K132934 · Siemens Medi Cal Solutions, Inc. · KPR · Apr 10, 2014 · Radiology
Device Facts
Record ID
K132934
Device Name
MULTIX SELECT DR
Applicant
Siemens Medi Cal Solutions, Inc.
Product Code
KPR · Radiology
Decision Date
Apr 10, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Multix Select DR is a digital system used for making X-ray exposures of the head, spinal column, abdomen, thorax (lungs), internal organs and extremities with/without using conventional film/screen and CR -systems, and may be used on pediatric, adult and bariatric patients. The Multix Select DR system is not meant for mammography.
Device Story
Multix Select DR is a stationary digital radiography (DR) system for low-end market segments; utilizes X-ray generator, tube, patient table, Bucky wall stand, and mobile (wired) flat panel detector (FPD). System captures X-ray signals, converts them to digital images via Fluorospot Compact High-Res imaging system. Operated by healthcare professionals in clinical settings. Provides diagnostic images for head, spine, abdomen, thorax, and extremities. Output viewed by clinicians to support diagnostic decision-making. Benefits include efficient digital imaging workflow compared to conventional film/screen or CR systems.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, verification, and validation activities, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance standards for diagnostic X-ray systems.
Indicated for radiographic X-ray exposures of head, spine, abdomen, thorax, internal organs, and extremities. Suitable for pediatric, adult, and bariatric patients. Contraindicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
APR 1 0 2014
# SIEMENS
### Multix Select DR 510(k) K132934 Al Response
## 510(k) Summary: Multix Select DR X-ray System
Siemens Medical Systems, Inc. Company: 1 Valley Stream Parkway Malvern, PA 19355
Date Prepared: March 12, 2014
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807,92.
#### 1. General Information:
Importer / Distributor: Siemens Medical Solutions, Inc. 51 Valley Stream Parkway, E-50 Malvern, PA 19355 Establishment Registration Number: 2240869
Location of Manufacturing Site: Siemens AG Medical Solutions X-Ray Products Henkestrasse 127 DE-91052 Erlangen Establishment Registration Number: 3002808157
### Manufacturer:
Siemens Shanghai Medical Equipment Ltd. 278 Zhou Zhu Road, Shanghai 201318, China Headquarters: Siemens AG Wittelsbacherplatz 2 D-80333 Munich 2, Germany Establishment Registration Number: 3003202425
### 2. Contact Person:
Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway D-02 Malvern, PA 19355 Fax: (610) 448-1787 Phone: (610) 448 -3536
Email: patricia.d.jones@siemens.com
{1}------------------------------------------------
- 3. Device Name and Classification: Trade Name: . Classification Name: Classification Panel: Radiology Classification Regulation: Device Class: Class II Product Code: 90 KPR
- Legally Marketed Predicate Device 4. Trade Name: 510(k) #: Clearance Date: Classification Name: Classification Panel: CFR Section: Device Class: Product Code:
Multix Select DR Stationary X-Ray System 21 CFR §892.1680
Multix Fusion K121513 August 10, 2012 Stationary X-Ray System Radiology 21 CFR §892.1680 Class II 90 KPR
### Device Description: ·
Select DR is a product sharing same image system platform with Multix Siemens' Multix Fusion x-ray system cleared under Premarket Notification K121513 on 08/10/2012, but target to low end DR market segment. The Multix Select DR system consists of radiologic table, x-ray generator, x-ray tube, flat panel detector (mobile (wired)), imaging system and Bucky-wall stand.
The Multix Select DR offers the following system configurations:
- A digital radiography system with a mobile (wired) flat panel detector, .
The key components are a Patient Table and a Bucky wall stand which are available in different configurations. The x-ray tube is a Single Tank Tube Assembly and mounted in a column integrated on the patient table. A manual movement of the x-ray tube is available.
Similar to the cleared Multix Fusion x-ray system, Multix Select DR has the same or similar comparable components. It does not affect the indication for use nor the intended use of the device nor does it alter its fundamental scientific technology.
#### 5. Indication for Use:
The Multix Select DR is a digital system used for making X-ray exposures of the head, spinal column, abdomen, thorax (lungs), internal organs and extremities with/without using conventional film/screen and CR -systems, and may be used on pediatric, adult and bariatric patients. The Multix Select DR system is not meant for mammography.
{2}------------------------------------------------
### 6. Substantial Equivalence:
- The Multix Select DR is substantially equivalent to the commercially available Siemens Multix Fusion (K121513) radiographic x-ray system with similar indication for use. The Multix Fusion was described in premarket notification K121513 which received FDA Clearance on August 10, 2013. (See Table 1 below).
| Predicate Device | 510(k) | Clearance | Comparable Properties |
|-----------------------|---------|------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Name & Manufacturer | Number | Date | |
| Multix Fusion<br>SSME | K121513 | 08/10/2012 | Indications for use<br>X-ray tube<br>Collimator<br>Table<br>X-ray Generator<br>Bucky wall stand<br>Digital Imaging system<br>Flat panel detector |
## Table 1: Predicate Device Comparable Properties
- Summary of Technological Characteristics of the Subject Device as 7. Compared with the Predicate Device:
The Multix Select DR uses the same Digital Imaging system as the predicate device. The differences in the Subject Device, such as Generator, X-ray Tube, Collimator, radiological table and Bucky Wall Stand, do not affect the safety or effectiveness of the device The Multix Select DR uses a flat panel detector similar to the predicate device, the difference does not affect the safety or effectiveness, which is supported by conforming to detector characteristics described per the FDA Guidance for Submission of 510(k)'s for Solid State X-ray Imaging Devices, (issued on Augus 6, 1999).
| Comparable | Subject Device Multix Select | Predicate Device Multix |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Properties | DR | Fusion |
| Indications for use | The Multix Select DR is a<br>digital system used for making<br>X-ray exposures of the head,<br>spinal column, abdomen,<br>thorax (lungs), internal organs<br>and extremities with/without<br>using conventional film/screen<br>and CR -systems, and may be<br>used on pediatric, adult and<br>bariatric patients. The Multix<br>Select DR system is not meant<br>for mammography. | The Multix Fusion system is a<br>radiographic system used in<br>hospitals, clinics, and medical<br>practices, Multix Fusion<br>enables radiographic<br>exposures of the whole body<br>including: skull,<br>chest, abdomen, and<br>extremities and may be used<br>on pediatric, adult and bariatric<br>patients.<br>Exposures may be taken with<br>the patient sitting, standing, or<br>in the prone position. The<br>Multix Fusion system is not<br>meant for mammography. The<br>Multix Fusion uses a mobile |
Table 2 Subject and Predicate Device Technical Properties
{3}------------------------------------------------
## SIEMENS
## Multix Select DR 510(k) K132934 Al Response
| Comparable<br>Properties | Subject Device Multix Select<br>DR | Predicate Device Multix<br>Fusion |
|--------------------------|--------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | (wired) |
| | | for generating diagnostic<br>images by converting x-rays<br>into image signals. The Multix<br>Fusion is also designed to be<br>used with conventional<br>film/screen or Computed<br>Radiography (CR) cassettes. |
| X-ray tube | OPTIPHOS 135/30/55R | OPTITOP 150/40/80HC-100 |
| Collimator | Manual | Manual or automatic* (ACSS) |
| Table | Floating table top | Lifting and floating table top |
| X-ray Generator | 55 kW | 55 kW/ 65 kW/ 80 kW |
| Bucky wall stand | Manual vertical movable Bucky<br>Wall stand, Non-tiltable Bucky<br>Tray. | Manual vertical movable<br>Bucky Wall stand, tiltable<br>Bucky Tray. |
| Digital Imaging system | Fluorospot Compact High - Res<br>Digital Imaging | Fluorospot Compact High -<br>Res Digital Imaging |
| Flat panel detector | DRZ+ (Gadox) with<br>amorphous silicon (a-Si)<br>technology | Cesium iodide scintillator (Csl)<br>with amorphous silicon (a-Si)<br>technology |
The modifications included in the comparison table above (see Table 2) and described throughout this submission do not alter the Indications for use or fundamental scientific technology of the legally marketed predicate device. The differences between the legally marketed predicate device and the subject device have been assessed via Verification and Validation as well as Risk Management. Any differences in technological characteristics are accompanied by information within this submission that demonstrates the device is as safe and effective as the predicate device and do not raise different questions of safety and effectiveness than the predicate.
#### 8. Performance Testing
Siemens claims conformance to a signed statements of performance standards and Federal X-ray Performance Standards. This submission contains performance data to demonstrate continued conformance with special controls for medical devices containing software for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005.
Detector of the Subject Device conforms to the guidance for the submission of 510(k) for Solid State X-ray Imaging Devices. Non-clinical Data is provided in Original 510k submission Appendix G.
Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
{4}------------------------------------------------
EMC Mechanical Safety electrical safety was evaluated according to the IEC Siemens certify to conformance Voluntary Standards Covering Standards. Electrical and Mechanical Safety." (See Table 3 below). In conclusion, the identified risk of EMC/Mechanical/Electrical hazards was mitigated and is substantially equivalent to the predicate device in terms of safety and effectiveness.
Siemens conforms to Voluntary and EPCR standard (see Table 3 below).
Siemens hereby certifies that the subject device the Multix Select DR will be in Clemono neroby ooking recognized consensus standards covering electrical and mechanical safety listed in Table 3 below.
| Recognition<br>Number | Product<br>Area | Title of Standard | Refer. No.&<br>Date | Standards<br>Dev. Org. |
|-----------------------|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|------------------------|
| 5-4 | General | Medical Electrical Equipment - Part 1:<br>General Requirements for Safety, 1988;<br>Amendment 1, 1991-11, Amendment 2, 1995 | 60601-1 | IEC |
| 5-27 | General | Medical Electrical Equipment - Part 1:<br>General Requirements for Safety Collateral<br>Standard: Safety requirements for medical<br>electrical system | 60601-1-1:<br>2000 | IEC |
| 5-34 | General | Medical Electrical Equipment - Part 1-2:<br>General Requirements for Safety - Collateral<br>standard: Electromagnetic Compatibility -<br>Requirements and Tests (Edition 2:2001 with<br>Amendment 1:2004; Edition 2.1 (Edition<br>2:2001 consolidated with Amendment<br>1:2004) | 60601-1-2 | IEC |
| 12-199 | General | Medical Electrical Equipment - Part 1:<br>General requirements for basic<br>safety3.Collateral Standard: General<br>requirements for Radiation protection in<br>diagnostic X-ray equipment. | 60601-1-3:First<br>Ed. 1994-07 | IEC |
| 12-210 | Radiology | Medical Electrical Equipment - Part 1-3:<br>General requirements for basic safety and<br>essential performance - Collateral Standard:<br>Radiation protection in diagnostic X-ray<br>equipment | 60601-1-3:Ed.<br>2.0 2008-01 | IEC |
| 5-41 | General | Medical Electrical Equipment - Part 1-4:<br>General requirements for safety - Collateral<br>standard: Programmable electrical medical<br>systems, edition 1.1 | 60601-1-4:2000<br>Consol. Ed. 1.1 | IEC |
| 5-40 | General | Medical devices - application of risk<br>management to medical devices | 14971 Second<br>edition 2007- | ISO |
Table 3: Conformance to Voluntary Standards
{5}------------------------------------------------
# SIEMENS
## Multix Select DR 510(k) K132934 Al Response
| Recognition<br>Number | Product<br>Area | Title of Standard | Refer. No.&<br>Date | Standards<br>Dev. Org. |
|-----------------------|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|------------------------|
| | | | 03-01 | |
| 13-8 | Software | Medical device software - Software life cycle<br>processes | 62304 Ed. 1.0 | IEC |
| 2-156 | Biocompati-<br>bility | Biological evaluation of medical devices --<br>part 1: evaluation and testing | 10993-1:2009 | AAMI ANSI<br>ISO |
| 12-34 | Radiology | Medical electrical equipment - Part 2-7:<br>Particular requirements for the safety of high-<br>voltage generators of diagnostic X-ray<br>generators | 60601-2-7<br>(1998) | IEC |
| 12-126 | Radiology | Medical electrical equipment -<br>Part 2: Particular requirements for the safety<br>of X-ray source assemblies and X-ray tube<br>assemblies for medical diagnosis | 60601-2-28:<br>2010 | IEC |
| 12-127 | Radiology | Medical Electrical Equipment -<br>Part 2: Particular requirements for the safety<br>of associated equipment of X-ray equipment | 60601-2-32:<br>1994 | IEC |
| 12-238 | Radiology | Digital Imaging and Communications in<br>Medicine | PS 3.1-3.20<br>(2011) | NEMA |
| N/A | Radiology | Medical Electrical Equipment - Part 1-6:<br>General requirements for safety - Collateral<br>standard: Usability ,First edition, 2004 | 60601-1-6<br>(2004) | IEC |
| 12-201 | Radiology | Medical electrical equipment - Part 2-54:<br>Particular requirements for the basic safety<br>and essential performance of X-ray<br>equipment for radiography and radioscopy | IEC 60601-2-54<br>(Edition 1.0) | IEC |
Siemens hereby certifies that the subject device the Multix Select DR will meet the applicable requirements of the FDA Performance Standards for lonizing Radiation applicatio roquets for diagnostic X-Ray systems and their major components, as listed Enliting i roduction into interstate commerce. All data will be available for inspection at the firm.
| | | | Table 3 (cont.): Required Performance Standards | |
|--|--|--|--------------------------------------------------|--|
|--|--|--|--------------------------------------------------|--|
| 21 CFR Title No. | Title of 21CFR Section |
|------------------|-----------------------------------------------|
| 1020.30(c) | Manufacturer's Responsibility (Certification) |
| 1020.30(e) | Identification of X-ray components |
| 1020.30(g) | Information to be provided to assemblers |
| 1020.30(h) | Information to be provided to users |
{6}------------------------------------------------
KB2934
Page 7 of 7
## SIEMENS
Multix Select DR 510(k) K132934 Al Response
| 21 CFR Title No. | Title of 21CFR Section |
|------------------|-----------------------------------------------------------------------------------------------------|
| 1020.30(j) | Warning label |
| 1020.30(k) | Leakage Radiation |
| 1020.30(l) | Radiation from components other than the diagnostic source<br>assembly |
| 1020.30(m) | Beam Quality |
| 1020.30(n) | Aluminum equivalent of material between patient and image<br>receptor |
| 1020.31(a) | Control and indication of technique factors |
| 1020.31(b) | Reproducibility |
| 1020.31(c) | Linearity |
| 1020.31(d) | Field limitation and alignment for mobile, portable and stationary<br>general-purpose x-ray systems |
| 1020.31(e) | Field indication and alignment on stationary general purpose x-<br>ray equipment |
| 1020.31(j) | Beam-on indicators |
### General Safety and Effectiveness Concerns: 9.
Instructions for use are included within the device labeling, and the information motions for acc are morage to operate the device in a safe and effective Several safety features including visual and audible warnings are manner. incorporated into the system design. In addition the Multix Select DR is intoliporulou- nio- in error occurs, the system functions will be blocked and an error message will be displayed.
Furthermore the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray examinations to be responsible for the over and mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing.
### Conclusion as to Substantial Equivalence: 10.
The Multix Select DR is intended for the same clinical use as Multix Fusion, and it uses the same or similar components as cleared in Multix Fusion.
The functionality of Multix Select DR is similar to the predicate device. It is Siemens opinion, that the Multix Select DR is substantially equivalent to the cleared predicate device the Multix Fusion (K121513) radiographic x-ray system.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in bold and is centered in the image. The text is likely the header of a document or website.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 10, 2014
Siemens Medical Solutions, Inc. % Ms. Patricia Jones Technical Specialist, Regulatory Submissions 51 Valley Stream Parkway MALVERN PA 19355
Re: K132934
Trade/Device Name: Multix Select DR Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: March 12, 2014 Received: March 13, 2014
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act that I DA has made a aothernations administered by other Federal agencies. You must Of any I cachi statutes and regaraments, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical CITY rail 807); labornig (21 CFR 803); good manufacturing practice requirements as set device-related adverse overse overse of the contract 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
{8}------------------------------------------------
Page 2-Ms. Jones
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Smh.7)
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# SIEMENS ______________________________________________________________________________________________________________________________________________________________________
.
. .
:
.
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | |
|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Food and Drug Administration | |
| Indications for Use | |
| | Form Approved: OMB No. 0910-0120 |
| | Expiration Date: January 31, 2017 |
| | See PRA Statement below. |
| 510(k) Number (if known) | K132934 |
| Device Name | Multix Select DR |
| Indications for Use (Describe) | The Multix Select DR is a digital system used for making X-ray exposures of the head, spinal column, abdomen, thorax (lungs), internal organs and extremities with/without using conventional film/screen and CR -systems, and may be used on pediatric, adult and baristric patients. The Multix Select DR system is not meant for mammography. |
| Type of Use (Select one or both, as applicable) | |
| <span style="text-decoration: underline;"><b>Prescription Use (Part 21 CFR 801 Subpart D)</b></span> | <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14) Page 1 of 1
.