← Product Code [KPS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS) · K023768

# ECAT PET/CT; ECAT LSO PET/CT; ECAT LSO PET/CT 16 (K023768)

_Cti Pet Systems, Inc. · KPS · Nov 25, 2002 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS/K023768

## Device Facts

- **Applicant:** Cti Pet Systems, Inc.
- **Product Code:** [KPS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS.md)
- **Decision Date:** Nov 25, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1200
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

The PET/CT Patient Handling System (PHS) is a system component of CPS ECAT PET/CT tomographic scanner systems, which are combined positron emission tomography (PET) and X-ray computed tomography (CT) scanners. The PET/CT Patient Handling System is used to support and move a patient through the PET/CT scanner while CT and PET diagnostic image data are being acquired from the patient. The ECAT PET/CT scanners are intended to be utilized by appropriately trained health care professionals to 1) image and measure the distribution of injected positron emitting radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body; and 2) to produce crosssectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. The PET and the CT functions of these systems can also be used in combination to provide high resolution, noise free CT attenuation correction maps for PET images and, additionally, utilized to produce fused CT and PET images, providing detailed anatomic and metabolic function information in a single image.

## Device Story

Patient handling system (PHS) for ECAT PET/CT scanners; supports and moves patients through scanner gantry. Modification adds two footswitches and updates controller firmware; enables vertical bed adjustment while patient is inside scanner field of view (FOV). Previously, vertical adjustment required retracting bed from port. Operated by healthcare professionals in clinical imaging settings. Allows more accurate patient positioning without full retraction; improves workflow efficiency and patient comfort.

## Clinical Evidence

Bench testing only. System tested by manufacturer to meet predetermined safety and performance requirements, including compliance with IEC 60601-1 and UL 187 standards.

## Technological Characteristics

Patient bed system for PET/CT scanners. Modification includes two footswitches and revised controller firmware. Complies with IEC 60601-1 (electrical safety) and UL 187 (X-ray equipment safety).

## Regulatory Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- PHS for ECAT PET/CT ([K002715](/device/K002715.md))
- PHS for ECAT LSO PET/CT ([K013504](/device/K013504.md))
- PHS for ECAT LSO PET/CT 16 ([K023518](/device/K023518.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

NOV 2 5 2002

K023768

## 510(k) SUMMARY

# [As required by 21 CFR 807.87(h)]

#### Identification of Submitter Submitter:

Contact Person:

Telephone No: Fax No: Date of preparation:

CTI PET Systems, Inc. 810 Innovation Drive Knoxville, TN 37932. William Skremsky Senior Regulatory Affairs Specialist (865) 218-2522 (865) 218-3000 November 8, 2002

### Identification of the Product

Device Proprietary Name: Common Name: Classification Name:

PET/CT Patient Handling System (PHS) Patient Bed Emission Computed Tomography System per 21 CFR 892.1200

#### Marketed Devices to Which Equivalence is Claimed

| Device                     | Manufacturer          | 510(k) Number |
|----------------------------|-----------------------|---------------|
| PHS for ECAT PET/CT        | CTI PET Systems (CPS) | K002715       |
| PHS for ECAT LSO PET/CT    | CTI PET Systems (CPS) | K013504       |
| PHS for ECAT LSO PET/CT 16 | CTI PET Systems (CPS) | K023518       |

#### Device Description and Comparison to the Unmodified Device

The subject device is the patient handling system (PHS) or patient bed that is used with the currently marketed CPS ECAT PET/CT scanners. The ECAT PET/CT, ECAT LSO PET/CT and ECAT LSO PET/CT 16 systems are combined position tomography (PET) and X-ray computed tomography (CT) scanners. These three dual modality PET/CT tomographs all utilize the same basic patient handling system (PHS) to support and move the patient through the patient port in the scanner gantry. Currently, vertical positioning of the patient on the bed of these PET/CT scanners is only possible when the PHS bed is fully retracted out of the patient port.

The proposed modification for the PHS used on these three PET/CT scanners, which consists of two added footswitches and revised PHS controller firmware, will enable the system operator to raise or lower the PHS bed a limited distance when the bed and patient are in the scanner field of view (FOV), inside the patient port. This modification will allow the operator to more accurately position the patient in the FOV without having to first retract the bed and patient completely out of the patient port.

#### Intended Use

The PET/CT Patient Handling System is used to support and move a patient through the PET/CT scanner while CT and PET diagnostic studies are performed on the patient.

{1}------------------------------------------------

## 510(k) Summary ECAT LSO PET/CT Scanner p. 2

#### Safety and Effectiveness

The PET/CT Patient Positioning System, with the proposed modification for the patient positioning enhancement described in this 510(k), has been designed to comply with applicable industry safety standards for this type of medical equipment including the international standard IEC 60601-1, General Requirements for the Safety Electrical Medical Equipment and UL 187, the X-Ray Equipment Standard for Safety. The modified PET/CT Patient Positioning System has been tested by CPS and found to meet its predetermined safety and performance requirements.

### Substantial Equivalence Determination

In the opinion of CPS, the PET/CT Patient Positioning System, incorporating the described patient positioning enhancement modification, utilizes the same scientific technology as the unmodified version and raises no new questions with regard to its safety and effectiveness. Therefore, we believe the modified Patient Positioning System is substantially equivalent to the unmodified version with respect to design, material and composition, energy source, and safety characteristics.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 2 5 2002

Mr. William Skremsky Senior Regulatory Affairs Specialist CTI, Inc. 810 Innovation Drive KNOXVILLE TN 37932

Re: K023768

Trade/Device Name: PET/CT Patient Handling System (PHS) Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II

Product Code: 90 KPS Dated: November 8, 2002 Received: November 12, 2002

Dear Mr. Skremsky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| . 8xx.1xxx                       | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page / of /

510(k) Number (if known):

Device Name: PET/CT Patient Handling System (PHS)

Indications for Use:

The PET/CT Patient Handling System (PHS) is a system component of CPS ECAT PET/CT tomographic scanner systems, which are combined positron emission tomography (PET) and X-ray computed tomography (CT) scanners. The PET/CT Patient Handling System is used to support and move a patient through the PET/CT scanner while CT and PET diagnostic image data are being acquired from the patient.

K023765

The ECAT PET/CT scanners are intended to be utilized by appropriately trained health care professionals to 1) image and measure the distribution of injected positron emitting radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body; and 2) to produce crosssectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.

The PET and the CT functions of these systems can also be used in combination to provide high resolution, noise free CT attenuation correction maps for PET images and, additionally, utilized to produce fused CT and PET images, providing detailed anatomic and metabolic function information in a single image.

(*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube and the simultaneous translation of the patient.)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David G. Seymour
(Division Sign-Off)

Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

Prescription Use √
(Per 21 CFR 801.109)

OF

Over-The-Counter Use

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS/K023768](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPS/K023768)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
