Pivotal validation dataset: 236 independent cases, sourced from diverse U.S. sites and scanner vendors.
Indications for Use
CT VScore+ is a software application intended for non-invasive evaluation of calcified lesions of the coronary arteries based on ECG-gated, non-contrast cardiac CT images for patients aged 30 years or older. The device automatically generates calcium scores for the coronary arteries (combined LM+LAD, RCA, LCX) and highlights the segmented calcium on the original CT image. The device also offers the option for the user to display the calcium scores in the context of reference data from the MESA and Hoff-Kondos databases. The segmented arteries include combined LM+LAD, RCA, and LCX. To obtain separate LM and LAD results, the user must perform manual segmentation. The segmentation map of calcifications is intended for informational use only and is not intended for detection or diagnostic purposes. The 3D Calcium View output is provided strictly as an informational and supplementary output and should never be used alone as the method of reviewing the calcium segmentation.
Device Story
Software application (SaMD) running on Vitrea platform; processes ECG-gated, non-contrast cardiac CT DICOM images. Uses deep learning (Attention UNet) to automatically segment coronary calcifications and calculate Agatston and volume scores for combined LM+LAD, RCA, and LCX arteries. Highlights segmented calcium on original CT images; provides optional 3D visualization and comparison against MESA/Hoff-Kondos reference databases. Operated by clinicians; allows manual editing of automated segmentations and anatomical assignments. Outputs exported via DICOM/DICOM SR for clinical review. Assists physicians in quantifying coronary artery calcium to support cardiovascular risk assessment. Benefits include standardized, automated quantification of calcified lesions, potentially improving efficiency and consistency in cardiac risk evaluation.
Clinical Evidence
Bench testing using 236 independent clinical cases; ground truth established by 2+1 consensus of board-certified physicians. Primary endpoints: Agatston/Volume score agreement (ICC >0.99), CAC-DRS/Standard risk classification (Kappa >0.95), and segmentation accuracy (Dice score ≥0.87). Subgroup analysis confirmed robust performance across scanner vendors, slice thicknesses, and patient demographics. No clinical prospective/retrospective patient outcome data; bench testing only.
Technological Characteristics
SaMD; deep learning-based segmentation using proprietary Attention UNet architecture. Inputs: ECG-gated, non-contrast cardiac CT DICOM images. Outputs: Agatston/Volume scores, segmented calcium overlays, 3D visualization. Connectivity: DICOM/DICOM SR export. Platform: Vitrea. No hardware components.
Indications for Use
Indicated for non-invasive evaluation of calcified coronary artery lesions in patients aged 30 years or older using ECG-gated, non-contrast cardiac CT images.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K221219 — syngo.CT CaScoring · Siemens Medical Solutions USA, Inc. · May 17, 2022
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 28, 2025
Canon Medical Informatics, Inc.
% Jay Vaishnav
Director, Regulatory Affairs
5850 Opus Parkway
Suite 300
MINNETONKA, MN 55343
Re: K252217
Trade/Device Name: CT VScore+
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: JAK
Dated: October 17, 2025
Received: October 27, 2025
Dear Jay Vaishnav:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K252217 - Jay Vaishnav
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252217 - Jay Vaishnav
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252217 | |
| Device Name CT VScore+ | |
| Indications for Use (Describe) CT VScore+ is a software application intended for non-invasive evaluation of calcified lesions of the coronary arteries based on ECG-gated, non-contrast cardiac CT images for patients aged 30 years or older. The device automatically generates calcium scores for the coronary arteries (combined LM+LAD, RCA, LCX) and highlights the segmented calcium on the original CT image. The device also offers the option for the user to display the calcium scores in the context of reference data from the MESA and Hoff-Kondos databases. The segmented arteries include combined LM+LAD, RCA, and LCX. To obtain separate LM and LAD results, the user must perform manual segmentation. The segmentation map of calcifications is intended for informational use only and is not intended for detection or diagnostic purposes. The 3D Calcium View output is provided strictly as an informational and supplementary output and should never be used alone as the method of reviewing the calcium segmentation. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Canon
Made For life
CT VScore+ Traditional 510(k) Submission
# CT VScore+ 510(k) Summary
510(k) #: K252217
Prepared on: 2025-11-21
# Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Canon Medical Informatics, Inc. |
| --- | --- |
| Applicant Address | 5850 Opus Parkway, Suite 300
Minnetonka MN 55343
United States of America |
| Applicant and Correspondent | Jay Vaishnav, PhD, RAC, FRAPS
Director, Regulatory Affairs
Phone 952-487-9530
E-mail Jay.Vaishnav@mi.medical.canon |
| Alternate Contacts | Vincent Swenson
Senior Director, Quality and Regulatory
Phone 952-487-9548
E-mail Vincent.Swenson@mi.medical.canon
Gargeyi Pavuluri
Senior Regulatory Affairs Specialist
Phone 952-487-9875
E-mail Gargeyi.Pavuluri@mi.medical.canon
Alisha Bouley
Associate Regulatory Affairs Specialist
Phone 952-487-9530
E-mail Alisha.Bouley@mi.medical.canon |
# Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | CT VScore+ |
| --- | --- |
| Classification Name | Computed tomography x-ray system |
| Common Name | System, X-Ray, Tomography, Computed |
| Regulation Number | 21 CFR 892.1750 |
| Product Code(s) | JAK |
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | +1 866.433.4624 |
{5}
Canon
Made For life
CT VScore+ Traditional 510(k) Submission
# Legally Marketed Predicate Device
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name | Product Code |
| --- | --- | --- |
| K213725 | GE CardIQ Suite | JAK |
# Device Description Summary
21 CFR 807.92(a)(4)
# How the Device Functions
CT VScore+ is a software application intended for non-invasive evaluation of calcified lesions of the coronary arteries based on ECG-gated, non-contrast cardiac CT images for patients aged 30 years or older. The application runs on the Vitrea platform.
The device automatically generates Agatston and volume calcium scores for each of the coronary arteries (combined LM+LAD, RCA, LCX) based on the volume and density of the calcium deposits and highlights the Segmented calcium on the original CT image. The device also offers the option for the user to display the calcium scores in the context of reference data from the MESA and Hoff-Kondos databases.
The software uses deep learning-based segmentation methods. Users can edit the automated segmentation, including manually assigning calcifications to anatomical structures.
The device automatically outputs a combined LM+LAD score as the final automated output. To obtain separate LM and LAD results, the user must perform manual segmentation using the provided editing tools.
# Scientific Concepts That Form The Basis Of The Device/Principle Of Operation For Achieving Intended Effect
The device reads DICOM images that meet input criteria. Using automatic processing, the device returns calcium segmentation and Agatston and volume scores for coronary arteries only. Optionally, and for reporting purposes, the device displays patient scores in the context of well-established population reference databases and calcium scoring reporting systems. Users can edit any automated outputs.
# Proposed Conditions for Use
The application is intended to operate on ECG-gated, non-contrast cardiac CT images. The device is indicated for patients aged 30 or older.
# Physical and Performance Characteristics of the Device
The device is Software as a Medical Device (SaMD) that operates on ECG-gated, non-contrast cardiac CT DICOM images.
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | +1 866.433.4624 |
{6}
Canon
Made For life
CT VScore+ Traditional 510(k) Submission
Device development used a total of 210 unique patient datasets adhering to the inclusion/exclusion criteria, gathered from four institutions (two US sites and two Japanese sites).
Device performance was validated on 236 cases, in a study by three US board-certified physicians. In addition to this testing, software verification and validation activities were completed to ensure that the software functions remained consistent with the software requirements. The software achieved all product release criteria.
The developmental and validation data are independent at both the patient level and the site level.
# How the Device Interacts With Other Devices and the Patient
The device does not interact directly with the patient. The device is a software application that runs on the Vitrea platform and processes ECG-gated non-contrast cardiac CT DICOM images. The device automatically generates Agatston and volume calcium scores for each of the coronary arteries (LAD+LM, RCA, LCX) based on the volume and density of the calcium deposits and highlights the segmented calcium on the original CT image. Results can be exported to image management, archival, or reporting systems that support DICOM standards for further review and interpretation.
Results can also be saved in DICOM Structured Reports (DICOM SR) format.
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
CT VScore+ is a software application intended for non-invasive evaluation of calcified lesions of the coronary arteries based on ECG-gated, non-contrast cardiac CT images for patients aged 30 years or older. The device automatically generates calcium scores for the coronary arteries (combined LM+LAD, RCA, LCX) and highlights the segmented calcium on the original CT image. The device also offers the option for the user to display the calcium scores in the context of reference data from the MESA and Hoff-Kondos databases.
The segmented arteries include combined LM+LAD, RCA, and LCX. To obtain separate LM and LAD results, the user must perform manual segmentation. The segmentation map of calcifications is intended for informational use only and is not intended for detection or diagnostic purposes. The 3D Calcium View output is provided strictly as an informational and supplementary output and should never be used alone as the method of reviewing the calcium segmentation.
# Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device has a more restrictive IFU and includes a subset of the predicate functionality, non-invasive evaluation of calcified lesions of the coronary arteries based on ECG-gated, non-contrast cardiac CT images for patients aged 30 years or older. The predicate device also performs other functions, including analysis of CCTA images. This change in the IFU does not create a new intended use.
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | +1 866.433.4624 |
{7}
Canon
Made For life
CT VScore+ Traditional 510(k) Submission
# Technological Comparison
21 CFR 807.92(a)(6)
CT VScore+ incorporates a machine learning algorithm (deep learning-based segmentation) for coronary calcification identification and quantification. The device uses a proprietary Attention UNet architecture for segmentation, trained on a diverse dataset of ECG-gated, non-contrast cardiac CT scans. The device outputs Agatston and volume scores for combined LM+LAD, LCX, and RCA, and also provides a 3D calcium view for informational, supplementary visualization.
The table below summarizes the key feature/technological differences and similarities between the predicate device and the proposed device:
| Characteristic | Subject Device (CT VScore+) | Predicate Device (CardIQ Suite K213725) | Analysis |
| --- | --- | --- | --- |
| Indications for Use | CT VScore+ is a software application intended for non-invasive evaluation of calcified lesions of the coronary arteries based on ECG-gated, non-contrast cardiac CT images for patients aged 30 years or older. The device automatically generates calcium scores for the coronary arteries (combined LM+LAD, RCA, LCX) and highlights the segmented calcium on the original CT image. The device also offers the option for the user to display the calcium scores in the context of reference data from the MESA and Hoff-Kondos databases. The segmented arteries include combined LM+LAD, RCA, and LCX. To obtain separate LM and LAD results, the user must perform manual segmentation. The segmentation map of calcifications is intended for informational use only and is not intended for detection or diagnostic purposes. The 3D Calcium View output is provided strictly as an informational and supplementary output and should never be used alone as the method of reviewing the calcium segmentation. | CardIQ Suite is a non-invasive software application designed to provide an optimized application to analyze cardiovascular anatomy and pathology based on 2D or 3D CT cardiac non contrast and angiography DICOM data from acquisitions of the heart. It provides capabilities for the visualization and measurement of vessels and visualization of chamber mobility. CardIQ Suite also aids in diagnosis and determination of treatment paths for cardiovascular diseases to include, coronary artery disease, functional parameters of the heart, heart structures and follow-up for stent placement, bypasses and plaque imaging. CardIQ Suite provides calcium scoring, a non-invasive software application, that can be used with non-contrasted cardiac images to evaluate calcified plaques in the coronary arteries, heart valves and great vessels such as the aorta. Calcium Scoring may be used to monitor the progression/regression of calcium in coronary arteries overtime, which may aid in the prognosis of cardiac disease. | The subject device has a more restrictive IFU and includes a subset of the predicate functionality, non-invasive evaluation of calcified lesions of the coronary arteries based on ECG-gated, non-contrast cardiac CT images for patients aged 30 years or older. The predicate device also performs other functions, including analysis of CCTA images. This change in the IFU does not create a new intended use. |
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | +1 866.433.4624 |
{8}
Canon
Made For life
CT VScore+ Traditional 510(k) Submission
| Intended Use | The device is intended for use in adults aged 30 years and older. It automatically generates calcium scores for the coronary arteries and highlights the segmented calcifications on the original CT images. The intended use is to quantify coronary artery calcium from ECG-gated, non-contrast cardiac CT images to support clinical interpretation of cardiovascular risk. | CardIQ Suite is a collection of non-invasive software features intended to analyze CT cardiovascular anatomy and pathology and aid in determining treatment paths. | Subject device has a more restrictive IFU, subset of predicate functionality. No new intended use. |
| --- | --- | --- | --- |
| Device Input | Non-contrast, ECG-gated cardiac CT scans | Non-contrast, ECG gated cardiac CT scans or coronary CT angiography images (CCTA) | The subject device processes noncontrast cardiac CT coronary artery calcium (CAC) scans only, and does not operate on CCTA images.
CAC and CCTA scans serve different clinical functions. This does not impact the safety or effectiveness of the subject device compared to the predicate. |
| Device Output | DICOM images with calcium highlighted. Total calcium scores as well as per-vessel scores for the LM+LAD, LCX, and RCA | Same | Same |
| Identification of calcium | Yes | Same | Same. |
| Comparison of calcium scores to cited literature | Yes | Yes | Similar. Both devices offer the option for the user to display the calcium scores in the context of data from well-established reference population databases (MESA; Hoff-Kondos). |
| Associates calcium with artery | Yes | Same | Similar. Both devices use deep learning methods to segment the coronary artery regions. |
| Methods | Deep learning methods used to segment coronary regions and assign calcium | Same | Exact predicate methods are unknown but likely similar. |
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | +1 866.433.4624 |
{9}
Canon
Made For life
CT VScore+ Traditional 510(k) Submission
| Automation | Semi-automatic; Software initializes coronary calcium results. | Same | Same |
| --- | --- | --- | --- |
| Interactive definition of ROIs and assignment of calcium | Users can manually edit results, including manually assigning calcifications to anatomical structures | Same | Same. Both devices allow users to manually edit calcifications and assign them to labeled anatomical structures |
| Export as DICOM SR | Yes | Same | Same |
| 3D Calcium View | A 3D rendered image, showing an optional supplemental view of segmented calcium in relation to full anatomy. | Not specified | Subject device output is supplementary only and should not be used for primary review. |
In summary, the subject and predicate devices use different proprietary deep learning algorithms for coronary region segmentation. These differences do not raise new questions of safety or effectiveness. The subject device's algorithms have been validated through performance testing, meeting pre-specified acceptance criteria and demonstrating accuracy and reliability.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
CT VScore+ is an automatic coronary artery calcium (CAC) scoring software that incorporates a machine learning algorithm for coronary calcification segmentation. Performance testing was conducted to validate analytical accuracy and demonstrate substantial equivalence to the predicate device. Testing included retrospective clinical datasets and followed FDA-recognized statistical methods.
## Truthing
The reference standard ground truth was established by consensus manual scoring on an FDA-cleared device (Vitrea CT VScore, K243240) and a 2+1 consensus process by three U.S. board-certified radiologists/cardiologists.
## Dataset Independence
The CT VScore+ deep learning model was developed and validated using separate, diverse datasets to ensure robust and unbiased performance. The following is information about the development and validation datasets:
- Development dataset: 210 cases (94 training, 40 validation, 76 internal testing).
- Pivotal validation dataset: 236 independent cases, sourced from diverse U.S. sites and scanner vendors.
There was no overlap in sites or cases between the development and pivotal validation datasets.
## Classification Performance-Based Summary
The following metrics were used to assess the device's overall performance:
- Total Agatston Score ICC(2,1): 0.997 [95% CI: 0.996–0.998] (Acceptance: >0.95)
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | +1 866.433.4624 |
{10}
Canon
Made For life
CT VScore+ Traditional 510(k) Submission
Total Volume Score ICC(2,1): 0.996 [95% CI: 0.995–0.997] (Acceptance: >0.95)
Per-Vessel ICCs: LCx: 0.937; RCA: 0.990; LM+LAD: 0.983 (Acceptance: >0.90)
CAC-DRS 4-Class Kappa: 0.959 [95% CI: 0.936–0.982] (Acceptance: >0.90)
CAC Standard 5-Class Kappa: 0.958 [95% CI: 0.938–0.978] (Acceptance: >0.90)
$\bigcirc$ Voxelwise Dice Score: 0.920 overall; per-vessel DS: LCx 0.874, RCA 0.883, LM+LAD 0.958 (Informational metric)
In addition, Bland-altman analysis was conducted to demonstrate the agreement between the device's outputs and ground truth. The device met all predefined acceptance criteria, confirming strong agreement with ground truth. Subgroup performance results follow:
# Technical Characteristics
- Scanner Vendors: Canon, GE, Siemens, Philips – ICC consistently >0.98 across all.
- Slice Thickness: 2.5 mm vs. 3.0 mm – ICC consistently >0.99 across all.
- Reconstruction Kernel (FBP) – ICC consistently >0.99 across all.
- Noise Levels: ICC consistently $>0.98$ across noise levels.
# Patient Demographics
- Patient Demographics: ICC>0.98 for all sex and age groups.
- Geographic Diversity: Data from six U.S. states across four U.S. regions and multiple institution types. ICC>0.98 for all.
# Disease Status
- Valve calcification severity: The device showed uniformly high agreement across "none," "moderate," and allowed "severe" categories (ICC ≈ 0.993–0.999), indicating that the presence of incidental valve calcification did not affect coronary calcium scoring performance. Severe mitral valve calcification cases were excluded.
- CAC Burden: Overall, subgroup testing confirmed robust performance across patient demographics, disease severity levels, and incidental cardiac findings. Detailed results are below:
○ Very high calcium ranges (CAC-DRS 3, ≥300 AU; CAC-5-STD-4, >400 AU): ICC values were extremely high (>0.99), confirming strong stability in higher-burden cases.
○ Moderate calcium ranges (CAC-DRS 1&2, 1-299 AU; CAC-5-STD-2&3, 11-400 AU): Agreement remained high (ICC ≈ 0.86–0.97), meeting expected acceptance levels for these clinically variable categories.
Low calcium ranges (CAC-5-STD-1, 1–10 AU) and Zero calcium (CAC-DRS-0 and CAC-5-STD-0, 0 AU): ICC is not applicable. These cases were handled per categorical scoring analysis: Among the fifty cases with confirmed zero calcium scores, nine (18%) were assigned nonzero scores by the software. No confirmed nonzero scores were assigned CAC=0.
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | +1 866.433.4624 |
{11}
Canon
Made For life
CT VScore+ Traditional 510(k) Submission
Users should note the possibility that CAC=0 cases may be misclassified as having low nonzero CAC. All results should be reviewed by a physician, particularly in score ranges of 0-10 AU.
## Conclusion
CT VScore+ demonstrated high analytical validity with ICC >0.99 for total scores, Kappa >0.95 for risk classification, and robust segmentation accuracy (Dice score ≥0.87 per vessel). Performance was consistent across subgroups, supporting generalizability. All acceptance criteria were met, and no new safety or effectiveness concerns were identified. The test results support the conclusion that CT VScore+ is as safe and as effective as the predicate device cleared under K213725.
Canon Medical Informatics, Inc. 5850 Opus Parkway, Suite 300 | Minnetonka, MN 55343 USA | +1 866.433.4624 |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.