To estimate bone mineral content (BMC, grams) and bone mineral density (BMD, grams/cm²).
Device Story
DTX-100 is a single energy X-ray absorptiometry (SXA) bone densitometer. Input: forearm submerged in a water bath (soft tissue equivalent). Operation: 40 kV X-ray generator (0.2 mAmps) with 1 mm source collimation; detector uses scintillating crystal and photomultiplier tube. Output: BMC and BMD estimates. Used in clinical settings by healthcare professionals. Device provides quantitative data to compare patient bone density against age-matched peers and young adult mean values, aiding in the assessment of bone mineral status.
Clinical Evidence
Clinical data used to establish a reference database for normal American female Caucasians (ages 20-80). Data collected from four clinical studies (Helen Hayes Hospital and New England Medical Center) involving a statistically valid sample size (at least 50 persons per ten-year age group). Results demonstrate the reference database is suitable for comparing patient BMC/BMD with age-matched peers and young adult means.
Technological Characteristics
Single energy X-ray absorptiometry (SXA) system. Components: computer, keyboard, monitor, printer, scanner. X-ray source: 40 kV, 0.2 mAmps, 1 mm collimation. Detector: scintillating crystal with photomultiplier tube. Soft tissue equivalent: water bath. Calibration: line-by-line. Reference location: 8 mm Ulna-Radius gap.
Indications for Use
Indicated for the estimation of bone mineral content (BMC) and bone mineral density (BMD) in the distal section of the forearm. Intended for prescription use by healthcare professionals.
Regulatory Classification
Identification
A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Osteometer MediTech A/S model DTX-100 bone densitometer
Related Devices
K961011 — DTX-200 BONE DENSITOMETER · Shotwell & Carr, Inc. · Jul 5, 1996
K964562 — DTX-200 BONE DENSITOMETER · Osteometer Meditech A/S · Aug 22, 1997
K021331 — DEXACARE G4 · Osteometer Meditech, Inc. · May 21, 2002
K983724 — ORCA-DX BONE DENSITOMETER · Lunar Corp. · Dec 8, 1998
K992125 — XR 46 DXA (CENTRAL) BONE DENSITOMETER, MODEL XR 46 · Norland Medical Systems, Inc. · Nov 22, 1999
Submission Summary (Full Text)
{0}
K 963789
Impressed
# Appendix E
510(k) Summary
[This 510(k) summary is furnished in accordance with 21 CFR 807.92]
JUL 14 1997
21 CFR 807.92(a):
21 CFR 807.92(a)(1):
* Submitter's name and address:
Osteometer Meditech A/S
Glerupvej 2, DK-2160 Rodovre
Denmark
* Submitter's telephone number: 011 45 44 92 42 00
* Contact person:
Ms. Anja Koue
Official Correspondent
Osteometer Meditech A/S
Glerupvej 2, DK-2160 Rodovre
Denmark
* Date this 510(k) summary was prepared: September 15, 1996
21 CFR 807.92(a)(2):
* Trade/proprietary name of the device: DTX-100 Bone Densitometer
* Classification name: Bone densitometer
21 CFR 807.92(a)(3); Legally marketed device (predicate device) to which equivalence is claimed:
* Osteometer MediTech A/S model DTX-100 bone densitometer
21 CFR 807.92(a)(4); Description of the device that is the subject of this premarket notification:
The DTX-100 candidate device is a single energy system, where the measured object, in this case the forearm, is submerged into a soft tissue equivalent water bath.
The X-Ray generator is driven at 40 kV, with a current of 0.2 mAmps. The detector uses a scintillating crystal and
OM53.6
299
{1}
OM53.6
2
301
# Appendix E
## 510(k) Summary
[This 510(k) summary is furnished in accordance with 21 CFR 807.92]
photomultiplier tube. A highly stable X-Ray output, together with a source collimation of 1 mm, and excellent pixel resolution, yields images of very high quality.
The typical time to estimate bone mineral content and bone mineral density is approximately 3.0 to 4.5 minutes. The effective patient dose is estimated to be 0.01 micro Sievert per scan.
## 21 CFR 807.92(a)(5); Intended use:
The intended use of the DTX-100 candidate device is the same as the labeled intended use of the predicate device to which equivalence is claimed; i.e., "To estimate bone mineral content (BMC, grams) and bone mineral density (BMD, grams/cm²).
## 21 CFR 807.92(a)(6); Technological characteristics:
The design, material, chemical composition, energy source and other technological characteristics of the subject device are considered to be the same as the technological characteristics of the predicate devices. A summary of the technological characteristics of the subject device in comparison to those of the predicate devices follows:
* X-ray transmission source:
Both the DTX-100 subject and the DTX-100 predicate bone densitometer devices consist of a computer, keyboard, monitor, printer, and scanner system using an X-ray transmission source. The DTX-100 candidate device and the DTX-100 predicate device are identical and use single X-ray photon absorptiometry (SXA) to estimate bone mineral content (BMC) and bone mineral density (BMD).
* Soft-tissue equivalent material:
Both the candidate and predicate DTX-100 devices are identical and use water as soft tissue equivalent material.
* Calibration:
Both the DTX-100 candidate device and the DTX-100 predicate device use a line-by-line calibration system, ensuring good precision.
* Imaging technique, data acquisition, and quality control phantom:
{2}
OM53.6
3
302
# Appendix E
## 510(k) Summary
[This 510(k) summary is furnished in accordance with 21 CFR 807.92]
These characteristics are the same on the DTX-100 candidate device as the DTX-100 predicate device.
* Source collimation:
The source collimation for both the DTX-100 candidate device and the DTX-100 predicate device is the 1 mm diameter.
* Reference location:
The reference location for both the DTX-100 candidate device and the DTX-100 predicate device is the 8 mm Ulna-Radius gap.
* Accuracy and precision error:
Accuracy and precision error of the DTX-100 candidate device is comparable to that of the DTX-100 predicate device.
## 21 CFR 807.92(b);
510(k) summaries for those premarket submissions in which a determination of substantial equivalence is also based on performance data shall also contain the following:
### 21 CFR 807.92(b)(1);
Brief discussion of the nonclinical tests submitted, referenced, or relied on in this premarket notification submission:
There were no nonclinical tests submitted, referenced, or relied on in this submission.
### 21 CFR 807.92(b)(2);
Brief discussion of the clinical tests submitted, referenced, or relied on in this premarket notification submission:
The results of four (4) clinical tests were used to obtain data to establish a reference data base for a reference population of normal American female caucasians ranging in age from 20 to 80 years. These clinical studies were conducted at:
1. Helen Hayes Hospital (West Haverstraw, NY); protocol # 94-12:
* Title: "Determination of Forearm Bone Density in a Normal US Population and Comparison of Two New Procedures to Measure Bone Mineral Density".
{3}
OM53.6
4
303
# Appendix E
## 510(k) Summary
[This 510(k) summary is furnished in accordance with 21 CFR 807.92]
* Conducted from May 1995 through May 1996.
2. New England Medical Center Hospitals (Boston, MA); protocol # 540:
* Title: "Is the Current Recommended Dietary Allowance of Vitamin D Sufficient for Minimizing Bone Loss?"
* Conducted from May 22, 1991 through August 17, 1993
3. New England Medical Center Hospitals (Boston, MA); protocol # 620:
* Title: "Effect of Calcium and Vitamin D on Bone Loss from the Hip"
* Conducted from February 6, 1992 through February 8, 1996.
4. New England Medical Center Hospitals (Boston, MA); protocol # 770:
* "Seasonal Changes in Bone Mass of Premenopausal Women"
* Conducted from January 31, 1993 through April 10, 1995.
Data in a reference population (normal, American, female, caucasians) of a statistically valid quantity of women participating in the four clinical studies were collected using a model DTX-100 bone densitometer to estimate bone mineral content and bone mineral density in the non-dominant forearm of a representative sample of the reference population and included at least 50 persons in ten year peer age groups. The mean and standard deviations of the data was calculated using appropriate statistical methods and was used to establish the reference data base for the reference population.
21 CFR 807.92(b)(3); The conclusions drawn from the nonclinical and clinical tests that demonstrate that the subject device is as safe, as effective, and performs as well as or better than the predicate device:
The standard deviation of estimated bone mineral content (BMC) and bone mineral density (BMD) of the reference population of participating female caucasians is statistically representative (as required by FDA's "Draft Guidance for Review of Bone Densitometer 510(k) Submissions"; i.e., at least 50 people in each ten year age
{4}
# Appendix E
## 510(k) Summary
[This 510(k) summary is furnished in accordance with 21 CFR 807.92]
group) of the reference population and the American reference data base (derived from this data) is suitable for use by a health care professional to compare BMC and BMD with age matched peers and for comparison with young adult mean value.
... END OF 510(k) SUMMARY ...
OM53.6
5
304
{5}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Osteometer Meditech A/S
c/o Mr. Bert Hudson
Regulatory Consultant
Shotwell & Carr, Inc.
3003 LBJ Freeway, Suite 100
Dallas, TX 75234-7755
JUL 14 1997
Re: K963789
DXT-100 Bone Densitometer, Part # 9SCA0006
Dated: May 28, 1997
Received: May 29, 1997
Regulatory Class: II
21 CFR 892.1170/Procode: 90 KGI
Dear Mr. Hudson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{6}
K 965 '89
# Appendix H1
FDA's New Form Describing "Indications for Use".
Page 1 of 1
510(k) Number (if known): Unknown; Not yet assigned by FDA.
Device Name: Model DTX-100 Bone Densitometer
Indications For Use for the device: Estimate bone mineral content (BMC in grams) and bone mineral density (BMD in grams/cm²) in the distal section of the forearm.
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
DM53.3
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K963789
311
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.