GC85A

K150165 · Samsung Electronics Co., Ltd. · KPR · Apr 2, 2015 · Radiology

Device Facts

Record IDK150165
Device NameGC85A
ApplicantSamsung Electronics Co., Ltd.
Product CodeKPR · Radiology
Decision DateApr 2, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1680
Device ClassClass 2

Intended Use

The GC85A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

Device Story

GC85A is a stationary digital X-ray imaging system used in clinical settings by physicians or technicians. It consists of a high-voltage generator, ceiling suspension, X-ray tube, collimator, detector, patient table, and wall stand. The system transmits X-rays through a patient's body; the detector captures the radiation and converts it into electrical signals. These signals are amplified and converted to digital data, then processed by the S-Station software. The resulting images are saved in DICOM format and transmitted to a PACS server for clinical review. The device assists healthcare providers in diagnostic imaging by providing high-quality radiographic images, facilitating clinical decision-making for various anatomical conditions.

Clinical Evidence

Clinical images were obtained and evaluated by a professional radiologist with U.S. board certification. The images were compared to those produced by the predicate device and found to be equivalent. Clinical data was provided as supplementary evidence to laboratory performance data.

Technological Characteristics

Stationary digital X-ray system. Components: High-frequency generator, ceiling suspension, wall stand, patient table, collimator, and CsI indirect detectors. Connectivity: DICOM-compliant, PACS integration. Standards: ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO 14971, 21 CFR 1020.30/31, IEC 62220-1 (MTF/DQE). Wireless functionality verified per FDA guidance.

Indications for Use

Indicated for generating radiographic images of human anatomy for use by qualified/trained doctors or technicians. Not for mammographic applications.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 2, 2015 Samsung Electronics Co., Ltd. % Chulsin Kim Regulatory Affairs Manager 129, Samsung-ro, Yeongtong-gu Suwon-si, Gyeonggi-do 443-742 REPUBLIC OF KOREA Re: K150165 Trade/Device Name: GC85A Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: January 23, 2015 Received: January 26, 2015 Dear Chulsin Kim: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert A Ochs Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K150165 Device Name GC85A Indications for Use (Describe) The GC85A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------| | <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, bold, sans-serif font. The word is set against a blue, oval-shaped background that is tilted slightly. The logo is simple and modern, and it is easily recognizable. 510(k) Premarket Notification - Traditional ### Section 5: 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92 - 1. Date: January 23, 2015 - 2. Submitter - Company Name: SAMSUNG ELECTRONICS Co., Ltd. A. - B. Address: 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 443-742, Republic of Korea ### 3. Primary Contact Person - A. Name: KIM, CHULSIN - B. Title: Regulatory Affairs Manager - C. Phone Number: +82-31-200-7661 - FAX Number: +82-31-200-1199 D. E-Mail: chulsin.kim@samsung.com ### 4. Secondary Contact Person - Name: Ninad Gujar A. - B. Title: Regulatory Affairs Manager - C. Phone Number: 978-564-8503 - D. FAX Number: 978-750-6677 E-Mail: ngujar@samsungneurologica.com ### 5. Proposed Device - A. Trade Name: GC85A - B. Device Name: GC85A - C. Common Name: Digital Diagnostic X-ray System - Classification Name: System, X-ray, Stationary D. - Product Code: KPR ட். - F. Regulation: 21 CFR 892.1680 ### 6. Predicate Device - A. Manufacturer: SAMSUNG ELECTRONICS Co., Ltd. - B. Trade Name: XGEO GC80 - C. Classification Name: System, X-ray, Stationary - D. Product Code: KPR - E. 510(k) Number: K140334 - 510(k) Decision Date: May 28, 2014 F. ### 7. Device Description The GC85A digital X-ray imaging system consists of High voltage generator (HVG), Ceiling {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The text is set against a blue, rounded, rectangular background. ### 510(k) Premarket Notification - Traditional Suspension, X-ray tube, Collimator, Detector, AEC, DAP, CIB(Control Interface Box), Wall Stand, Patient Table, Collimator, Detector, Remote controller, Grid, Foot switch, Barcode scanner and Auto-stitching stand. This system is used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process device before being sent to the S-Station (Operation Software) and saved in DICOM file, a standard for medical imaging. The captured images are sent to the Picture Archiving & Communication System (PACS) server, and can be used for reading images. ### 8. Intended Use The GC85A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. ### 9. Summary of Technological characteristic of the proposed device compared with the predicate device The proposed GC85A adds new detectors and modify wall stand to the predicate device, and it does not have significant change in materials, energy source or technological characteristics compared to the predicate device, XGEO GC80 (K140334). Comparisons of the following technological characteristics were assessed and the results demonstrate the substantial equivalence to the predicates. | Specification | Predicate Device | Proposed Device | Discussion | |---------------|--------------------------------|--------------------------------|------------| | Device Name | XGEO GC80 | GC85A | | | Manufacturer | SAMSUNG ELECTRONICS | SAMSUNG ELECTRONICS | | | 510(k) Number | K140334 | N/A | | | Appearances | | | Same | | Intended Use | The XGEO GC80 digital X-ray | The GC85A digital X-ray | Same | | | imaging system is intended for | imaging system is intended for | | | | use in generating radiographic | use in generating radiographic | | | | images of human anatomy by a | images of human anatomy by | | | | qualified/trained doctor or | a qualified/trained doctor or | | | | technician. This device is not | technician. This device is not | | | | intended for mammographic | intended for mammographic | | | | applications. | applications. | | | Manufacturer Contents | XGEO GC80 (K140334) | GC85A | Discussion | |-----------------------|---------------------|-------|------------| |-----------------------|---------------------|-------|------------| {5}------------------------------------------------ Image /page/5/Picture/1 description: The image features the SAMSUNG logo. The logo is displayed in white text against a blue background. The text is bold and slightly angled. 510(k) Premarket Notification - Traditional | Manufacturer Contents | | XGEO GC80 (K140334) | GC85A | Discussion | |----------------------------------------------|-------------------------------------|---------------------|----------------|---------------| | (1)High Voltage Generator | | | | | | Type | | High Frequency | High Frequency | Same | | Max. Power | | 80kW | 82kW | Difference(1) | | Output<br>RANGE | Tube Voltage | 40-150kV | 40-150kV | Same | | | Tube Current | 10-1000mA | 10-1000mA | Same | | | Exposure<br>Time | 1msec-6.3sec | 1msec-10sec | Difference(2) | | | AEC<br>(Automatic Exposure Control) | | Yes | Yes | | APR (Anatomically<br>Programmed Radiography) | | Yes | Yes | Same | | Manufacturer Contents | XGEO GC80 (K140334) | GC85A | Discussion | | |-----------------------------|---------------------|--------------------------------------|-----------------------------------------|---------------| | (2)Ceiling Suspension | | | | | | Moving<br>Range<br>(mm) | Longitudinal | 1680~4180<br>(Varies with room size) | 1680~4180<br>(Varies with room<br>size) | Same | | | Lateral | 1030~3030<br>(Varies with room size) | 1030~3030<br>(Varies with room<br>size) | Same | | | Vertical | 1840 | 1800 | Difference(3) | | Vertical Tube Moving method | | Motorized | Motorized | Same | | Tube Assembly Rotation | | -157 ~ +183 | -157 ~ +183 | Same | | Brake locking Method | | Electromagnetic | Electromagnetic | Same | | Automatic Centering | | O | O | Same | | Moving Rail Type | | Al Extrusion | Al Extrusion | Same | | Image Preview | | O | O | Same | | Display Type | | Color LCD | Color LCD | Same | | Control Switch Type | | Button<br>+ Touch Screen | Button<br>+ Touch Screen | Same | | | With Table | O | O | Same | | Vertical Sync. | With Stand | O | O | Same | | Manufacturer Contents | XGEO GC80 (K140334) | GC85A | Discussion | | | |------------------------------|---------------------|------------------|------------|---------------|------| | (3) Wall Stand | | | | | | | Vertical<br>Movement | Mechanism | Motorized/Manual | Motorized | Difference(4) | | | | Range(mm) | 400~1800 | 280~1850 | Difference(5) | | | Detector/tube servo coupling | | Yes | Yes | Same | | | Detector | Tilting | Mechanism | Motorized | Motorized | Same | {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the SAMSUNG logo in white text on a blue background. The logo is stylized with a slightly italicized font. The blue background is in the shape of an oval. 510(k) Premarket Notification - Traditional | Manufacturer Contents | | XGEO GC80 (K140334) | GC85A | Discussion | |---------------------------|----------------|----------------------------------------|----------------------------------------|---------------| | | Range | -20~+90 | -20~+90 | Same | | | AEC | Conventional | Conventional | Same | | Grid | Lines/cm | 84.6 | 85/92 | Difference(6) | | | Grid mechanism | Stationary | Stationary | Same | | | Removability | Removable | Removable | Same | | Detector Support Mounting | | Floor | Floor | Same | | Patient Support Device | | Patient handgrips, lateral support bar | Patient handgrips, lateral support bar | Same | | Manufacturer Contents | | | XGEO GC80 (K140334) | GC85A | Discussion | |----------------------------------|----------------|------------------|------------------------------|------------------------------|---------------| | (4)Patient Table | | | | | | | Table Top | Size(mm) | | 2410 X 812 | 2410 X 812 | Same | | | Rang | Lateral | ±140 | ±140 | Same | | | e<br>(mm) | Longitudi<br>nal | +480 | +480 | Same | | Table height | Mechanism | | DC Motor, Ball screw | DC Motor, Ball screw | Same | | | Range(mm) | | 545 ~ 900 | 545 ~ 900 | Same | | Horizontal range of detector(mm) | | | 590 | 688 | Difference(7) | | AEC | | | Conventional | Conventional | Same | | Grid | Lines/cm | | 84.6 | 85/92 | Difference(8) | | | Grid mechanism | | Stationary | Stationary | Same | | | Removability | | Removable | Removable | Same | | Vertical Sync. | | | O | O | Same | | Control Switch Type | | | Foot switch | Foot switch | Same | | Maximum Patient Weight(kg) | | | 350<br>(Static, Center load) | 350<br>(Static, Center load) | Same | | Manufacturer Contents | XGEO GC80 (K140334) | GC85A | Discussion | |--------------------------------------|-----------------------|-----------------------|------------| | (5)Collimator | | | | | Overall Size(mm) | H212 X W300<br>X D179 | H212 X W300<br>X D179 | Same | | Beam Limiting Blade<br>Moving Method | Motorized<br>/Manual | Motorized<br>/Manual | Same | | Manual Operation Method | Volume | Volume | Same | | Collimator Rotation | ±45 | ±45 | Same | | Beam Light Source | LED | LED | Same | | Light Field Indicator Timer | O | O | Same | | Side Lamp | O | O | Same | {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the word "SAMSUNG" in white letters against a blue background. The font is bold and sans-serif. The logo is simple and recognizable, and the blue color is a signature of the brand. 510(k) Premarket Notification - Traditional | Manufacturer Contents | XGEO GC80 (K140334) | GC85A | Discussion | |--------------------------|---------------------|--------------|------------| | | Laser Module | Laser Module | Same | | Field Size / SID Display | Color LCD | Color LCD | Same | | Manufacturer<br>Contents | XGEO GC80 (K140334) | GC85A | | | Discussion | | |--------------------------------------------------------------------|------------------------------|------------------------------|------------------------------|------------------------------|------------------------------|--------------------| | (6) Detector<br>*NOTE: S4335-W, S4343-W were cleared with K140334. | | | | | | | | Name | S4335-W | S4343-W | S4335-W | S4343-W | S3025-W | | | Detector Type | CsI | CsI | CsI | CsI | CsI | Same | | | Indirect | Indirect | Indirect | Indirect | Indirect | Same | | Detector Area | 14"X17"<br>(345mmX42<br>5mm) | 17"X17"<br>(425mmX42<br>5mm) | 14"X17"<br>(345mmX42<br>5mm) | 17"X17"<br>(425mmX42<br>5mm) | 10"X12"<br>(245mmX29<br>5mm) | Difference(9) | | | Number of pixels | 2466X3040 | 3036X3040 | 2466X3040 | 3036X3040 | 1750X2108 | | Pixel Pitch(um) | 140 | 140 | 140 | 140 | 140 | | | High Contrast<br>Limiting<br>Resolution<br>(LP/mm) | 3.57 | 3.57 | 3.57 | 3.57 | 3.57 | Difference<br>(10) | | Communication | Wired /<br>Wireless | Wired /<br>Wireless | Wired /<br>Wireless | Wired /<br>Wireless | Wired /<br>Wireless | Same | | No | Differences | Explanation | |-----|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | (1) | HVG Max Power | Proposed medical device's HVG has higher max power than the predicate device's max power, and the higher max power does not contribute any adverse impacts to the device's safety and performance. | | (2) | HVG Exposure Time Range | Proposed medical device's HVG maximum exposure time longer than the predicate device's one, and the longer exposure time does not contribute any adverse impacts to the device's safety and performance. | | (3) | Vertical moving range of Ceiling Suspension | Proposed medical device's Vertical moving range of Ceiling Suspension has shorter than the predicate device's moving range, but the shorter moving range does not contribute any adverse impacts to the device's safety and performance. | | (4) | Vertical Movement Mechanism of Wall Stand | Proposed medical device's Vertical Movement Mechanism of Wall Stand has a motorized method, while the predicate device's Vertical Movement Mechanism of Wall Stand has two methods as manual & motorized. The movement mechanism does not contribute any adverse impacts to the device's safety and performance. | | (5) | Vertical Moving range of | Proposed medical device's Vertical Moving range of Wall Stand has | {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue, rounded rectangular background. The font is sans-serif and bold, giving the logo a clean and modern appearance. The blue background provides a strong contrast, making the white text stand out. 510(k) Premarket Notification - Traditional | | Wall Stand | longer than the predicate device's one, and the longer moving range<br>does not contribute any adverse impacts to the device's safety and<br>performance. | |------|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | (6) | Grid line ( Wall stand ) | Proposed medical device's Line Pair of Grid installed in Wall Stand<br>has higher than the predicate device's one, and the higher line pair of<br>grid does not contribute any adverse impacts to the device's safety<br>and performance. | | (7) | Horizontal range of<br>detector (Patient Table) | Proposed medical device's Horizontal Moving range of Detector<br>installed Patient Table has longer than the predicate device's one,<br>and the longer moving range does not contribute any adverse impacts<br>to the device's safety and performance. | | (8) | Grid line (Patient Table) | Proposed medical device's Line Pair of Grid installed in Patient Table<br>has higher than the predicate device's one, and the higher line pair of<br>grid does not contribute any adverse impacts to the device's safety<br>and performance. | | (9) | Detector Area | Proposed medical device's S3025-W detector has smaller area than<br>the predicate device's detectors while technical specification is<br>identical among them such as type & pixel pitch, and the smaller area<br>does not contribute any adverse impacts to the device's safety and<br>performance. | | (10) | Number of pixels<br>Resolution and pixel pitch<br>of detector | Proposed medical device's S3025-W detector has smaller detector<br>area. Therefore, the proposed medical device's numbers of pixels is<br>smaller than the predicate device's detectors while pixel pitch is<br>identical among them, and the smaller number of pixels does not<br>contribute any adverse impacts to the device's safety and<br>performance. | In non-clinical data, the proposed detectors show curves and measurements of MTF and DQE that do not differ from the predicate device. In clinical data, the proposed GC85A has been shown a substantially equivalent to the predicate device. ### 10. Safety, EMC and Performance Data Electrical, mechanical, environmental safety and performance testing according to standard ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO14971, 21CFR1020.30 and 21CFR1020.31 were performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2. Wireless function was tested and verified followed by guidance, Radio frequency Wireless Technology in Medical Devices. All test results were satisfying the standards. ### 11. Non-clinical data Non-clinical testing data was provided in conformance to the FDA "Guidance for the Submission of 510(k)'s for Solid-State X-ray Imaging Devices", which includes MTF and DQE measurements as tested by IEC 62220-1. The proposed device shows no difference in non-clinical testing data such as MTF and DQE measurements from the predicate device. {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, sans-serif font, set against a blue, elliptical background. The ellipse is tilted slightly, giving the logo a dynamic appearance. ### 510(k) Premarket Notification - Traditional ### 12. Clinical data In clinical data, clinical images were obtained in accordance with FDA guidance for the submission of 510(k)'s for Solid State X-ray Imaging Devices. They were evaluated by a professional radiologist and found to be equivalent to the predicate device. ### 13. Conclusions Clinical images were provided; these images were not necessary to establish substantial equivalence based on the modifications to the device but they provide further evidence in addition to the laboratory performance data to show that the complete system works as intended. These images were evaluated by a radiologist with equivalent U.S. board certification and found to be equivalent to the predicate device. - 14. Samsung Electronics Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...