzes standard "On the Sheat" portuges and windows 2000 operating system, and a Pentium III - class processor or equivalent.
The software designed to control and manipulate the diagnostic images follows the international standard ISO/TEC 12207: 1995 Information Technology - Software Life Cycle Processes Incoment wind standard "ISO/ILC" 12207. 1775 "mormation" Toothers relative to this software has been determined using the decision tree provided in Version 1 of the FDA Software Guidance.
| Feature Comparison Table: | | | | |
|---------------------------|--|--|--|--|
|---------------------------|--|--|--|--|
| Feature | AquariusNET Server | Imatron Ultra<br>Access 972903 | TeraRecon IiVS<br>K994329 |
|----------------------------------------------------------------------------|--------------------|--------------------------------|---------------------------|
| 2D Image Review | Yes | Yes | Yes |
| Multiplanar reformatting | Yes | Yes | Yes |
| 3D Volume Rendering | Yes | Yes | Yes |
| Maximum Intensity<br>Projection | Yes | Yes | Yes |
| Image Archiving | Yes | Yes | Yes |
| Image Filming | Yes | Yes | Yes |
| Image Transfer or<br>Network Connectivity | Yes | Yes | Yes |
| Examination of 2D image<br>data from a calcium scan | Yes | Yes | Yes |
| Examination of calcium scan<br>as a 3D volume | Yes | Yes | Yes |
| Semi-automated identification<br>of regions that are considered<br>calcium | Yes | Yes | No |
| User override of<br>automatically identified<br>regions | Yes | Yes | No |
| Automatic calculation of<br>calcium score | Yes | Yes | No |
| Ability to measure CT<br>numbers on a 2D image | Yes | Yes | Yes |
| Saving of calcium data with<br>patient exam data | Yes | Yes | No |
| Creation of a paper calcium<br>report | Yes | Yes | No |
| Comparison of multiple scans | Yes | Yes | Yes |
| Indications for use - general<br>medical imaging workstations | Yes | Yes | Yes |
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
TeraRecon, Inc. % Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 PLEASANT HILL CA 94523-3389 Re: K012086
Trade/Device Name: AguariusNET Server Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System
Regulatory Class: II Product Code: 90 LLZ Dated: June 26, 2001 Received: July 3, 2001
Dear Mr. Schlerf:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
SEP 1 3 2001
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket
notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
# 510(k) NUMBER: _______________________________________________________________________________________________________________________________________________________________
DEVICE NAME : AquariusNET Server
### INDICATIONS FOR USE:
The AquariusNET Server acquires, stores, transmits, and enables compatible The AquariasNET - Bervor "acquires, dical images from medical scanning devices computers on a fictwork to display tient reports of various types. Teleradiology, such as EDT, CT of NIN and parchiving, image manipulation, 3D and 4D Calcium scoring from whole body computed visualization are supported. visualization are Supported.
tomography derived measurements, for non-invasive detection and quantification of atherosclerotic plaque. Tools for histogram analysis of the density distribution of certain regions of interest are provided. A database management and report generation tool is included.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Snowdon
(Division Sign-Off)/
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K012086
**Prescription Use**
(Per 21 CFR 801.109) √
OR
Over-The-Counter Use (Optional format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.