uDR Aurora CX is intended to acquire X-ray images of the human body by a qualified technician, examples include acquiring two-dimensional X-ray images of the skull, spinal column, chest, abdomen, extremities, limbs and trunk. The visualization of such anatomical structures provide visual evidence to radiologists and clinicians in making diagnostic decisions. This device is not intended for mammography.
Device Story
Stationary X-ray system; includes high-voltage generator, X-ray tube, and flat-panel detectors. Captures X-ray photons; converts to electrical signals; generates DICOM images. Operated by qualified technicians in clinical settings. Features include uVision (camera-based patient positioning/collimation planning) and uAid (AI-based image quality assessment for chest X-rays). Output viewed by radiologists/clinicians for diagnostic decision-making. Benefits include standardized image quality, reduced retake rates, and optimized workflow.
Clinical Evidence
Bench testing only. Performance of uVision (positioning) validated over 5 months with 95% compliance. uAid (image quality) validated using independent dataset (n=thousands); sensitivity/specificity >0.9 for foreign object detection, lung field integrity, scapula positioning, and centerline deviation. Clinical image evaluation by board-certified radiologist confirmed diagnostic quality.
Technological Characteristics
Stationary X-ray system; 65kW/80kW generator; 150kV max voltage; CsI flat-panel detectors. Includes 2D camera for positioning. Connectivity: DICOM 3.0. Safety standards: ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-54, IEC 60601-2-28. Biocompatibility per ISO 10993-5/10.
Indications for Use
Indicated for acquiring 2D X-ray images of skull, spinal column, chest, abdomen, extremities, limbs, and trunk in adult and pediatric patients. Not for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 19, 2025
Shanghai United Imaging Healthcare Co., Ltd.
% Xin Gao
Regulatory Affairs Manager
No.2258 Chengbei Rd. Jiading District
SHANGHAI, 201807
CHINA
Re: K251167
Trade/Device Name: uDR Aurora CX
Regulation Number: 21 CFR 892.1680
Regulation Name: Stationary X-Ray System
Regulatory Class: Class II
Product Code: KPR
Dated: April 15, 2025
Received: August 19, 2025
Dear Xin Gao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251167 - Xin Gao
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251167 - Xin Gao
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251167 | |
| Device Name uDR Aurora CX | |
| Indications for Use (Describe) uDR Aurora CX is intended to acquire X-ray images of the human body by a qualified technician, examples include acquiring two-dimensional X-ray images of the skull, spinal column, chest, abdomen, extremities, limbs and trunk. The visualization of such anatomical structures provide visual evidence to radiologists and clinicians in making diagnostic decisions. This device is not intended for mammography. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED IMAGING
1
K251167
510 (k) SUMMARY
1. Date of Preparation:
April 15, 2025
2. Sponsor Identification
Shanghai United Imaging Healthcare Co., Ltd.
No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China
Establishment Registration Number: 3011015597
3. Contact Person
Name: Xin Gao
Tel: +86-021-67076888-5386
Fax: +86-021-67076889
Email: xin.gao@united-imaging.com
4. Subject Device Name and Classification
Trade Name: uDR Aurora CX
Common Name: Digital Medical X-ray Imaging System
Model(s): uDR Aurora CX
Regulatory Information
Classification Name: Stationary X-Ray System
Device Classification: II
Product Code: KPR
Regulation Number: 21 CFR 892.1680
Review Panel: Radiology
5. Identification of Predicate Device(s)
Predicate Device:
Trade Name: uDR 780i
510(k) Number: K241068
Classification Name: Stationary X-Ray System
Classification Panel: Radiology
Regulation Number: 21 CFR 892.1680
Classification: II
Product Code: KPR
{5}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888 Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED IMAGING
6
Reference Device:
Trade Name: MULTIX Impact C
510(k) Number: K213700
Classification Name: Stationary X-Ray System
Classification Panel: Radiology
Classification Regulation: 21 CFR §892.1680
Device Class: Class II
Product Code: KPR
6. Device Description:
uDR Aurora CX is a model of Digital Medical X-ray Imaging System developed and manufactured by Shanghai United Imaging Healthcare Co., Ltd(UIH). It includes X-ray Generator, X-ray Imaging System. The X-ray Generator produces controlled X-rays by high-voltage generator and X-ray tube assembly, ensuring stable energy output for human body penetration. The X-ray Imaging System converts X-ray photons into electrical signals by detectors, and generates DICOM-standard images by workstation to reflecting density variations of human body.
7. Intended Use Statement:
uDR Aurora CX is intended to acquire X-ray images of the human body by a qualified technician, examples include acquiring two-dimensional X-ray images of the skull, spinal column, chest, abdomen, extremities, limbs and trunk. The visualization of such anatomical structures provide visual evidence to radiologists and clinicians in making diagnostic decisions. This device is not intended for mammography.
2
{6}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED IMAGING
3
# 8. Substantially Equivalent (SE) Comparison
A comparison between the technological characteristics of proposed and predicate devices is provided as below.
Table 1 Comparison of Technology Characteristics to predicate device
| Item | Proposed Device uDR Aurora CX | Predicate Device uDR 780i (K241068) | Remark |
| --- | --- | --- | --- |
| General | | | |
| Product Code | KPR | KPR | Same |
| Regulation No. | 892.1680 | 896.1680 | Same |
| Class | II | II | Same |
| Intended Use | uDR Aurora CX is intended to acquire X-ray images of the human body by a qualified technician, examples include acquiring two-dimensional X-ray images of the skull, spinal column, chest, abdomen, extremities, limbs and trunk. The visualization of such anatomical structures provide visual evidence to radiologists and clinicians in making diagnostic decisions. This device is not intended for mammography. | The uDR 780i Digital Medical X-ray system is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other anatomic sites. Applications can be performed with the subject sitting, standing, or lying in the prone or supine position. Not for mammography. | Note1 |
| Specifications | | | |
{7}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED
IMAGING
4
| Item | Proposed Device
uDR Aurora CX | Predicate Device
uDR 780i (K241068) | Remark |
| --- | --- | --- | --- |
| High Voltage Generator | | | |
| Max. Power/kW | 65kW/80kW | 65kW/80kW | Same |
| Max. tube Voltage(kV) | 150kV | 150kV | Same |
| Shortest exposure time | 1ms | 1ms | Same |
| X-Ray Tube Assemble | | | |
| Focus Nominal Value | 0.6/1.2 | 0.6/1.2 | Same |
| Maximum peak voltage | 150kV | 150kV | Same |
| Anode Heat Content | 65kw: ≥300kHU
80kw: ≥400kHU | 65kw: ≥300kHU
80kw: ≥400kHU | Same |
| Anode Target Angle | 12° | 12° | Same |
| X-ray tube assembly Heat content | 65kw: ≥1250KHU | 65kw: ≥1250KHU | Same |
{8}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED
IMAGING
E
| Item | Proposed Device
uDR Aurora CX | Predicate Device
uDR 780i (K241068) | Remark |
| --- | --- | --- | --- |
| | 80kw: ≥1500KHU | 80kw: ≥1500KHU | |
| Flat Panel Detector-Config.1 | | | |
| Model | uFPD1717-100 | AR-C4343W | |
| Power | Battery or DC operated | Battery or DC operated | Same |
| Scintillator | Cesium iodide (CsI) | Cesium iodide (CsI) | Same |
| Image Matrix Size | 4267x4267
100 μm | 3320x3408
125μm | Note2 |
| Effective radiographic size | 42.67cm x 42.67cm | 41.5cm x 42.6cm | Note 3 |
| Flat Panel Detector-Config.2 | | | |
| Model | uFPD1417-100 | AR-C4343W | |
| Power | Battery or DC operated | Battery or DC operated | Same |
| Scintillator | Cesium iodide (CsI) | Cesium iodide (CsI) | Same |
| Image Matrix Size | 3500x4300 | 3320x3408 | Note 4 |
{9}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED
IMAGING
6
| Item | Proposed Device
uDR Aurora CX | Predicate Device
uDR 780i (K241068) | Remark |
| --- | --- | --- | --- |
| | 100 μm | 125μm | |
| Effective Radiographic Size | 35cm x 43cm | 41.5cm x 42.6cm | Note 5 |
| Collimator | | | |
| Inherent filtration | 1mm Al | 1mm Al | Same |
| Copper prefilter | without filter,
0.1 mm,
0.2 mm,
0.3 mm; | without filter,
0.1 mm,
0.2 mm,
0.3 mm; | Same |
| Bulit-in camera | Live 2D Camera for patient positioning and collimation | N/A | Note 6 |
| Elevating table | | | |
| Motorized vertical travel | ≥38.2cm | ≥38.2cm | Same |
| Max. patient weight | 320kg | 320kg | Same |
{10}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED
IMAGING
7
| Item | Proposed Device
uDR Aurora CX | Predicate Device
uDR 780i (K241068) | Remark |
| --- | --- | --- | --- |
| Detector travel range | ≥90cm | ≥67cm | Note 7 |
| Auto tracking for adjusting the table height is maintained | Yes, X-ray tube follows table height adjustment; source-image distance is maintained. | Yes, X-ray tube follows table height adjustment; source-image distance is maintained. | Same |
| Auto tracking for longitudinal tube travel | Yes, detector follows tube movement; centering maintained. | Yes, detector follows tube movement; centering maintained. | Same |
| Auto tracking for tube rotation | Yes, detector follows tube movement; centering maintained. | Yes, detector follows tube movement; centering maintained. | Same |
| Software function | | | |
| uAid | Yes | No | Note8 |
| Stitching | Yes | Yes | Same |
| AEC | Yes | Yes | Same |
| Safety | | | |
{11}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED
IMAGING
8
| Item | Proposed Device
uDR Aurora CX | Predicate Device
uDR 780i (K241068) | Remark |
| --- | --- | --- | --- |
| Electrical Safety | ANSI/AAMI ES 60601-1:2005 & A1:2012 & A2:2021 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance | ANSI/AAMI ES 60601-1:2005 & A1:2012 & A2:2021 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance | Same |
| EMC | Comply with IEC 60601-1-2:2014+A1:2020 | Comply with IEC60601-1-2 | Same |
| Biocompatibility | Patient Contact Materials were tested and demonstrated no cytotoxicity (ISO 10993-5), no evidence for irritation and sensitization (ISO 10993-10). | Patient Contact Materials were tested and demonstrated no cytotoxicity (ISO 10993-5), no evidence for irritation and sensitization (ISO 10993-10). | Same |
| Clinical Image Evaluation | Clinical Image Evaluation for the proposed device are provided in Section 11.5 Clinical Image Evaluation. | | |
| Standards | | | |
| DICOM | DICOM3 | DICOM3 | Same |
| Power Source | AC Line, Various voltages available | AC Line, Various voltages available | Same |
{12}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED IMAGING
Table2: Comparison of new features to predicate device
| Item | Proposed Device uDR Aurora CX | Reference Device MULTIX Impact C(K213700) | Remark |
| --- | --- | --- | --- |
| uVision Function (optional) | Users can manually adjust FOV and stitching range on the workstation. To assist the users with setting the FOV and stitching range, exam range is automatically planned for chest and stitching range is automatically planned for WholeSpine & WholeLowerExtremity. | - Virtual Collimation Manually adjust collimation size on imaging system by 3D camera - Smart Virtual Ortho: Ortho range set by 2D camera in the image system manually - Auto Thorax Collimation Exam range automatically planned for Thorax by 3D camera with manual adjustment - Auto Full-Spine& Long-Leg Collimation: Ortho range automatically planned for Full-Spine &Long-Leg by 2D camera with manual adjustment | Note 9 |
{13}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED IMAGING
10
| Justification | |
| --- | --- |
| Note 1 | Minor grammatical changes only. |
| Note 2 | The larger the image matrix size, the higher the resolution, and the greater the capability to capture fine structures, which means the ability to identify smaller lesions and tissues. The difference does not introduce safety and effectiveness issues. |
| Note 3 | Effective Radiographic Size refers to the actual area size of the detector panel that can be utilized in practical imaging. A larger Effective Radiographic Size indicates that the detector can cover a larger range of anatomical areas in practical use. The difference does not introduce safety and effectiveness issues. |
| Note 4 | The larger the image matrix size, the higher the resolution, and the greater the capability to capture fine structures, which means the ability to identify smaller lesions and tissues. The difference does not introduce safety and effectiveness issues. |
| Note 5 | Only the different of size specification, does not affect safety and effectiveness. The 35cm x 43cm normally used in free exam mode which is different with 41.5cm x 42.6cm which used in the tray in table. |
| Note 6 | 2D camera only introduced to capture optical information and support more clinical operational possibilities., does not affect safety and effectiveness. |
| Note 7 | Detector travel range refers to the movement range of the tray in Table. A larger travel range indicates a larger coverage on table and more operational possibilities. The difference does not introduce safety and effectiveness issues. |
{14}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED IMAGING
11
| Note 8 | uAid evaluates the positioning quality of chest images with deep learning methods. It efficiently and objectively categorizes images into three levels according to four criteria, to assist standardizing the management of image quality/technologist skills. The difference does not introduce safety and effectiveness issues. |
| --- | --- |
| Note 9 | For manually adjust collimation size and automatically planned for chest and stitching range these two functions are same with predicate device MULTIX Impact C, the difference is the description way, the difference does not introduce safety and effectiveness issues. |
# 9. Non-Clinical Test Conclusion
## 9.1 Performance Evaluation
Non clinical tests were conducted to verify that the proposed device met all design specifications as it is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ANSI/AAMI ES 60601-1:2005 & A1:2012 & A2:2021 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests.
- IEC 60601-1-3:2008+A1:2013+A2:2021 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral standard: Radiation protection in diagnostic X-ray equipment.
- IEC 60601-2-54:2022 Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy.
- IEC 60601-2-28:2017 Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
Additional non-clinical tests are conducted for key features to ensure safe and effectiveness when integrated into the system:
| Feature | Bench Testing Performed |
| --- | --- |
| uVision | Introduction |
{15}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
K251167
UNITED IMAGING
www.united-imaging.com
The uVision algorithm in the digital medical X-ray imaging system (uDR Aurora CX) aims to optimize the radiographic scanning workflow through patient positioning recognition technology. This algorithm utilizes cameras to capture natural images of the human body, achieving multi-modal real-time automatic localization of key anatomical points, body modeling, and pose estimation for patients. It provides the system with scanning positions, ranges, and generates motion trajectory plans for DR racks.
## Acceptance Criteria
As an auxiliary function designed to enhance clinical workflow efficiency, uVision is expected to assist users in completing pre-exposure positioning tasks. In the context of chest X-ray imaging, the retake rate due to incorrect positioning is a critical quality control metric. According to relevant studies and literature, incorrect positioning is one of the primary causes of retakes. By a 5-month-long observation experiment, positioning error results in approximately 9% rejection in DR. Furthermore, some literature indicates that positioning errors contribute to 28% of rejections. The specific figures may vary depending on the healthcare institution, equipment type, and technician experience. Considering the impact of camera specifications and gantry control accuracy on the application of uAI vision algorithm to DR equipment, we expect that when users employ the uVision function for automatic positioning, the automatically set system position and field size will meet clinical technicians' criteria with 95% compliance, thereby demonstrating that uVision can effectively assist clinical technicians in positioning tasks.
## Testing Data Information
The device with uVision function has been installed with equipment serial number 11XT7E0001.
Since the installation and commissioning over a year ago, the average daily imaging volume on the device has been around 80 patients, with approximately 45 chest X-rays per day and about 10 to 20 stitching cases per week. After receiving specialized training prior to use, the technicians operating this equipment utilize the uVision function to set the FOV and system position when conducting chest PA, whole-spine, and whole-lower-limb stitching exams.
12
{16}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED IMAGING
13
The results automatically set by the system are then statistically analyzed by clinical experts .
Testing data includes individuals of all genders and varying heights (capable of standing independently)
| Height (m) | ≤1.25 | 1.25~1.5 | 1.5~1.75 | ≥1.75 |
| --- | --- | --- | --- | --- |
| Percentage | 3% | 7% | 58% | 32% |
Table presents the evaluation results of the imaging positioning sampled randomly over a period since the equipment was put into use.
Table . The evaluation results of uVision automatically system positioning and FOV setting for chest PA、WholeSpine and WholeLowerExtremity
| Date | Chest/case | Case of Non-Compliant Cases in System-Automatically Set Results | Full Spine or Full Lower Limb Stitching/case | Case of Non-Compliant Cases in System-Automatically Set Results |
| --- | --- | --- | --- | --- |
| 2024.12.17 | 62 | 3 | 2 | 0 |
| 2024.12.18 | 44 | 3 | 2 | 0 |
| 2024.12.19. | 35 | 2 | 5 | 0 |
| 2024.12.20 | 18 | 1 | 5 | 0 |
| 2024.12.21 | 59 | 2 | 2 | 0 |
| 2024.12.22 | 47 | 1 | 1 | 0 |
| 2024.12.23 | 63 | 2 | 3 | 0 |
{17}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED IMAGING
| | Total number of cases in a week | 328 | 14 | 20 | 0 | |
| --- | --- | --- | --- | --- | --- | --- |
| | **Equipment and Protocols**
The test data was collected, and the testing protocol included chest, Whole-spine stitching, Whole-Lower-extremity stitching.
**Clinical Subgroups**
No clinical subgroups and confounders have been defined for the datasets.
**Testing & Training Data Independence**
The testing dataset was collected independently from the training dataset, with separated subjects and during different time periods. Therefore, the testing data is entirely independent and does not share any overlap with the training data.
**Summary**
According to the results of the current equipment statistics, in 95% of patient positioning processes, the light field and equipment position automatically set by uVision can meet the clinical positioning and shooting requirements. In the remaining 5% of cases, based on the light field and system position automatically set by the equipment, technicians still need to make manual adjustments | | | | | |
| uAid | **Introduction**
uAid is used for checking the quality of examination and positioning. The results can help to assist with departmental management functions. uAid is triggered after the acquisition of chest X-ray images in patients aged over 20 years, which automatically evaluates image characteristics against four criteria, namely whether there is a foreign object, whether the lung field is complete, whether the scapula is open, and whether the spine is located on the center line, categorizing images into one of three quality levels. The outcome of the evaluation is instantly accessible to radiologic technologists, reminding them to verify that the image meets the image quality control. It bears emphasis that the result of image quality control is for reference only and cannot be used as the basis for clinical diagnosis.
**Acceptance Criteria**
uAid is designed to provide an objective image evaluation method, offering hospitals a unified assessment tool to manage images/technicians. The accuracy of non-standard image | | | | | |
14
{18}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
K251167
UNITED IMAGING
www.united-imaging.com
recognition is a key quality control metric. According to relevant research and literature, the rate of Grade A clinical images is typically influenced by factors such as the technician's operational standardization, equipment performance, and quality control processes, with variations observed across different levels of medical institutions and equipment types. Mature industry guidelines and standards, such as those from European Radiology and the ACR-AAPM-SPR Practice Parameter, indicate that the Grade A image rate in public hospitals generally ranges between 80% and 90%. To ensure uAid's functionality meets clinical requirements, we referenced these guidelines and set a 90% pass rate, aligning with industry standards. This demonstrates that uAid can effectively assist clinical technicians in managing standardized image quality.
## Testing Data Information
The data collection started in October 2017, with a wide range of data sources. Some of the data come from different cooperative hospitals. After multiple cleaning and sorting, the data is stored in DICOM format. The study was approved by the institutional review board of the hospitals.
It does not include data on DR-sensitive groups such as infants and young children, and is only applicable to frontal chest images.
> Age and gender distribution of data sets for uAid:
| Age | Male | Female |
| --- | --- | --- |
| 20-29 | 310 | 698 |
| 30-39 | 308 | 744 |
| 40-49 | 298 | 798 |
| 50-59 | 385 | 801 |
| 60-69 | 320 | 799 |
| 70-79 | 200 | 472 |
| 80-89 | 97 | 210 |
| 90-99 | 21 | 46 |
| No Age | 97 | 187 |
| No Age, No Gender | 45 | |
> Distribution of negative and positive data for uAid:
{19}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED IMAGING
| | Negative | Positive |
| --- | --- | --- |
| lung field segmentation | 465 | 31 |
| Spinal centerline segmentation | 815 | 68 |
| Shoulder blades segmentation | 210 | 1089 |
| Foreign object | 1078 | 3080 |
**Equipment and Protocols:**
The data collection started in October 2017 on the uDR 780i, with a wide range of data sources. Some of the data come from different cooperative hospitals.
**Clinical Subgroups:**
No clinical subgroups and confounders have been defined for the datasets.
**Testing & Training Data Independence**
The testing dataset was collected independently from the training dataset, with separated subjects and during different time periods. Therefore, the testing data is entirely independent and does not share any overlap with the training data.
**Summary:**
Test dataset analysis results are summarized as below:
1. The average time of the uAid algorithm is 1.359 seconds, and the longest does not exceed 2 seconds;
2. The maximum memory occupation of uAid algorithm is not more than 2G;
3. For uAid, the sensitivity and specificity of whether there is a foreign body, whether the lung field is intact, and whether the scapula is open all exceed 0.9;
The uAid function can correctly identify four types of results: Foreign object, Incomplete lung fields, Unexposed shoulder blades, and Centerline deviation and make classification after the exposure image is generated: Green (qualified image), yellow (secondary image), red (waste image).
uAid can meet the requirement which is used for checking the quality of examination and position for institutions. The results can assist the image quality assessment with departmental management functions.
16
{20}
Shanghai United Imaging Healthcare Co., Ltd.
Tel: +86 (21) 67076888
Fax: +86 (21) 67076889
www.united-imaging.com
K251167
UNITED IMAGING
17
## 9.2 Clinical Image Evaluation
The clinical image evaluation was performed under the proposed device. Sample images of chest, abdomen, spine, pelvis, upper extremity and lower extremity were provided with a board certified radiologist to evaluate the image quality in this submission. Each image was reviewed with a statement indicating that image quality is sufficient for clinical diagnosis.
## 10. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the technology characteristics of the modified uDR Aurora CX, reflected in this 510(k) submission, do not alter the scientific technology of the devices and are substantially equivalent to those of the predicate devices.
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, we conclude that the uDR Aurora CX Stationary X-Ray Systems are substantially equivalent to the predicate devices. It does not introduce new indications for use, and has the same technological characteristics and does not introduce new potential hazards or safety risks.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.