← Product Code [KPR](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR) · K150663

# Diagnostic X-ray System (K150663)

_Osko, Inc. · KPR · May 7, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K150663

## Device Facts

- **Applicant:** Osko, Inc.
- **Product Code:** [KPR](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR.md)
- **Decision Date:** May 7, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The XR5 diagnostic X-ray system is intended for use on adult and pediatric patients for taking diagnostics radiographic exposure of all body parts and operated by a qualified/trained doctor or technician. The XR5 diagnostic X-ray system is designed to be used with conventional film/screen, CR cassettes or digital detectors. NOT intended for Mammography use.

## Device Story

XR5 is a stationary diagnostic X-ray system for general radiographic imaging. It utilizes a high-frequency generator and X-ray tube to produce X-ray beams for image capture on conventional film/screen, CR cassettes, or digital flat panel detectors. Operated by qualified doctors or technicians in clinical settings, the system includes a 4-way floating patient table, tube stand, and manual collimator. The device does not include AEC or integrated imaging software; it relies on user-provided, FDA-cleared imaging software associated with the chosen detector. The system synchronizes X-ray tube and bucky positions based on patient positioning. By providing high-quality diagnostic images, the device assists clinicians in evaluating various body parts, facilitating accurate diagnosis and treatment planning.

## Clinical Evidence

No clinical data. Substantial equivalence established via bench testing, including detector performance parameters (DQE, MTF) per IEC 62220-1:2003, and system safety/EMC testing per IEC 60601-1 and IEC 60601-2-54.

## Technological Characteristics

Stationary X-ray system; 40kW high-frequency generator; rhenium-tungsten-faced molybdenum target X-ray tube (E7239x); 4-way floating table; manual collimator. Connectivity: supports conventional film, CR, or digital flat panel detectors. Software: embedded firmware (ANSI C). Standards: IEC 60601-1, IEC 60601-2-54, IEC 60601-1-2 (EMC), 21 CFR 1020.30.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Multix Fusion by Siemens Medical Solutions USA, Inc ([K121513](/device/K121513.md))

## Reference Devices

- 1717SCC ([K122173](/device/K122173.md))
- 1417WCC ([K141566](/device/K141566.md))
- FLAATZ600 ([K132842](/device/K132842.md))
- FEEL-DRCS ([K110033](/device/K110033.md))
- XmaruView V1 ([K102078](/device/K102078.md))
- Manual collimator ([K062788](/device/K062788.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of a human figure in profile, with three overlapping heads suggesting a sense of community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure, emphasizing the organization's name and national affiliation. The logo is simple, clean, and easily recognizable.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 7, 2015

OSKO, INC. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025

Re: K150663

Trade/Device Name: XR5 Diagnostic X-ray System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: March 13, 2015 Received: March 16, 2015

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K150663

Device Name

XR5 Diagnostic X-ray System

#### Indications for Use (Describe)

The XR5 diagnostic X-ray system is intended for use on adult and pediatric patients for taking diagnostics radiographic exposure of all body parts and operated by a qualified/trained doctor or technician. The XR5 diagnostic X-ray system is designed to be used with conventional film/screen, CR cassettes or digital detectors. NOT intended for Mammography use.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)                                                     |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo and contact information for OSKO, INC. The logo features the company name in bold blue letters next to a stylized blue graphic. The contact information includes the address 7260 NW 58th Street, Miami, FL, 33166 USA, as well as the phone number +1-305-599-7161, fax number +1-305-599-7144, and website www.oskomedical.com.

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date 510k summary prepared: May 1, 2015

Submitter's Name, address, telephone number, a contact person:

| 1. Submitter      | OSKO, Inc.<br>7260 NW 58th Street, Miami, Florida 33166, USA<br>www.oskomedical.com                             |
|-------------------|-----------------------------------------------------------------------------------------------------------------|
| 2. Contact Person | Name: Wang Choi<br>Position: Vice President<br>E-mail: bioking@medisoneconet.com<br>Telephone: + 1 305-599-7161 |

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

| 3. Regulation Name  | Stationary X-ray System                                                                                                                          |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| 4. Device Name      | XR5                                                                                                                                              |
| 5. Classification   | 21 CFR 892.1680                                                                                                                                  |
| 6. Product Code     | KPR                                                                                                                                              |
| 7. Device Class     | 2                                                                                                                                                |
| 8. Predicate Device | Multix Fusion by Siemens Medical Solutions USA, Inc<br>K121513 (Decision Date – Aug 10, 2012)<br>Stationary X-ray System<br>21CFR892.1680<br>KPR |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo and contact information for OSKO, INC. The logo features the company name in blue, with a stylized blue graphic to the left. The contact information includes the address 7260 NW 58th Street, Miami, FL, 33166 USA, as well as the phone number +1-305-599-7161, fax number +1-305-599-7144, and website www.oskomedical.com.

#### 9. Device Description

The XR5 diagnostic X-ray system is used for image capture by using X-rays on a patient's body. The XR5 is a conventional X-ray machine, with an intuitive operation console to provide the user an easy way to manage optimal conditions for quality images. The high frequency generator is capable of delivering the exposition dose appropriate for general X-ray and diagnostic images of a patient. This system is made to use other detector options including: conventional films, CR, or Digital Flat Panel Detectors. Please note that the quality of the image in any detector depends on the manufacture of the receptor device. The XR5 diagnostic X-ray system doesn't provide an AEC feature.

#### 10. Indications for Use

The XR5 diagnostic X-ray system is intended for use on adult and pediatric patients for taking diagnostics radiographic exposure of all body parts and operated by a qualified/trained doctor or technician. The XR5 diagnostic X-ray system is designed to be used with conventional film/screen, CR cassettes or digital detectors. NOT intended for Mammography use.

#### 11. Specifications of Device

- 11.1 System Environment Requirement
	- a) ENVIRONMENTAL SPECIFICATIONS

### OPERATING

| Ambient temperature range | 10 to 40 °C (50 to 104 °F).                                 |
|---------------------------|-------------------------------------------------------------|
| Relative humidity         | 20 to 80%, non-condensing.                                  |
| Altitude                  | -700 to 3000 meters (1100 to 700 hPa, 825 to 525 mm<br>Hg). |

### Electrical Requirement

| System Part     | Specifications                               |
|-----------------|----------------------------------------------|
| Generator       | 220 ~ 240 VAC monophasic, 50/60 Hz @<br>40kW |
| Mechanical Part | 110~220VAC                                   |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo and contact information for OSKO, INC. The logo features the company name in blue, with a stylized graphic of intersecting lines. The contact information includes the address 7260 NW 58th Street, Miami, FL, 33166 USA, as well as the telephone number +1-305-599-7161, fax number +1-305-599-7144, and website www.oskomedical.com.

- 11.2 Movement Specifications Required power : 24V, 3A Tube stand height : 1,940 mm / 73" SID range : 1,000 ~ 1,800 mm / 39" ~ 70" Vertical movement : 1,480 mm / 58" Longitudinal movement : 2,540 mm / 100" Tube rotation range : - 135° ~ +135° Forward/backward tube head range : ± 120 mm / ± 4.7"
- 11.3 Patient Table Specifications

Type : 4-Way floating table Dimensions : 220 X 806 X 728 X 655 mm / 9" x 32" x 26" (Rounded to nearest inch) Backward/forward movement : ± 105 mm / ± 4.1" Left/right Movement: ± 450 mm / ± 18" Patient Weight Limit: Patient Weight(215 kg) + Accessory Weight(15kg) / Patient Weight(473lb) + Accessory Weight (33lb)

- 11.4Collimator Specifications
Type : Manual Power Output : DC 24V Projecting lamp : 150W Halogen lamp Lighting type : Automatic illumination Lighting time: press button, 30 sec APR memory : No Laser guide : No Max. window : 48 x 48 cm (SID 100 cm) / 19"x 19" ( SID 39")

- Min. window : 5 X 5 cm (SID 100 cm) / 2"x 2" ( SID 39")
- 11.5X-ray Tube Specifications

Model : E7239x Focal Spot : 1.0 / 2.0 mm Operation Voltage : 40 ~ 125 kVp Target angle : 16° Composition of the target : Rhenium - Tungsten faced molybdenum Permanent filtration : 0.9mmAl/75kV Anode Heat capacity : 140 KHU Additional Filter : 1.5mmAl Target : rhenium-tungsten-faced molybdenum Permanent filtration : 0.9mm Al at 75 kV Cooling method : Natural / Forced air Operation Temp. : 10 ~ 40°C / 50 ~ 104°F Operation Hum. : 30 ~ 85%

- 11.6Grid Specifications

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo and contact information for OSKO, INC. The logo features the company name in blue, with a stylized graphic to the left. The contact information includes the address 7260 NW 58th Street, Miami, FL, 33166 USA, as well as the phone number +1-305-599-7161, fax number +1-305-599-7144, and website www.oskomedical.com.

Composition : interspaced carbon with carbon fiber cover Dimensions : 46 x 46,8 cm / 18"x 18"x 3" Density : 103 lp/cm Ratio : 10:1 FID : 100 and 180 cm / 39" and 71" 12.HF Generator Specifications

| OUTPUT PARAMETER                                                                                    | GENERATOR SERIES | LOADING FACTOR                   |
|-----------------------------------------------------------------------------------------------------|------------------|----------------------------------|
| Maximum X-ray tube voltage<br>and highest X-ray tube<br>current at that voltage                     | 40 kW            | 125 kV, 320 mA / 150 kV, 250 mA  |
| Maximum X-ray tube current<br>and highest X-ray tube<br>voltage at that current                     | 40 kW            | 500 mA, 80 kV                    |
| Combination of X-ray tube<br>current and X-ray tube<br>voltage resulting in highest<br>output power | 40 kW            | 400 mA, 100 kV                   |
| Highest constant output<br>power at 100 kV, 0.1 sec                                                 | 40 kW            | 40 kW (400 mA,<br>100 kV, 0.1 s) |

13. Summary of the Technological Characteristics of the device compared to the Predicate Device

The XR5 diagnostic X-ray system described in this 501K has the similar indications for use and technical characteristics as the predicate device, Multix Fusion of Siemens Medical Solution USA, Inc. (K121513)

| Feature                | New Device: XR5<br>Product Code: KPR                                                                                                                                                                                                                                                                    | Predicate: Multix Fusion (K121513)<br>Product Code: KPR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer           | OSKO, Inc                                                                                                                                                                                                                                                                                               | Siemens Medical Solution USA, Inc                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Indications<br>for Use | The XR5 diagnostic X-ray system is intended<br>for use on adult and pediatric patients for<br>taking diagnostics radiographic exposure of<br>all body parts and operated by a<br>qualified/trained doctor or technician. The XR5<br>diagnostic X-ray system is designed to be used<br>with conventional | The Multix Fusion system is a radiographic<br>system used in hospitals, clinics, and<br>medical practices. Multix Fusion enables<br>radiographic exposures of the whole body<br>including: skull, chest, abdomen, and<br>extremities and may be used on pediatric,<br>adult and bariatric patients. It can also be<br>used for intravenous, small interventions<br>(like biopsy, punctures, etc.) and<br>emergency (trauma, critical ill)<br>applications.<br>Exposures may be taken with the patient<br>sitting, standing, or in the prone position. |

{7}------------------------------------------------

|                         | film/screen, CR<br>cassettes or digital<br>detectors. NOT intended<br>for Mammography use.                                                                                                                                                                                                                                                                 | The Multix Fusion system is not meant for<br>mammography.<br>The Multix Fusion uses a mobile (wired) or<br>portable (wireless) digital detector<br>for generating diagnostic images by<br>converting x-rays into electronic signals.<br>The Multix Fusion is also designed to be<br>used with conventional film/screen<br>or Computed Radiography (CR) cassettes. |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Configuration           | Patient sitting, standing,<br>in the prone position                                                                                                                                                                                                                                                                                                        | Patient sitting, standing, in the prone<br>position                                                                                                                                                                                                                                                                                                               |
| Performance<br>Standard | 21 CFR 1020.30                                                                                                                                                                                                                                                                                                                                             | 21 CFR 1020.30                                                                                                                                                                                                                                                                                                                                                    |
| Generator               | High frequency                                                                                                                                                                                                                                                                                                                                             | High frequency                                                                                                                                                                                                                                                                                                                                                    |
| Collimator              | Manual (K062788)                                                                                                                                                                                                                                                                                                                                           | ACSS collimator (Automatic Cassette Size<br>Sensing)                                                                                                                                                                                                                                                                                                              |
| Image<br>acquisition    | Film/Screen, CR<br>Cassettes, or digital flat<br>panel detectors                                                                                                                                                                                                                                                                                           | Film/Screen, CR Cassettes, or digital flat<br>panel detector                                                                                                                                                                                                                                                                                                      |
| Flat Panel<br>Detector  | 1717SCC(K122173):<br>43cm x 43 cm<br>1417WCC (K141566):<br>43 cm x 36cm<br>FLAATZ600 (K132842):<br>43cm x 36cm                                                                                                                                                                                                                                             | 43cm X 36cm (4336X)                                                                                                                                                                                                                                                                                                                                               |
| Imaging<br>Software     | By default, XR5 does not provide the imaging software with image<br>processing features. Separate imaging software with a FDA<br>clearance is available for the user provided by the manufacturer of the<br>detector of choice; FEEL-DRCS (K110033) for FLAATZ 600<br>(K132842) and XmaruView V1(K102078) for 1417 WCC (K141566)<br>and 1717SCC (K122173). |                                                                                                                                                                                                                                                                                                                                                                   |

14. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:

The XR5 diagnostic X-ray system does not have significant changes in energy source, or technological characteristics when compared to the predicate device. It uses the components similar to the predicate device (tube, generator, floor anchored tube support, table, bucky wall stand). The indications for use and fundamental scientific technology are similar to the predicate device, with the exception of bariatric patients and tomography application.

The main differences are the digital detectors used for XR5 X-ray system. However, the fundamental imaging technology is the same. The safety and effectiveness of the

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for OSKO. The logo consists of the word "OSKO" in a bold, sans-serif font, with the letters in a dark blue color. To the right of the word is a stylized graphic element, also in shades of blue, that resembles a waveform or a stylized butterfly. The graphic element is composed of two curved shapes that intersect in the middle.

digital detectors for XR5 have been reviewed and cleared by FDA as indicated by the 510K number of each detector.

1717SCC (K122173) 1417WCC (K141566) FLAATZ600 (K132842)

By default, XR5 does not provide the imaging software with image processing features. Separate imaging software with a FDA clearance is available for the user provided by the manufacturer of the detector of choice; FEEL-DRCS (K110033) for FLAATZ 600 (K132842) and XmaruView V1(K102078) for 1417 WCC (K141566) and 1717SCC (K122173).

## 15. Summary of Software Validation

XR5 firmware is a software embedded device programmed with ANSI C language. It controls the hardware operating module to run the equipment and synchronize the positions of the X-ray tube and buck according to the patient's position. It also displays information necessary for taking an X-ray image including SID.

The complete system configuration including the firmware function has been assessed and tested by the manufacturer and passed all in-house testing criteria. The firmware validation test was designed to evaluate for the X-ray system including X-ray exposure control and exposure switch performed by XR5.

The level of concern is moderate and the firmware is not based on any predicate device. The validation testing verified and validated the risk analysis and individual performance results were within the predetermined acceptance criteria. Both subject device and predicate device are categorized in product code KPR ; equivalence between these models is evident.

In conclusion, the XR5 is substantially equivalent to predicate devices.

### 16. Summary of Non-Clinical Testing

The detectors listed have been cleared by FDA's 510(k) process. The detector performance parameters including DQE and MTF are evaluated according to IEC 62220-1:2003 Standard to demonstrate the safety and effectiveness performance. MTF of 1717SCC (K122173) and 1417 WCC (K141566) and FLAATZ600 (K132842) are 13%, 12.1% and 50.02%, respectively at 3.5 lp/mm. DQE of 1717SCC (K122173) and 1417 WCC (K141566) and FLAATZ600 (K132842) are 54%, 76.7% and 50%, respectively at 0 spatial frequency,. The line resolution of 1717SCC (K122173) and 1417 WCC (K141566) and FLAATZ600 (K132842) are similar between 3.0~3.5 lp/mm.

The risk analysis is performed to identify additional hazards, conduct risk mitigation measures and evaluate the acceptance criteria for residual risks.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo and contact information for OSKO, INC. The logo features the company name in blue, along with a blue abstract graphic. The contact information includes the address 7260 NW 58th Street, Miami, FL, 33166 USA, as well as the telephone number +1-305-599-7161, fax number +1-305-599-7144, and website address www.oskomedical.com.

#### 17. Summary of Clinical Testing

Bench performance parameters for the digital detectors are sufficient to demonstrate substantial equivalence to a predicate device. While the sample clinical images may be helpful for demonstrating the clinical use of your device and that it performs adequately in the general use radiographic imaging for which your device is indicated, bench performance data should be sufficient for determining substantial equivalence. Clinical images are not necessary to establish substantial equivalence based on the previously FDA cleared detectors.

#### 18. Functional and Safety Testing

The XR5 diagnostic X-ray system has been tested in accordance with Safety standard of IEC 60601-1; 2005 + A1 (2012) and IEC 60601-2-54; Edition 1.0.2009. The test reports have shown good performance, substantially equivalent to the predicate device.

The XR5 has also met applicable Electro Magnetic Compatibility (EMC) requirements; EN60601-1-2:2007 / AC : 2010 (IEC60601-1-2:2007).

#### 19. Conclusion

The XR5 diagnostic X-ray system is intended for the similar use as Multix Fusion (with the exception of bariatric patients and tomography application). It uses similar components cleared for the Multix Fusion (e.g. tube, generator, ceiling-mounted tube support, table, bucky wall stand). The XR5 is substantially equivalent to the already cleared predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K150663](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K150663)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
