EXPERT-KL AND HAND ACQUISITION AND ANALYSIS SOFTWARE
Device Facts
| Record ID | K964263 |
|---|---|
| Device Name | EXPERT-KL AND HAND ACQUISITION AND ANALYSIS SOFTWARE |
| Applicant | Lunar Corp. |
| Product Code | KGI · Radiology |
| Decision Date | Nov 7, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1170 |
| Device Class | Class 2 |
Intended Use
The EXPERT-XL Hand and/or Forearm Acquisition and Analysis Software is an accessory software option for estimation of Bone Mineral Density BMD (in g/cm²) of the forearm and/or hand region provided for by the EXPERT-XL Bone Densitometer.
Device Story
Accessory software for EXPERT-XL Bone Densitometer; estimates bone mineral density (BMD) in g/cm² for forearm and hand regions. Input: X-ray scan data from densitometer; 10-second scan duration. Processing: software analysis of scan data to calculate BMD. Output: BMD values displayed to clinician. Used in clinical settings by trained operators. Results assist clinicians in assessing bone health. Benefits: provides quantitative bone density measurements for diagnostic assessment.
Clinical Evidence
Bench testing and comparative analysis. Results correlate highly (r=0.96) with predicate DPX series measurements. In vivo short-term precision (CV) is 2% or less for all measured sites, comparable to DPX series densitometers.
Technological Characteristics
Software accessory for bone densitometry. Operates on EXPERT-XL Bone Densitometer hardware. Scan time: 10 seconds. Radiation exposure: 12 mrem.
Indications for Use
Indicated for the estimation of bone mineral density (BMD) in the forearm and hand regions using the EXPERT-XL Bone Densitometer.
Regulatory Classification
Identification
A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Related Devices
- K965226 — EXPERT-XL TOTAL BODY ACQUISITION AND ANALYSIS SOFTWARE · Lunar Corp. · Oct 3, 1997
- K983271 — FRACTURE RISK ASSESSMENT OPTION FOR DPX SERIES BONE DENSITOMETERS · Lunar Corp. · Nov 18, 1998
- K983269 — FRACTURE RISK ASSESSMENT OPTION FOR EXPERT-XL BONE DENSITOMETER · Lunar Corp. · Nov 18, 1998
- K992677 — EAGLE OPTION FOR THE QDR BONE DENSITOMETER, MODELS QDR4500, QDR2000, QDR1500, QDR4000, QDR1000, QDR1000W · Hologic, Inc. · Sep 3, 1999
- K133664 — ENCORE VERSION 16 SOFTWARE FOR GE LUNAR DXA BONE DENSITOMETERS · Ge Medical Systems Ultrasound & Primary Care Diagn · May 15, 2014