The EXSYS DEXi is a diagnostic X-ray system intended for use in generating radiographic images of human anatomy for general purpose. The system obtains necessary information of patient's anatomical structure by an image processing (workstation) after process of examination using radiation exposure with DR. This system is not intended for mammography applications.
Device Story
EXSYS DEXi is a stationary diagnostic X-ray system used in clinical settings to generate radiographic images. Input consists of radiation exposure through human anatomy captured by a flat panel detector. The system includes an X-ray generator, tube, collimator, tube stand, bucky stand, patient table, and console. An image processing workstation transforms raw detector data into diagnostic images. Operated by trained healthcare professionals (radiologic technologists/physicians), the system provides images for clinical review, aiding in anatomical assessment and diagnostic decision-making. Benefits include high-quality radiographic imaging for general diagnostic purposes. The device supports manual and automated operation modes for positioning and exposure control.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing and compliance with international consensus standards (IEC 60601 series, IEC 62304, IEC 81001-5-1, ANSI UL 2900-1).
Indicated for generating radiographic images of human anatomy for general diagnostic purposes. Not intended for mammography. Applicable to general patient population.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K150816 — Jumong Series · Sg Healthcare Co, Ltd. · May 8, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 23, 2024
DRTECH Corporation % Hanbyul Kim Assistant Manager Suite No.1, 2 Floor /Suite No. 2, 3 Floor, 29, Dunchon-Daero 541 Beon-Gil, Jungwon-Gu Seongnam-si. Gyeonggi-do 13216 SOUTH KOREA
#### Re: K242847
Trade/Device Name: EXSYS DEXi (EXSYS DEXi-D401S-FRM); EXSYS DEXi (EXSYS DEXi-A401S-FRM); EXSYS DEXi (EXSYS DEXi-D402S-FRM); EXSYS DEXi (EXSYS DEXi-A402S-FRM); EXSYS DEXi (EXSYS DEXi-D401P-FRM); EXSYS DEXi (EXSYS DEXi-A401P-FRM); EXSYS DEXi (EXSYS DEXi-D402P-FRM); EXSYS DEXi (EXSYS DEXi-A402P-FRM); EXSYS DEXi (EXSYS DEXi-D503T-FRM); EXSYS DEXi (EXSYS DEXi-A503T-FRM); EXSYS DEXi (EXSYS DEXi-D401S-FRA); EXSYS DEXi (EXSYS DEXi-D402S-FRA); EXSYS DEXi (EXSYS DEXi-D401P-FRA); EXSYS DEXi (EXSYS DEXi-D402P-FRA); EXSYS DEXi (EXSYS DEXi-D503T-FRA) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: KPR Dated: September 20, 2024 Received: September 20, 2024
#### Dear Hanbyul Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and
{1}------------------------------------------------
adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K242847
Device Name
EXSYS DEXi (EXSYS DEXi-D401S-FRM); EXSYS DEXi (EXSYS DEXi-A401S-FRM): EXSYS DEXi (EXSYS DEXi-D402S-FRM); EXSYS DEXi (EXSYS DEXi-A402S-FRM): EXSYS DEXi (EXSYS DEXi-D401P-FRM): EXSYS DEXi (EXSYS DEXi-A401P-FRM); EXSYS DEXi (EXSYS DEXi-D402P-FRM); EXSYS DEXi (EXSYS DEXi-A402P-FRM); EXSYS DEXi (EXSYS DEXi-D503T-FRM); EXSYS DEXi (EXSYS DEXi-A503T-FRM): EXSYS DEXi (EXSYS DEXi-D401S-FRA); EXSYS DEXI (EXSYS DEXi-D402S-FRA); EXSYS DEXi (EXSYS DEXi-D401P-FRA); EXSYS DEXi (EXSYS DEXi-D402P-FRA); EXSYS DEXi (EXSYS DEXi-D503T-FRA)
#### Indications for Use (Describe)
The EXSYS DEXi is a diagnostic X-ray system intended for use in generating radiographic images of human anatomy for general purpose. The system obtains necessary information of patient's anatomical structure by an image processing (workstation) after process of examination using radiation exposure with DR. This system is not intended for mammography applications.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
[As required by 21 CFR 807.92]
510(k) Number: K242847
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92
### 1. Date Prepared [21 CFR 807.92(a) (1)]
09/20/2024
●
#### 2. Submitter's Information [21 CFR 807.92(a) (1)]
- Name of Sponsor: DRTECH Corporation ●
- Address: Suite No.1, 2 Floor / Suite No. 2, 3 Floor, 29, Dunchon-daero541 beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, 13216, Republic of Korea
- Contact Name: Hanbyul Kim
- Telephone No.: + 82-31-779-7720 ●
- Fax No.: + 82-31-779-7790
- Email Address : ● drtechra@drtech.com
- Registration Number: 3005172103 ●
- Name of Manufacturer: Same as Sponsor .
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]
- Trade(Brand) Name: ● EXSYS DEXi Device(Model) Name : ● EXSYS DEXi (EXSYS DEXi-D401S-FRM); EXSYS DEXi (EXSYS DEXi-A401S-FRM); EXSYS DEXi (EXSYS DEXi-D402S-FRM); EXSYS DEXi (EXSYS DEXi-A402S-FRM); EXSYS DEXi (EXSYS DEXi-D401P-FRM); EXSYS DEXi (EXSYS DEXi-A401P-FRM); EXSYS DEXi (EXSYS DEXi-D402P-FRM); EXSYS DEXi (EXSYS DEXi-A402P-FRM); EXSYS DEXi (EXSYS DEXi-D503T-FRM); EXSYS DEXi (EXSYS DEXi-A503T-FRM); EXSYS DEXi (EXSYS DEXi-D401S-FRA); EXSYS DEXi (EXSYS DEXi-D402S-FRA); EXSYS DEXi (EXSYS DEXi-D401P-FRA); EXSYS DEXi (EXSYS DEXi-D402P-FRA); EXSYS DEXi (EXSYS DEXi-D503T-FRA);
- Common Name: Diagnosis X-Ray System ●
- . Classification Name: Stationary x-ray system
- Classification Panel: ● Radiology
- Classification Regulation: 21 CFR 892.1680 ●
- . Product Code: KPR
- Device Class: II ●
{5}------------------------------------------------
#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
- 510(k) Number: ● K233530
- Applicant: DRTECH Corporation
- Trade(Brand) Name: EXSYS DEXi
- Device(Model) Name : EXSYS DEXi (EXSYS DEXi-D401S-FRM); ● EXSYS DEXi (EXSYS DEXi-A401S-FRM); EXSYS DEXi (EXSYS DEXi-D402S-FRM); EXSYS DEXi (EXSYS DEXi-A402S-FRM); EXSYS DEXi (EXSYS DEXi-D401P-FRM); EXSYS DEXi (EXSYS DEXi-A401P-FRM); EXSYS DEXi (EXSYS DEXi-D402P-FRM); EXSYS DEXi (EXSYS DEXi-A402P-FRM); EXSYS DEXi (EXSYS DEXi-D503T-FRM); EXSYS DEXi (EXSYS DEXi-A503T-FRM);
- Classification Name: Stationary x-ray system .
- Classification Panel: Radiology
- Classification Regulation: 21 CFR 892.1680 ●
- Product Code: KPR
- Device Class: . II
#### 5. Description of the Device [21 CFR 807.92(a) (4)]
The EXSYS DEXi is a diagnostic X-ray system intended for use in generating radiographic images of human anatomy for general purpose. The system obtains necessary information of patient's anatomical structure by an image processing (workstation) after process of examination using radiation exposure with DR. This system is not intended for mammography applications.
The EXSYS DEXi composed of a x-ray generator, tube, collimator, tube stand, bucky stand, patient table, flat panel detector and console.
Although it shares the same brand name as the predicate device EXSYS DEXi (K233530), a model name has been added with the following modifications:
- 1) Addition of collimator
- 2) Addition of mechanical parts (tube stand, bucky stand, patient table)
- 3) Addition of a detector
- 4) Addition of software
- 5) Update of software
#### 6. Indication for Use [21 CFR 807.92(a)(5)]
The EXSYS DEXi is a diagnostic X-ray system intended for use in generating radiographic images of human anatomy for general purpose. The system obtains necessary information of patient's anatomical structure by an image processing (workstation) after process of examination using radiation exposure with DR. This system is not intended for mammography applications.
{6}------------------------------------------------
#### 7. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device [21 CFR 807.92(a)(6), 21 CFR 807.92(b)]
The EXSYS DEXi composed of a x-ray generator, tube stand, bucky stand, bucky stand, patient table, flat panel detector and console.
It is a diagnostic X-ray system intended for use in generating radiographic images of human anatomy for general purpose. The system obtains necessary information of patient's anatomical structure by an image processing (workstation) after process of examination using radiation exposure with DR. This system is not intended for mammography applications.
The subject device has the same fundamental scientific technologies as the predicate devices. The technological comparison table below demonstrates the comparability of the technological characteristics of the new device and the currently cleared predicate devices. The Technological differences do not affect the intended use of the device.
The table 1 below compares the main performance data of the subject device with the predicate devices to substantiate equivalence of the subject device and predicates.
| Parameter | Subject Device | Predicate Device | Discussion |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | Unknown | K233530 | N/A |
| Manufacturer | DRTECH Corporation | DRTECH Corporation | N/A |
| Device/<br>Brand Name | EXSYS DEXi | EXSYS DEXi | N/A |
| Model Name | EXSYS DEXi-A401S-FRM<br>EXSYS DEXi-A402S-FRM<br>EXSYS DEXi-A401P-FRM<br>EXSYS DEXi-A402P-FRM<br>EXSYS DEXi-A503T-FRM<br>EXSYS DEXi-D401S-FRM<br>EXSYS DEXi-D402S-FRM<br>EXSYS DEXi-D401P-FRM<br>EXSYS DEXi-D402P-FRM<br>EXSYS DEXi-D503T-FRM<br>EXSYS DEXi-D401S-FRA<br>EXSYS DEXi-D402S-FRA<br>EXSYS DEXi-D401P-FRA<br>EXSYS DEXi-D402P-FRA<br>EXSYS DEXi-D503T-FRA | EXSYS DEXi-A401S-FRM<br>EXSYS DEXi-A402S-FRM<br>EXSYS DEXi-A401P-FRM<br>EXSYS DEXi-A402P-FRM<br>EXSYS DEXi-A503T-FRM<br>EXSYS DEXi-D401S-FRM<br>EXSYS DEXi-D402S-FRM<br>EXSYS DEXi-D401P-FRM<br>EXSYS DEXi-D402P-FRM<br>EXSYS DEXi-D503T-FRM | Model names have<br>been added due to the<br>addition of product<br>configurations.<br>The system has been<br>tested and there is "No<br>negative impact on<br>safety or efficacy" and<br>there are no new<br>potential or increased<br>safety risks concerning<br>this difference. |
| Classification<br>Name | Stationary x-ray system | Stationary x-ray system | Identical |
| Classification<br>Regulation | 21 CFR 892.1680 | 21 CFR 892.1680 | Identical |
| Product Code | KPR | KPR | Identical |
| Device Class | Class II | Class II | Identical |
| Intended Use | The EXSYS DEXi is a<br>diagnostic X-ray system<br>intended for use in generating<br>radiographic images of<br>human anatomy for general<br>purpose. The system obtains<br>necessary information of<br>patient's anatomical structure<br>by an image processing | The EXSYS DEXi is a<br>diagnostic X-ray system<br>intended for use in generating<br>radiographic images of<br>human anatomy for general<br>purpose. The system obtains<br>necessary information of<br>patient's anatomical structure<br>by an image processing | Identical |
| | (workstation) after process of<br>examination using radiation<br>exposure with DR. This<br>system is not intended for<br>mammography applications. | (workstation) after process of<br>examination using radiation<br>exposure with DR. This<br>system is not intended for<br>mammography applications. | |
| High Frequency X-ray Generator | | | |
| Output Power | 40kW, 50kW | 40kW, 50kW | Identical |
| Generator<br>models | DXG-40S<br>DXG-40P<br>DXG-50T | DXG-40S<br>DXG-40P<br>DXG-50T | Identical |
| Line voltage | DXG-40S, DXG-40P :<br>220~230VAC<br>DXG-50T :<br>230/380/400/480VAC | DXG-40S, DXG-40P :<br>220~230VAC<br>DXG-50T :<br>230/380/400/480VAC | Identical |
| kV Range | DXG-40S, DXG-40P :<br>40~125kV, 1kV step<br>DXG-50T :<br>40~150kV, 1kV step | DXG-40S, DXG-40P :<br>40~125kV, 1kV step<br>DXG-50T :<br>40~150kV, 1kV step | Identical |
| mA Range | DXG-40S, DXG-40P :<br>10 to 500mA<br>DXG-50T :<br>10 to 630mA | DXG-40S, DXG-40P :<br>10 to 500mA<br>DXG-50T :<br>10 to 630mA | Identical |
| Image Acquisition | | | |
| Detector | Used with DRTECH<br>Detector<br><br>K193017 :<br>EVS 4343W, EVS 4343WP,<br>EVS 3643W, EVS 3643WP<br>K192400 :<br>EVS 4343A, EVS 3643A<br>K193031 :<br>EXPD 4343P, EXPD 3643P<br>K223124 :<br>EXPD 86P, EXPD 86PG,<br>EXPD 129P, EXPD 129PG<br>K231959 :<br>EXPD 4357, EXPD 4357P<br>K232082 :<br>EXPD 4343S<br>K232753 :<br>EXPD 3643D, EXPD 4343D | Used with DRTECH<br>Detector<br><br>K193017 :<br>EVS 4343W, EVS 4343WP,<br>EVS 3643W, EVS 3643WP<br>K192400 :<br>EVS 4343A, EVS 3643A<br>K193031 :<br>EX PD 4343P, EXPD 3643P<br>K223124 :<br>EXPD 86P, EXPD 86PG,<br>EXPD 129P, EXPD 129PG | 510(k) cleared<br>detectors have been<br>added.<br><br>The system has been<br>tested and there is “No<br>negative impact on<br>safety or efficacy” and<br>there are no new<br>potential or increased<br>safety risks concerning<br>this difference.<br><br>All the flat panel<br>detectors have been<br>previously cleared by<br>510(k). |
| Image Management Software | | | |
| Software | EConsole1 (K231225)<br>EConsole2 (K240243) | EConsole1 (K231225) | Added our software,<br>EConsole2 (K240243),<br>which was previously<br>cleared with 510(k).<br><br>The system has been<br>tested and there is “No<br>negative impact on<br>safety or efficacy” and<br>there are no new<br>potential or increased |
| | | | safety risks concerning |
| | | | this difference. |
| Image viewing | Available | Available | Identical |
| Image search | Available | Available | Identical |
| Image storage | Available | Available | Identical |
| Image<br>annotation | Available | Available | Identical |
| Image<br>measurement | Available | Available | Identical |
| Image<br>processing | Available | Available | Identical |
| Image stitch | Available | Available | Identical |
| Generator<br>Control | Available | Available | Identical |
| X-ray Tube | | | |
| Tube models | E7239X, E7242X, E7884X | E7239X, E7242X, E7884X | Identical |
| Max. Tube<br>Voltage | E7239X : 125 kV<br>E7242X : 125 kV<br>E7884X : 150 kV | E7239X : 125 kV<br>E7242X : 125 kV<br>E7884X : 150 kV | Identical |
| Anode Heat<br>Capacity | E7239X : 140 kHU<br>E7242X : 200 kHU<br>E7884X : 300 kHU | E7239X : 140 kHU<br>E7242X : 200 kHU<br>E7884X : 300 kHU | Identical |
| Collimator | | | |
| Collimator<br>Models | M-38<br>R 221/A DHHS<br>R 225 ACS DHHS | M-38 | A new collimator that<br>operates manually and<br>another type that<br>operates automatically<br>have been added.<br>We have conducted<br>tests regarding the |
| Adjustment<br>method | Manual, Auto | Manual | addition of these<br>collimators, and there<br>is "no negative impact<br>on safety or efficacy."<br>Furthermore, there are<br>no new potential risks<br>or increased safety<br>hazards related to these<br>differences. |
| Tube Stand | | | |
| Models | DTS-M<br>DTS-A | DTS-M | Motorized tube stand<br>has been added.<br>We have conducted<br>tests regarding this<br>addition, and there is |
| Operation<br>Method | Manual, Auto | Manual | "no negative impact on<br>safety or efficacy."<br>Furthermore, there are<br>no new potential risks<br>or increased safety<br>hazards related to these<br>differences. |
| | | | |
| Type | Floor-supported type | Floor-supported type | Identical |
| Tube Head<br>Console | Membrane OP<br>LCD Console | Membrane OP<br>LCD Console | Identical |
| Bucky Stand | | | |
| Models | DBS-43M<br>LLD 129 With Rotation<br>LLD 129 Compact<br>DBS-57M<br>DBS-43A<br>DBS-57A | DBS-43M<br>LLD 129 With Rotation<br>LLD 129 Compact | Three Types of bucky<br>stands have been<br>added.<br>We have conducted<br>tests regarding this<br>addition, and there is |
| Operation<br>Method | Manual, Auto | Manual | "no negative impact on<br>safety or efficacy."<br>Furthermore, there are<br>no new potential risks<br>or increased safety<br>hazards related to these<br>differences. |
| Type | Floor-supported type | Floor-supported type | Identical |
| Patient Table | | | |
| Models | DPT-4A<br>DPT-6A<br>DPT-4B<br>DPT-6B<br>DPT-6AT<br>DPT-6BT | DPT-4A<br>DPT-6A | Four Types of patient<br>tables have been added.<br>We have conducted<br>tests regarding this<br>addition, and there is |
| Type | 4 way : DPT-4A, DPT-4B<br>6 way : DPT-6A, DPT-6B,<br>DPT-6AT, DPT-6BT | 4 way : DPT-4A<br>6 way : DPT-6A | "no negative impact on<br>safety or efficacy."<br>Furthermore, there are<br>no new potential risks<br>or increased safety<br>hazards related to these<br>differences. |
| Fixed method | Electromagnet | Electromagnet | Identical |
#### Table 1. Comparison of the Subject Device to the Predicate Device Substantial Equivalence
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
There are no significant differences between the subject Device and the predicate device that would have a negative impact on the product's use. Therefore, the subject device substantially equivalent to the predicate device.
{10}------------------------------------------------
### 9. Summary of Non-Clinical Data [21 CFR 807.92(b)(1)]
The EXSYS DEXi comply with the following international and FDA-recognized consensus standards list in Table 2.
| Standards development<br>organization, reference<br>number, and date | Standard name |
|----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 14971: Third Edition<br>2019-12 | Medical devices - Application of risk management to medical devices |
| IEC 60601-1 Edition 3.2<br>2020-08<br>CONSOLIDATED<br>VERSION | Medical electrical equipment - Part 1: General requirements for basic<br>safety and essential performance |
| IEC 60601-1-2 Edition 4.1<br>2020-09<br>CONSOLIDATED<br>VERSION | Medical electrical equipment - Part 1-2: General requirements for basic<br>safety and essential performance - Collateral Standard: Electromagnetic<br>disturbances - Requirements and tests |
| IEC 60601-1-6 Edition 3.2<br>2020-07<br>CONSOLIDATED<br>VERSION | Medical electrical equipment - Part 1-6: General requirements for basic<br>safety and essential performance - Collateral standard: Usability |
| IEC 62366-1 Edition 1.1<br>2020-06<br>CONSOLIDATED<br>VERSION | Medical devices - Part 1: Application of usability engineering to<br>medical devices |
| IEC 60601-1-3 Edition 2.2<br>2021-01<br>CONSOLIDATED<br>VERSION | Medical electrical equipment - Part 1-3: General requirements for basic<br>safety and essential performance - Collateral Standard: Radiation<br>protection in diagnostic X-ray equipment |
| IEC 60601-2-28 Edition<br>3.0 2017-06 | Medical electrical equipment - Part 2-28: Particular requirements for the<br>basic safety and essential performance of X-ray tube assemblies for<br>medical diagnosis |
| IEC 60601-2-54 Edition<br>1.2 2018-06<br>CONSOLIDATED<br>VERSION | Medical electrical equipment - Part 2-54: Particular requirements for the<br>basic safety and essential performance of X-ray equipment for<br>radiography and radioscopy |
| IEC 62304 Edition 1.1<br>2015-06<br>CONSOLIDATED<br>VERSION | Medical device software - Software life cycle processes |
| UL ANSI 2900-1 First<br>Edition 2017 | Standard for Safety, Standard for Software Cybersecurity Network-<br>Connectable Products, Part 1: General Requirements |
| IEC 81001-5-1 Edition<br>1.0 2021-12 | Health software and health IT systems safety, effectiveness and security<br>Part 5.1: Security - Activities in the product life cycle |
Table 2. International and FDA-recognized consensus standards
{11}------------------------------------------------
And EXSYS DEXi comply with the FDA guidance documents listed in Table 3.
Table 3. FDA Guidance Documents
| Title of Guidance Document | Issue Date |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Guidance for Industry and Food and Drug Administration Staff: The 510(k)<br>Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] | July 28, 2014 |
| Pediatric Information for X-ray Imaging Device Premarket Notifications,<br>Guidance for Industry and Food and Drug Administration Staff | Nov 28, 2017 |
| Guidance for Industry and Food and Drug Administration Staff: Medical X-Ray<br>Imaging Devices Conformance with IEC Standards | May 8, 2019 |
| Content of Premarket Submissions for Device Software Functions, Guidance for<br>Industry and Food and Drug Administration Staff | June 14, 2023 |
| Cybersecurity in Medical Devices Quality System Considerations and Content of<br>Premarket Submissions | Sep 27, 2023 |
The risk analysis was completed and risk controls were implemented to mitigate identified hazards. The test results support that all the specifications have met the acceptance criteria. Verification and validation testing were found acceptable to support the claim of substantial equivalence.
#### 10. Summary of Clinical Data [21 CFR 807.92(b)(2)]
#### Not Applicable
Clinical studies are unnecessary to validate the safety and effectiveness of the Stationary x-ray system, EXSYS DEXi, the subject of this 510(k) notification.
#### 11. Conclusion [21 CFR 807.92(b)(3)]
The EXSYS DEXi is substantially equivalent to the currently marketed predicate device (K233530) in terms of technical characteristics, general function, applications for use, safety, and effectiveness.
Substantial equivalence for Stationary X-ray System(EXSYS DEXi) was demonstrated through the non-clinical performance in compliance with the requirements specified in the international and FDA recognized consensus standards, IEC 60601-1, IEC 60601-1-6, IEC 60601-1-6, IEC 62366-1, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, IEC 62304, IEC 81001-5-1 and ANSI UL 2900-1.
The comparison of technological characteristics, non-clinical performance data and safety testing demonstrate that the EXSYS DEXi is substantially equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.