K020150 · Capintec, Inc. · KPT · Aug 2, 2002 · Radiology
Device Facts
Record ID
K020150
Device Name
CRC-15BT
Applicant
Capintec, Inc.
Product Code
KPT · Radiology
Decision Date
Aug 2, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1360
Device Class
Class 2
Intended Use
The CRC 15BT dose calibrator is intended to be used by trained medical physicists, or qualified technologists to measure the activity or output of radioactive brachytherapy sources, including HDR, LDR and IVBT sources.
Device Story
The CRC 15BT is a well-type ionization chamber dose calibrator designed to measure radioactive brachytherapy sources (HDR, LDR, IVBT). It accepts various configurations, including seeds, ribbons, and source trains, using appropriate source holders. The device functions by measuring current within a sealed, pressurized UPC Argon ionization chamber, which is then converted by a readout unit into activity (Ci) or air kerma strength (U). It is intended for use by medical physicists or technologists in clinical settings. The device provides quantitative measurements of source output, which healthcare providers use to verify brachytherapy source activity before clinical application, ensuring accurate dosing for patient treatment.
Clinical Evidence
Bench testing only. The device was evaluated by the University of Wisconsin ADCL and found suitable for all clinically available gamma and high-energy beta brachytherapy sources. Compliance with safety standards EN60601-1, IEC60601-1, EN60601-1-2, IEC60601-1-4, and IEC 60601-1-1 was demonstrated.
Technological Characteristics
Well-type ionization chamber; pressurized with UPC Argon gas; readout unit converts current to Ci or U; minor modifications to fill gas pressure and bias voltage compared to predicate; complies with EN60601-1, IEC60601-1, EN60601-1-2, IEC60601-1-4, and IEC 60601-1-1 standards.
Indications for Use
Indicated for use by trained medical physicists or qualified technologists to measure activity or output of radioactive brachytherapy sources, including HDR, LDR, and IVBT sources.
Regulatory Classification
Identification
A radionuclide dose calibrator is a radiation detection device intended to assay radionuclides before their administration to patients.
Predicate Devices
Capintec CRC 15R dose calibrator
Related Devices
K013548 — STANDARD IMAGING IVB 1000 WELL CHAMBER · Standard Imaging, Inc. · Nov 9, 2001
K060532 — HDR 1000 PLUS WELL CHAMBER, MODEL 90008 · Standard Imaging, Inc. · Apr 11, 2006
K071396 — CRC 25 SERIES (CRC 25R, CRC 25PET, CRC 25W) DOSE CALIBRATORS · Capintec, Inc. · Jun 28, 2007
K052595 — CRC 15ULTRA · Capintec, Inc. · Nov 8, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
AUG 0 2 2002
Image /page/0/Picture/1 description: The image shows the logo for CAPINTEC, INC. The logo consists of the letters "CII" stacked on top of each other inside of a square. Below the square is the company name, CAPINTEC, INC. To the left of the logo is the number 2002.
January 10, 2001
RE: Summary of Safety and Effectiveness Information for the Capintec CRC 15BT calibrator.
The CRC 15BT dose calibrator is a well type ionization chamber specifically designed to accurately and rapidly measure all types of brachytherapy sources with appropriate calibration and source positioning device. The ionization chamber is connected to a readout, which converts measured current into a displayed value in units of either Ci (activity) or U (air kerma strength). The unit is sealed and pressurized with UPC Argon. which improves sensitivity and eliminates the need for temperature and pressure corrections.
The CRC 15BT has been evaluated by the University of Wisconsin ADCL, and found to be a suitable instrument for all clinically available gamma and high-energy beta brachytherapy sources, including HDR, LDR, and IVBT in any number of configurations including, seeds, ribbons, and source trains with appropriate source holders.
The CRC 15BT design is a minor modification to the Capintec CRC 15R dose calibrator to increase the activity range to accommodate HDR sources. The CRC 15R has been commercially available for over 10 years and has an excellent reputation for long term stability, reliability, safety and effectiveness. The modifications to the 15BT are limited to reduced fill gas pressure and in increase bias voltage to accommodate a higher activity range required for HDR sources. The basic chamber design and readout are the same as the CRC 15R. Consequently, the CRC 15BT is expected to maintain the same excellent quality as the CRC 15R. with the same level of safety and effectiveness and long term reliability.
The CRC 15BT has also been tested and approved to the following safety standards for medical equipment:
Medical Electrical Equipment, Part 1 General Requirements - EN60601-1: 1990+A1+A2 Medical Electrical Equipment, Part 1 General Reguirements - IEC60601-1: 1998+A1+A2 Medical Electrical Equipment, Part 1 General Requirements Section 1.2 Collateral standard: Electromagnetic Compatibility-EN60601-1-2 Medical Electrical Equipment, Part 1 General Requirements for Safety Section 1.4 Collateral Standard: Programmable Electrical Medical Systems-IEC60601-1-4:1996 Medical Electrical Equipment, Part 1 General Requirements-Can/CSA C22.2 No.606101-M90 Medical Electrical Systems-IEC 606601-1-1:1993+A1
{1}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## AUG 0 2 2002
Ms. Mary Anne Dell CAPINTEC, Inc. 540 Alpha Drive PITTSBURGH PA 15238 USA Re: K020150
Trade/Device Name: CRC 15BT Calibrator Regulation Number: 21 CFR 892.1360 Regulation Name: Radionuclide dose calibrator Regulatory Class: II Product Code: 90 KPT Dated: May 13, 2002 Received: May 14, 2002
Dear Ms. Dell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-459 |
|----------------------------------|---------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-461 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-461 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-465 |
| Other | (301) 594-469 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Henry C. Snogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Appendix 2
Indications for Use Form
Page
510(k) Number (if known):
Device Name: CRC 15BT
Indications For Use: The CRC 15BT dose calibrator is intended to be used by trained medical physicists, or qualified technologists to measure the activity or output of radioactive brachytherapy sources, including HDR, LDR and IVBT sources.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Nancy Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K020150
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.