← Product Code [KPS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS) · K241364

# Hybrid Viewer (00859873006189) (K241364)

_Hermes Medical Solutions AB · KPS · Nov 8, 2024 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS/K241364

## Device Facts

- **Applicant:** Hermes Medical Solutions AB
- **Product Code:** [KPS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS.md)
- **Decision Date:** Nov 8, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1200
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

Hybrid Viewer is a software application for nuclear medicine and radiology. Based on user input, Hybrid Viewer processes, displays and analyzes nuclear medicine and radiology imaging data and presents the result to the user. The result can be stored for future analysis. Hybrid Viewer is equipped with dedicated workflows which have predefined settings and layouts optimized for specific nuclear medicine investigations, The software application can be configured based on user needs. The investigation of physiological or pathological states using measurement and analysis functionality provided by Hybrid Viewer is not intended to replace visual assessment. The information obtained from viewing and/or performing quantitative analysis on the images is used, in conjunction with other patient related data, to inform clinical management,

## Device Story

Hybrid Viewer is a software application for nuclear medicine and radiology imaging. It processes, displays, and analyzes imaging data based on user input. The device features dedicated workflows with predefined settings and layouts for specific nuclear medicine investigations. It supports various DICOM formats, including Segmentation (SEG), RT Dose, and CT Fluoroscopy. Key functionalities include Three Energy Window (TEW) correction, motion correction, SUV unit display for SPECT, and specialized processing workflows for lung-liver shunting, heart-to-mediastinum ratios (e.g., for AdreView imaging), and DMSA SPECT kidney uptake. The software is used by clinicians to assist in clinical management; it does not replace visual assessment. Results are stored for future analysis. The device is configured based on user needs and does not require physical hardware keys, utilizing a ZIP-based deployment model.

## Clinical Evidence

Bench testing only. No clinical data was required. Substantial equivalence was established through analytical verification of new processing workflows against relevant scientific literature and scripted verification testing of new software features against predefined acceptance criteria. Usability testing confirmed the device is equivalent to the predicate.

## Technological Characteristics

Software application for nuclear medicine and radiology. Features include DICOM support (SEG, RT Dose, CT Fluoroscopy), TEW correction, motion correction, SUV quantification, and specialized NM processing workflows. Deployed as a ZIP file; licensing is software-based (no HASP key). Cybersecurity updates applied to SOUP components. Operates on standard computing hardware.

## Regulatory Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- Hermes Medical Imaging Suite v5.7 ([K171681](/device/K171681.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 8, 2024

Hermes Medical Solutions AB Hanne Grinaker Chief Quality and Regulatory Affairs Officer Strandbergsgatan 16 Stockholm, 11251 Sweden

Re: K241364

Trade/Device Name: Hybrid Viewer (00859873006189) Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: Class II Product Code: KPS Dated: October 9, 2024 Received: October 9, 2024

Dear Hanne Grinaker:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

#### Indications for Use

Submission Number (if known)

K241364

Device Name

Hybrid Viewer (00859873006189)

#### Indications for Use (Describe)

Hybrid Viewer is a software application for nuclear medicine and radiology. Based on user input, Hybrid Viewer processes, displays and analyzes nuclear medicine and radiology imaging data and presents the result to the user. The result can be stored for future analysis.

Hybrid Viewer is equipped with dedicated workflows which have predefined settings and layouts optimized for specific nuclear medicine investigations,

The software application can be configured based on user needs.

The investigation of physiological or pathological states using measurement and analysis functionality provided by Hybrid Viewer is not intended to replace visual assessment. The information obtained from viewing and/or performing quantitative analysis on the images is used, in conjunction with other patient related data, to inform clinical management,

Type of Use (Select one or both, as applicable)

|X | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| 510(k) #: | K241364        |
|-----------|----------------|
|           | 510(k) Summary |

|                 | Prepared on: 2024-11-06 |
|-----------------|-------------------------|
| Contact Details | 21 CFR 807.92(a)(1)     |

| Applicant Name              | Hermes Medical Solutions AB                |
|-----------------------------|--------------------------------------------|
| Applicant Address           | Strandbergsgatan 16 Stockholm 11251 Sweden |
| Applicant Contact Telephone | +468190325                                 |
| Applicant Contact           | Ms. Grinaker Hanne                         |
| Applicant Contact Email     | Hanne.grinaker@hermesmedical.com           |

| Device Name | 21 CFR 807.92(a)(2) |
|-------------|---------------------|
|-------------|---------------------|

| Device Trade Name   | Hybrid Viewer (00859873006189)         |
|---------------------|----------------------------------------|
| Common Name         | Emission computed tomography system    |
| Classification Name | System, Tomography, Computed, Emission |
| Regulation Number   | 892.1200                               |
| Product Code(s)     | KPS                                    |

| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
|------------------------------------|---------------------|
|------------------------------------|---------------------|

| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K171681     | Hermes Medical Imaging Suite v5.7                        | KPS          |

| Device Description Summary | 21 CFR 807.92(a)(4) |
|----------------------------|---------------------|
|----------------------------|---------------------|

Hybrid Viewer provides general tools for viewing and processing nuclear medicine and radiology images. It includes software fornuclear medicine (NM) processing studies for specific darts of the body and specific diseases using predefined workflows. Hybrid Viewer 7.0 includes the following additional clinical features compared to Hybrid Viewer 2.8:

• Additional DICOM file support for Segmentation (SEG), RT Dose and CT Fluoroscopy

· Three energy window (TEW) correction for whole body studies

• Automatic motion correction and additional motion correction option for dual isotope

• Display quantitative SPECT studies in SUV units

• Additional NM processing workflows for:

- Assessment of the percentage of activity which is shunted to the lung prior to Y90 microsphere treatment planning.

- Assessment of the ratio of activity in the heart compared to the mediastinum. The workflow contains specific options for the GE Healthcare product AdreView™, a radiopharmaceutical agent used in the detection of primary or secondary pheochromocytoma or neuroblastoma as an adjunct to other diagnostic tests.

- Assessment of the relative uptake in right, left and duplex kidneys in DMSA™ (dimercaptosuccinic acid) SPECT studies. This is an extension of an existing workflow for planar DMSA studies.

{5}------------------------------------------------

### 21 CFR 807.92(a)(5)

Hybrid Viewer is a software application for nuclear medicine and radiology. Based on user input. Hybrid Viewer processes, displays and analyzes nuclear medicine and radiology imaging data and presents the result can be stored for future analysis. Hybrid Viewer is equipped with dedicated workflows which have predefined settings and layouts optimized for specific nuclear medicine investigations.

The software application can be configured based on user needs.

The investigation of physiological or pathological states using measurement and analysis functionality provided by Hybrid Viewer is not intended to replace visual assessment. The information obtained from viewing quantitative analysis on the images is used, in conjunction with other patient related data, to inform clinical management.

## Indications for Use Comparison

The statement of intended use of the Hybrid Viewer 7.0 device has been rewritten for clarification but there are no underlying differences in intended use between 7.0 and that of the Hermes Medical Imaging Suite 5.7, which includes Hybrid Viewer 2.8.

### Technological Comparison

There are no differences between 7.0 and the predicate device 2.8 in design, principles of operation and critical performance requirements.

For the new NM Processing workflows that include new clinical features (DMSA SPECT, Lung Liver Shunt and DICOM SEG file support) calculations have been validated using analytical verification, where the results from using the new workflow have been based on relevant publications, referenced in the verification documents. All analytical verifications concluded that the new workflows introduced since the predicate device fulfill the acceptance criteria and are therefore safe to use.

The new features that did not include advanced calculations (support for RT Dose and CT Fluoroscopy, new motion correction options, SUV display, TEW correction and Heart Mediatinum) were verified by comparing their acceptance criteria against the test results from the scripted verification testing during the development process.

Hybrid Viewer 7.0 and predicate device Hybrid Viewer 2.8 include the same Software of Unknown Provenance items (SOUPs). In 7.0 they have been updated to the latest versions for cybersecurity.

There are differences in deployment between 7.0 and the predicate device which are:

-The Instructions for Use/User manual can be downloaded on-line in 7.0

- Licensing in 7.0 is handled as pure software so a HASP key is no longer required

- The software is deployed as a ZIP file in 7.0 instead of an installer

In conclusion, there is no change in the overall safety and effectiveness of Hybrid Viewer version 7.0 compared to predicate Hybrid Viewer version 2.8.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The conclusion of usability testing and validation is that Hybrid Viewer 7.0 is substantially equivalent to the predicate device Hybrid Viewer 2.8 (which is included in Hermes Medical Imaging Suite 5.7). There are no differences in the intended use of the 7.0 device and the technological characteristics of the device related to safety and effectiveness are unchanged. No new questions of safety and effectiveness have been raised.

### 21 CFR 807.92(a)(5)

### 21 CFR 807.92(a)(6)

# Intended Use/Indications for Use

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS/K241364](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS/K241364)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
