← Product Code [KPR](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR) · K200944

# UNIEXPERT 2 PLUS (K200944)

_Adani · KPR · May 26, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K200944

## Device Facts

- **Applicant:** Adani
- **Product Code:** [KPR](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR.md)
- **Decision Date:** May 26, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

The UNIEXPERT 2 PLUS is a stationary general-purpose radiographic imaging system indicated for use in acquiring diagnostic X-Ray images of osseous structures and soft tissues of the human body to aid physicians with patient diagnosis. The system has analogue, computed radiography (CR) as well as digital radiography (DR) imaging capabilities. The UNIEXPERT 2 PLUS is not indicated for use in mammography.

## Device Story

Stationary general-purpose radiographic imaging system; acquires diagnostic X-ray images of osseous structures and soft tissues. Comprised of X-ray generator, tube, beam limiting device, tube stand, radiographic table, wall stand, and digital detector. Operated by clinicians in a clinical setting. System controls X-ray generation, patient registration, image acquisition, visualization, and transfer to PACS or external media. Provides diagnostic images to physicians to aid in patient diagnosis. Benefits include standardized radiographic imaging capabilities for various body parts.

## Clinical Evidence

Bench testing only. No clinical data. Verification and validation testing confirmed system performance, safety, and EMC compliance per IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 6234, and 60601-2-54. Sample clinical images provided to demonstrate imaging capabilities.

## Technological Characteristics

Stationary X-ray system. Components: High voltage generator (40-100kW), X-ray tube, beam limiting device, vertical tube stand, radiographic table, wall stand, digital detector. Connectivity: DICOM for PACS/external communication. Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 6234, 60601-2-54. Software: StudyWorks Software Complex.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Q-Rad Radiographic System ([K011486](/device/K011486.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is an emblem of the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.

May 26, 2020

Adani % Vladimir Linev CEO & General Director 7 Selitsky str. Minsk, 220075 BELARUS

Re: K200944

Trade/Device Name: UNIEXPERT 2 PLUS Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR Dated: April 8, 2020 Received: April 8, 2020

Dear Vladimir Linev:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K200944

Device Name UNIEXPERT 2 PLUS

#### Indications for Use (Describe)

The UNIEXPERT 2 PLUS is a stationary general-purpose radiographic imaging system indicated for use in acquiring diagnostic X-Ray images of osseous structures and soft tissues of the human body to aid physicians with patient diagnosis. The system has analogue, computed radiography (CR) as well as digital radiography (DR) imaging capabilities. The UNIEXPERT 2 PLUS is not indicated for use in mammography.

| Type of Use (Select one or both, as applicable)                                      |                                                                                     |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Adani. The logo consists of a blue triangle with a curved line in the lower left corner, followed by the word "ADANI" in blue, and the tagline "from ideas to solutions" in a smaller, lighter blue font. The logo is simple and modern, and the colors are clean and professional.

# 510(k) SUMMARY K200944

## The following 510(k) Summary is submitted in accordance with the requirements as set forth in 21 CFR Part 807.92.

| Submitter Name:   | Adani                                     | Establishment Registration Number: 3004718499 |
|-------------------|-------------------------------------------|-----------------------------------------------|
| Address:          | Adani<br>7 Selitsky Str.<br>Minsk, 220075 |                                               |
| Telephone Number: | +375 (17) 349-00-00                       |                                               |
| Contact Person:   | Vladimir Linev, CEO & General Manager     |                                               |
| Date Prepared:    | March 6, 2020                             |                                               |

# Subject Devic

| Device Trade Name:    | UNIEXPERT 2 PLUS             |
|-----------------------|------------------------------|
| Device Common Name:   | Digital X-Ray Imaging System |
| Device Classification | Stationary X-Ray System      |
| Regulation            | 892.1680                     |
| Name: Device Class:   | Class II                     |
| Product Code:         | 90 KPR                       |

### Predicate Device:

| Device Trade Name:          | Q-Rad Radiographic System (K011486) |
|-----------------------------|-------------------------------------|
| Device Common Name:         | Digital X-Ray Imaging System        |
| Device Classification Name: | Stationary X-Ray System             |
| Regulation                  | 892.1680                            |
| Device Class:               | Class II                            |
| Product Code:               | 90 KPR                              |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Adani. The logo consists of a blue triangle with a curved line in the lower left corner, followed by the word "ADANI" in blue, and the tagline "from ideas to solutions" in a smaller, lighter blue font. The logo is simple and modern, and the colors are clean and professional.

#### Device Description:

The UNIEXPERT 2 PLUS is a diagnostic X-Ray system intended for radiographic imaging of osseous structures and soft tissues of the human body. The system may be used for analogue film imaging as well as digital image acquisition.

The UNIEXPERT 2 PLUS is comprised of an X-Ray generator with user interface, X-Ray tube, beam limiting device, vertical column tube stand, radiographic table, vertical wall stand, digital X-Ray image detector.

The following table contains the specific components for the UNIEXPERT 2 PLUS. A complete system is comprised of a component from each category and may include more than one grid and more than one detector. All detectors are previously-cleared by FDA.

| Generator      | X-Ray Power Supply (high voltage generator) CMP200, model: VZW2556RE2-90    |
|----------------|-----------------------------------------------------------------------------|
|                | X-Ray Power Supply (high voltage generator) CMP200 DR, model: VZW2556RB2-KK |
|                | X-Ray Power Supply (high voltage generator) CMP200 DR, model: VZW2556RE2-96 |
|                | RAD/FLUORO INDICO IQ 50 kW 150 kV 400VAC                                    |
|                | RAD/FLUORO INDICO IQ 65 kW 150 kV 400VAC                                    |
|                | RAD/FLUORO INDICO IQ 80 kW 150 kV 400VAC                                    |
| User Interface | USER AUTOMATED WORK STATION (AWS)                                           |
| Tube Stand     | ADN114.13.00.000                                                            |
| Wall Stand     | ADN114.14.00.000                                                            |
| Patient Table  | ADN114.11.00.000                                                            |
| AEC            | Solid State Measuring Chamber 610                                           |
| Grids          | MG05870002AP135AS                                                           |
|                | MG05870002AP200AS                                                           |
|                | MG05870002AR200AS                                                           |
|                | ACS AL (48 см x 44 см; 80 л/см; 12:1; f=100 см)                             |
|                | ACS AL (48 см x 44 см; 80 л/см; 12:1; f=150 см)                             |
|                | G_480X440MM_40L/CM_12_100CM_AS                                              |
| Detectors      | Varex PaxScan 4336W v4 (K171138); PaxScan 4343R v3 (K172951)                |
|                | Perkin Elmer, XPRAD2 4336 (K161966)                                         |
|                | Vieworks FXRD-1417NAW (K163703)                                             |
| Beam Limiting  | Ralco R 225                                                                 |
| Device         | Ralco R 302/A                                                               |
|                | Siemens ML03                                                                |
|                | Siemens ML04                                                                |
| X-Ray Tube     | X-Ray Tube C352-RTM 78 HS F:0.6/1.2                                         |
|                | X-Ray Tube C352-RTM 782 HS F:0.6/1.2                                        |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Adani. The logo consists of a blue triangle with a smaller triangle cut out of the bottom left corner. To the right of the triangle is the word "ADANI" in blue, and below that is the tagline "from ideas to solutions" in a smaller, lighter blue font.

#### Indications for Use:

The UNIEXPERT 2 PLUS is a stationary general-purpose radiographic imaging system indicated for use in acquiring diagnostic X-Ray images of osseous structures and soft tissues of the human body to aid physicians with patient diagnosis. The system has analogue, computed radiography (CR) as well as digital radiography (DR) imaging capabilities. The UNIEXPERT 2 PLUS is not indicated for use in mammography.

#### Summary of Technological Characteristics Comparison to Predicate Device:

The UNIEXPERT 2 PLUS and the predicate device incorporate very similar technological characteristics and indications for use are essentially the same. They are comprised of a similar set of components with comparable construction and are configured essentially the same. The following table provides information for comparison of both systems.

| Feature                             | UNIEXPERT 2 PLUS                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Predicate<br>Q-Rad Radiographic System                                                                                                                                                                                                                                                                                                                                                                                           | Comparison to<br>Predicate                                                                                                                                                                                                                                     |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number                    | K200944                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K011486                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                |
| Indications<br>For Use              | The UNIEXPERT 2 PLUS<br>is a stationary general-<br>purpose radiographic<br>imaging system indicated<br>for use in acquiring<br>diagnostic X-Ray images<br>of osseous structures and<br>soft tissues of the human<br>body to aid physicians<br>with patient diagnosis.<br>The system has<br>analogue, computed<br>radiography (CR) as well<br>as digital radiography<br>(DR) imaging capabilities.<br>The UNIEXPERT 2 PLUS<br>is not indicated for use in<br>mammography. | The Q-Rad Radiographic<br>System is indicated for use in<br>obtaining diagnostic quality<br>radiographic images to aid the<br>physician with diagnosis. The<br>system can be used to perform<br>radiographic imaging of various<br>portions of the human body,<br>including the skull, spinal<br>column, extremities, chest,<br>abdomen and other body parts.<br>The Q-Rad System is not<br>indicated for use in<br>mammography. | Indications for use<br>for the<br>UNIEXPERT 2<br>PLUS are the<br>same as the<br>predicate.                                                                                                                                                                     |
| X-Ray<br>Generator                  | Input power: 3 Phase 380-<br>480 VAC, Output power<br>options: 40kW, 50kW,<br>65kW, 80kW and 100kW.<br>All are manufactured by:<br>Communications and<br>Power Industries (CPI)<br>Available Models:<br>CMP200, CMP200 DR,<br>Indico IQ                                                                                                                                                                                                                                   | Input Power Choices - Three<br>Phase 380-480 VAC, 208-240<br>Single Phase 208-240VAC, or<br>Stored Energy for low input<br>power applications.<br>Output power ranging from 32<br>kW to 80 kW                                                                                                                                                                                                                                    | Both the<br>UNIEXPERT 2<br>PLUS and the<br>predicate<br>incorporate X-Ray<br>Generator system<br>options with<br>essentially the<br>same technological<br>characteristics with<br>the exception of a<br>Stored Energy<br>option available for<br>the predicate |
| Feature                             | UNIEXPERT 2 PLUS                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Predicate<br>Q-Rad Radiographic System                                                                                                                                                                                                                                                                                                                                                                                           | Comparison to<br>Predicate                                                                                                                                                                                                                                     |
| Tube Stand                          | Adani Model<br>ADN114.13.00.000<br>Vertical Travel:<br>61.02"/1550mm<br>Longitudinal Travel:<br>78.74"/1080mm<br>Transverse Travel:<br>7.48"/190mm<br>Rotational Capability: Yes                                                                                                                                                                                                                                                                                          | Model QS-550<br>Vertical Travel: 60.5"/1537mm<br>Longitudinal Travel:<br>98"/2489mm<br>Transverse Travel: 10"/254mm<br>Rotational Capability: Yes                                                                                                                                                                                                                                                                                | Both the<br>UNIEXPERT 2<br>PLUS and the<br>predicate<br>incorporate a floor<br>mounted vertical<br>tube support<br>structure with<br>vertical, horizontal<br>and transverse<br>travel as well as<br>rotational<br>capability.                                  |
| FDA<br>Performance<br>Standard      | all applicable performance<br>standards under 21 CFR<br>1020                                                                                                                                                                                                                                                                                                                                                                                                              | all applicable performance<br>standards under 21 CFR 1020                                                                                                                                                                                                                                                                                                                                                                        | Both the<br>UNIEXPERT 2<br>PLUS and the<br>predicate comply<br>with the same FDA<br>Performance<br>Standard                                                                                                                                                    |
| Wall Stand                          | Adani Model<br>ADN114.14.00.000<br>Vertical Travel:<br>56.69"/1440mm                                                                                                                                                                                                                                                                                                                                                                                                      | Model QW-420<br>Vertical Travel: 60.5"/1537mm                                                                                                                                                                                                                                                                                                                                                                                    | The UNIEXPERT 2<br>PLUS and the<br>predicate provide<br>wall stands with<br>similar designs and<br>range of motion<br>capabilities for<br>patient positioning.                                                                                                 |
| Patient Table                       | Adani Model<br>ADN114.11.00.000<br>Vertical Travel:<br>14.37"/365mm<br>Longitudinal Travel:<br>42.52"/1080mm<br>Transverse Travel:<br>9.84"/250mm                                                                                                                                                                                                                                                                                                                         | Model QT-750<br>Vertical Travel: 11.5"/292mm<br>Longitudinal Travel:<br>30.5"/775mm<br>Transverse Travel:<br>11.5"/292mm                                                                                                                                                                                                                                                                                                         | The UNIEXPERT 2<br>PLUS and the<br>predicate provide<br>patient tables with<br>similar adjustability<br>and range of<br>motion capabilities<br>for patient transfer<br>and positioning.                                                                        |
| Detector                            | Varex PaxScan 4336W v4<br>Varex PaxScan 4343R v3<br>PerkinElmer XRpad2 4336<br>Vieworks FXRD-1417N                                                                                                                                                                                                                                                                                                                                                                        | DRX Wireless Detector<br>DRX Fixed Detector<br>DRX 2530 Wireless Detector                                                                                                                                                                                                                                                                                                                                                        | Detectors used<br>with the<br>UNIEXPERT 2<br>PLUS and the<br>predicate device<br>incorporate<br>comparable design<br>and technology for<br>diagnostic image<br>acquisition and<br>processing.                                                                  |
| Feature                             | UNIEXPERT 2 PLUS                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Predicate Q-Rad Radiographic System                                                                                                                                                                                                                                                                                                                                                                                              | Comparison to Predicate                                                                                                                                                                                                                                        |
| Beam Limiting Device                | Ralco R 225<br>Ralco R 302A                                                                                                                                                                                                                                                                                                                                                                                                                                               | The same Manual or Automatic Collimators as well as others with equivalent design characteristics                                                                                                                                                                                                                                                                                                                                | Beam Limiting Devices available with the UNIEXPERT 2 PLUS are equivalent to those provided with the predicate device                                                                                                                                           |
| X-Ray Tube                          | X-Ray Tube C352-RTM 78 HS F:0.6/1.2<br>X-Ray Tube C352-RTM 782 HS F:0.6/1.2                                                                                                                                                                                                                                                                                                                                                                                               | X-Ray Tubes ranging from 140kHu to 600 kHU                                                                                                                                                                                                                                                                                                                                                                                       | X-Ray tubes available with the UNIEXPERT 2 PLUS are equivalent in design and functionality to those provided with the predicate device                                                                                                                         |
| User Interface and Imaging Software | Automated Work Stations (AWS) and StudyWorks Control Software Complex For The Radiography Room System                                                                                                                                                                                                                                                                                                                                                                     | Q-Rad - DRX/TechVision/QVision                                                                                                                                                                                                                                                                                                                                                                                                   | The UNIEXPERT 2 PLUS and the predicate incorporate similar image acquisition and management software                                                                                                                                                           |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Adani. The logo consists of a blue triangle with a curved line at the bottom left corner. To the right of the triangle is the word "ADANI" in blue, and below that is the phrase "from ideas to solutions" in a smaller, gray font.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for Adani. The logo consists of a blue triangle with a curved line at the bottom left corner, followed by the word "ADANI" in blue, bold letters. Below the word "ADANI" is the tagline "from ideas to solutions" in a smaller, lighter font.

## Summary of Technological Characteristics of the UNIEXPERT 2 PLUS as compared to the Predicate Device:

The technological characteristics of the UNIEXPERT 2 PLUS present no significant differences with respect to overall design and intended use when compared to the predicate device. The UNIEXPERT 2 PLUS incorporates components similar to the predicate device, including; an X-Ray generator, vertical floor mounted tubestand, patient table, wall stand, analog, CR as well as digital image capability, a beam limiting device and an X-Ray tube. The fundamental scientific technology of the UNIEXPERT 2 PLUS is the same as the predicate device.

### Non-Clinical and Clinical Testing:

Non-clinical testing included verification and validation testing and image evaluation have been performed for the UNIEXPERT 2 PLUS system. Sample clinical images covering a range of anatomical areas were provided to demonstrate the overall imaging capabilities of the device. Risk analysis was performed on the entire system. All devices not manufactured by Adani and subject to CDRH performance standards are certified to comply with the standard by their respective manufacturers.

No new or modified indications for use have been introduced by the UNIEXPERT 2 PLUS system. It has the same intended use and is technologically, the same as the predicate device. It is the determination of Adani that non-clinical testing is sufficient and that clinical testing is not required to support a determination of substantial equivalence.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Adani. The logo consists of a blue triangle on the left side of the image. To the right of the triangle is the word "ADANI" in large, blue, sans-serif font. Below the word "ADANI" is the phrase "from ideas to solutions" in a smaller, lighter blue font.

### Summary of Performance and Safety Testing:

System performance testing including functional testing of all mechanical and electrical functions, imaging performance, safety and EMC testing of the UNIEXPERT 2 PLUS system was conducted in accordance with IEC 60601-1-2:2014, IEC 60601-1:2012, IEC 60601-1-3:2013, IEC 60601-1-6:2013, IEC 6234:2006 and IEC 60601-2-54:2015. Verification and validation, as well as acceptance testing were conducted to verify that all features of the UNIEXPERT 2 PLUS functioned as intended. Results confirmed all performance criteria were met satisfactorily and fell within predetermined acceptance criteria. The output of the verification and validation process confirmed functionality and safety to be within acceptable limits.

#### Summary of Software Verification/Validation and Acceptance Testing:

The StudyWorks Software Complex enables operation of the UNIEXPERT 2 PLUS system by way of controlling the components and ensures control of the entire image acquisition cycle, starting from the input of the patient registration data to visualization of the acquired image and transfer to a PACS server or saving to CD or DVD. The StudyWorks Software Complex uses DICOM as the interface for external communication. Software/Firmware verification and validation, as well as acceptance testing were conducted to verify that all features of the StudyWorks Software Complex functioned as intended and that results fell within pre-determined acceptance criteria.

Risk analysis was performed to identify potential hazards, mitigate risks and evaluate acceptance criteria for residual risks. Since a latent design flaw in the software tool could result in a misdiagnosis or delayed medical assistance, thus potentially causing a minor injury to the patient or operator, the level of concern for this software has been determined to be "moderate". A summary of Verification and validation activities was prepared in accordance with the FDA document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

#### Substantial Equivalence Discussion:

Intended use and indications for use of the Adani UNIEXPERT 2 PLUS system and the predicate, Q-Rad Radiographic System are the same. Both the UNIEXPERT 2 PLUS and the predicate perform the same functions utilizing the same technology. They include components of comparable design with similar technological characteristics and a similar operator interface. Imaging procedures are conducted in a similar manor on both the Adani UNIEXPERT 2 PLUS and the predicate Q- Rad Rdiographic System.

#### Substatial Equivalence Conclusion:

The Adani UNIEXPERT 2 PLUS system does not introduce any new indications for use, nor does the use of the systems introduce any new potential safety or effectiveness concerns when compared to the predicate, Q-Rad Radiographic System. Therefore, Adani concludes that the UNIEXPERT 2 PLUS diagnostic X-Ray systems is substantially equivalent with the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K200944](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K200944)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
