← Product Code [KPR](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR) · K012546

# OPTIMA URS UNIVERSAL RADIOGRAPHIC SYSTEM (K012546)

_Sedecal USA, Inc. · KPR · Nov 2, 2001 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K012546

## Device Facts

- **Applicant:** Sedecal USA, Inc.
- **Product Code:** [KPR](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR.md)
- **Decision Date:** Nov 2, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Intended Use

The "Optima URS" Universal Radiographic System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.

## Device Story

Optima URS is a stationary universal swivel arm X-ray system; operates on 120V AC. System features floor-to-wall column and turnable arm with variable height center. Used in clinical settings by qualified doctors or technicians to capture diagnostic radiographic images of various body parts. Compact design facilitates quick installation and space efficiency. Provides conventional exposure techniques for patients in standing, sitting, or recumbent positions. Output is radiographic exposure used by clinicians for diagnostic decision-making.

## Clinical Evidence

Bench testing only. No clinical data provided. Safety and effectiveness established through comparison of performance standards and user testing against the predicate device.

## Technological Characteristics

Stationary X-ray system; floor-to-wall column with turnable swivel arm; variable height center. Operates on 120V 50-60Hz AC. Complies with 21 CFR 1020.30, UL-2601 (IEC-60601), UL187, EN50081-1, EN50082-1, and EN60950. ISO 9001 certified.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Siemens Multix Stationary X-Ray System ([K001201](/device/K001201.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

EXHIBIT 2

K012546 Page 1 of 2

SEDECAL SA Pelaya 9- Poligono Industrial Rio De Janeiro 28110 -Algete Madrid Spain Tel (34) 91-628 0544/91-628 1592 Fax (34) 91-628 0574 (Foreign Manufacturer)

SEDECAL USA. Inc. 2910 N. Arlington Heights Rd. Arlington Heights Illinois 60006 Tel 847-394-6960 Fax 847-394-6966 (Initial Importer)

## August 2, 2001 Contact: Gary Fromberg, Official Correspondent 510(k) Summary of Safety and Effectiveness

- 1. Identification of the Device: Proprietary-Trade Name: "Optima URS" Universal Radiographic System Classification Name: Stationary X-Ray System, Product Code 90 KPR Common/Usual Name: Stationary X-Ray System
- 2. Equivalent legally marketed devices This product is similar in function to the Siemens Multix Stationary X-Ray System (K001201)
- 3. Indications for Use (intended use) The "Optima URS" Universal Radiographic System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
- 4. Description of the Device: The "Optima URS" is a stationary unit which operates from 120 V 50-60~ AC. Optima URS is a universal swivel arm X-ray system. It is easy to operate and permits a swift radiographic procedure, a feature which applies to all conventional exposure techniques on all parts of the body. The system is composed of a floor-to wall column and a turnable arm with variable high center. It allows one to take exposures of patients in standing, sitting or laying position. Owing to its compact design Optima URS is a low-cost radiography system which takes up little space and is quick to install.
- 5. Safety and Effectiveness, comparison to predicate device. The results of bench and user testing indicates that the new device is as safe and effective as the predicate devices.

{1}------------------------------------------------

K012596
page 2 of 2

| Characteristic          | Siemens Multix<br>Stationary X-Ray<br>System (K001201)                                                                                                                                                                                                                                                                                                                      | "Optima URS"<br>Universal<br>Radiographic<br>System                                                                                                                                                                                                    |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use:           | Intended for use by a<br>qualified/trained doctor or<br>technician on both adult and<br>pediatric subjects for taking<br>diagnostic radiographic<br>exposures of the skull, spinal<br>column, chest, abdomen,<br>extremities, and other body<br>parts. Applications can be<br>performed with the patient<br>sitting, standing, or lying in<br>the prone or supine position. | SAME                                                                                                                                                                                                                                                   |
| Performance<br>Standard | 21 CFR 1020.30                                                                                                                                                                                                                                                                                                                                                              | SAME                                                                                                                                                                                                                                                   |
| Electrical safety       | Electrical Safety per<br>Underwriters Laboratories<br>Standard UL-2601(IEC-<br>60601) and IEC 60601,<br>Underwriters Laboratories<br>Standard UL187:  UL<br>Standard for Safety for X-<br>Ray Equipment, CE<br>Marking Requirements, ISO<br>9001.                                                                                                                           | SAME, plus EMC:<br>EN50081-1 Residential<br>commercial, light<br>industry general - EMC<br>emission EN50082-1<br>Residential, commercial,<br>light industry general -<br>EMC immunity,<br>EN60950 Safety of IT<br>and electrical business<br>equipment |

## 6. Substantial Equivalence Chart, "Optima URS"

## 7. · Conclusion

After analyzing both bench and user testing data, it is the conclusion of Sedecal USA that the "Optima URS" Universal Radiographic System is as safe and effective as the predicate device, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird-like figure with three curved lines representing wings or streams.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 0 2 2001

Sedecal USA. Inc. % Mr. Daniel Kamm Regulatory Engineer Kamm & Associates P.O. Box 7007 DEERFIELD IL 60015 Re: K012546

Trade/Device Name: Optima URS Universal Radiographic System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: 90 KPR Dated: August 6, 2001 Received: August 7, 2001

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

NOV 0222001

K012546

## i) Indications for Use

510(k) Number KO12546

Device Name: "Optima URS" Universal Radiographic System

Indications for Use: The "Optima URS" Universal Radiographic System is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

Over the Counter Use (Per 21 CFR 801.109)

Nancyc Brozdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K012546

OR

7

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K012546](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K012546)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
