← Product Code [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK) · K991764

# SOMATOM PROJECT 10-CT SCANNERS (K991764)

_Siemens Medical Solutions USA, Inc. · JAK · Aug 17, 1999 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K991764

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK.md)
- **Decision Date:** Aug 17, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The SOMATOM P10 systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes* taken at different angles. (*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)

## Device Story

SOMATOM Project 10 is a whole-body X-ray computed tomography (CT) scanner; utilizes continuously rotating tube-detector system; operates on fan beam principle. Inputs: X-ray transmission data from axial or spiral planes. Transformation: Computer reconstruction of transmission data into cross-sectional images. Output: Cross-sectional body images for clinical evaluation. Used in clinical settings by trained medical professionals. Healthcare providers use images for diagnostic assessment and clinical decision-making. Benefits: Non-invasive visualization of internal body structures.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Whole-body X-ray CT scanner; fan beam geometry; continuously rotating tube-detector system. Operates with SOMARIS/4.5 software. Complies with 21 CFR Subchapter J, IEC 601-1, and UL 187 safety standards.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Siemens SOMATOM Plus 4
- General Electrics Hispeed (DX/i and LX/i)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SIEMENS

AUG 17 1999

K991764

.

# Attachment 1

# Summary of Safety and Effectiveness

This summary of 510(k)-safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

.

:

.

- General Information I.

| Device Name:          | SOMATOM Project 10 Computed Tomography X-ray<br>Systems                      |
|-----------------------|------------------------------------------------------------------------------|
| Classification Name:  | & 892.1750:<br>Computed tomography X-ray system                              |
| Propriety Trade Name: | SOMATOM Star / SOMATOM Sprinter                                              |
| Classification:       | Class II                                                                     |
| Performance Standard: | 21 CFR Subchapter J,<br>Federal Diagnostic X-ray Equipment Standard          |
| Registration Number:  | 2240869                                                                      |
| Address:              | Siemens Medical Systems, Inc.<br>186 Wood Avenue South<br>Iselin, N.J. 08830 |
| Contact Person:       | Kathleen M. Rutherford<br>Manager, Regulatory Submissions<br>(732) 321-4779  |

Date of Summary Preparation: 5/20/99

{1}------------------------------------------------

# II. Safety and Effectiveness Information Supporting the Substantial Equivalence Determination

# Device Description:

The Siemens SOMATOM Project 10 systems are a whole body X-ray computed tomography scanners, which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system software is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image . . archive/evaluation.

#### Intended Use:

The SOMATOM P10 systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes* taken at different angles.

(*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)

#### Technological Characteristics:

The SOMATOM Project 10 systems are whole body X-ray computed tomography scanners, which features a continuously rotating tube-detector system and functions according to the fan beam principle. The systems are based on the existing SOMATOM Plus 4 system (for further details see chapter 2). The system will operate with SOMARIS/4.5 software.

#### General Safety and Effectiveness Concerns:

All components of the SOMATOM Project 10 systems subject to the Federal Diagnostic Equipment Performance Standard and applicable regulations of 21CFR § 1020.30 and § 1020.33 are certified to meet those requirements; and an initial report as per 21 CFR § 1002.10 will be filed with the Center for Devices and Radiological Health (CDRH). To minimize electrical, mechanical, and radiation hazards. Siemens adheres to recognized and established industry The SOMATOM is designed to meet the ELECTRICAL AND MECHANICAL practice. SAFETY STANDARD IEC 601-1 and UL 187 X-RAY EQUIPMENT STANDARD FOR SAFETY.

#### Substantial Equivalence:

The SOMATOM Project 10 systems operating with SOMARIS/4.5 software are substantially equivalent to the Siemens SOMATOM Plus 4 and General Electrics Hispeed (DX/i and LX/i) systems in commercial distribution.

Kattlen Rutherford

Kathleen Rutherford Manager, Regulatory Submissions

5/21/99
Date

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 7 1999

Alicia Juergensen Technical Specialist Siemens Medical Systems, Inc. 186 Wood Avenue South Iselin, NJ 08830

RE:

K991764 Somatom Project 10-CT Scanners Dated: May 20, 1999 Received: May 24, 1999 Regulatory Class: II 21 CFR 892.1750/Procode: 90 JAK

Dear Ms. Juergensen:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been redassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

[signature]

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

#### Attachment 2

# Indication for use

510(k) Number (if known):

K991764

Device Name:

SOMATOM Project 10 systems

Indication for use:

The SOMATOM Project 10 systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes* taken at different angles.

(*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)

Daniel A. Szymon
(Division Sign-Off)

Division of Reproductive, Abdominal, ENT, and Radiological Devi 510(k) Number

Prescription Use
(Per 21 CFR 801.109)

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K991764](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K991764)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
