DXAVIEW SPINE AND HIP, MODEL QCT-3D PLUS

K031991 · Image Analysis, Inc. · KGI · Jul 30, 2003 · Radiology

Device Facts

Record IDK031991
Device NameDXAVIEW SPINE AND HIP, MODEL QCT-3D PLUS
ApplicantImage Analysis, Inc.
Product CodeKGI · Radiology
Decision DateJul 30, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1170
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The DXAView™ Hip and Spine bone densitometry Options are intended to provide estimates of bone mineral density (BMD) and bone mineral content (BMC) in the proximal femur and spine. Young normal reference data taken from caucasian populations can be used to calculate T-Scores. Age matched normal reference data arc used to calculate Z-Scores. The BMD results, along with the T- and Z-Scores, can be used by physicians in their determination of fracture risk and therapy effectivencss.

Device Story

DXAView Hip and Spine are software options for Image Analysis QCT-3D Plus product; processes 3D CT volumetric image sets obtained via conventional CT protocols. Device performs phantom calibration to estimate BMC (grams) and BMD (g/cm2 of calcium hydroxyapatite) of proximal femur and spine. Used in clinical settings by physicians to calculate T-Scores and Z-Scores using NHANES III reference data. Output assists in identifying low bone density, estimating fracture risk, and monitoring therapy effectiveness. Provides results comparable to conventional DXA densitometry.

Clinical Evidence

Clinical comparison studies against DXA densitometry. Sample size not specified. Primary endpoints: BMD correlation and software precision. Results: Correlation coefficients (R) of 0.93-0.94 for total femur and femoral neck. Software precision: 0.012 g/cm² (total femur) and 0.017 g/cm² (femoral neck).

Technological Characteristics

Software-based bone densitometry. Utilizes phantom calibration for CT scanners. Computes BMC and BMD from 3D CT volumetric image sets. Uses NHANES III standardized reference data for T-Score and Z-Score calculations.

Indications for Use

Indicated for patients requiring bone mineral density (BMD) and bone mineral content (BMC) estimation in the proximal femur and spine to assess fracture risk and therapy effectiveness.

Regulatory Classification

Identification

A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows handwritten text on a white background. The text at the top reads "K03/991". Below this, the text reads "Page 1 of 2". The handwriting is somewhat stylized and appears to be from a document or page numbering system. # 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS (Submitted in accordance with the requirements of 21 CFR 807.3) | Contact information: | Ben Arnold, Ph.D. | |----------------------|-----------------------------------------------------------------------------| | | President | | | Image Analysis, Inc. | | | 1380 Burkesville Street | | | Columbia, KY 42728 | | | PH: 270-384-6400 | | | FAX: 270-384-6405 | | | Email: <a href="mailto:info@image-analysis.com">info@image-analysis.com</a> | | Date: | June 4, 2003 | | Device/Trade Name: | DXAVIEW <sup>TM</sup> Hip and Spine | | Common/Usual Name: | QCT Bone Densitometry | | Classification Name: | Bone Densitometer, 21 CFR 892.1170, Class II | Predicate Devices: K992246: OCT Bone Mineral Analysis Software Intended Use: OCT Bone Densitometry to estimate bone density and bone density changes, for monitoring therapy, and to estimate fracture risk. K941127: OCT Bone Mineral Analysis Software Intended Use: Estimation of bone density in the spine. K943505: Hologic ODR 3000 X-Ray Bone Densitometer Intended Use: Estimate bone mineral density and bone mineral content at various anatomical sites. K002113: CTXA Hip Intended Use: Estimate bone density and bone mineral content in the hip. Compare to reference data to determine T-Scores and fracture risk. Device Description: The DXAView" Hip and Spine are software Options to the Image Analysis QCT-3D Plus product, which provides 3D projection measurements similar to conventional DXA. The product Options provide estimates of bone mineral content (BMC) and bone density (BMD) of the proximal femur and spine. BMC is expressed in grams and BMD in g/cm2 of calcium hydroxyapatite. DXAView™ uses phantom calibration by Quantitative Computed Tomography (QCT) on all compatible CT scanners. The computations are preformed on 3D CT volumetric image sets obtained with conventional CT protocols and radiation doses. {1}------------------------------------------------ KOZIRAL Rogozloff ### Intended Use: The DXAView™ Hip and Spine bone densitometry options are intended to estimate bone mineral content (BMC) and bone mineral density (BMD) in the proximal femur and spine. The BMD may be compared to DXA reference data. T-Scores and Z-Scores are calculated with respect to young normal and age matched reference data, and can be used by the physician in determining fracture risk. Summary of Fundamental Scientific Technology and Comparison with Predicate Devices: The DXAVIEW™ bone densitometry Options provide estimates of bone mineral content (BMC) and bone mineral density (BMD) values similar to those obtained from the predicate DXA device (K943505: Hologic ODR 3000 X-Ray Bone Densitometer) for regions of interest in the proximal femur and spine. The predicate device, CTXA Hip K002113 measures bone density in the hip using calibrated CT volumetric images, and is highly similar to DXAVIEW™. DXAVIEW™ uses the same scanning and calibration procedures to acquire and calibrate the CT image as are used for the predicate device K992246: QCT Bone Mineral Analysis Software. NHANES III Standardized Reference Data are used to calculate T-Scores and Z-Scores. Pearson's R values between DXA and DXAVIEW™ on clinical comparison studies has been shown to be 0.93 to 0.94 for the neck and total femur regions. The uncertainty between the two results are about 1/3 of 1 population Standard Deviation (T=1/3), and are quite acceptable. The DXAVIEW™ BMD estimates are compared to reference populations to determine T-Scores, which assists physicians in identifying patients with low bone density and estimating fracture risk. ## Summary of Clinical Performance Data: Patient validation studies indicate a software precision of 0.012 g/cm² for total femur and 0.017 g/cm² for the femoral neck. Clinical studies comparing DXA and DXAVIEW™ for BMD of the hip showed correlation coefficients R of 0.93-0.94 for the total femur and femoral neck. #### Conclusions: The DXAVIEW Hip and Spine Bone Densitometry Options are substantially equivalent to the predicate devices while producing comparable results. Correlations between DXAVIEW™ and DXA are on the order of 0.93 and sufficiently high to warrant the use of the NHANES III Reference Data. The radiation dose resulting from the CT Scans used for analysis by the DXAVIEWTM software is comparable to other CT studies. Ben Arnold Signature Ben Arnold, Ph.D. Printed Name President Title {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the edge. Inside the circle is a stylized eagle with its wings spread. The eagle is facing left and has three horizontal lines above it, representing the stripes of the American flag. The seal is black and white and appears to be a logo or emblem. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 3 0 2003 Ben Arnold, Ph.D. President Image Analysis, Inc. 1380 Burkesville Street COLUMBIA KY 42728 Re: K031991 Trade/Device Name: DXAVIEW™ Regulation Number: 21 CFR 892.1170 Regulation Name: Radiographic Bone Densitomer Regulatory Class: II Product Code: 90 KGI Dated: June 19, 2003 Received: July 30, 2003 Dear Dr. Arnold: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATIONS FOR USE K 031991 510 (k) Number (if known): Device Name: QCT DXAVIEW™ The DXAVIEW™ Hip and Spine bone densitometry Options are intended to provide estimates of bone mineral density (BMD) and bone mineral content (BMC) in the proximal femur and spine. Young normal reference data taken from caucasian populations can be used to calculate T-Scores. Age matched normal reference data arc used to calculate Z-Scores. The BMD results, along with the T- and Z-Scores, can be used by physicians in their determination of fracture risk and therapy effectivencss. NancyC Broadon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K031441 Prescription Use (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...